The largest database of trusted experimental protocols

12 protocols using kα spectrometer

1

Crystallographic and Electrochemical Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
The crystallographic data for BO/BWO and Au–BO/BWO were obtained using a Rigaku MicroMax 007 HF X-ray generator (Cu Kα, λ = 1.54 Å) operated at 40 kV and 30 mA. For morphological analyses, field emission scanning electron microscopy (FESEM, FEI Nova NanoSEM 450 with accelerating voltage 18 kV) was used. Detailed structural and morphological analyses were carried out using transmission electron microscopy (TEM, JEOL JEM F200). X-ray photoelectron spectroscopy was performed on a Thermo Kα+ spectrometer (Al Kα X-rays with energy ∼ 1486.6 eV). The electrochemical characteristics of the prepared electrodes, including CV, GCD, and impedance spectroscopy measurements, were obtained using a PG262A Potentiostat/Galvanostat (PGSTAT) (Techno Science Ltd. Bangalore, India).
+ Open protocol
+ Expand
2

XPS Analysis of Perovskite Quantum Dots

Check if the same lab product or an alternative is used in the 5 most similar protocols
Spectra from X-ray photoelectron spectroscopy (XPS) of perovskite quantum dots were captured with an AXIS SUPRA spectrometer from Kratos Analytical, part of the Shimadzu Group, located in Manchester, UK. This device features a Thermo Scientific Kα spectrometer and utilizes a monochromatic Al-Kα X-ray for illumination. Data collected were then transformed into VGD format and analyzed using Avantage software (599-31), release 5.9922. Calibration of the binding energy scale was aligned using the C 1 s peak, attributing a binding energy of 284.8 eV specifically to the carbon–carbon (C–C) bond.
+ Open protocol
+ Expand
3

Surface Chemistry Analysis of Carbon Fibers

Check if the same lab product or an alternative is used in the 5 most similar protocols
The surface chemistry of the carbon fibers was analyzed by X-ray photoelectron spectroscopy (XPS) using a Thermo Scientific Kα spectrometer with an Al Kα monochromatic source (1486.6 eV) and a multidetection analyser under a residual pressure of 5 × 10–8 mbar. Temperature programmed desorption (TPD) experiments were carried out in a TGA-DSC instrument (TA Instruments, SDT Q600 Simultaneous) coupled to a mass spectrometer (Thermostar, Balzers, GSD 300 T3) by heating the samples (approx. 4 mg) up to 950 °C (heating rate: 20 °C min–1) in a helium atmosphere (flow rate: 100 ml min–1). The thermobalance was purged for 2 hours in a helium atmosphere prior to the heating of the sample. The calibration of the equipment for H2O, CO and CO2 evolved gases was carried out using the decomposition of a calcium oxalate (99.999%, Sigma Aldrich) standard.
+ Open protocol
+ Expand
4

Comprehensive Materials Characterization Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Raman studies were carried out using a Renishaw 1000 spectrometer equipped with a He-Ne laser (λ = 633 nm) and coupled to a microscope with a 50× objective. The Raman system was calibrated using a silicon reference. Acquisition time was 10 s. Raman scattering spectra were recorded in the range of 150–2000 cm−1. X-ray diffraction (XRD) analysis was carried out using a Bruker AXS D8 (Lynxeye XE) diffractometer with a monochromated copper X-ray source (Kα1, λ = 1.54 Å) under a glancing incident angle (θ) of 1°. UV–Vis spectroscopy was performed using a Perkin Elmer Lambda 950 UV/Vis/NIR Spectrophotometer. A Labsphere reflectance standard was used as a reference. X-Ray photoelectron spectroscopy (XPS) was performed using a Thermo K-α spectrometer with monochromated aluminium Kα radiation, a dual beam charge compensation system and constant pass energy of 50 eV. Survey scans were collected in the range of 0–1200 eV. High-resolution peaks were used for the principal peaks of Ti 2p, O 2p and C 1s. The peaks were modelled using sensitivity factors to calculate the film composition. The area underneath these bands is an indication of the element concentration within the region of analysis (spot size: 400 µm).
+ Open protocol
+ Expand
5

Comprehensive Structural Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
Powder X-ray diffraction (XRD) was measured by Bruker D2 phaser diffractometer with Cu Kα radiation (λ = 1.541 Å, V = 30 KV and I = 10 mA) with a step of 0.2 °/s. Field-emission scanning electron microscopy (SEM, JEOL JEM-7100F) and high-resolution transmission electron microscopy (HRTEM, JEOL JEM-2100) were utilized to characterize the morphology of specimens. X-ray photoelectron spectroscopy (XPS) was measured by a Thermo Scientific Kα spectrometer with a monochromatic Al Kα source at 1486.6 eV, the binding energy was corrected based on the C 1s photoelectron peak at 284.6 eV. The energy dispersive X-ray spectroscopy (FEI QUANTA 400FEG) was measured.
+ Open protocol
+ Expand
6

