The largest database of trusted experimental protocols

Beh c8 column

Manufactured by Waters Corporation
Sourced in United States

The BEH C8 column is a liquid chromatography column manufactured by Waters Corporation. It features a stationary phase comprising 1.7 μm particles with a C8 bonded ligand. The column is designed for reversed-phase liquid chromatography separations.

Automatically generated - may contain errors

28 protocols using beh c8 column

1

Quantitative Analysis of Cannabis Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
The product profile of AtaPT was identified and quantified using LC–MS using a 1290 Infinity II UHPLC coupled to a 6560 Ion Mobility Q-TOF (Agilent). The quantitative analysis of product formation was performed by using a Xevo TQ-S triple quadrupole tandem mass spectrometer (Waters MS Technologies) connected to an Acquity UPLC system (H-Class; Waters). Olivetol and CBG separation was performed by LCMS on a BEH C8 column (1.7 µm, 2.1 × 50 mm, Waters). Linalool quantitation was performed by GCMS using an Agilent Technologies 7890B GC equipped with a 5977A MSD detector. Product concentrations were determined by comparing peak areas to a standard curve generated from authentic standards run under identical conditions. Further details of the running conditions for compound identification and quantitation are found in Supplementary Methods S4.
+ Open protocol
+ Expand
2

LC-MS/MS Metabolite Profiling Workflow

Check if the same lab product or an alternative is used in the 5 most similar protocols
Under positive-ion mode, the chromatography was performed on the BEH C8 column (2.1×100 mm×1.7 μm, Waters, Milford, MA, United States). The flow rate was 0.35 mL/min. The column temperature was 50°C. The injection volume was 5 μL. The mobile phase was composed of solvent A (water with 0.1% formic acid) and solvent B (acetonitrile with 0.1% formic acid). Gradient elution procedure: 0–1 min, 5%B; 1.1–11 min, 5–100%B; 11–13 min,100%B; 13.1–15 min, 5%B(Table S1-1). Under negative-ion mode, a 5µl reconstitution sample was injected for chromatography separation in the HSS T3 column (2.1×100 mm×1.8μm, Waters, Milford, MA, United States). The other chromatographic conditions Under negative-ion mode are the same as positive-ion mode (Table S1-2).
+ Open protocol
+ Expand
3

Lipidomic Analysis of RYGB and Sham Lymph Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
Seven lipid categories were analyzed: TG (n = 81), DG (n = 22), Phosphatidylinositol (PI, n = 12), Phosphatidylethanolamines (PE, n = 16), Phosphatidylcholines (PC, n = 28), Lysophosphatidyl-choline (LPC, n = 11) and Sphingomyelins (SM, n = 13) from RYGB (n = 8) and Sham (n = 6) lymph samples. Lipids were extracted by a methyl-tert-butyl ether (MTBE) protocol as previously described [22] (link). Data acquisition was performed on an LTQ Orbitrap Velos Pro instrument (Thermo Scientific) coupled to a Dionex Ultimate 3000 UHPLC (Thermo Scientific) according to previously published protocols [23] (link), [24] (link). Briefly, chromatographic separation was performed on a Waters (Waters, Milford, MA, USA) BEH C8 column (100 × 1 mm, 1.7 μm), and the mass spectrometer was operated in Data Dependent Acquisition (DDA) mode. Full scan profile spectra were acquired in the Orbitrap mass analyzer at a resolution setting of 100,000. For MS/MS experiments, the 10 most abundant ions of the full scan spectrum were sequentially fragmented. Data analysis was performed by Lipid Data Analyzer, a custom software described in Hartler et al. [25] (link), [26] (link).
+ Open protocol
+ Expand
4

