The largest database of trusted experimental protocols

Agilent 5975c mass selective detector

Manufactured by Agilent Technologies
Sourced in United States

The Agilent 5975C mass selective detector is a laboratory instrument used for the identification and quantification of chemical compounds. It functions by ionizing and separating molecules based on their mass-to-charge ratio, providing detailed information about the composition of a sample.

Automatically generated - may contain errors

15 protocols using agilent 5975c mass selective detector

1

GC-MS Analysis of Root Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC-MS analyses were performed using Agilent 5975C mass selective detector coupled to a 7890A gas chromatograph (Agilent Technologies, Santa Clara, CA, United States) with 7693 autosampler and a DB-5MS DG capillary column (30 m plus 10 m Duraguard®) with a diameter of 0.25 mm, film thickness of 0.25 μm (Agilent J &W, Santa Clara, CA, United States) as described by Mamer et al. (2013) (link). The GC-MS was run in electron ionization mode (70 eV) and Selected Ion Monitoring (SIM) mode. Data acquisition was done in Scan and SIM modes using MassHunter (Agilent) software. The spectra obtained were compared against the NIST (National Institute of Standards) database. The root samples had large amounts of malic and citric acids, and were diluted 1:40 before being run again in Scan mode. While root exudates did not require dilution. Data were represented as normalized area which is the area of peak divided by amount of sample in mg (roots) or μL (root exudates).
+ Open protocol
+ Expand
2

GC-MS Analysis of Essential Oils and Diterpenes

Check if the same lab product or an alternative is used in the 5 most similar protocols
Relative abundances of essential oil components and esterified diterpene acids were studied using gas chromatography with mass spectrometric detection (GC-MS). GC-MS analyses were performed using an Agilent Technologies 7890A GC-System coupled with an Agilent 5975C mass selective detector (triple-Axis detector, Agilent Technologies, Wilmington, DE, USA). An autosampler unit (Agilent Technologies 7693-100 positions) held samples. Separation of 1-μL injections used an HP-5MS Agilent column (30 m × 250 μm × 0.25 μm). Operating conditions were as follows: injector split ratio 25:1, temperature 250 °C, carrier gas helium, 1.0 mL/min, and constant flow. Column temperature was 50 °C (no hold) and 5 °C per minute. Then, at 280 °C, it was held at 5 min. Mass fragmentation patterns were acquired at −70 eV using a mass scan range of m/z 30–400.
Primary identifications were performed by comparison of mass spectra with an electronic library database [54 ] and confirmed using arithmetic indices, calculated relative to n-alkanes, when compared with values published in Adams [55 ] by visual comparison against mass spectral images [55 ]. Semi-quantification was achieved by the GC-MS operating software, using data with a minimum peak area of 0.1%, by calculating the area under the curve.
HRESIMS spectra were recorded using an AB Sciex 5600 TripleTOF mass spectrometer in positive mode.
+ Open protocol
+ Expand
3

GC-FID and GC-MS Analysis of Essential Oil

Check if the same lab product or an alternative is used in the 5 most similar protocols
The EO was analyzed by an Agilent 6890 gas chromatograph (GC) equipped with a flame ionization detector (FID). Capillary column: HP-5MS (60 m × 0.25 mm, 0.25 μm film thickness). The injection volume was 1 μL and split injection was used (split ratio 1:20). The flow rate of carrier gas helium was set at 1 mL/min. The following GC oven temperature was used: kept at 70 °C (2 min), 2 °C per min to 180 °C (55 min), 10 °C per min to 310 °C (13 min), and held at 310 °C (4 min). The GC-MS analysis was carried out using an Agilent 6890 gas chromatograph equipped with an Agilent 5975C mass selective detector (Agilent Technologies Inc., CA, USA). GC column and parameters were the same as in GC-FID. The mass spectra operated in the mass range (m/z 29 to 500) and EI mode (70 eV). The interface temperature and ion source temperature were 280 °C and 230 °C, respectively. The relative percentage of chemical constituents was determined by the peak area. The retention index (RI) was determined by referring to a series of n-alkanes (C8–C21). The constituents of the EO were identified by comparison of their retention index and mass spectrum with those listed in NIST 2017 and Wiley 275 databases.
+ Open protocol
+ Expand
4

