The largest database of trusted experimental protocols

J w hp innowax gc column

Manufactured by Agilent Technologies
Sourced in United States

The J&W HP-INNOWax GC column is a gas chromatography column designed for a wide range of applications. It features a polyethylene glycol stationary phase that provides excellent separation for polar analytes. The column dimensions and specifications are tailored to meet the needs of various analytical workflows.

Automatically generated - may contain errors

3 protocols using j w hp innowax gc column

1

GC-FID Analysis of Essential Oils

Check if the same lab product or an alternative is used in the 5 most similar protocols
For the quantitative analysis, an Agilent gas chromatograph (GC) (model 6890N series, Agilent Technologies, Santa Clara, CA, USA) equipped with a flame ionization detector (FID) was used. The GC-FID analyses were performed using a nonpolar Agilent J&W DB-5ms Ultra Inert GC column (30 m, 0.25 mm, 0.25 µm) (5%-phenyl-methylpolyxilosane), a polar Agilent J&W HP-INNOWax GC column (30 m, 0.25 mm, 0.25 µm) (polyethylene glycol), and an automatic injector (Agilent 7683 automatic liquid sampler, Agilent Technologies, Santa Clara, CA, USA) in split mode. The samples, 1 µL of solution (1/100, v/v, EO/DCM), were injected with a split ratio of 1:50. Helium was used as a carrier gas at 1 mL/min in constant flow mode and an average speed of 25 cm/s. The initial oven temperature was maintained at 50 °C for 3 min. and then heated to 230 °C with a ramp of 3 °C/min, and the temperature was maintained for 3 min until the end. The injector and detector temperatures were 250 °C. Quantification was done by the external standard method using calibration curves generated by running a GC analysis of representative compounds.
+ Open protocol
+ Expand
2

GC-FID and GC-MS Analysis of Essential Oils

Check if the same lab product or an alternative is used in the 5 most similar protocols
The dentification of the chemical constituents of the essential oil was carried out using an Agilent gas chromatograph (GC) (6890N series, Agilent Technologies, Santa Clara, CA, USA). For the quantitative analysis gas chromatograph was equipped with a flame ionization detector (FID) and for qualitative analysis gas chromatograph was coupled to a mass spectrometer (quadrupole) detector (MS) (model Agilent 5973 inert series, Agilent Technologies, Santa Clara, CA, USA). The GC-FID and GC-MS analyses were performed according to the procedure described by Valarezo et al. [35 (link)]. The injection of the samples was carried out by an automatic injector (Agilent 7683 automatic liquid sampler, Agilent Technologies, Santa Clara, CA, USA) in split mode. Chromatographic runs were performed using a nonpolar and a polar column. The nonpolar was an Agilent J&W DB-5ms Ultra Inert GC column with stationary phase 5%-phenyl-methylpolyxilosane and the polar was an Agilent J&W HP-INNOWax GC column with stationary phase polyeth-ylene glycol. Both columns with a length of 30 m, an outer diameter of 0.25 mm and a stationary phase thickness of 0.25 µm. Identification of the EO compounds was based on a comparison of relative retention indices (RIs) and mass spectra data with those of the published literature [38 ,39 ] according as described by Valarezo et al. [36 (link)].
+ Open protocol
+ Expand
3

Quantitative Analysis of Essential Oils

Check if the same lab product or an alternative is used in the 5 most similar protocols
For quantitative analysis, an Agilent gas chromatograph (GC) (model 6890N series, Agilent Technologies, Santa Clara, CA, USA) equipped with a flame ionization detector (FID) was used. The GC-FID analyses were performed using a nonpolar Agilent J&W DB-5 ms Ultra Inert GC column (30 m, 0.25 mm, 0.25 µm) (5%-phenyl-methylpolyxilosane), a polar Agilent J&W HP-INNOWax GC column (30 m, 0.25 mm, 0.25 µm) (polyethylene glycol), and an automatic injector (Agilent 7683 automatic liquid sampler, Agilent Technologies, Santa Clara, CA, USA) in split mode. The samples, 1 µL of solution (1/100, v/v, EO/DCM), were injected with a split ratio of 1:50. Helium was used as a carrier gas at 1 mL/min in constant flow mode and at an average velocity of 25 cm/s. The initial oven temperature was held at 50 °C for 3 min and then it was heated to 230 °C with a ramp of 3 °C/min, and the temperature was maintained for 3 min until the end. The injector and detector temperatures were 250 °C. Quantification was done by external standard method using calibration curves generated by running GC analysis of representative compounds.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!