Comprehensive UPLC-MS/MS Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The sample was analyzed using an Ultimate 3000UPLC system (Dionex, United States) combined with a BEH C8 column (2.1 × 100 mm, 1.7 µm) (Waters, United States) and a HSS T3 column (2.1 × 100 mm, 1.8 µm) (Waters, United States). The temperature of the column was kept at 30°C, the flow rate was 0.35 mL/min, and the sample injection volume was 5 μL. In the positive ion mode, the mobile phase was consisted of water with 0.1% formic acid (solvent A) and 0.1% formic acid in acetonitrile (solvent B). The gradient profile was performed as follows: 5% B (0–1 min); 5%–100% B (1–24 min); 100% B (24.1–27.5 min); 5% B (7.6–30 min). In the negative ion mode, the mobile phase was a mixture of water with 6.5 mM ammonium bicarbonate (solvent A) and 95% methanol with 6.5 mM ammonium bicarbonate (solvent B). Gradient elution condition: 5% B (0–1 min); 5% B (1–18 min); 100% B (18.1–22 min); 5% B (22.1–25 min). The electrospray ionization (ESI) source was operated and optimized and the MS parameters were as follows: aux gas flow rate: 8 Arb; sheath gas flow rate: 35 Arb; capillary temperature: 320°C; aux gas heater temperature: 350°C; mass range: 70–1,050 m/z, the full MS resolution: 70000; MS/MS resolution: 17500. The collision energy was 20, 40% in NCE mode; the spray voltage was +3.8 kV (positive) or −3.0 kV (negative).
+ Open protocol
+ Expand
5

UPLC-Q Exactive HF Lipidomics Profiling

Check if the same lab product or an alternative is used in the 5 most similar protocols
Lipidomics profiling was performed using a UPLC system (Waters) coupled to a Q Exactive HF mass spectrometer (Thermo Scientific, Rockford, IL, USA). The lipids were separated using a BEH C8 column (2.1×100 mm, 1.7 µm; Waters) at 60 °C and the flow rate was set at 0.3 mL/min. The mobile phase A was 60% acetonitrile/water (v/v) with 10 mM ammonium acetate and phase B was 90% isopropanol/acetonitrile (v/v) containing 10 mM ammonium acetate. The elution gradient conditions were 50% B for 1.5 min, linearly increased to 85% B at 9 min and further linearly increased to 100% B in 0.1 min, and maintained at 100% B for 1.9 min. The elution gradient then returned to 50% B in 0.1 min, followed by equilibration with 50% B in the next 1.9 min. The scan range of precursor ion was set at 300–1,100 m/z (ESI+) and 160–1,600 m/z (ESI−) in full-scan mode, while the sheath gas flow rate was 45 arb and aux gas flow rate was 10 arb. The spray voltages were 3.5 kV (ESI+) and −3.0 kV (ESI−), the capillary temperature was 320 °C, the aux gas heater temperature was 350 °C, and the resolution was set at 140,000 (full width at half maximum at 200 m/z).
+ Open protocol
+ Expand
6

Quantification of Imperatorin by UPLC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The 2′- and 4′-hydroxy-esfenvalerate, xanthotoxin, indoxacarb, and abamectin were monitored as previously described (24 (link), 36 (link), 40 (link)). For imperatorin, 1 μL samples were separated on a BEH C8 column (2.1 × 50 mm, 1.7 μm particle size; Waters, Milford, Massachusetts, USA) using an Acquity UPLC system (I-Class; Waters) and eluted with acetonitrile (A) and 1 mM ammonium acetate (B). The gradient elution program was 0 min A:B 10:90; 0.3 min A:B 10:90; 2 min A:B 95:5; 3 min A:B 95:5; 3.1 min A:B 10:90; and 5 min A:B 10:90, and the flow rate was 0.3 mL/min. The imperatorin was detected on a tandem triple quadrupole mass spectrometer (Xevo TQ-S micro; Waters) run in positive ESI and multiple reaction monitoring (MRM) modes, with the MRM transitions 271 > 203.2 and 271 > 147.1 used for quantification of imperatorin and confirmation of its identity. The LOD for imperatorin conversion under these conditions was 0.06 pmol/min/pmol P450.
+ Open protocol
+ Expand
7

