The largest database of trusted experimental protocols

4 protocols using torus diol column

1

SFC-MS/MS Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Sample analysis was performed by using a supercritical fluid chromatography (SFC, Acquity UPC 2 system) coupled to a tandem mass spectrometry (TQXS) both from Waters (Milford, MA, USA). The analytical method was based on Schulze et al. (2020) (link). The compounds were separated using a Waters Acquity UPC 2 BEH (3.0 × 100 mm, 1.7 µm, 130 Å) or Waters Torus Diol column (3.0 × 150 mm, 1.7 µm, 130 Å) which were operated at 55 °C. Table S4 lists the compounds of interest analyzed with which column. Both methods run with an eluent flow rate of 1.5 mL min -1 and a make-up flow rate of 0.3 mL min -1 . The gradients for both methods are listed in Table S5. The method using the Waters Acquity UPC 2 BEH column runs for 17.2 min and with a mobile phase of eluent A (CO 2 ) and eluent B (95% methanol, 5% Milli-Q water, 10 mM ammonium formate). The second method (Waters Torus Diol column) runs for 9 min and with a mobile phase of eluent A (CO 2 ) and eluent B (90% methanol, 10% Milli-Q water, 0.05% of a 25% ammonium hydroxide solution).
+ Open protocol
+ Expand
2

UHPLC and UHPSFC/HRMS Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
The UHPLC experiments were performed on a Waters H-Class UHPLC system using a modified method described by Krithika et al. [37 (link)]. The UHPLC system was equipped with a Waters BEH C18 column (100 × 2.1 mm, 1.7 µm) and run at 50 °C. Detection was performed using a PDA detector at 280 nm, with the baseline-corrected by 350–400 nm. The injection volume varied between 3 and 5 µL, and the flow rate was 0.6 mL/min. The mobile phase was composed of: (A) Milli-Q water with 0.5% acetic acid or 10 mM formic acid, and (B) acetonitrile with 0.5% acetic acid or 10 mM formic acid.
All UHPSFC/HRMS experiments were performed using a Waters Ultra Performance Convergence Chromatography System (Waters Milford, MA, USA) connected via a flow splitter (ACQUITY UPC2 splitter, Waters) to an LTQ Orbitrap Velos Pro mass spectrometer (Thermo Scientific, Waltham, MA, USA), which was equipped with a heated electrospray ionization source (HESI, Thermo Scientific). Chromatographic separation of the samples was performed using a Torus DIOL column (3 mm × 100 mm, 1.7 µm, Waters) protected with a Torus DIOL VanGuard pre-column (2.1 mm × 5 mm, 1.7 µm, Waters). The samples were centrifuged in a 5424R Eppendorf centrifuge (Eppendorf, Hamburg, Germany).
+ Open protocol
+ Expand
3

UPC2-HRMS Analysis of Complex Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
An Acquity UPC2 hyphenated to a G2‐Si Synapt HRMS from Waters (Milford, MA, USA) was used. The separation was performed on a Torus DIOL column (3.00×100mm, 1.7μm) with a VanGuard pre‐column (diol, 2.1×5.0mm, 1.7μm) from Waters. The choice of the column was based on literature [12 (link), 13 (link)]. The post‐column splitter interface from Waters was used to connect the UPC2 and MS, with a make‐up solvent flow via a quaternary solvent manager (Waters). Methanol as a modifier (B‐solvent), column temperature (40°C), automated backpressure (140bar), flow rate (1.5ml/min), and injection volume (3μl) were kept constant. Strong and weak washing solvents were methanol:water [4:1] and isopropanol, respectively. Both positive (ESI+) and negative (ESI) ionization modes were applied. If not stated otherwise, the MS parameters were set as follows: source temperature 120°C, desolvation temperature 600°C, desolvation gas flow 1000L/h, cone gas flow 100L/h, scan rate 0.2scans/s, scan type ‘centroid’, and mass range 50–1200Da. Lockspray calibration with leucine‐enkephalin (0.2ng/ml, Waters) was performed every 30s. The data was acquired with MassLynx; TargetLynx (v4.1; Waters) and R software (RStudio 2022.02) were used to process the data.
+ Open protocol
+ Expand
4

UHPSFC-MS Analysis of Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
All experiments were performed on a Waters Ultra Performance Convergence Chromatography System (Waters, Milford, MA, USA) equipped with a diode array detector (ACQUIY UPC2 PDA detector, Waters). The UHPSFC system was connected via a flow splitter (ACQUITY UPC2 splitter, Waters) to an LTQ Orbitrap Velos Pro mass spectrometer (Thermo Scientific, Waltham, MA, USA). The MS system was equipped with a heated electrospray ionization source (HESI; Thermo Scientific). A Torus DIOL column (3 mm×100 mm, 1.7 μm) equipped with a Torus DIOL VanGuard pre‐column (2.1 mm×5 mm, 1.7 μm) was purchased from Waters. A 5424R Eppendorf centrifuge (Eppendorf, Hamburg, Germany) was used for sample centrifugation.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!