The largest database of trusted experimental protocols

10 protocols using s 4800

1

Characterizing Ti Surface Morphology

Check if the same lab product or an alternative is used in the 5 most similar protocols
The surface morphology of Ti disks, contaminated Tallow-Ti disks, and Ti disks cleaned using the three methods (Finevo-Ti, UV-Ti, and Plasma-Ti) was examined by scanning electron microscopy (SEM; S-4800; Shimadzu, Kyoto, Japan) at a 5 kV accelerating voltage. The mean average surface roughness (Ra) and surface topography were assessed using a scanning probe microscope (SPM; SPM-9600; Shimadzu, Kyoto, Japan). The range of analysis was 125 μm × 125 μm. To compare the elemental composition of the Ti surface before and after cleaning using the three methods, the samples were analyzed using X-ray photoelectron spectroscopy (XPS; PHI X-tool; ULVAC-PHI, Kanagawa, Japan) equipped with a monochromatic X-ray source (Al Kα anode) operating at 15 kV and 13 W. The diameter of the analysis point was about 55 μm, and the angle between the electronic analyzer and the sample surface was 45 degrees.
+ Open protocol
+ Expand
2

Characterization of Dye-Sensitized Solar Cells

Check if the same lab product or an alternative is used in the 5 most similar protocols
The structure and morphology properties of the samples were measured by X-ray diffraction (XRD; XRD-6000, Shimadzu Corp., Kyoto, Japan) and scanning electron microscopy (SEM; S-4800, Ltd., Tokyo, Japan). The element distribution was tested by X-ray photoelectron spectroscopy (XPS) and electron diffraction spectroscopy (EDS). The photovoltaic performance of DSCs was characterized using a source meter (2400, Keithley Instruments, Inc., Beijing, China) under AM 1.5G irradiation (100 mW/cm2) generated by a solar simulator (XES-301S + EL-100, San-ei Electric Co., Ltd., Osaka, Japan). Electrochemical impedance spectroscopy (EIS) was carried out using the electrochemical workstation (CHI660D), performed on symmetric cells.
+ Open protocol
+ Expand
3

Characterization of EGCG Nanoparticles

Check if the same lab product or an alternative is used in the 5 most similar protocols
The zeta potentials of EGCG nanoparticles were determined using Zetasizer Nano-ZS90 (Malvern instruments, England) on the basis of Dynamic light scattering DLS techniques. FTIR spectra of EGCG nanoparticles were measured over 4000–400 cm−1 with a resolution of 4 cm−1 on a PerkinElmer Spectrum 100 spectrometer (PerkinElmer, USA). 1H NMR spectra of FA-NPS-PEG and FA-PEG-CS were measured in D2O containing 0.175% C2D4O2 using a 300 MHz spectrometer (Bruker Avance DPX 300, Switzerland)35 (link). The morphologies and particles sizes of EGCG nanoparticles were characterized through SEM (S-4800, Shimadzu, Tokyo, Japan), operating at an accelerating voltage of 30.0 kV.
In order to determine EEs of EGCG in EGCG nanoparticles, the suspensions of EGCG nanoparticles were centrifuged at 10 000 g for 30 min at 25 °C, and free EGCG in the supernatant was determined by HPLC (LC-20A, Shimadzu, Tokyo, Japan) according to the previous method59 (link). The EE of nanoparticles was calculated as equation (1).

+ Open protocol
+ Expand
4

Surface Topography and Wettability Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
TNS and plasma-TNS surface topography were evaluated by SEM (S-4800, Shimadzu, Kyoto, Japan) with 10 kV accelerating voltage. AFM (SPM-9600, Shimadzu Co., Tokyo, Japan) was utilized to assay the mean average surface roughness (Ra), mean peak-to-valley height (Rz), surface profiles and three-dimensional surface topography of the samples. The sample chemical composition was determined by XPS (Kratos Axis Ultra, Shimadzu, Japan). Sample surface wettability was evaluated using a contact angle measurement system (VSA 2500 XE; AST Products, Billerica, MA, USA).
+ Open protocol
+ Expand
5

Characterization of Surface Topography and Composition

Check if the same lab product or an alternative is used in the 5 most similar protocols
The surface topography of TNS samples was qualitatively evaluated by SEM (S-4800 Shimadzu, Kyoto, Japan) and SPM (SPM-9600; Shimadzu). The composition of the coating was analyzed by XPS using a Kratos Analytical Axis Ultra DLD electron spectrometer (Kratos Instruments, Manchester, UK) with a monochromatic Al Kα X-ray source. Each sample was etched with Ar ions for 2 min (evaporation rate of 5 nm/min) to remove surface contaminants. The surface phase properties were investigated by TF-XRD (RINT-2500; Rigaku, Tokyo, Japan). Spectra were recorded in the range of 2θ = 20°–50°, operating at 50 kV and 300 mA, using a Cu-Kα radiation source, scanning speed of 2°/min, and incident angle of 1°. Contact-angle measurements were performed using a video contact-angle measurement system (VSA 2500 XE; AST Products, Tokyo, Japan) at room temperature. Ultrapure water was used for measurements.
+ Open protocol
+ Expand
6

