The largest database of trusted experimental protocols

Trace 1310 gc

Manufactured by Thermo Fisher Scientific
Sourced in United States, Italy, Germany, France

The Thermo Scientific Trace 1310 GC is a gas chromatography system designed for a variety of analytical applications. It features a compact and modular design, offering reliable performance and flexibility for diverse sample types and laboratory requirements.

Automatically generated - may contain errors

61 protocols using trace 1310 gc

1

GC-MS/FID Analysis of Essential Oil

Check if the same lab product or an alternative is used in the 5 most similar protocols
The extract was dried over anhydrous CaCl2, and the essential oil was stored at 4 °C until the GC-MS/FID measurements. The yield of oil was 1.52%. The GC-MS analysis was carried out using a Thermo Scientific Trace GC 1310 connected to a Thermo TSQ 9610 MS system on a DB-5 capillary column (60 m × 0.25 mm, 0.25 mm film thickness) with helium as the carrier gas (0.8 mL/min). GC oven temperature was maintained at 80 °C for 10 min, programmed to 280 °C at a rate of 4 °C/min, and kept constant at 280 °C for 5 min. The split ratio was adjusted to 1:20. The injector temperature was set to 250 °C. The mass spectra were recorded at 70 eV. The mass range was m/z 35–650. GC-FID analysis was performed using a Thermo Scientific Trace GC 1310 instrument. The FID detector temperature was 280 °C. To obtain the same elution order as GC-MS, simultaneous auto-injection was performed in duplicate on the same column under the same operational conditions. The relative percentage of the separated compounds was calculated from the FID chromatograms [53 ,54 (link),55 (link)]. Alkanes were used as reference points in the calculation of Kovats Indices (KI). Compounds were identified by comparing their retention times and mass spectra with those obtained from authentic samples and/or the NIST and Wiley spectra, as well as literature data [54 (link),55 (link),56 ,57 (link)].
+ Open protocol
+ Expand
2

GC-MS/FID Analysis of Essential Oil

Check if the same lab product or an alternative is used in the 5 most similar protocols
The essential oil was kept at 4 °C until the GC-MS/FID computations, and the extract was dried over anhydrous CaCl2. The oil yield was 1.52%. On a DB-5 capillary column (60 m × 0.25 mm, 0.25 mm film thickness) with helium as the carrier gas (0.8 mL/min), GC-MS analysis was performed using a Thermo Scientific Trace GC 1310 linked to a Thermo TSQ 9610 MS system. After being held at 80 °C for 10 min, the GC oven’s temperature was scheduled to rise to 280 °C at a rate of 4 °C per minute, and it remained at 280 °C for 5 min. It was changed to a 1:20 split ratio. The injector was adjusted at a temperature of 250 °C. At a mass of 70 eV, the spectra were captured. The m/z 35–650 was the mass range. A Thermo Scientific Trace GC 1310 equipment was used to conduct the GC-FID analysis. The FID detector was held at a temperature of 280 °C. Parallel auto-injection was carried out twice on the same column with the same operating settings in order to achieve the same elution sequence as GC-MS. From the FID chromatograms, the relative proportion of the isolated chemicals was determined [29 (link)]. When calculating the Kovats Indices (KIs), alkanes were employed as reference points. Authentic samples, the NIST and Wiley spectra, as well as information from the literature, were used to compare retention durations and mass spectra, which helped identify the compounds [30 (link),31 ,32 (link)].
+ Open protocol
+ Expand
3

