The largest database of trusted experimental protocols

Acquity uplc protein beh sec column

Manufactured by Waters Corporation
Sourced in United States

The Acquity UPLC Protein BEH SEC column is a size exclusion chromatography column designed for the separation and analysis of proteins and protein complexes. It utilizes Waters' BEH (Ethylene Bridged Hybrid) particle technology to provide high-resolution separation and robust performance.

Automatically generated - may contain errors

15 protocols using acquity uplc protein beh sec column

1

Characterizing PD-L1 Binding Interactions

Check if the same lab product or an alternative is used in the 5 most similar protocols
PD-L1 (aa 19–239) expressed in Escherichia coli (BPS, 100443) was exchanged into assay buffer (20 mmol/L phosphate buffer, 2.7 mmol/L KCl, 137 mmol/L NaCl, 0.05% Tween-20 at pH 7.4) using Zeba Spin Desalting Columns (Thermo Fisher, 7K MWCO, 0.5 mL). A 30-µL sample was made in assay buffer with 15 µmol/L of PD-L1 and 100 µmol/L of INCB086550 or peptide 71. Then 10 µL of the sample was loaded on an SEC multiangle light scattering instrument (Wyatt Technology) at a flow rate of 0.5 mL/minute on an ACQUITY UPLC Protein BEH SEC Column (Waters Corporation; particle size 1.7 µm, pore size 200 Å, diameter 2.1 mm, and length 150 mm). The elution was monitored using ultraviolet/visible trace at 280 nm. Approximate molecular weights of the eluted peaks were calculated using the MALS data (Wyatt QELS DLS, Wyatt Technology) and compared with apo PD-L1 with 2% DMSO.
+ Open protocol
+ Expand
2

Serum Separation Using SEC-ICP-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The separation of human serum was carried out using an Acquity UPLC Protein BEH SEC column (1.7 µm, 4.8×300 mm, 200 Å, Waters, Massachusetts, USA). The HPLC parameters are summarized in Table 3.
For the initial SEC-ICP-MS optimization, the following columns were tested to get an ideal separation efficiency (Table 4).
+ Open protocol
+ Expand
3

Assessing Purity of LR004 and LR004-VC-MMAE

Check if the same lab product or an alternative is used in the 5 most similar protocols
Size‐exclusion chromatography analyses were used to detect purity of LR004 and LR004‐VC‐MMAE. 1200 HPLC (Agilent Technologies); Acquity UPLC protein BEH SEC column (4.6 × 150 mm, particle size 1.7 μm, Aperture 200 Å; Waters); mobile phase 100 mm ammonium acetate (pH = 8); 0.3 mL·min−1 flow rate; column temperature was 30 °C; UV detection wavelength was 280 nm.
+ Open protocol
+ Expand
4

SEC-UPLC Method Development and Optimization

Check if the same lab product or an alternative is used in the 5 most similar protocols
Method development, optimization, and qualification were carried out on Primary Sample 8670 (PS 8670) derived from a single production lot of the NISTmAb. Acquity UPLC Protein BEH SEC Column (1.7 μm particle size, 200 Å pore size, 4.6 mm × 150 mm length) was obtained from Waters Corporation (PN 186005225). Sodium phosphate monobasic (PN 17844-250G), Sodium phosphate dibasic dihydrate (PN 71633-250G), sodium chloride (73575-250G–F) were from Sigma Aldrich. IQ gel filtration standard #1 lot 220,004,819 and #2 lot 64,038,897 were from Bio-Rad (PN 151–1901). 0.22 μm cellulose acetate filters were from Corning (PN 430015). Vial inserts (PN 5182–0549), vials (PN 5182–0716), and vial caps (PN 5182–0717) were from Agilent Technologies.
+ Open protocol
+ Expand
5

SEC Analysis of Antibody Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
SEC analysis was performed according to a previously developed method (Turner et al., 2018 (link)). Briefly, all samples were analyzed on an Agilent high pressure liquid chromatography system using isocratic elution (100 mmol/L sodium phosphate supplemented with 250 mmol/L sodium chloride, pH 6.8) at 0.30 ml/min and monitored at 280 nm. 60 µg of antibody sample was injected onto a Waters Acquity UPLC Protein BEH SEC column (1.7 μm particle size, 200 Å pore size, 4.6 × 150 mm length).
+ Open protocol
+ Expand
6

