The largest database of trusted experimental protocols

Accela 600 hplc

Manufactured by Thermo Fisher Scientific
Sourced in Germany, United States

The Accela 600 HPLC is a high-performance liquid chromatography system designed for efficient and reliable separation and analysis of a wide range of chemical compounds. It features a high-pressure pump capable of delivering mobile phases at flow rates up to 10 mL/min, and a variable wavelength UV-Vis detector for sensitive detection of analytes.

Automatically generated - may contain errors

5 protocols using accela 600 hplc

1

LC-MS Analysis of Small Molecules

Check if the same lab product or an alternative is used in the 5 most similar protocols
For LC-MS analysis, a 1 mg/mL sample was prepared in a 1:1 MeOH–ACN solvent system. Experiments were carried out using an Exactive mass spectrometer with an electrospray ionization source attached to an Accela 600 HPLC pump with an Accela autosampler and UV/Vis detector (Thermo Scientific, Bremen, Germany). The mass accuracy was set to less than 3.0 ppm. The Orbitrap mass analyzer can limit the mass error to ±3.0 ppm. Mass spectrometry was carried out over a mass range of 100–2000 m/z in positive and negative ionization modes with a spray voltage of 4.5 kV and capillary temperature of 270 °C. About 10 μL was injected from each vial, at a flow rate of 300 μL/min. The column used was an ACE5 C18 column (5 μm × 75 mm × 3 mm) (Hichrom Limited, Reading, UK). A binary gradient method was utilized using the two solvents, A (water and 0.1% formic acid) and B (MeCN and 0.1% formic acid). The gradient was carried out for 45 min and the program followed; at zero minutes A = 90% and B = 10%, at 30 min A = 0% and B = 100%, at 36 min A = 90% and B = 10% until end at 45 min. The UV absorption wavelength was set at 254 nm, the sample tray temperature was maintained at 4 °C, and the column at 20 °C. The samples were run sequentially, with solvent and media blanks analyzed first. LC-MS data were acquired using Xcalibur version 2.2.
+ Open protocol
+ Expand
2

Quantification of Vitamin A in Oils

Check if the same lab product or an alternative is used in the 5 most similar protocols
Portions of all oil samples (~ 20 mL) were shipped to a commercial, independent, laboratory (Intertek, Germany). These oils were analyzed by HPLC, with method replication carried out every ten samples on a different sequence (these replicates were used to calculate the inter-assay precision, “Procedure for Assessment of the Methods”). Briefly, oil samples (1 g) were filled up to 10 mL with methyl tert-butyl ether in dibuthylhydroxytoluene (1 g/L). After thoroughly shaking the solution, methanol was added (1:1) and injected (20 µL) to the HPLC. Vitamin A was analyzed on a Thermo Scientific Accela 600 HPLC (Waltham, USA), coupled to a fluorescence detector (λex, 325 nm; λem, 480 nm, Thermo Scientific Surveyor FL Plus, Serial number 650212). The vitamin A was separated using a Thermo hypersil GOLD C18 (150 × 4.6 mm, 3 µm). The mobile phase included methanol (solvent A) and methyl tert-butyl ether (solvent B), programmed as follows: 0–10.0 min: 0% B, 10–14 min: 100% B, 14–19 min: 0% B, at a flow rate of 1.0 mL/min. Retinyl palmitate in methanol was used as an external calibrator. The limit of quantification was 0.55 mg RE/kg as defined by the commercial independent laboratory.
+ Open protocol
+ Expand
3

Quantitative Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Liquid chromatography–mass spectrometry (LC-MS) analyses were performed by electrospray ionization (ESI)-MS with an Exactive mass spectrometer (Thermo Scientific) interfaced with an Accela 600 HPLC (Thermo Scientific). A portion of the sample was loaded onto a Hydrosphere C18 column (4.6 mm, 250 mm, 5 µm; YMC, Kyoto, Japan). Separation was achieved by elution with acetonitrile/0.5% acetic acid aqueous solution (1:1, v/v) at a constant temperature of 30 °C and a flow rate of 1.0 mL/min.
Nuclear magnetic resonance (NMR) spectra were recorded on an ANANCE 400 NMR spectrometer (Bruker Biospin, Billerica, MA, USA). Chemical shifts were reported as δ values (ppm) relative to internal tetramethylsilane (TMS) and the coupling constants (J) were given in Hz.
+ Open protocol
+ Expand
4

SCFA Analysis via HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
SCFA identification and quantification was performed as in Panzella et al. (2017) (link). The equipment used was an Accela 600 HPLC (Thermo Scientific) with a quaternary pump, an autosampler and a UV-VIS PDA detector set at 210 nm. We used 0.1 M phosphate buffer (pH 2.8)/acetonitrile 99:1 v/v with a flow rate of 1.25 mL/min as mobile phase. An Aquasil C18 reverse phase (Thermo Scientific) (150 × 4.6 mm, 5 μm) was used with a run time of 30 min. The samples did not require any pretreatment before injecting except for centrifugation at 13000 rpm and further filtration through 0.22 μm nylon filter. One mL of fermentation supernatant was centrifuged and filtered prior injection of 20 μL. The SCFA standards (acetate, propionate and butyrate) were prepared in the mobile phase at concentrations ranging from 5 to 10000 ppm. Results were expressed as mM.
+ Open protocol
+ Expand
5

HPLC Analysis of Polyphenol Bioaccessibility

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC analysis of EA during in vitro digestion was run on an Agilent 1100 Series instrument using a reverse phase Phenomenex Sphereclone 250 × 4.60 mm column (particle size 5 µm). A gradient elution at a flow rate of 0.7 mL/min was performed using 0.1% formic acid in water (eluant A) and methanol (eluant B): 5% B, 0–10 min; 5–80% B (10–45 min); detection wavelength was fixed at 254 nm.
Short chain fatty acids (SCFAs) determination was carried out on an Accela 600 HPLC (Thermo Scientific, Munich, Germany) equipped with a pump, an autosampler, and a UV–VIS PDA detector set at 210 nm; the mobile phase used was 0.1 M phosphate buffer (pH 2.8)/acetonitrile 99:1 v/v delivered at a 1 mL/min flow rate; the column used was an Aquasil C18 reversed phase (Thermo Scientific, Munich, Germany) (150 × 4.6 mm, 5 µm), with a total run time of 20 min.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!