The largest database of trusted experimental protocols

Exactive mass spectrometer

Manufactured by Thermo Fisher Scientific
Sourced in United States, Germany

The Exactive mass spectrometer is a high-resolution, accurate-mass (HRAM) instrument designed for applications that require sensitive and selective analysis. It utilizes Orbitrap technology to achieve accurate mass measurements, providing reliable qualitative and quantitative data.

Automatically generated - may contain errors

79 protocols using exactive mass spectrometer

1

Stability Assessment of SSRI Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
In order to assess the stability of the SSRIs over the 5-day exposure period chosen for the static exposure water tank renewal period, a nominal concentration of 1.5 ng/mL of each SSRI was monitored over the 5-day period (n = 3 tanks per SSRI). The 20 mL samples were removed from 3 L tanks daily and the SSRIs were extracted using Oasis HLB solid phase extraction columns (Waters Canada, Mississauga, ON, Canada) following the manufacturer’s instructions. SSRIs were eluted from the columns with 100% methanol and eluates were analyzed on an ACQUITY UPLC BEHC18 column (2.1 × 100 mm, 1.7 µm) using an UltiMateTM 3000 LC pump coupled to an Exactive™ mass spectrometer (Thermo Fisher Scientific, Waltham, MA, USA). The limit of quantitation (LOQ) was found to be 0.1 ng/mL. Mixtures were not tested for SSRI stability since loss (which is either due to degradation, or adsorption to tank material, or both) is assumed to be independent of the presence of other SSRIs. It was found that all three of the compounds were stable over the exposure period with the maximum reduction of 20% found for paroxetine at day 5 (Supplemental Scheme S1).
+ Open protocol
+ Expand
2

Spectroscopic Characterization of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The IR spectra were obtained with a Nicolet iS10 spectrophotometer (Thermo Scientific). 1H NMR spectra were measured on a Bruker AVANCE III HD 600 MHz spectrometer. The ESI-MS spectra were measured on an Exactive mass spectrometer (Thermo Scientific). Elemental analyses (C, H, N) were performed on a Perkin-Elmer 2400 series II analyzer. DNA concentration was measured by a NanoDrop 8000 spectrophotometer (Thermo Scientific).
+ Open protocol
+ Expand
3

Lipid Extract Analysis by ESI-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Dried lipid extracts (as above) were analysed by the University of St Andrews Mass Spectrometry service ESI-MS upon a Thermo-scientific Exactive mass spectrometer tuned to an accuracy of > 1 ppm. Survey scans were acquired in both the positive and negative ion modes over a range of 80-1600 m/z (or a portion thereof) with a cone voltage of 3.7kV. Each spectrum consists of a minimum of 20 consecutive scans.
+ Open protocol
+ Expand
4

Lipid Profiling by UPLC-FT-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
For lipid extraction, the lipophilic phase (see extraction protocol above) was collected and vacuum-dried. Samples were processed using ultra-performance liquid chromatography coupled with Fourier transform mass spectrometry (UPLC-FT-MS, Hummel et al., 2011 (link)), on a C8 reverse phase column coupled with an Exactive mass spectrometer (Thermo-Fisher)2 in positive and negative ionization mode. Processing of chromatograms, peak detection and integration were performed using REFINER MS® 6.0 (GeneData)3. Processing of mass spectrometry data included the removal of the fragmentation information, isotopic peaks, as well as chemical noise. Obtained features (m/z at a certain retention time) were queried against an in-house lipid database for further annotation.
+ Open protocol
+ Expand
5