Plasma Probe B-1 Binding Assay

Check if the same lab product or an alternative is used in the 5 most similar protocols
Plasma (45 μL) was incubated with probe B-1 (20 μM; 3.2 μL of probe B-1 from a 300 μM stock solution in DMSO) or 3.2 μL of DMSO for 24 hours at 37°C. Before injecting onto the size exclusion column (Acquity UPLC protein BEH SEC Column, 200Å, 1.7 μm, 4.6 mm × 150 mm, Waters), each plasma sample was filtered through a P30 gel filtration column (Bio-Spin Columns with Bio-Gel P-30) to remove excess unbound peptide and preserve the lifespan of the SEC column (final volume = 50 μL). 5 μL of each sample was then injected onto the column and separated using a constant flow of 0.2 mL/min of 10 mM phosphate buffer, 1 mM EDTA, 100 mM KCl, pH 7.8, for 30 minutes. To account for plasma auto-fluorescence, the resulting signal from the samples with no probe B-1 (DMSO only) was subtracted from the signal from the corresponding sample with the probe B-1. At the start and the end of each sample set, four injections of different concentrations of fluorescein in the running buffer (freshly prepared each day) were analyzed by the same instrument, in order to make a standard curve for day-to-day comparison. All data were then normalized to the total protein concentration as calculated from the integrated area in the 280 nm absorbance channel. The results of the three groups were compared using a one-way analysis of variance (ANOVA) test.
+ Open protocol
+ Expand
7

Characterization of Humanized Monoclonal Antibody

Check if the same lab product or an alternative is used in the 5 most similar protocols
The sample used was NIST standard reference material 8671 (Lot No. 14HB-D-001), a recombinant humanized IgG1κ monoclonal antibody.
SE-UHPLC was performed with an Acquity UPLC pump and autosampler (Waters Corporation) and Acquity UPLC® Protein BEH SEC column (200 Å, 1.7 µm, 4.6 mm × 150 mm, Waters Corporation). The mobile phase was 100 mM sodium phosphate pH 6.8, supplemented with 250 mM NaCl. Samples were injected onto a pre-conditioned column at a flow rate of 0.300 mL/min. Triplicate injections of 1 µL, 2 µL and 6 µL of samples were made to confirm reproducibility. The effluent of the SEC column flowed through an inline UV/Vis detector (Waters Corporation), the new MALS/DLS detector and an Optilab UT-rEX dRI detector (Wyatt Technology Corporation). For the thermal stress study, triplicate injections of 1 µL of the sample were made. Unstressed sample kept at 4 °C is measured as a control each time measurements are made of the heat-stressed sample.
Data collection and analysis was performed with ASTRA software, version 7.1.1 (Wyatt Technology Corporation). ASTRA software utilizes a patented method14 for the interdetector delay and band broadening correction. First-order fit Zimm formalism is used for all molar mass calculation.
+ Open protocol
+ Expand
8

SEC-nMS Analysis of Protein-DNA Complexes

Check if the same lab product or an alternative is used in the 5 most similar protocols
For SEC-nMS, an ACQUITY UPLC H-class system (Waters, Manchester, UK) comprising a quaternary solvent manager, a sample manager cooled at 10°C, a column oven at ambient temperature, and a TUV detector operating at 280 and 260 nm was coupled to the Synapt G2 HDMS mass spectrometer (Waters, Manchester, UK). All analyses were realized with a baseline response element concentration of 5 μM. Additionally, NR7:RXR:DNA analyses involved 10μM of each protein while single protein:DNA analyses involved 20μM of NR7 or RXR. The different samples were loaded (25 μL) on an ACQUITY UPLC Protein BEH SEC column (4.6 × 150 mm, 1.7 μm particle size, 200 Å pore size from Waters, Manchester, UK) using an isocratic elution with pH 6.8, 150 mM ammonium acetate solvent at a flow rate of 0.25 mL/min over 3 min. The flow rate was then decreased to 0.10 mL/min over 10 min and finally increased to 0.25 mL/min over 5 min. The Synapt G2 HDMS was operated in positive ion mode and instrumental parameters were optimized as follows: capillary voltage, 3 kV; interface pressure, 6 mbar; cone voltage, 180 V; m/z range, 1000–10,000; scan time, 1.5 s. SEC-nMS data interpretations were performed using Mass Lynx V4.1 (Waters, Manchester, UK).
+ Open protocol
+ Expand
9

Characterization of Protein Aggregates

Check if the same lab product or an alternative is used in the 5 most similar protocols
Cell free samples collected from each bioreactor were evaluated for total aggregation by ultra-performance size exclusion chromatography with an ACQUITY UPLC Protein BEH SEC column, 200 Å, 1.7 µm, 4.6 mm × 300 mm (Waters Corp., MA) and an Agilent 1200 HPLC system (Agilent Technologies, CA). Intracellular aggregate levels were determined using western blot analysis under reducing and non-reducing conditions as described below.
+ Open protocol
+ Expand
10

UPLC-SEC Analysis of Oligosaccharides

Check if the same lab product or an alternative is used in the 5 most similar protocols
The oligosaccharides present in CWEs or released upon digestion were separated and analyzed according to (57 (link)). The chromatographic separation was performed on an ACQUITY UPLC Protein BEH SEC Column (125A, 1.7 μm, 4.6 mm × 300 mm; Waters Corporation, Milford, MA, USA) coupled with BEH SEC Guard Column (125A, 1.7 μm, 4.6 mm × 30 mm).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!