The largest database of trusted experimental protocols

4 protocols using dgu 20as degasser

1

HPLC Analysis of Microcystins

Check if the same lab product or an alternative is used in the 5 most similar protocols
MCs analyses were performed on a Shimadzu HPLC System with a DGU-20As Degasser, LC-20AD Pump, CTO-20AC Oven, and a SPD-M20A Diode Array Detector (Shimadzu do Brasil, Barueri, SP, Brazil). Chromatographic separation was achieved at 40 °C using a Luna C18 column (Phenomenex; 4.6 × 250 mm × 5 μm). The mobile phase consisted of an isocratic mixture of 20 mmol.L−1 ammonium acetate and Acetonitrile (70:30; v:v). The flow rate was 1.0 mL.min−1 and 20 μL of samples were injected into the HPLC System. MCs were detected at their maximum absorbance of 238 nm. The full chromatographic run was 10 min and the peak corresponding to MCs was observed at 6.36 min. All samples were analyzed in duplicate.
+ Open protocol
+ Expand
2

HPLC Analysis of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC was conducted on a Shimadzu system consisting of a DGU-20AS degasser, SIL-20A HT autosampler, and a SPD-M20A UV–vis detector set at a wavelength of 230 nm. The system was fitted with an Eclipse XDB C18 column (4.6 × 150 mm, 3.5 μm particle size) and the oven temperature was set to 25 °C. The mobile phase consisted of ammonium acetate buffer (0.02 M, pH 5) and acetonitrile (43:57% v/v) at a flow rate of 1 mL/min, and the injection volume was 20 μL.
+ Open protocol
+ Expand
3

Comprehensive LC-MS/MS Analysis of OMW

Check if the same lab product or an alternative is used in the 5 most similar protocols
The OMW extracts were analysed using a Shimadzu UFLC XR system consisted of an electrospray ionization source (ESI) equipped with two LC-20ADXR solvent delivery units, a SIL-20AXR autosampler, an SCL-10A system controller, a CTO-20 AC column oven, a DGU-20AS degasser (Shimadzu, Kyoto, Japan). A volume of 5 μL each extract was injected at 0.5 mL/min to a Discover BIO Wide Pore C18 column (150 mm × 3 mm, 3 μm) at 40 °C and separated with two mobile phases: A (0.1% formic acid in water v/v) and B (0.1% formic acid in methanol v/v) following the programmed linear gradient elution : 0–14 min, from 10 to 20% B; 14–27 min, from 20 to 55% B; 27–37 min, from 55 to 100% B; 37–45 min, 100% B; and 45–50 min 10% B48 (link),49 (link). The ionization mode was negative and The ESI conditions were set as follows: capillary voltage of − 3.5 v, a nebulizing gas flow of 1.5 L/min, a dry gas flow rate of 15 L/min, a DL (dissolving line) temperature of 280 °C, a block source temperature of 400 °C, and a voltage detector of 1.35 V. Compounds were identified by comparing their retention time and mass spectra with those of reference standards. The validation of the method was determined as detailed in50 (link).
+ Open protocol
+ Expand
4

Quantifying Drug Release from Stents

Check if the same lab product or an alternative is used in the 5 most similar protocols
Drug-loaded stents were cut into small pieces (30.95 ± 0.15 mg) using a punch, accurately weighed (n = 3) and then the polymer coating and DTX were dissolved in THF (1 mL). The solution was then added to 9 mL mixture of ammonium acetate buffer (0.02 M, 25 mL, pH 5) and acetonitrile (43:57% v/v). The samples were sonicated for 30 min, filtered through a 0.45 µm polyvinylidene difluoride (PVDF) syringe filter and the DTX concentration was determined via HPLC. Examples of calibration curves and DTX chromatograms are provided in the SI (Figure S4).
HPLC was conducted using a method described by Shaikh et al. [17 (link)] on a Shimadzu system consisting of a DGU-20AS degasser, SIL-20A HT autosampler, LC-20AD pump unit and a SPD-M20A UV–vis detector (λ = 230 nm). The system was fitted with an Agilent Eclipse XDB C18 column (4.6 × 150 mm, 3.5 μm particle size) and the oven temperature was set to 25 °C. The mobile phase consisted of ammonium acetate buffer (0.02 M, pH 5) and acetonitrile (43:57% v/v) at a flow rate of 1 mL/min, and the injection volume was 20 μL.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!