Characterization of Polyurethane Polymer Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
The following samples were prepared for antimicrobial testing: control polyurethane polymer samples (solvent treated only), glutaraldehyde‐impregnated samples (biocidal swelling solution treated), and glutaraldehyde‐coated samples (biocidal solution treated only). The infrared absorbance spectra of the polyurethane polymer samples were measured using a Bruker Platinum ATR, within the range 4000–400 cm−1 with an accumulation of 15 scans per sample. X‐ray photoelectron spectroscopy (XPS) analysis of the solvent‐treated and glutaraldehyde‐impregnated polyurethane polymer samples was carried out using a Thermo Scientific, (United Kingdom) K‐α spectrometer to classify the different elements present as a function of polymer depth. All binding energies were calibrated to the C 1s peak at 284.5 eV.
+ Open protocol
+ Expand
7

Comprehensive Characterization of Nanomaterials

Check if the same lab product or an alternative is used in the 5 most similar protocols
SEM and TEM images were recorded using an
EVO-18 microscope (ZEISS, Oberkochen, Germany) and an FEI Tecnai-G2
F30 microscope (FEI Co., Hillsboro, OR) spectrometer, respectively.
XPS spectra were obtained using a K-α spectrometer (Thermo Fisher
Scientific Co., Waltham, MA). The FTIR spectra were collected with
a Spectrum 65 FTIR spectrophotometer (PerkinElmer Co., Ltd., Waltham,
MA). The UV–vis absorption spectra were obtained using a Shimadzu
UV-6100 UV–vis-NIR spectrophotometer (Shanghai Mapada Instruments
Co., Ltd., Shanghai, China). Red excitation light was provided by
a PEAC 200A system (Ada Hengsheng Technology Development Co., Ltd.,
Tianjin, China). The distance between the illumination source and
the sample cell was maintained at 10 cm. PEC measurements were performed
using an electrochemical workstation (CHI760e, Chenhua Instrument
Co., Ltd., Shanghai, China). ITO slices (≤6 Ω, South
China Xiangcheng Technology Co., Ltd., Shenzhen, China) with an active
surface area of 0.25 cm2 were used as the working electrode
vs Ag/AgCl as the reference electrode.
+ Open protocol
+ Expand
8

X-Ray Photoelectron Spectroscopy Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
XPS was performed
on a Thermo Fisher Scientific K-α+ spectrometer. Samples were
analyzed using a microfocused monochromatic Al X-ray source (72 W)
over an area of ∼400 μm. Data were recorded at pass energies
of 150 eV for survey scans and 40 eV for high-resolution scans with
1 and 0.1 eV step sizes, respectively. Charge neutralization of the
sample was achieved through a combination of both low-energy electrons
and argon ions. Three well-separated areas were selected on each sample
for analysis to examine any surface heterogeneity. Data analysis was
performed in CasaXPS using a Shirley-type background and Scofield
cross sections, with an energy dependence of −0.6.
+ Open protocol
+ Expand
9

Comprehensive Materials Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
The morphology of the synthesized materials was examined using scanning electron microscopy (SEM, VERIOS 460, FEI) operating at 10 kV and high-resolution transmission electron microscope (HR-TEM, ARM300, JEOL) operating at 300 kV. Elemental mapping was attained using energy-dispersive X-ray spectroscopy (EDX) equipped with the SEM and TEM. The crystal structure analysis was performed using an X-ray diffractometer (XRD, D/Max2000, Rigaku) with Cu-Ka radiation. Surface chemistry was analyzed using X-ray photoelectron spectroscopy (XPS, Thermo Scientific Kα spectrometer, 1486.6 eV)
+ Open protocol
+ Expand
10

Characterizing Titanium Implant Surfaces

Check if the same lab product or an alternative is used in the 5 most similar protocols
Field-emission scanning electron microscopy. Two Ti implants from each surface group were removed from the sterile packs with plastic tweezers and fixed with carbon tape to SEM stubs, taking care to not to damage or contaminate the surfaces. Both the lateral and flat surfaces of the implants were imaged non-coated at an accelerating voltage of 5 keV and increasing magnifications (up to 50,000×) by FE-SEM (Hitachi S-5200, Japan).
X-ray photoelectron spectroscopy. Implants were analyzed by XPS using a Thermo Fisher Scientific Kα spectrometer (E. Grinstead, UK). A monochromatic Al Kα X-ray source was used with a nominal 400 µm spot size. Survey spectra were obtained (200 eV pass energy (PE)) followed by an examination at 150 eV PE of spectral regions of interest from which the relative atomic percentage composition was obtained. High-resolution spectra (25 eV PE) were also obtained for the Ti envelope. Charge compensation was applied for all spectra using a combined e/Ar + floodgun, and the energy scale was shifted to place the C1s peak at 284.6 eV. All data processing was performed using the Avantage 5.926 software supplied by the manufacturer.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!