The largest database of trusted experimental protocols

Etc 273t water cooled peltier thermostated cell holder

Manufactured by Jasco

The ETC-273T is a water-cooled Peltier thermostated cell holder. It is designed to maintain a stable temperature for samples during analysis.

Automatically generated - may contain errors

3 protocols using etc 273t water cooled peltier thermostated cell holder

1

Thermoresponsivity Test of Compound 8

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 27

A thermoresponsivity test of compound 8 in 150 mM KCl was performed by the following procedure. A JASCO FP-6500 spectrofluorometer was used, and using ultrapure water obtained from a Milli-Q reagent system made in Millipore Corporation, KCl purchased from Wako Pure Chemical Industries, Ltd. dissolved to make a 150 mM concentration was used as a solvent. The initial concentration of the compound in this experiment was 0.01 w/v %, and the excitation wavelength was 456 nm. A JASCO ETC-273T water-cooled Peltier thermostated cell holder was used for temperature control of the solution, and temperature was measured with its accompanying thermocouple. Solution temperature was raised in steps of 2° C., and the fluorescence intensity at each temperature was measured.

The test results are shown in FIG. 19. Even when VBTMA was used as a cationic unit, change in fluorescence intensity in response to temperature increase in a 150 mM potassium chloride solution could be confirmed. With regard to temperature responsivity, it is desirable that acrylamide, methacrylamide, or acrylic acid is contained in all units, however, it was revealed that if the mole fraction in a copolymer is about 10% as in this example, the probe functions as a probe that well responds to temperature even when a usual vinyl-type monomer such as styrene group is comprised in a unit.

+ Open protocol
+ Expand
2

Thermoresponsive Fluorescence Probe Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 27

A thermoresponsivity test of compound 8 in 150 mM KCl was performed by the following procedure. A JASCO FP-6500 spectrofluorometer was used, and using ultrapure water obtained from a Milli-Q reagent system made in Millipore Corporation, KCl purchased from Wako Pure Chemical Industries, Ltd. dissolved to make a 150 mM concentration was used as a solvent. The initial concentration of the compound in this experiment was 0.01 w/v %, and the excitation wavelength was 456 nm. A JASCO ETC-273T water-cooled Peltier thermostated cell holder was used for temperature control of the solution, and temperature was measured with its accompanying thermocouple. Solution temperature was raised in steps of 2° C., and the fluorescence intensity at each temperature was measured.

The test results are shown in FIG. 19. Even when VBTMA was used as a cationic unit, change in fluorescence intensity in response to temperature increase in a 150 mM potassium chloride solution could be confirmed. With regard to temperature responsivity, it is desirable that acrylamide, methacrylamide, or acrylic acid is contained in all units, however, it was revealed that if the mole fraction in a copolymer is about 10% as in this example, the probe functions as a probe that well responds to temperature even when a usual vinyl-type monomer such as styrene group is comprised in a unit.

+ Open protocol
+ Expand
3

Thermoresponsive Compound 8 in KCl

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 27

A thermoresponsivity test of compound 8 in 150 mM KCl was performed by the following procedure. A JASCO FP-6500 spectrofluorimeter was used, and using ultrapure water obtained from a Milli-Q reagent system made in Millipore Corporation, KCl purchased from Wako Pure Chemical Industries, Ltd. dissolved to make a 150 mM concentration was used as a solvent. The initial concentration of the compound in this experiment was 0.01 w/v %, and the excitation wavelength was 456 nm. A JASCO ETC-273T water-cooled Peltier thermostated cell holder was used for temperature control of the solution, and temperature was measured with its accompanying thermocouple. Solution temperature was raised in steps of 2° C., and the fluorescence intensity at each temperature was measured.

The test results are shown in FIG. 19. Even when VBTMA was used as a cationic unit, change in fluorescence intensity in response to temperature increase in a 150 mM potassium chloride solution could be confirmed. With regard to temperature responsivity, it is desirable that acrylamide, methacrylamide, or acrylic acid is contained in all units, however, it was revealed that if the mole fraction in a copolymer is about 10% as in this example, the probe functions as a probe that well responds to temperature even when a usual vinyl-type monomer such as styrene group is comprised in a unit.

+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!