The largest database of trusted experimental protocols

5 protocols using dg 2080 53

1

Chromatographic Analysis and Spectroscopic Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
Thin layer chromatography (TLC) was conducted by using precoated silica gel 600 F254 plates (Art. 5715, Merck, Darmstadt, Germany) or reverse phase C18 F254 plates (Art. 15389, Merck, Darmstadt, Germany). Flash column chromatography was carried out on silica gel by a medium-pressure gradient system equipped with a Pump Module C-605 and a Pump Manager C-615 (BÜCHI, Flawil, Switzerland). High resolution ESIMS were recorded on a Mariner Biospectrometry Workstation (Applied Biosystems of Thermo Fisher Scientific, Waltham, MA, USA) equipped with an electrospray ion source in the positive mode. High performance liquid chromatography (HPLC) was performed on a high pressure gradient system that is composed of pumps PU-2087, a degasser DG-2080-53, a mixer MX-2080-32, and a detector UV-2075 (JASCO, Tokyo, Japan). FT-IR spectra were recorded on an FT/IR-400 spectrometer instrument (JASCO). Specific rotations were observed on a DIP-370 polarimeter (JASCO). NMR spectra were investigated on an Avance ARX400 (400 MHz for 1H) or Avance III HD 600 MHz Cryo-probe spectrometer (600 MHz for 1H) (Bruker Bio Spin, Yokohama, Japan). The chemical shifts (ppm) were referenced to the solvent residual peak at δH 7.26 ppm (CDCl3), δH 3.30/δC 49.0 ppm (CD3OD), or δH 2.06/δC 29.9 ppm (acetone-d6).
+ Open protocol
+ Expand
2

Polymer MW Determination via SEC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Size Exclusion Chromatography (SEC) has been used to determinate the molecular weights of polymer samples. A SEC system from JASCO was used including a Degasser DG-2080-53, a pump PU-980, an autosampler AS-2051 Plus, a column oven, a UV-detector UV-975, and a refractive index detector RI-2031 Plus. NaNO3 (0.1 M) containing NaN3 was used as an eluent at 30 °C. A SUPREMA columns guard/1000/30 Å from Polymer Standards Service was used. Pullulan standards were used for calibration.
+ Open protocol
+ Expand
3

HPLC-FD Anion-Exchange Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC-FD system consists of a Model Ternary Gradient Unit LG2080-02, a Model 3-Line Degasser DG-2080-53, and a Model FB-1520S fluorescence detector (excitation wavelength, 280 nm; emission wavelength, 340 nm) together with a Model Intelligent HPLC Pump PU-2080-Plus (JASCO, Tokyo, Japan). A Shodex Asahipak ES-502N 7C column (7.5 mm ID × 100 mm, Showa Denko Co., Tokyo, Japan) was used for the anion-exchange HPLC with an ethanol gradient on 0.05 M sodium acetate, 0.40 M Na2SO4 (acetate-sulfate buffer), pH 4.85, at the flow-rate of 1.00 mL/min15 (link),19 (link). The following ethanol concentrations were used for the analysis: 0–1 min, 0%; 1–50 min, linear increase from 0 to 10%; 50–55 min, linear decrease from 10 to 0%; 55–60 min, 0%. All solvents were filtered through a filter unit (0.22 μm, pore size, TPP) and the samples were filtered through filter units (Millex-GV 0.22 μm Filter Unit, Millipore, Bedford, MA, USA) before use.
+ Open protocol
+ Expand
4

HPLC Analysis of Pharmaceutical Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC equipment of ChromNAV Chromatography Data System, with an autosampler (AS-2055plus), pump (PU-2080plus), ultraviolet-visible detector (UV-2075plus), column oven (CO-2060plus), and solvent deaerator (DG-2080-53), manufactured by Jasco Corporation (Tokyo, Japan), was used. The mobile phase comprised a phosphate buffer at pH 5.9: acetonitrile = 65:35, and the stationary phase was an Inertsil® ODS-3 (4.6 × 150 mm, 5 µm, GL Sciences Co., Ltd., Tokyo, Japan). The measurement conditions were 40 °C for the column oven, 1 mL/min for the flow rate, 20 μL for the injection volume, and 254 nm for the measurement wavelength. Phosphate buffer pH 5.9 was prepared by dissolving 6.8 g of potassium dihydrogen phosphate in 800 mL of water, adding diluted potassium hydroxide (0.1 g/mL) to adjust the pH to 5.9, and then adding water to 1000 mL.
+ Open protocol
+ Expand
5

HPLC Analysis of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC instrumentation and analytical conditions An HPLC system from Jasco (Como, Italy) was used, consisting of a ternary gradient system (PU 980), in line degasser (DG-2080-53) , autosampler (AS-2055) and an ultraviolet, multiple wavelength detector (MD-1510). The chromatographic separation assay was performed with a Gemini C18 analytical column (250 × 4.6 mm inner diameter, 5 μm particle size; Phenomenex, Torrance, CA, USA) maintained at 30°C using a Peltier system (CO-4062; Jasco). The mobile phase consisted of acetonitrile:water (90:10% v:v) at a flow rate of 1 mL/min and the optimal wavelength for the quantification was set at 242 nm.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!