Sphingolipid Separation by UHPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Chromatographic separation
of sphingolipids
was performed as previously described by Triebl et al.24 (link) Briefly, a BEH C8 column (100 × 1 mm, 1.7
μm; Waters, Milford, MA) thermostated at 50 °C was used
in a Dionex Ultimate 3000 RS UHPLC system. The mobile phase A consisted
of deionized water containing 1 vol % of 1 M aqueous ammonium formate
(final concentration: 10 mmol/L) and 0.1 vol % of formic acid as additives.
The mobile phase B consisted of a mixture of acetonitrile/isopropanol
5:2 (v/v) containing the same additives. The gradient elution started
at 50% mobile phase B, rising to 100% B over 15 min, held at 100%
B for 10 min, and the column was then re-equilibrated with 50% B for
8 min before the next injection. The flow rate was 150 μL/min.
The samples were kept at 8 °C, and the injection volume was 2
μL.
+ Open protocol
+ Expand
8

Quantification of Acylcarnitines via UPLC-TOF

Check if the same lab product or an alternative is used in the 5 most similar protocols
Tissues for analysis were removed quickly, frozen in liquid nitrogen, and stored at − 80 °C prior to quantification. Acylcarnitines were analyzed using an Acquity UPLC-TOF system (Waters) with a BEH C8 column (1.7 μm particle size, 100 mm × 2.1 mm, Waters). The two mobile phases were 1 mM ammonium formate in methanol (phase A) and 2 mM ammonium formate in H2O (phase B), both phases with 0.05 mM formic acid. The following gradient was programmed: 0 min, 65% A; 10 min, 90% A; 15 min, 99% A; 17 min, 99% A; 20 min, 65% A, and a flow rate of 0.3 mL min−1. Quantification was carried out using the extracted ion chromatogram of each compound, with 50-mDa windows. The linear dynamic range was determined by injecting standard mixtures. Positive identification of compounds was based on the accurate mass measurement with an error < 5 ppm and their LC retention time was compared to that of a standard (± 2%).
+ Open protocol
+ Expand
9

Peptide Composition Analysis of Transferon

Check if the same lab product or an alternative is used in the 5 most similar protocols
Peptide composition and homogeneity of Transferon batches were determined by MS (mass spectrometry) using an ESI-Q-TOF (electrospray ionization-quadrupole-time of flying) mass detector, coupled to an Acquity chromatographic system (Waters). Transferon samples were filtered through a 0.10 µm PVDF membrane (Merck Millipore Co.), and 5 μL was injected into the chromatographic system. Sample desalination and separation were performed on a BEH C8 column (130 Å, 1.7 μm, 2.1 × 100 mm) (Waters) using aqueous formic acid (0.1%) (Sigma-Aldrich) (Solution A) and acetonitrile (Mallinckrodt Baker) (Solution B). Column temperature was maintained at 30°C, and the flow rate was set at 0.2 mL/min using a gradient configuration (1-2 min, 60% A; 3–40 min, 90% A; 41–45 min, 60% A). Mass spectrometric analysis was performed on a Xevo G2-SQ-TOF mass detector (Waters). Electrospray mass spectra data were recorded in a positive ion mode for a mass range from 100 m/z to 2000 m/z with a scan time of 0.1 s. The source temperature was set at 150°C with a cone gas flow of 20 L/h. Desolvation gas flow was set at 600 L/h and 250°C. The capillary was maintained at 3.5 kV, and the cone voltage was set to 20 V. Mass spectra data were analyzed using Mass Lynx (Waters).
+ Open protocol
+ Expand
10

Lipid Extraction and UPLC-MS Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The lipids extracted from the previous step were solubilised in 300 μl of a mixture of acetonitrile:isopropanol (7:3 v/v), vortexed and incubated 10 min in an ultrasonication bath. The samples were centrifuged for 5 min at 10,000 × g, and 100 μl aliquots were transferred into glass vials for further analysis (Salem et al., 2016 (link)). Lipids in 2 μl injection samples were separated on a reversed-phase Bridged Ethyl Hybrid (BEH) C8 column (100 mm × 2.1 mm containing 1.7 μm diameter particles, Waters, Manchester, United Kingdom) using a Waters Acquity UPLC system (Waters, Manchester, United Kingdom) (Hummel et al., 2011 (link)). The UPLC–FT–MS raw chromatograms were analysed and processed either by using Xcalibur (Version 2.10, Thermo-Fisher, Bremen, Germany) or ToxID (Version 2.1.1, Thermo-Fisher) as described in Hummel et al. (2011) (link).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!