The largest database of trusted experimental protocols

Zetasizer 3000

Manufactured by Malvern Panalytical
Sourced in United Kingdom, United States, Australia, Germany

The Zetasizer 3000 is a laboratory instrument designed to measure the size, size distribution, and zeta potential of particles and molecules in a liquid. It uses the technique of dynamic light scattering to determine particle size, and electrophoretic light scattering to measure zeta potential. The Zetasizer 3000 can analyze a wide range of sample types, including proteins, polymers, emulsions, and suspensions.

Automatically generated - may contain errors

74 protocols using zetasizer 3000

1

Zeta Potential and Size Analysis of IFNβ-1a

Check if the same lab product or an alternative is used in the 5 most similar protocols
The average zeta potential of the IFNβ-1a ReciGen and IFNβ-1a Rebif was evaluated without preliminary sample preparation (0.088 mg/mL of IFNβ-1a in acetate buffer pH 4.2) using a zeta potentiometer (Zetasizer 3000, Malvern Instruments, Warriewood, Australia). The following setting used 5 scans for each run of the sample and 3 runs were collected. The experiment was repeated at least 6 times. DLS analysis was performed using Zetasizer 3000 (Malvern Instruments, Warriewood, Australia). The size distribution was determined applying 10 scans for each run of the spectrogram and at least 10 runs were recorded.
+ Open protocol
+ Expand
2

Nanoparticle Characterization by DLS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Photon correlation spectroscopy (Zetasizer 3000; Malvern Instruments, Malvern UK) was used to measure the particle size and zeta potential of all drug-loaded nanoparticles [26 ]. Before analysis, all samples were diluted with appropriate distilled water. Z-average particle size, polydispersity index (PDI), and zeta potential were all measured in triplicates at room temperature.
+ Open protocol
+ Expand
3

Morphological Characterization of Cationic Liposomes

Check if the same lab product or an alternative is used in the 5 most similar protocols
Cationic liposomes were diluted and dropped onto copper gauze with a membrane for morphological characterization. The particle size was measured using a Zeta Sizer 3000 laser particle size analyzer and processed using Dynamic Light Scattering software (Malvern Instruments Ltd.). The average particle size and polydispersity index (PDI) were recorded. No aggregation or deposition were observed in the cationic liposomes on being maintained at RT for 2 months, and the stability was further examined using TEM. Briefly, a few drops of the liposome solution were dropped onto a TEM grid, dried and absorbed the extra solution with filter paper prior to analysis. TEM images were photographed using a field emission JEM-200CX TEM equipped with a charge-coupled device camera.
+ Open protocol
+ Expand
4

Nanoparticle Characterization in Physiological Buffers

Check if the same lab product or an alternative is used in the 5 most similar protocols
The particle size and zeta potential of each NP sample (0.1 mg/mL) was measured using a Zetasizer 3000 instrument (Malvern Instruments, Malvern, UK) after being stabilized in 150 mM PBS (pH 7.4, 6.5, or 6.0) at 25 °C for 4 h. The morphology of each NP sample was confirmed using a field-emission scanning electron microscopy (FE-SEM, Hitach S-400, Fukuoka, Japan) after being stabilized in 150 mM PBS (pH 7.4, 6.5, or 6.0) at 25 °C for 4 h [32 (link),33 (link),34 (link),35 (link),36 (link),37 (link)].
+ Open protocol
+ Expand
5

Peptide-Functionalized Nanodiamonds Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
The RRRRRRRCLPFFD peptide was dissolved in ultra pure Milli Q water at a final concentration of 0.05 mg/ml. The nanodiamonds are added to this peptide solution remaining at a concentration of 0.8 nM in the final solution, and then they are incubated with vigorous stirring for 2 h. The adsorption of the peptide on the nanocrystal surface was evaluated by the change in the Zeta potential (pZ) and hydrodynamic diameter (Dh) (Zeta sizer 3000, Malvern Instruments, UK). The colloidal suspension was centrifuged and washed three times. The washed fNDs were reevaluated by Zp and Dh to ensure that the functionalization remained. Finally, we analyzed the functionalization of the nanodiamond by High Resolution Transmission Electron Microscopy (HR-TEM) staining the samples with phosphotungstic acid (1%) in order to evaluate the presence of the peptide surrounding the nanodiamond.
+ Open protocol
+ Expand
6