Lipid Extraction and Fatty Acid Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Total lipids from cortex, plasma and livers were extracted using Folch’s procedure [41 (link)]. Boron trifluoride in methanol was used for transmethylation [42 (link)]. Hexane was used to extract fatty acid methyl esters (FAMEs) and dimethyl acetals (DMAs). Analyses were performed on a GC Trace 1310 (Thermo Scientific, Illkirch, France) gas chromatograph (GC) using a CPSIL-88 column (100 m × 0.25 mm inside diameter, film thickness 0.20 μm; Agilent, CA, USA). This device was coupled to a flame ionization detector (FID). The configuration was: inlet pressure of hydrogen 210 kPa, oven temperature 60 °C for 5 min + 165 °C at 15 °C per min and upholding for 1 min, +225 °C at 2 °C per min and upholding at 225 °C for 17 min. The injector and the detector were maintained at 250 °C. Comparisons with commercial and synthetic standards enabled the identification of FAMEs and DMAs. The ChromQuest 5.0 version 3.2.1 software (Thermo Scientific, Illkirch, France) was used to process the data.
+ Open protocol
+ Expand
4

Quantitative Fatty Acid Profiling

Check if the same lab product or an alternative is used in the 5 most similar protocols
Total lipids were transmethylated using boron trifluoride in methanol according to Morrison and Smith (Morrison and Smith, 1964 (link)). Fatty acid methyl esters (FAMEs) were subsequently extracted with hexane and analyzed on a GC Trace 1310 (Thermo Scientific, Les Ulis, France) gas chromatograph (Palo Alto, CA, USA) using a CPSIL-88 column (100 ×0.25 mm i.d., film thickness 0.20 μm; Varian, Les Ulis, France) equipped with a flame ionization detector. Hydrogen was used as carrier gas (inlet pressure 210 kPa). The oven temperature was held at 60°C for 5 min, increased to 165°C at 15°C/min and held for 1 min, and then to 225°C at 2°C/min and finally held at 225°C for 17 min. The injector and the detector were maintained at 250°C. FAMEs were identified by comparison with commercial and synthetic standards. The data were processed using the ChromQuest software (Thermo Scientific) and reported as a percentage relative to total fatty acids (considered as 100%).
+ Open protocol
+ Expand
5

Gas Flux Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Gas samples were analyzed by gas chromatography (Thermo Scientific, GC TRACE 1310), and the concentration of CO2 and CH4 were determined by a flame ionization detector (250 °C) and N2O by a Ni-electron capture detector (320 °C). Injector was setting to 250 °C and Porapak columns to 70 °C of temperature. The GC was calibrated with standard gases (White Martins, Praxair), and standard gases were read periodically (between every 20 samples), as quality control. The gas molar volume (Vm) was corrected for the headspace chamber air temperature (K) as measured at sampling time. And the gas fluxes (f) were calculated by each gas considering the change in gas concentration in the chamber during the incubation time (ΔCt), the chamber volume (V), the soil area covered by the chamber (A), and the molecular weight of the gas (m), by the equation: f = ΔCt × V/A × m/Vm [36 (link)–38 ].
+ Open protocol
+ Expand
6

Fatty Acid Profiling by GC-FID

Check if the same lab product or an alternative is used in the 5 most similar protocols
Total lipids were extracted as described above88 (link). Boron trifluoride in methanol was used for transmethylation90 (link). Hexane was used to extract fatty methyl esters (FAMEs) and dimethyl acetals (DMAs). Analyses were performed on a GC Trace 1310 (Thermo Scientific) gas chromatograph (GC) using a CPSIL-88 column (100 × 0.25 mm i.d., film thickness 0.20 μm; Varian). This device was coupled to a flame ionization detector (FID). The configuration was: inlet pressure of hydrogen 210 kPa, oven temperature 60 °C for 5 min + 165 °C at 15 °C per min and upholding for 1 min, + 225 °C at 2 °C per min and upholding at 225 °C for 17 min. The injector and the detector were maintained at 250 °C. Comparisons with commercial and synthetic standards enabled the identification of FAMEs and DMAs. The ChromQuest software (Thermo Scientific) was used to process the data.
+ Open protocol
+ Expand
7