Comprehensive Material Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
The morphologies of the samples were investigated by SEM (S4800, Japan), TEM (JEM-2100F, Japan), AFM (Veeco dimension V, USA), XRD (D8 Advance, Germany), UV-vis diffuse reflectance spectra (Shimadzu UV2550, Japan), XPS (Thermo ESCALAB 250, USA), inductively coupled plasma-atomic (ICP) mass spectrometry (Atomscan Advantage, Germany), superconducting quantum interference device (Quantum Design MPMS–XL, USA), X-ray absorption near-edge structure (Beamline of BL12B of Nation al Syncrotron Radiation Laboratory, China), Theoretical calculations (National Super Computing Centre in Jinan, China).
+ Open protocol
+ Expand
7

Surface Characterization of Ti-QCM Sensors

Check if the same lab product or an alternative is used in the 5 most similar protocols
Scanning electron microscopy (SEM; S-4800, Shimadzu, Kyoto, Japan) and scanning probe microscopy (SPM; SPM-9600, Shimadzu, Kyoto, Japan) were used to analyze the surface topology and roughness of the Ti-QCM sensors over an area of interest of 2.0 × 2.0 μm2. X-ray photoelectron spectroscopy (XPS; ESCA 5600, Ulvac-Phi Inc., Kanagawa, Japan) was used to analyze the chemical composition of the surface of the Ti-QCM sensors. AlKα radiation (15 kV, 300 W) was used as the X-ray source. During XPS, Ar ion sputtering was used to determine the composition of the surface layer.
Contact angle measurements of the Ti-QCM sensors were performed using a video contact angle measurement system (SImage Entry 6; Excimer Inc., Kanagawa, Japan). The measurement was performed after 2.6 μL of distilled water was dropped immediately after the surface treatment on the Ti-QCM sensors. The contact angle was used as a simple measure of the surface energy and hydrophilic nature of the sensor surfaces.
+ Open protocol
+ Expand
8

Multimodal Characterization of Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
The samples were characterized by OM (Olympus BX51), SEM (Hitachi S-4800, 2 kV), XPS (Kratos Analytical AXIS-Ultra with monochromatic Al Kα X-ray), XRD (Shimadzu Thin Film, using Cu Kα radiation at room temperature in the 2θ range of 10 ~ 90°), Raman spectroscopy (Renishaw, Invia Reflex, excitation light of ~514 nm), TEM (JEOL JEM-2100F LaB6; acceleration voltage, 200 kV), and AFM (Dimension Icon, Bruker). The LEEM and μ-XPS elemental mapping data were acquired at the X-ray photoemission electron microscopy end station of the 09U (Dreamline) beamline of the Shanghai Synchrotron Radiation Facility. High resolution STEM-HAADF images were obtained on an aberration corrected transmission electron microscope JEM-ARM200F equipped with cold field emission gun with acceleration voltage of 200 kV.
+ Open protocol
+ Expand
9

Comprehensive Material Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
The samples were characterized by Raman spectroscopy (Horiba, LabRAM HR‐800, 532 nm laser excitation, 100× objective lens), SEM (Hitachi S‐4800, operating at 1 kV), XPS (Kratos Analytical Axis‐Ultra spectrometer using a monochromatic Al Kα X‐ray source), optical microscopy (Olympus DX51), Keithley 2001 multimeter (Tektronix, Inc.), and Fourier transform infrared (Nicolet 6700) spectrometer equipped with integrating sphere.
+ Open protocol
+ Expand
10

Characterization of C-PANI Electrode

Check if the same lab product or an alternative is used in the 5 most similar protocols
The samples were characterized by SEM (Hitachi S-4800, 5 kV), XPS (Kratos Analytical AXIS-Ultra with monochromatic Al Kα X-ray) and TEM (JEM-F200). C-PANI and PANI were polymerized directly onto the surface of carbon cloth as a working electrode. Samples for ex situ XPS measurements at different voltages were prepared in a nitrogen-filled glovebox, kept dry and removed for testing without the need for transport procedures. For in situ Raman testing, the Fe||C-PANI cell was assembled inside a nitrogen-filled glovebox using an in situ Raman spectroscopy electrochemical cell (C031-1, GaossUnion). All binding energy values of the XPS results were referenced to the C 1s peak of carbon at 284.8 eV. Raman spectroscopy (HORIBA iHR550, excitation light of ~633 nm) and FTIR spectra (Bruker Hyperion FTIR spectrometer). The spectral emittance of the devices was tested by an INVENIO-R (Bruker) FTIR spectrometer in the spectral range of 2.5–25 µm (equipped with an integrating sphere). The C-PANI for tTEM was scraped off the carbon cloth with a knife.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!