Phytohormone Extraction and Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Phytohormones were extracted and analysed according to [41 (link)]. Briefly, ~100 mg of frozen tissue from the same batches used for the RNA-Seq experiment were extracted twice with 1 ml of methanol/water 80 %, centrifuged at 20,000 g for 15 min. at 4 °C, the supernatant was passed through a C18 cartridge, and the samples were collected in a 5-ml tube for speed-Vac evaporation to dryness. The residue was resuspended in 1 ml methanol/water 20 %. Ten μl of filtrated extract were injected in a U-HPLC-MS system consisting of an Accela Series U-HPLC (ThermoFisher Scientific, USA) coupled to an Exactive mass spectrometer (ThermoFisher Scientific) using a heated electrospray ionization interface. Mass spectra were obtained using the Xcalibur software version 2.2 (ThermoFisher Scientific). For quantification of the plant hormones, calibration curves were constructed for each analysed component (1, 10, 50, and 100 μg l−1) and corrected for 10 μg l−1 deuterated internal standards. Recovery percentages ranged between 92 and 95 %.
+ Open protocol
+ Expand
6

Characterization of Chemical Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Optical rotations were measured on a Jasco P-1030 polarimeter (JASCO, Tokyo, Japan). Infrared spectra were measured on a Jasco FT/IR 4100 (JASCO, Japan). NMR spectra were recorded on a JEOL ECX 500 (500 MHz) or a Bruker DRX (500 MHz) spectrometer (Bruker, Billerica, MA, USA). Chemical shifts are denoted in δ (ppm) relative to residual solvent peaks as internal standard (CDCl3, 1H δ 7.24, 13C δ 77.0). ESI–MS spectra were recorded on a Thermo Scientific Exactive mass spectrometer (Thermo Fisher Scientific, Waltham, MA, USA) or a SHIMADZU LCMS-2020 spectrometer (Shimadzu, Kyoto, Japan). High performance liquid chromatography (HPLC) experiments were performed with a SHIMADZU HPLC system equipped with a LC-20AD intelligent pump. LC–MS experiments were performed with amaZon SL (Bruker Daltonics, Bremen, Germany). Cell density for cytotoxic and anti-microbial assay was recorded on Tecan infinite® M200 plate reader (Tecan, Salzburg, Austria) at Drug Discovery Scientific Research and Education Center-Open Lab. facility, Faculty of Pharmaceutical Sciences, Hokkaido University. All reagents were used as supplied unless otherwise stated.
+ Open protocol
+ Expand
7

Quantitative HILIC-LCMS Metabolomics

Check if the same lab product or an alternative is used in the 5 most similar protocols
Prepared samples were analysed on a LCMS platform consisting of an Accela 600 LC system and an Exactive mass spectrometer (Thermo Scientific). A Sequant ZIC-HILIC column (4.6mm x 150mm, 3.5μm) (Merck) was used to separate the metabolites with the mobile phase mixed by A=0.1% (v/v) formic acid in water and B=0.1% (v/v) formic acid in acetonitrile. A gradient program starting at 20% of A and linearly increasing to 80% at 30 min was used followed by washing (92% of A for 5 mins) and re-equilibration (20% of A for 10min) steps. The total run time of the method was 45 min. The LC stream was desolvated and ionised in the HESI probe. The Exactive mass spectrometer was operated in full scan mode over a mass range of 70–1,200 m/z at a resolution of 50,000 with polarity switching. The LCMS raw data was converted into mzML files by using ProteoWizard and imported to MZMine 2.10 for peak extraction and sample alignment. An in-house database including all possible 13C and 15N isotopic m/z values of the relevant metabolites was used for the assignment of LCMS signals. Finally the peak areas were used for comparative quantification.
+ Open protocol
+ Expand
8

