The largest database of trusted experimental protocols

13 protocols using amira avizo software

1

High-Resolution FIB-SEM Imaging of Resin-Embedded Tissues

Check if the same lab product or an alternative is used in the 5 most similar protocols
Resin-embedded tissue blocks were trimmed, mounted on SEM stubs, and then coated with a 5 nm platinum layer using a Q150T-ES sputter coater (Quorum Technologies, UK) before FIB-SEM volume imaging. Data was acquired using Scios Dual beam microscope (FIB-SEM) (Thermo Fischer Scientific). Electron beam imaging was acquired at 2 kV, 0.1 nA current, 1.9 × 1.9 nm pixel spacing, 7 mm working distance, 10 µs acquisition time, and 3072 × 2048 resolution using a T1 detector. SEM images were acquired every 20 nm. The working voltage of gallium ion beam was set at 30 kV, and 0.5 nA current was used for FIB milling. The specimens were imaged at 5 × 5 µm block face and 5 µm depth. FIB milling and SEM imaging were automated using the Auto Slice and View software (v4.1.1.1582, Thermo Fischer Scientific). SEM volume images were aligned and reconstructed using ImageJ (v. 1.8.0_172, NIH) with linear stack alignment, with SIFT and MultiStackRegistration plugins77 ,78 . Analysis and segmentation of SEM volume images were done using Amira-Avizo software (v2020.3.1, Thermo Fisher Scientific).
+ Open protocol
+ Expand
2

Quantifying Thigh Muscle Composition

Check if the same lab product or an alternative is used in the 5 most similar protocols
The distal muscle volumes were evaluated by manual segmentation of muscle regions of interest (ROIs) using the Amira-Avizo Software (V6; Thermo Scientific, Waltham, MA, United States). ROIs included the four heads of the quadriceps femoris (vastus medialis, vastus lateralis, vastus intermedius, rectus femoris), the hamstrings (biceps femoris, semimembranosus, semitendinosus), gracilis and sartorius muscles. The segmented muscle volume spanned a distal thigh section of 87.7 ± 2.8 mm (18 ± 0.57 slices) and was standardized relative to the patients’ femur length, defined as distance between the top of the greater trochanter and the lower border of the lateral femoral condyle. This ensured that a comparable muscle volume was analysed in each patient, regardless of body size or leg length. Intramuscular fat content of the muscle ROIs was quantified on the four most superior contiguous axial CT-slices of the investigated thigh region with the software Image J (V1.51; U.S. National Institutes of Health, Bethesda, MD, United States). A voxel-based analysis of muscle density was performed using predefined Hounsfield unit (HU) threshold values to differentiate intramuscular fat (−190 to −30 HU) and skeletal muscle tissue (−29 to 150 HU) (Aubrey et al., 2014 (link)).
+ Open protocol
+ Expand
3

Viral Signal Mapping in Rodent Brains

Check if the same lab product or an alternative is used in the 5 most similar protocols
OPT images with viral signal and autofluorescence signal were reconstructed in DICOM format using NRecon software (v.1.7.0.4, Bruker) followed by their conversion into NifTi using PMOD VIEW tool (v.4.2, PMOD Technologies Inc., Switzerland) or the dcm2nii tool in MRIcron software for OPT and MR images, respectively. Coregistration and normalization of OPT with the OCUM MRI template was performed using the toolbox SPMmouse in SPM8. All transformation matrices were calculated using individual tissue autofluorescence from OPT images and applied to the corresponding viral OPT images21 . The accuracy of all transformations were assessed visually by two individual readers in the check registration tool of SPM8. Fusion images of viral OPT signal were created for each individual brain using its own MRI and with the OCUM template21  using the PMOD VIEW tool or Amira-Avizo software (v6.3.0, Thermo Fisher Scientific) for 3D renderings. Finally, brain areas with viral signal were identified according to the OCUM atlas21 .
+ Open protocol
+ Expand
4