Electrokinetic Potential of Chromium(III) Oxide

Check if the same lab product or an alternative is used in the 5 most similar protocols
Electrokinetic potential measurement started with the preparation of 500 ml chromium(III) oxide suspension by adding 0.03 g of solid to the solution containing 0.01 M supporting electrolyte (NaCl) and appropriate EPS amount giving its final concentration of 0.1, 1, 10 or 50 ppm. After 3-min sonication suspensions were poured into five Erlenmeyer flasks and their pH values were adjusted to 3, 4.6, 6, 7.6 and 9 ± 0.05. Then the zeta potential was measured using Zetasizer 3000 (Malvern Instruments); the apparatus was washed twice before each measurement. One result was the average of five measurements. The measurement error did not exceed 5 %.
In the same way, the zeta potential of chromium(III) oxide particles in the supporting electrolyte solution (0.01 M NaCl) without the bacterial polymer was measured.
+ Open protocol
+ Expand
7

Characterization of AuNPs Coated with BSA-r8

Check if the same lab product or an alternative is used in the 5 most similar protocols
AuNRs and AuNPrs were characterized before and after BSA-r8 coating by UV-Vis-NIR absorption spectra, using a Lambda 25 spectrophotometer (Perkin Elmer, Waltham, MA, USA). Dynamic light scattering (DLS) and Z potential measurements were acquired in PBS pH = 7, at 25 °C, using a Zetasizer 3000 (Malvern Instruments, Malvern, UK), as triplicates in aqueous solution at 25 °C.
TEM images of AuNRs were acquired with a JEOL JEM-1010 microscope (JEOL USA, Peabody, MA, USA), using Formvar carbon-coated copper microgrids (200 mesh; Ted Pella, Redding, CA, USA). For AuNPrs, TEM images were obtained using a Philips CM 120 transmission electron microscope with an accelerating voltage of 120 kV and a 300 mesh Formvar/Carbon-Coated Copper grid. For both AuNPs, liquid suspensions were deposited on the microgrid and allowed to stand overnight before TEM image acquisition.
+ Open protocol
+ Expand
8

Nanogel Hydrodynamic Diameter and Size

Check if the same lab product or an alternative is used in the 5 most similar protocols
The hydrodynamic diameter (HD) and size distribution of the nanogels at a concentration of 10 mg/mL were determined in a Zetasizer 3000 (Malvern Instruments, Worcestershire, UK) with a scattering angle of 173°.
+ Open protocol
+ Expand
9

Zeta-potential analysis of PECNP

Check if the same lab product or an alternative is used in the 5 most similar protocols
The Zeta-potential measurements on PECNP dispersions in order to determine the net charge of PECNP were performed at neutral pH values (pH = 7.4) using a Zetasizer 3000 (Malvern Instruments Ltd., Worchestershire, UK).
+ Open protocol
+ Expand
10

Characterization of FA-DDP/PTX Nanoparticles

Check if the same lab product or an alternative is used in the 5 most similar protocols
Particle morphology of FA-DDP/PTX NLCs was observed by a Transmission Electron Microscope (TEM, Hitachi, Tokyo, Japan). Diluted FA-DDP/PTX NLCs were placed on a carbon-coated copper grid, negatively stained with 2% phosphotungstic acid, and then observed with TEM.
The mean particle size, polydispersity index (PDI), and zeta potential were analyzed by photon correlation spectroscopy (PCS), with a Zetasizer 3000 (Malvern Instruments, Malvern, England). The average particle size was expressed as volume mean diameter (Wang et al., 2012 ).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!