The largest database of trusted experimental protocols

Su8020 ultra high resolution field emission scanning electron microscope

Manufactured by Hitachi
Sourced in Japan

The SU8020 is an Ultra-High Resolution Field Emission Scanning Electron Microscope (FE-SEM) manufactured by Hitachi. It is designed to provide high-resolution imaging and analysis of a wide range of samples. The SU8020 utilizes a field emission electron gun to generate a high-brightness electron beam, enabling the microscope to achieve a high resolution at low accelerating voltages.

Automatically generated - may contain errors

3 protocols using su8020 ultra high resolution field emission scanning electron microscope

1

Comprehensive Material Characterization Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
The X-ray diffraction (XRD) patterns were recorded on X’Pert–PRO, Panalytical, Almelo, The Netherlands, 2012 using Cu Kα radiation.
The N2 sorption isotherms were measured at −196 °C using an ASAP Sorption Surface Area and Pore Size Analyzer (ASAP 2460, Micrometrics, Norcross, GA, USA 2018).
Scanning electron microscopy was performed with an SU8020 Ultra-High Resolution Field Emission Scanning Electron Microscope (Hitachi Ltd., Ibaraki, Japan, 2012). The elemental analysis was conducted using Energy-Dispersive X-ray spectrometers (EDX) with the same instrument. The samples for SEM were sputter-coated with 40 nm of chromium in order to reduce charging.
FT-IR spectra of the catalyst were obtained using a Thermo Finnigan Nicolet 380 FT-IR instrument with an ATR Smart iTX attachment (Thermo Fisher Scientific Inc., Waltham, MA, USA) in the wavenumber range from 490 to 4000 cm−1.
Diffuse reflectance UV-Vis spectra of the catalyst in the wavelength range from 190 to 900 nm were obtained using a Jasco 650 (V-650, Jasco, Tokyo, Japan) spectrometer with a PIV-756 horizontal integrating sphere.
+ Open protocol
+ Expand
2

Spectroscopic Characterization of Metal-Organic Interactions

Check if the same lab product or an alternative is used in the 5 most similar protocols
Sodium hydroxide, CuSO4, 3,4-Dihydroxy-phenylalanine, and other metal ions were purchased from Sigma-Aldrich reagent company (Shanghai, China). N-2-hydroxyethylpiperazine-N-ethane-sulphonic acid (HEPES) was purchased from Aladdin chemistry company (Shanghai, China). All reagents were of analytical grade and used without further purification. Water was purified by a Milli-Q system. UV–vis absorbance spectrum measurements were performed on a PerkinElmer Lambda 750S UV–visible Spectrophotometer (PerkinElmer, Waltham, MA, USA). Fluorescent spectrum and lifetime were measured on a HORIBA Spectro fluorophotometer (Horiba, Kyoto, Japan). X-ray photoelectron spectroscopy (XPS) was conducted on a Thermo Fisher Scientific Escalab 250Xi photoelectron spectrometer (Thermo Fisher, Waltham, MA, USA). The scanning electron microscope images were captured on a SU8020 ultra-high resolution field emission scanning electron microscope (Hitachi, Tokyo, Japan).
+ Open protocol
+ Expand
3

Characterizing Thin Metallic Films by SEM

Check if the same lab product or an alternative is used in the 5 most similar protocols
The measurements for thin films were performed with the use of a high-resolution scanning electron microscope (SU8020 Ultra-High Resolution Field Emission Scanning Electron Microscope; Hitachi Ltd., Tokyo, Japan, 2012) equipped with a cold field electron emitter. The setup was screened against changes of magnetic field enabling high-quality ultra-high magnification of 200,000 times. The samples in the form of glass plates covered with thin metallic film were mounted over the sample holder with conductive carbon tape. Additionally, in order to avoid charging effects, the metal surface was connected to the ground, also with conductive carbon tape. A 5 kV accelerating voltage was applied for the SEM images.
The elemental composition of the thin metallic surfaces was characterized with an analyzer of X-ray photons emitted from the sample during electron bombardment. The Energy Dispersive Spectroscopy system (Thermo Scientific Ultra Dry, Pittsburgh, PA, USA) was applied. The samples were excited with 20 keV electrons during elemental analysis.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!