Visualizing Collagen Intermolecular Channels

Check if the same lab product or an alternative is used in the 5 most similar protocols
The electron density map of 8 × 8 × 1 collagen unit cells was built using the UCSF Chimera package76 (link), based on the structure obtained by a previous X-ray study32 (link). The map was generated with a resolution of 0.3 nm and voxel size of 0.1 nm. From this electron density map we segmented the intermolecular channels within the collagen structure by applying in-house Matlab scripts. The edges of the channels were defined by considering the diameter of each collagen triple-helix molecule as 1.5 nm77 (link). Channels with diameters <8 pixels (0.8 nm) were removed from the segmentation by eroding and then dilating the segmented channels for 4 pixels for an enhanced separation of the main channels. The channels on the model surface were also removed for improved visualization of the channels inside the collagen structure. The remaining channels were then labeled with different colors according to their connectivity, and visualized in 3D using Amira-Avizo software (Thermo ScientificTM).
+ Open protocol
+ Expand
5

High-resolution CT Segmentation and 3D Modeling

Check if the same lab product or an alternative is used in the 5 most similar protocols
High-resolution CT scan (Philips Brilliance 64 CT scanner, Cleveland, Ohio, USA; voxel size 0.2 × 0.2 × 0.3 mm3, 120 kV, 150 mAs, pitch 0.5) were performed. The similar resolution values had been previously described for the virtual bone digital model reconstructions in several studies [20 (link)–22 (link)]. A level of accuracy and reliability of measurement capability that approaches the higher resolution using microCT were reported [23 (link)]. The DICOM (Digital Imaging and Communications in Medicine) files obtained from CT scan were used to create the 3D model using Amira-Avizo software (Thermo Fisher Scientific, MA, USA). Segmentation of bone was applied based on Hounsfield unit (range: 250 to 2500). This is the standard value described for segmenting bone in the several studies [14 , 19 (link), 24 (link)].
+ Open protocol
+ Expand
6

3D Visualization and Segmentation of Middle Ear

Check if the same lab product or an alternative is used in the 5 most similar protocols
The Amira-Avizo software (Thermo Scientific Co., version 2020.3.1) with the XImagePAQ - Advanced Image processing and quantification extension was used for the 3D and 4D visualization of the datasets. The software has previously proven its capability of handling big data10 . With the extension, we can profit from additional advanced image processing and quantification tools, e.g., the watershed segmentation tool.
Because image files in the standard tiff file format are limited to 4 GB, the reconstructed tiff files were converted to the AmiraMesh file format (.am), which is the native file format of Amira. For the 3D visualization of the middle ear (see Fig. 3), the five sub-scans containing the ossicular chain and the tympanic membrane were loaded into Amira and registered one by one with the Register Images module. The mutual information metric was chosen as the similarity measure38 –40 . The merge module finally allowed us to end up with only one image containing the five sub-scans and the entire ossicular chain. In the Segmentation Editor, the watershed segmentation41 was initiated by placing markers on every 50th slides in the region of the stapes and every 100th in the remaining area. The stapes, the incus, the malleus, and the tympanic membrane were segmented the same way.
+ Open protocol
+ Expand
7

3D Imaging of Cleared Brain with TSPO Staining

Check if the same lab product or an alternative is used in the 5 most similar protocols
The brain stained for TSPO and imaged by NiR-OPT was consequently rescanned using an UltraMicroscope II (Miltenyi Biotec, Germany) including a 1x Olympus objective (Olympus PLAPO 2XC) coupled to an Olympus MVX10 zoom body, providing 0.63x up to 6.3x magnification with a lens corrected dipping cap MVPLAPO 2x DC DBE objective (Olympus). The UltraMicroscope II is equipped with the white laser SuperK EXTREME (NKT photonics Birkerød, Denmark). The cleared brain was immerged in BABB and magnification was set to 0.63x. For image acquisition, left and right light sheets were blend merged with a 0.14 numerical aperture, resulting in a light sheet z-thickness of 3.87 μm and 80% width, while using a 12-step contrast adaptive dynamic focus across the field of view. LSFM images were acquired with the following settings: Ex: 250/25, Em: 625/30 (for AlexaFluor-594 channel) and Ex: 470/40, Em: 525/50 (for the autofluorescence channel). Image sections with a step size of 10 μm were generated by Imspector Pro software (v7.1.15, Miltenyi Biotec Gmbh, Germany) and stitched together using the implemented TeraStitcher script (v.9). The obtained images were then converted into NifTi files using Amira Avizo software (version 6.3.0, Thermo Fisher Scientific, Waltham, MA, United States) and resampled prior to co-registration.
+ Open protocol
+ Expand
8