Quantitative Analysis of Ascorbic Acid

Check if the same lab product or an alternative is used in the 5 most similar protocols
Acetonitrile, ethyl acetate, n-hexane (all GC grade), and sodium chloride (analytical grade) were supplied by Merck KGaA, (Darmstadt, Germany); AA (purity ≥ 99.9%), and the internal standard (IS) AA−13C3, hydrobromic acid (48% w/w aq.), sodium thiosulfate pentahydrate, potassium bromide (both analytical grade) were purchased from Sigma Aldrich (St. Louis, MO, USA); bulk C18 (size 125 Å 55–105 µm) supplied by Waters (Milford, MA, USA); saturated bromine solution was supplied by Titolchimica (Pontecchio Polesine, RO, Italy). Agilent 7890A GC-MS system coupled to an Agilent 5975C mass selective detector (MSD) (Agilent Technologies, Santa Clara, CA, USA); the column was a Restek Rxi®-XLB GC (proprietary phase; length × I.D.: 30 m × 0.25 mm; df: 0.25 µm) (Restek, Bellefonte, PA, USA).
+ Open protocol
+ Expand
5

Volatile Compound Analysis by SPME/GC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
All samples were analyzed three times with solid-phase microextraction fiber and assayed with a gas chromatography–mass spectrometer (SPME/GC-MS) and a carboxen/polydimethylsiloxane fiber. A 6 mL solution with 4 µL of a 1,2-dichlorobenzene methanol solution (100 mg/L) used as an internal standard was added into a 15 mL amber vial closed by a PTFE/silicone septum.
Analysis of volatile compounds was performed on a GC-MS using an Agilent 5975C mass selective detector coupled with an Agilent 7890A GC (Agilent, Santa Clara, CA, USA), equipped with a DB-5MS capillary column (a 60 m × 0.25 mm inner diameter and a 0.25 µm film thickness). Helium was used as a carrier gas, and the flow rate was 0.8 mL/min. Subsequent headspace equilibration occurred over a period of 20 min at 60 °, and the SPME fiber was exposed in the headspace for 30 min. GC oven temperature was programmed at 40 °C (held for 3 min) and increased to 240 °C at 4 °C/min (held for 2 min). The MS conditions were as follows: the transfer line temperature was 280 °C; the ion source temperature was 230 °C; ionization energy was 70 eV, and mass range was 40–400 a.m.u. Desorption was in splitless mode at 250 °C for 5 min.
+ Open protocol
+ Expand
6

GC-MS Analysis of Tobacco Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC–MS analysis of the metabolites in the tobacco leaves was carried out using an Agilent 7890A gas chromatograph (GC) interfaced to an Agilent 5975C mass-selective detector (Agilent, USA), controlled by an Agilent G1701EA GC-MSD ChemStation. Samples were tested in random order with quality control (QC) samples being inserted with every six samples in the running sequence. QC samples were mixed with the same amount of each sample. Chromatographic separations were achieved in an DB-5 (30 m × 0.250 mm, 0.25 µm film thickness) capillary column (Agilent Technologies Inc., USA). The column flow, using helium as the carrier gas, was held constant at 1.0 ml min−1. The column temperature was set to 70°C for 4 min, and programmed to 310°C at 5°C min−1 and kept at this temperature for 10 min. The injector temperature was 290°C, and the injector was set in split mode (10 : 1) with an injection volume of 1 μl. The interface temperature was 230°C and the ion source temperature was 280°C. Mass spectra were recorded at 70 eV, and used both full scan and selected ion monitoring (SIM) mode with scanning from 40 to 510 amu. Ions were acquired with a solvent cut time of 8.0 min [1 (link)].
+ Open protocol
+ Expand
7