GC-MS Analysis of Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC–MS data was recorded with Single Quadrupole Mass Spectrometer and GC (Trace 1,310; Thermo Fisher Scientific, United States). The column was 30 m long dimethyl (95%)/diphenyl polysiloxane (5%) RTX-5MS with 0.25 mm ID, and guard column (10 m; Restek). The initial temperature of GC oven was 60°C for 60 s and the temperature increased to 325°C (@10°C/min). Metabolites were identified using NIST library with Xcalibur software.
+ Open protocol
+ Expand
8

Determination of Cyclic Compounds by GC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Reactants were extracted from samples (1 mL) collected from the outlet with ice‐cold diethyl ether (500 μL, containing 0.2 mM decane as internal standard) by vigorous mixing for 2 min followed by centrifugation (17,000 g, 2 min, 4°C). The ether phase was separated and dried over anhydrous Na2SO4. Cyclohexanol, cyclohexanone, and ε‐caprolactone were quantified using a GC Trace 1310 (Thermo Fisher Scientific) equipped with a TG‐5MS capillary column (5% diphenyl/95% dimethyl polysiloxane, 30 m, i.d., 0.25 mm, film thickness: 0.25 μm, Thermo Fisher Scientific) equipped with a TG‐5MS capillary column (5% diphenyl/95% dimethyl polysiloxane, 30 m, I.D., 0.25 mm, film thickness: 0.25 μm, Thermo Fisher Scientific) and a flame ionization detector operated at 320°C, 350 mL min−1 air, 30 mL min−1 makeup gas (N2), and 35 mL min−1 hydrogen gas flow rates. N2 was applied as carrier gas at a constant flow of 1.5 mL min−1. An injection volume of 1 μL sample was applied onto the column using a PTV injector (40°C), programmed with a temperature gradient of 2°C s−1 from 40 to 250°C (1 min hold at 250°C), followed by a cleaning phase of 5 min at 450°C. A split ratio of 7 was applied (split‐flow 11 mL min–1). The oven temperature profile was: 40°C for 1 min, 40–80°C at 10°C min−1, 80–320°C at 100°C min−1, and 320°C for 3 min.
+ Open protocol
+ Expand
9

GC-FID Analysis of Ether Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
Ether samples (1 μL) were injected by a PTV injector, programmed with a temperature gradient of 10°C s–1 from 90–300°C into a GC Trace 1310 (Thermo Fisher Scientific) equipped with a TG-5MS capillary column (5% diphenyl/95% dimethyl polysiloxane, 30 m, i.d. 0.25 mm, film thickness: 0.25 μm, Thermo Fisher Scientific). A split ratio of 11 was applied, and the oven temperature profile was set to 40°C for 1 min, 40–80°C at 10°C min–1, 80–320°C at 100°C min–1, and 320°C for 10 min, N2 was applied as carrier gas (flow rate: 1.5 mL min–1). The flame ionization detector was operated at 320°C, 350 mL min–1 air, 30 mL min–1 makeup gas, and 35 mL min–1 hydrogen gas flow.
+ Open protocol
+ Expand
10

Quantifying Gut Short-Chain Fatty Acids

Check if the same lab product or an alternative is used in the 5 most similar protocols
The concentration of short-chain fatty acids (SCFAs) in the colons of mice from each intervention group was measured using gas chromatography-mass spectrometry (GC-Trace 1310, MS-ISQ 7000; Thermo Fisher Scientific, Waltham, MA, USA). Mouse fecal samples weighing 50 mg were homogenized in 1 mL of 6% phosphoric acid solution, with 4-methylvaleric acid serving as an internal standard. The column temperature was programmed to start at 90 °C, increase to 120 °C at a rate of 10 °C/min, then to 150 °C at 5 °C/min, and finally to 250 °C at 25 °C/min, where it was held for 2 min. The ion source and MS transfer line temperature were 300 °C and 250 °C, respectively, with an injection volume of 1 μL. Helium was used as the carrier gas with a follow rate of 1 mL/min and a split ratio of 10:1.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!