Multi-modal image registration and fusion

Check if the same lab product or an alternative is used in the 5 most similar protocols
Initially, both the autofluorescence image and the OPT image displaying specific signals were reoriented manually in SPM8 to the templates orientation and their origins were set tangent to the upper edge of the brain at Bregma. For co-registration of OPT and MR images, voxel-to-voxel affine transformation matrices were calculated using the autofluorescence OPT images and applied to those displaying the specifically targeted signals. To further improve the fusion images and enable image warping, binary masks of the autofluorescence OPT images were created in ITK-SNAP version 3.8.0 (Yushkevich et al., 2006 (link)) which were consequently normalized to a binary OCUM template mask. Both co-registration and normalization were performed in SPM8, using the SPMmouse toolbox. Finally, fusion images were created in the PMOD view tool and 3D images were created in Amira-Avizo software (version 6.3.0, Thermo Fisher Scientific).
+ Open protocol
+ Expand
9

Quantitative Myelin Imaging in Mouse Brain

Check if the same lab product or an alternative is used in the 5 most similar protocols
Sections ranging from Bregma AP –0.1 mm to AP –0.7 mm for each animal (n = 84 animals; n = 211 sections) were selected for fluorescence-based microscopy imaging. Images were acquired using a TxRed filter (excitation = 540–580 nm; emission = 600–660 nm) on a Nikon Eclipse Ti-E inverted microscope with a DU897 ANDOR EMCCD camera controlled by Nikon NIS Elements interface, equipped with Nikon CFI Plan Apochromat ×20 (N.A 0.75) objective. Prior to analysis, all the images were aligned using Amira-Avizo Software (version 6.3.0, Thermo Fisher Scientific). A region of interest (ROI) was selected based upon a significant VBM cluster in SSp-ul described in this study. The ROIs were manually positioned and saved for each section using FIJI (Schindelin et al., 2012 (link)). Signal-to-noise (specific myelin immunoreactivity versus background fluorescence) was determined by the segmentation of each image using FIJI’s Multi Otsu Threshold plugin using three different levels of classification. This plugin is based on Otsu’s original method but also implements an algorithm described by Liao and Chung (Liao et al., 2001 (link)). The specific signal was quantified for each section per individual to calculate a mean immunoreactive value for each subject (n = 12 per time point).
+ Open protocol
+ Expand
10

Micro-CT Imaging of Lung Structures

Check if the same lab product or an alternative is used in the 5 most similar protocols
MicroCT images were obtained on a Bruker Skyscan 1176 at the Ralph H. Johnson VA Medical Center or on a Skyscan 1276 at National Jewish Heatlh (indicated in figure legends, Bruker MicroCT, Kontich, Belgium). Longitudinal and, respiration-controlled studies images were acquired at 35 μm and post-mortem studies images were acquired at 9 μm resolution, with a 0.5 mm Aluminum filter, and 0.7° rotation step. Scans were reconstructed using NRecon software (Bruker, v1.17.7). Analysis and generation of VOIs was performed using CTAn software (Bruker) or by Amira-Avizo software (Thermo Fisher Scientific, version 2022.1) as indicated in the text. CTVol software (Bruker) was used to create 3D surface renderings from surface model files. Method-specific task lists and considerations are provided (S1 Methods).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!