The largest database of trusted experimental protocols

6 protocols using acquity uplc flr detector

1

ABEE-Labeled Monosaccharides Separation and Detection

Check if the same lab product or an alternative is used in the 5 most similar protocols
Chromatographic separation and detection of monosaccharides labeled with ABEE was performed with a modified method described by Kumagai et al. (2012) . The ABEE-labeled monosaccharides were subjected to liquid chromatography (ACQUITY UPLC H-Class system, Waters Inc.) equipped with an ACQUITY UPLC BEH C18 column (Waters) and fluorescence detector (ACQUITY UPLC FLR Detector, Waters). The mixture of 200 mM potassium borate buffer (pH 8.9) and pure acetonitrile was used as the mobile phase. The ABEE-labeled monosaccharides were detected by a fluorescence detector using emission and excitation wavelengths of 305 and 360 nm, respectively. The peak assignment and quantification were made with a mixture of 10 monosaccharides as an external standard. The detailed procedure will be published elsewhere.
+ Open protocol
+ Expand
2

UPLC-MS Analysis of N-glycans

Check if the same lab product or an alternative is used in the 5 most similar protocols
Ten microlitres (10 μL) of reconstituted released N-glycans were injected into an ACQUITY H-Class UPLC (Waters Corporation, United States) coupled to a Xevo G2S QTof mass spectrometer (Waters Corporation, United States). Samples were separated using an ACQUITY UPLC Glycan BEH amide column (130 Å, 1.7 μm, 2.1 mm × 150 mm, Waters Corporation, United States) at 60°C and 400 μL/min, with a 40 min gradient from 25 to 49% of 50 mM Ammonium Formate (mobile phase A). 100% ACN was used as mobile phase B. RFMS-labelled glycans were excited at 265 nm and measured at 425 nm with an ACQUITY UPLC FLR detector (Waters Corporation, United States). The MS1 profile scans of m/z 400–2,000 were acquired using the Xevo G2S-QTof in positive mode with an acquisition rate of 1 Hz. The electrospray ionisation capillary voltage was set at 2.75 kV, cone voltage at 15 V, desolvation gas flow at 800 L/h, ion source temperature and desolvation temperature were kept at 120°C and 300°C, respectively. Glu1-fibrinopeptide B (Waters Corporation, United States) was used as the LockSpray compound for real-time mass correction. RapiFluor-MS Dextran Calibration ladder (Waters Corporation, United States) was also injected into LC-MS to calibrate the retention time of sample peaks. The retention times were normalised using the dextran calibration curve to Glucose Units (GU).
+ Open protocol
+ Expand
3

Quantification of (S)-Warfarin 7-Hydroxylation

Check if the same lab product or an alternative is used in the 5 most similar protocols
(S)-Warfarin 7-hydroxylation by CYP2C9 was measured as previously reported, with several modifications [23 (link)]. The reaction mixture, in a total volume of 150 μL, consisted of the following components: the microsomal fraction (25 μg), (S)-warfarin (0.2, 0.5, 1, 2, 5, 10, 20, or 40 μM), and 100 mM potassium phosphate buffer (pH 7.4). Following pre-incubation at 37 °C for 3 min, reactions were initiated by the addition of 10 mM NADPH, with incubation at 37 °C for 60 min. Reactions were terminated by adding 150 μL of acetonitrile containing 25 nM 7-ethoxycoumarin as an internal standard. After protein removal by centrifugation at 15,400× g for 10 min, 10 μL of the supernatant was injected into an ultra-high performance liquid chromatography (UPLC)-fluorescence system consisting of an ACQUITY UPLC H-Class PLUS (Waters, Milford, MA, USA), ACQUITY UPLC FLR Detector (Waters), and an ACQUITY UPLC HSS C18 column (2.1 × 50 mm, 1.8-μm particle size; Waters) maintained at 40 °C. The mobile phase was a mixture of acetonitrile and water (40:60, v/v) containing 0.1% formic acid at a flow rate of 0.5 mL/min. (S)-7-Hydroxywarfarin content was measured at an excitation wavelength of 320 nm and an emission wavelength of 415 nm. Standard curves were constructed in the 12.5–6400 nM range using metabolite standards, with a quantification limit of 10 nM for (S)-7-hydroxywarfarin.
+ Open protocol
+ Expand
4

Monosaccharide Composition Analysis by UPLC-ABEE

Check if the same lab product or an alternative is used in the 5 most similar protocols
Monosaccharide composition was analyzed based on the UPLC-ABEE system as previously described [34 ]. AIR powder was hydrolyzed by two-step sulfuric acid hydrolysis, and the hydrolysate was neutralized with calcium carbonate. The neutralized supernatant was labeled with ethyl p-aminobenzoate (ABEE) reagent. Chromatographic separation and detection of labeled monosaccharides were performed using an ACQUITY UPLC H-Class system (Waters Inc.) equipped with an ACQUITY UPLC BEH C18 column (100 × 2.0 mm, 1.7 μm particle size, Waters Inc.) and fluorescence detector (ACQUITY UPLC FLR detector, Waters Inc.). The eluents used for the chromatographic separation were 200 mM potassium borate buffer (pH 8.9) and 100% acetonitrile.
+ Open protocol
+ Expand
5

Glycan Analysis of Sortilin Protein

Check if the same lab product or an alternative is used in the 5 most similar protocols
The glycan analysis was performed as described for the GlycoWorks RapiFluor-MS N-Glycan Kit (Waters Inc.). In short Sortilin was denatured, reduced and treated with PNGase F. The released glycans were labelled with an easily ionized fluorophore and extracted by solid-phase-extraction. Following extraction the labelled glycans were separated by hydrophilic interaction liquid chromatography (HILIC) (nanoAcquity, Waters Inc.) and detected by fluorescence (Acquity UPLC FLR Detector, Waters Inc.) and ESI-MS (Xevo G2-XS QTof, Waters Inc.).
+ Open protocol
+ Expand
6

UPLC-FLR Analysis of Ochratoxin A

Check if the same lab product or an alternative is used in the 5 most similar protocols
OTA was commercially purchased from Sigma Chemical Cp. (St Louis, MO, USA). Standard stock solution (1000 µg/g) was prepared in 1 mL methanol and stored at 8 °C until used. From this, five working stock solutions (2, 4, 6, 8 and 10 µg/g) were prepared in methanol and stored at 4 °C until analysed using UPLC.
UPLC-FLR ANALYSIS UPLC system consisting of Waters Acquity UPLC ® binary pump equipped with Waters Acquity UPLC ® FLR detector was used for quantification of OTA produced by isolates. For detection, the excitation and emission maxima were set at 330 and 460 nm wavelength, respectively. Chromatographic separations were carried out on a C18 reversed-phase column (2.1 × 100 mm, 1.7 µm) (Waters, USA) with isocratic programme of 57% acetonitrile (CH 3 CN), 41% water and 2% acetic acid (filtered through a PTFE syringe filter). The separation was performed at a flow rate of 0.2 mL/min with total run time of 4 min. For OTA quantification, peak height of the sample was compared with the calibration curve of the standards.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!