GC-MS Analysis of Volatile Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples were analyzed and identified using an Agilent 7890A Series GC system coupled with an Agilent 5975C Mass Selective Detector (Agilent Technologies, Santa Clara, CA, USA). Helium was used as the carrier gas at a flow rate of 1.0 mL/min with an HP-INNOWAX capillary column (30 m × 250 µm × 0.25 µm). The inlet temperature was 200 °C. The column temperature program of the GC was initially set at 40 °C for 3 min, heated to 100 °C at 3 °C/min, maintained for 5 min, increased to 200 °C at 20 °C/min, and maintained for 7 min. Mass spectrometry detection used an electron impact (EI) ionization system at 70 eV, the ionization source temperature was 280 °C, the quadrupole temperature was 150 °C, the full-scan acquisition mode was performed with a mass range of 33–350 (m/z), and the tuning file included standard tuning. Constituents were identified by comparing the mass spectra with the National Institute of Standards and Technology (NIST08, http://webbook.nist.gov/chemistry/ (accessed on 29 November 2022)) library and published data. Furthermore, the retention time (RT) of the standard substances for the representative compounds was measured according to the above experimental conditions. Relative percentages (RPs) were calculated to determine the proportion of each component.
+ Open protocol
+ Expand
8

GC-MS Analysis of Derivatized Pleural Fluid

Check if the same lab product or an alternative is used in the 5 most similar protocols
The pleural fluid samples were chemically derivatized and then analyzed by GC‐MS.24, 25 The sample processing diagram is shown in Figure 1. 2.0 μL solution was injected in splitless mode into the system after derivatization. The system was an Agilent 7693 Series Autosampler (Agilent Technologies) and an Agilent 7890A GC System equipped with a fused silica capillary column (30 m × 0.25 mm i.d.) chemically bonded to a 0.25 m HP‐5MS stationary phase (Phenomenex). The carrier gas helium was used with a constant flow rate of 1.0 mL/min. The injection, ion source, and quadrupole temperatures were 270℃, 230℃, and 150℃, respectively. Optimized conditions are as follows: The initial column temperature was held at 70℃ for 5 min and then ramped at a rate of 15℃/min to 280℃, and maintained for 9 min. The column effluent was introduced into the ion source of an Agilent 5975C Mass Selective Detector (Agilent Technologies). Masses were acquired from m/z 60 to 800.
+ Open protocol
+ Expand
9

GC-FID and GC-MS Analysis of Essential Oil

Check if the same lab product or an alternative is used in the 5 most similar protocols
The EO was analyzed by an Agilent 6890 gas chromatograph equipped with a flame ionization detector (FID). Capillary column: HP-5MS (60 m × 0.25 mm, 0.25 μm film thickness). The flow rate of carrier gas helium was set at 1 mL/min and the split ratio was 1:20. The following GC oven temperature was used with the injection volume of 1 μL: held at 70°C (2 min), 2°C per min to 180°C (55 min), 10°C per min to 310°C (13 min), and kept at 310°C (12 min). The GC-MS analysis was performed using an Agilent 6890 gas chromatograph equipped with an Agilent 5975C mass selective detector (Agilent Technologies Inc., CA, USA). GC column and parameters were the same as in GC-FID. The mass spectra operated in the mass range (m/z 29 to 500) and EI mode (70 eV). The interface temperature and ion source temperature were 280°C and 230°C, respectively. The relative percentage of chemical constituents was determined by the peak area. The retention index (RI) is determined by referring to a series of n-alkanes (C8–C22). The identification of chemical constituents was based on comparison of their retention index and mass spectrum with those in Wiley 275 and NIST 2017 databases.
+ Open protocol
+ Expand
10

GC-MS Analysis of Epiandrosterone

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC-MS analyses were carried out as previously described [49 (link)] following derivatisation of samples and standards with N-methyl-N-trimethylsilyl-trifluoroacetamide (MSTFA, CS-Chromatographie Service, Langerwehe, Germany), trimethylsilyl iodide (TMSI), and dithioerythritol (DTE, both from Sigma Chemie GmbH, Deisenhofen, Germany) according to Magnisali et al. (2008) allowing quantification of the resulting trimethylsilyl derivatives by GC–MS. Sample analysis was performed on an Agilent 7890A Gas Chromatograph (Agilent Technologies, Böblingen, Germany) equipped with an Agilent HP-5MS (5% phenyl-, 95% methylsiloxane) fused silica capillary column (30 m x 0.25 mm i.d. x 0.25μm) interfaced with an Agilent 5975C mass selective detector.
The MS was operated in the EI mode with the electron voltage set to autotune value. MS acquisition was performed in selected ion monitoring mode (SIM) by monitoring the ions m/z 434 and 419 for epiandrosterone. The Agilent MSD chemstation data chemstation software was used for peak integration and library searches.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!