The largest database of trusted experimental protocols

5 protocols using empower software version 3

1

UPLC Analysis of Chemical Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
UPLC analysis was performed using a Waters ACQUITY UPLCTM system (Waters Corporation) equipped with a quaternary solvent manager system, autosampler and a PDA detector. The samples were eluted using a Waters ACQUITYUPLC HSS T3 column (100 × 2.1 mm, 1.7 μm). The column and autosampler temperatures were maintained at 40 °C and 25 °C, respectively. A mobile phase containing water (A) and ACN (B) was used with the optimized gradient program as follows: 10%–25% B (0–0.8 min), 25%–35% B (0.8–2.4 min), 35%–95% B (2.4–2.8 min), 95% B (2.8–4 min). The flow rate was set at 0.5 mL/min. The sample injection volume was 1 μL. The wavelength of PDA detector was set to 260 nm. Empower software Version 3.0 (Waters Corporation) was used for system control and data acquisition [32 (link)].
+ Open protocol
+ Expand
2

UPLC Analysis of IGF Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The chemical composition of the IGF was analyzed using UPLC. Crude extracts were dissolved in methanol to produce a final concentration of 0.5 mg/mL. The standards (verproside, catalposide, and amphicoside) were dissolved in methanol to produce a final concentration of 0.1 mg/mL, respectively. Our UPLC-photodiode array (PDA) used a Waters Acquity System (Waters Co., Milford, MA, USA) that consisted of a PDA, binary autosampler, online degasser, and column oven. The UPLC system was fitted with an Acquity UPLC HSS T3 column (100 × 2.1 mm, 1.7 μm; Waters Co.). The column and autosampler temperatures were maintained at 40°C and 25°C, respectively. The mobile phases A and B were ultrapure water and acetonitrile, respectively. The following gradient profile was used: 10–25% B (0–0.8 min), 25–35% B (0.8–2.4 min), 35–95% B (2.4–2.8 min), 95% B (2.8–4 min), and 95–100% B (4–8 min). The flow rate was 0.5 mL/min and the injection volume was 1.0 µL. Compounds were identified by a comparison of the retention times between samples and standards. The PDA detection was conducted at 260 nm. Levels of the IGF were quantitated using Empower software version 3.0 (Waters) [10 (link), 16 (link)].
+ Open protocol
+ Expand
3

HPLC Quantitative Analysis of Marker Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
A Waters Alliance e2695 HPLC system (Waters Corp., Milford, MA, USA) equipped with a pump, degasser, column oven, auto sample injector, and photodiode array (PDA) detector (#2998, Waters Corp. Milford, MA, USA) was used in the quantitative analysis and Empower software (version 3; Waters Corp, Milford, MA, USA) was used to data processing. The chromatographic separation of the seven marker compounds was performed at 30 °C using a Gemini C-18 analytical column (250 × 4.6 mm, 5 μm; Phenomenex, Torrance, CA, USA) with a gradient solvent system of 1.0% (v/v) aqueous acetic acid (A) and acetonitrile (B). The elution conditions were as follows: 12–42% B for 0–25 min, 42–52% B for 25–30 min, 52–65% B for 30–55 min, 65–100% B for 55–56 min, and 100% B for 56–63 min. The flow rate and injection volume were 1.0 mL/min and 10 μL, respectively. The wavelength range of the PDA detector was 190 nm to 400 nm.
+ Open protocol
+ Expand
4

HPLC Analysis of Test Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Chromatographic analyses were performed using an HPLC Alliance system (Waters, Milford, MA, USA) equipped with a gradient quaternary pump, an on-line degasser apparatus, an autosampler and a 2996-photodiode array detector. Data were collected and processed on a computer running with Empower software (version 3) from Waters Corporation (Milford, CT, USA). Separations were carried out with the same stationary and mobile phases described in Section 2.2.3 except for the flow rate fixed at 0.20 mL/min and the 20 μL injection volume. The autosampler temperature was 37 °C. Detection was performed at 254 nm or 360 nm. Serial dilutions of EDA, the test compounds and 5-formyluracil stock solutions (10–100 mM) were prepared using DMSO, prior to dilutions in the mixture of methanol and ammonium formate buffer 5 mM, pH 4.5.
+ Open protocol
+ Expand
5

HPLC Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The samples were analyzed using a HPLC system (Waters 2695 Separations Module, Milford, USA) coupled to a PDA detector (Waters 2998 Photodiode Array Detector, Milford, USA). The analysis was performed on an Alltima C18 column (250 mm × 3 mm; 5 μm particle size) (Grace, Columbia, USA). The mobile phase consisted of a gradient with an ammonium formate buffer (0.020 M) pH = 3 and methanol. First, a ratio of 90 % buffer and 10 % methanol was held for 2 min. During the next 5 min, the ratio changed to 50 % buffer and 50 % methanol. This ratio was kept for 7 min. The next 6 min, the gradient altered to 10 % buffer and 90 % methanol, which was held for 5 min. During the last 5 min, the gradient returned to its starting condition, making a total run of 30 min for each sample. This gradient was run at a flow rate of 0.5 mL min -1 . Five microliters of each sample was injected at a temperature of 15 °C, while the column temperature was set at 30 °C. PDA signals were measured in the range of 210 to 400 nm. Data acquisition was achieved using the Empower software version 3 (Waters, Milford, USA).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!