AMT Tag Database Generation for LC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS/MS analysis of the pooled fractions described above was used to generate a reference database of peptide markers defined by accurate masses and elution times, i.e., AMT tag (described in the paragraph below and refs. [46] (link)–[48] (link)). The AMT tag database then served as a comprehensive “look up” table for the subsequent higher throughput LC-MS analyses described below. For the AMT tag database generation, each of the 24 fractions corresponding to the purine-replete and purine-starved cells were analyzed using a 4-column, custom-built, capillary LC system coupled online via an in-house manufactured electrospray ionization (ESI) interface to an LTQ-Orbitrap mass spectrometer (Thermo Scientific, San Jose, CA) [134] (link). To identify the eluting peptides, the LTQ-Orbitrap instrument was operated in a data-dependent MS/MS mode for the top six abundant precursor ions in each full MS scan. Following the AMT tag database generation, LC-MS analyses with full MS scan (400–2,000 m/z range) were performed on the 45 unfractionated purine-replete and purine-starved peptide samples described above to generate quantitative data. For this, samples were analyzed in random order using the same chromatographic and electrospray conditions as described for the LC-MS/MS analyses except that the LC system was interfaced to an Exactive Mass Spectrometer (Thermo Scientific).
+ Open protocol
+ Expand
9

Isotope Tracing of Cellular Metabolism

Check if the same lab product or an alternative is used in the 5 most similar protocols
For isotope tracing, cells were cultured in a 3 mL M9 medium supplied with the labeled/unlabeled carbon sources described in the main text. For 13CO2 labeling, the experiment was performed in a 10 L desiccator that was first purged twice of the contained ambient air with a vacuum pump and refilled with an atmosphere of 80% air and 20% 13CO2 (Cambridge Isotope Laboratories Inc., MA USA). All cultures were inoculated to an OD600 0.02 and grown at 37 °C until the stationary phase. Then, ~109 cells (1 mL of culture with OD600 = 1) were pelleted, washed once with ddH2O, and hydrolyzed in 1 mL hydrochloric acid (6 M) at 95 °C for a duration of 24 h. Subsequently, the acid was evaporated by heating at 95 °C and the hydrolyzed biomass was re-suspended in ddH2O.
Hydrolyzed amino acids were separated using ultra-performance liquid chromatography (Acquity, Waters, Milford, MA, USA) using a C18-reversed-phase column (Waters, Eschborn, Germany)90 (link). Mass spectra were acquired using an Exactive mass spectrometer (ThermoScientific, Dreieich, Germany). Data analysis was performed using Xcalibur (ThermoScientific, Dreieich, Germany). Prior to analysis, amino-acid standards (Merck, Darmstadt, Germany) were analyzed under the same conditions in order to determine typical retention times.
+ Open protocol
+ Expand
10

Metabolomic Analysis of PDT-Treated Cells

Check if the same lab product or an alternative is used in the 5 most similar protocols
SK-ChA-1 cells were seeded in 6-wells plates and cultured until confluence. Cells were treated using the PDT protocol as described in “PDT protocol” (n = 3 per group). After 90 min, the cells were washed with 1 mL cold PBS and the cells were lysed in 1 mL lysis buffer (40% acetonitrile, 40% methanol, 20% water). The cells were scraped and transferred to 2-mL centrifuge tubes that were shaken for 10 min at 4 °C. Next, the samples were centrifuged for 15 min at 20,000×g (4 °C), after which the supernatant was aspirated and stored at −80 °C. LC-MS analysis was performed on an Exactive mass spectrometer (Thermo Scientific) coupled to a Dionex Ultimate 3000 autosampler and pump (Thermo Scientific). The MS operated in polarity-switching mode with spray voltages of 4.5  and −3.5  kV. Metabolites were separated using a Sequant ZIC-pHILIC column [2.1 × 150 mm, 5 µm, guard column 2.1 × 20 mm, 5 µm (Merck)] using a linear gradient of acetonitrile and eluent A (20 mM (NH4)2CO3, 0.1% NH4OH in ULC/MS grade water [Biosolve, Valkenswaard, the Netherlands)]. The flow rate was set to 150 µL/min. Metabolites were identified and quantified using LCquan software (Thermo Scientific) on the basis of exact mass within 5 ppm and further validated in accordance with the retention times of standards. Peak intensities were normalized based on total ion count.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!