The largest database of trusted experimental protocols

Zetasizer 3000 hs particle size analyzer

Manufactured by Malvern Panalytical
Sourced in United Kingdom

The Zetasizer 3000 HS is a particle size analyzer that uses dynamic light scattering (DLS) technology to measure the size distribution of particles in a sample. The instrument can measure particle sizes ranging from 0.6 nanometers to 6 micrometers. It provides detailed information about the size and shape of particles in a sample.

Automatically generated - may contain errors

4 protocols using zetasizer 3000 hs particle size analyzer

1

Preparation of BSA Nanoparticles Encapsulating hMDA-7 Plasmid

Check if the same lab product or an alternative is used in the 5 most similar protocols
The pcDNA3.1-hMDA-7 plasmid (a generous gift from Professor Jicheng Yang of the Department of Molecular Biology, Suzhou University) was constructed by inserting the hMDA-7 DNA fragment from the Puc19- hMDA-7 plasmid into pcDNA3.1 through Kpn I and Xba I restriction sites. The recombinant plasmid was transformed into DH5α competent cells by the CaCl2 method.
A total of 0.1 mg pcDNA3.1-hMDA-7 plasmid DNA dissolved in Tris-EDTA buffer was mixed with 3 ml of 2% BSA (Sijiqing Company, China), and water was added to a total mixture volume of 5 mL. Absolute ethyl alcohol (6.5 mL) was gradually added as well as 40% glutaraldehyde, and incubated at room temperature overnight. After centrifugation, the supernatant was discarded and the precipitate dissolved in water after ethanol evaporation. The solution was submitted to high speed centrifugation on an ultracentrifuge from BECKMAN (USA), and BSA nanoparticles obtained in the supernatant were stored at -20°C. Blank BSA nanoparticles (BSA-NP) were prepared as described above but without the hMDA-7 plasmid DNA. BSA-NP morphology and size were examined on a JEM-1200EX transmission electron microscope (HITACH, Japan). Zeta-potential was measured on a Zetasizer 3000HS Particle size analyzer (Malvern Instruments, UK).
+ Open protocol
+ Expand
2

Detailed Materials Characterization Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Transmission electron microscopy (TEM) images were obtained in a JEOL JEM-2100F electron microscope, at an acceleration voltage of 120 kV. Samples for TEM analysis were prepared by adding a single drop (2 µL) of the aqueous solution (ca. 0.1 mg/mL in milliQ water) of particles onto a copper grid coated with a carbon film (Electron Microscopy Sciences). Further details on the preparation of grids can be found in the SI, Section A4.
UV-Vis spectra were measured in an Agilent 8453 UV-Vis diode-array spectrophotometer. Zeta potential measurements were performed in a Malvern Zetasizer 3000 HS particle size analyzer (Malvern Instruments, UK). Fluorescence measurements were carried out with a PerkinElmer LS55 Fluorescence Spectrometer. ICP-MS measurements were performed on a Thermo iCAP Q ICP-MS (Thermo Fisher Scientific GmbH, Bremen, Germany). A ASX560 autosampler was coupled to the ICP-MS (CETAC Tech, Omaha, NE, USA). In order to digest the different materials for ICP experiments, 100 µL of CaF2 particles aqueous solution was digested overnight in 10 mL HNO3 (65%, Scharlau) at 70 oC. Full beads were digested overnight in aqua regia. Spheroids were digested in aqua regia as well, but they required sonication before and after a 24h digestion time.
+ Open protocol
+ Expand
3

Characterization of SERS Nanotags by TEM, Raman, and DLS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Transmission electron microscopy (TEM) images were obtained in a field-emission high-resolution Philips JEOL JEM-2100F electron microscope, at an acceleration voltage of 200 kV. Samples for TEM analysis were prepared by drop-casting the diluted solutions of SERS nanotags on holey carbon-coated Cu mesh grids (400 mesh) and air-dried at room temperature. SERS measurements were performed by means of an InVia Reflex Raman system (Renishaw) comprising a microscope (Leica), a 785 nm laser excitation source (nominal output 260 mW) and a spectrometer equipped with a 1200 grooves/mm diffraction grating and a front-illuminated Peltier-cooled CCD detector (1024 pixels × 512 pixels). Dynamic light scattering (DLS) measurements were performed in a Malvern Zetasizer 3000 HS particle size analyzer (Malvern Instruments, UK). UV-vis absorption spectra were recorded at room temperature, using a Beckman Coulter DU 800 spectrometer.
+ Open protocol
+ Expand
4

Characterization of Gold Nanoparticles

Check if the same lab product or an alternative is used in the 5 most similar protocols
Transmission electron microscopy (TEM) images were obtained in a JEOL JEM-2100F electron microscope, at an acceleration voltage of 200 kV. Samples for TEM analysis were prepared by adding a single drop (2 µL) of the aqueous solution (ca. 0.1 mg/mL in milliQ water) of gold nanoparticles onto a copper grid coated with a carbon film (Electron Microscopy Sciences). The grid was left to dry in air for several hours at room temperature. Zeta potential measurements were performed in a Malvern Zetasizer 3000 HS particle size analyzer (Malvern Instruments, UK). Cell microscopy experiments were carried out using a Zeiss 880 Confocal Fluorescence microscope using Ibidi clear-bottomed 96-well microscopy plates. Milli-Q water (resistivity 18.2 MΩ•cm at 25 °C) was used in all experiments. ICP-MS measurements were performed on a Thermo iCAP Q ICP-MS (Thermo Fisher Scientific GmbH, Bremen, Germany). A ASX-560 autosampler was coupled to the ICP-MS (CETAC Tech, Omaha,NE, USA).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!