The largest database of trusted experimental protocols

Waters 2414 differential refractive index detector

Manufactured by Waters Corporation
Sourced in United States

The Waters 2414 differential refractive index detector is a laboratory instrument used to measure the refractive index of liquid samples. It is designed to detect differences in the refractive index between a sample and a reference liquid, providing information about the composition and concentration of the sample.

Automatically generated - may contain errors

5 protocols using waters 2414 differential refractive index detector

1

Comprehensive Material Characterization Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fluorescence spectra were obtained on a Shimadzu RF-7600 fluorescence spectrometer (Hitachi, Tokyo, Japan). UV-Vis absorption spectra were obtained using a TU-1901 spectrophotometer (Beijing general instrument, Beijing, China). Fourier transform infrared (FT-IR) spectra (4000–400 cm−1) in KBr were recorded in a Vector 22 FT-IR spectrophotometer (Bruker, Karlsruhe, Germany). Morphological characterizations were obtained by scanning electron microscope (SEM) (LEO 1530VP, Carl Zeiss, Oberkochen, Germany) with an attached energy-dispersive X-ray spectroscope (EDS). Thermal characterizations were carried out by STA 449 F5 Jupiter Netzsch thermogravimeter (Netzsch, Selb, Germany). Zeta potentials were measured at room temperature in a neutral water solution with a Zetasizer Nano ZS90 (Malvern, Worcestershire, U.K.). Molecular weights and molecular weight distributions were measured by waters 515 gel permeation chromatography (GPC) equipped with a Waters 2414 differential refractive index detector (set at 35 °C) (Waters, Milford, Massachusetts, USA).
+ Open protocol
+ Expand
2

Polymer Molar Mass Determination by SEC

Check if the same lab product or an alternative is used in the 5 most similar protocols
A Waters size exclusion chromatography (SEC) instrument (Waters Corporation, Milford, MA, USA) was utilized for the determination of the molar mass (Mw), molar mass distributions, and dispersity index (Mw/Mn) of the synthesized P(SMA-co-OEGMA) copolymer. The chromatography system is equipped with a Waters 1515 isocratic pump (Waters Corporation, Milford, MA, USA), a set of three µ-Styragel mixed pore separation columns (pore size 102–106 Å), and a Waters 2414 differential refractive index detector (equilibrated at 40 °C) (Waters Corporation, Milford, MA, USA). The measurements and data analysis were conducted using the Breeze software (Waters Corporation, Milford, MA, USA). Tetrahydrofuran (THF) containing 5% v/v trimethylamine was the mobile phase, at a flow rate of 1 mL/min and temperature set at 30 °C. The calibration curve was set by utilizing linear polystyrene standards with average molecular mass in the range of 1200–152,000 g·mol−1 and narrow molecular mass distributions. The copolymer was dissolved in the mobile phase and measured at concentration of 1 mg mL−1.
+ Open protocol
+ Expand
3

Molecular Weight Determination of Polysaccharides

Check if the same lab product or an alternative is used in the 5 most similar protocols
The molecular weight of polysaccharides was measured by high-performance gel permeation chromatography (HPGPC). The HPGPC system (Shimadzu LC-20A, Shimadzu Instrument Co., Ltd., Tokyo, Japan) was equipped with a TSK-GEL G-3000PWXL column (7.8 mm × 300 mm i.d., 7 μm, Tosoh Corporation, Tokyo, Japan) and a TSK-GEL G-6000 PWXL column (7.8 mm × 300 mm i.d.,13 μm, Tosoh Corporation, Tokyo, Japan) that were linked in series, eluted with 0.02 M potassium dihydrogen phosphate (KH2PO4) at a flow rate of 0.5 mL/min and detected by a Waters 2414 differential refractive index detector (Waters Co. Ltd., Milford, MA, USA). The column temperature was kept at 35 ± 1 °C. The polysaccharides were dissolved in 0.02 M KH2PO4 and filtered through a 0.22 µm filter membrane. An aliquot of 25 μL of the sample was injected into the system. The molecular weight of polysaccharides was determined by the calibration curve made from dextran standards with known molecular weights (4.66, 12.6, 63.3, 126, and 556 kDa).
+ Open protocol
+ Expand
4

Monosaccharide Composition and Relative Mw of Polysaccharides

Check if the same lab product or an alternative is used in the 5 most similar protocols
The precolumn derivatization approach was used to determine the monosaccharide composition of polysaccharides [17 (link)]. The sample was dissolved by adding 3 mg of polysaccharide to 1 mL of distilled water and hydrolyzed with 1 mL (4 mol/L) of trifluoroacetic acid (TFA) at 110 °C for 6 h. The residual TFA was then removed by adding methanol. 1 mL of hydrolysis product was added to distilled water for derivatization. After derivatization, the derivatives filtered through a 0.22-μm membrane were detected by an Agilent Eclipse XBD-C18.
The relative Mw of polysaccharides was measured using a high-performance liquid chromatography system equipped with a Waters-2414 differential refractive index detector (Waters Co., Milford, MA, USA). A G4000SWXL column (7.8 × 300 mm, Tosoh Co., Ltd. Tokyo, Japan) was used. The column temperature was 40 °C, and 0.1 M sodium nitrate solution was used as the eluent. The relative Mw of polysaccharides was calculated according to calibration curves (Log Mol Wt = 1.27e − 8.10e−1 T, R2 = 0.9943) of the dextran standards (Sigma Co., St Louis, MI, USA), including 667,000 Da, 413,000 Da, 76,900 Da, 43,500 Da, 10,500 Da, and 5000 Da.
+ Open protocol
+ Expand
5

Characterization of P(DMAEMA-co-OEGMA) Copolymers

Check if the same lab product or an alternative is used in the 5 most similar protocols
The molar masses (Mw), molar mass distributions, and polydispersity index (Mw/Mn) of the synthesized P(DMAEMA-co-OEGMA) copolymers were determined by SEC, using a Waters instrument (Waters Corporation, Milford, MA, USA). The chromatography system consists of a Waters 1515 isocratic pump (Waters Corporatione, Milford, MA, USA), a set of three m-Styragel mixed pore separation columns (pore size 102–106 Å), and a Waters 2414 differential refractive index detector (equilibrated at 40 °C) (Waters Corporation, Milford, MA, USA). The measurements and data analysis were performed using the Breeze software (version 3.2) (Waters Corporation, Milford, MA, USA). Tetrahydrofuran containing 5% v/v trimethylamine was the mobile phase, at a flow rate of 1 mL/min, and the temperature was set at 30 °C. Linear polystyrene standards with average molecular mass in the range of 1200–929,000 g·mol−1 and narrow molecular mass distributions were utilized for setting up the calibration curve. P(DMAEMA-co-OEGMA) copolymers were soluble in the mobile phase and were measured at concentrations in the range of 2–4 mg mL−1.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!