The largest database of trusted experimental protocols

67 protocols using dsc25

1

Thermal Analysis of PLGA Nanofibers

Check if the same lab product or an alternative is used in the 5 most similar protocols
The thermal characteristics of pristine PLGA and PLGA nanofibers embedded with ibuprofen were evaluated through differential scanning calorimetry (DSC), employing a DSC25 instrument manufactured by TA Instruments in New Castle, DE, USA. The temperature range employed was 30 to 250 °C, with a heating speed of 10 °C/min applied to the specimens.
+ Open protocol
+ Expand
2

Characterization of SiNPs using FTIR-ATR and DSC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fourier-transform Infrared spectroscopy (FTIR-ATR) analysis using a Thermo Fisher Scientific Nicolet 6700 / Smart iTR device was performed to characterize the SiNPs. A TA Instruments DSC 25 was used to perform Differential Scanning Calorimetry (DSC) on the samples that were positioned within hermetically sealed pans made of aluminum-lead alloy. A scanning rate of 10 °C per minute was used. Zeta potential and particle size of the samples were measured on a ZetaView NTA device.
+ Open protocol
+ Expand
3

Differential Scanning Calorimetry of Compound 1

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 2

To evaluate the physical characteristics of the free base crystalline form of Compound 1, differential scanning calorimetry data was collected. Differential scanning calorimeter model DSC25 from TA Instruments ˜Waters LLC was used. Sample was weighed into a standard aluminum pan and sealed by a standard aluminum lid with pinhole. The measurement was completed by using 10° C./min scanning rate, under a nitrogen purge. The DSC analysis determined that the free base crystalline form exhibits a melting (on set) of ˜213° C. and a endothermic peak at around 216° C. A plot of the DSC thermogram is shown in FIG. 2.

Although the foregoing invention has been described in some detail by way of illustration and example for purposes of clarity of understanding, one of skill in the art will appreciate that certain changes and modifications may be practiced within the scope of the appended claims. In addition, each reference provided herein is incorporated by reference in its entirety to the same extent as if each reference was individually incorporated by reference. Where a conflict exists between the instant application and a reference provided herein, the instant application shall dominate.

+ Open protocol
+ Expand
4

Anti-Freezing Gel Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
The anti-freezing ability of the gel samples was characterized via differential scanning calorimetry (DSC) (DSC 25, TA Instruments) with a refrigerated cooling system (TA Instruments, RCS90), the cooling cycle was executed from 20°C to −70°C at a rate of 2°C min−1. Rheological test was performed on a rotational rheometer (DHR-2, TA Instruments) under a fixed strain of 0.1% and fixed frequency of 1 Hz, this temperature-sweep was cooled from 25°C to −40°C at a rate of 2°C min−1. The storage modulus (G′) and loss modulus (G″) were recorded. Besides, the stretching behavior of different gels at −40°C was tested using a tensile tester (Instron 3369, United States) with a separate frozen chamber.
+ Open protocol
+ Expand
5

Thermal and Decomposition Analysis of Semi-IPNs

Check if the same lab product or an alternative is used in the 5 most similar protocols
The thermal and decomposition (TGA) characteristics of different semi-IPNs were determined on a Setaram Setsys Evolution 16 apparatus by heating the samples at a rate of 10 °C.min−1 from 20 to 600 °C under atmospheric air and normalized to the C=O band at 1732 cm−1.
Differential scanning calorimetry (DSC) measurements were recorded on TA Instruments DSC25. A first heating run from −80 °C to 200 °C with a heating rate of 10 °C/min was performed to determine the melting temperature and the fusion enthalpy. This was followed by a cooling run to −80 °C with a cooling ramp of 200 °C/min. The Tg was obtained in a second heating run from −80 °C to 200 °C at 10 °C/min.
+ Open protocol
+ Expand
6

Thermal Analysis of Sample Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
DSC analysis was performed using a thermal analyzer (DSC25, TA instruments, Inc., New Castle, DE, USA). Prior to each analysis, temperature and heat capacity calibration were performed using high purity indium and aluminum oxide sapphire, respectively, with a temperature range of 0 to 390 °C, a modulation rate of 0.6 °C/min every 40 s, and a scan speed of 5 °C/min. Next, 2–4 mg of sample was weighed and placed in a pre-weighed aluminum hermetic pan, then sealed with an aluminum cover. Closed, empty aluminum pans were used as reference samples. Analysis was carried out by heating samples from 0 to 350 °C at a heating rate of 10 °C/min under a nitrogen purge of 400 mL/min.
+ Open protocol
+ Expand
7

Thermal Analysis of Biopolymer Crystallinity

Check if the same lab product or an alternative is used in the 5 most similar protocols
The thermal properties of the polymers were analysed using differential scanning colorimetry (DSC) using a TA Instruments DSC 25 (New Castle, DE, USA). Approximately 9–11 mg of resin or polymer powder sealed in Tzero Pans with Tzero lid with pin hole (TA instruments, USA) were subjected to a heating-cooling-heating cycle at a rate of 10 °C/min. The measurement for the polymers PCL, PBAT, PBS and PHBH was performed from 0 to 180 °C, starting for 5 min at 0 °C, heating up to 180 °C and kept for 5 min for the first run, cooling down to 0 °C and kept for 5 min and heated up to 180 °C for the second heating run. For the polymers with a higher Tm, PLLA, PHB, PHBV and PLLA/PCL blend, the measurement was conducted from 0 to 220 °C.
To determine the degree of crystallinity of the sample, the following formula was used: Xc = ΔHm0 × 100% [24 ]. Xc is the degree of crystallinity. ΔHm is the enthalpy during the melting. ΔHm0 is the enthalpy for the normalized enthalpy values (J/g) for 100% crystalline polymer, as reported in the literature: for PHB = 146 J/g [25 (link)], PLLA = 93.1 J/g [26 (link)], PBAT = 114 J/g [27 (link)], PBS = 110.3 J/g [28 (link)], PCL = 139.5 J/g [29 (link)] and PHBH = 146 J/g [29 (link)]. PHBV was assumed to have the same enthalpy as PHB [30 (link)].
+ Open protocol
+ Expand
8

Thermal Analysis of Materials by DSC

Check if the same lab product or an alternative is used in the 5 most similar protocols
The calorimetric measurements of the samples were carried out using a differential scanning calorimeter DSC25 (TA Instruments, New Castle, DE, USA) and sealed aluminum Tzero pans. The measurements were carried out in a nitrogen atmosphere at a flow rate of 70 mL/min in the temperature range from −20 to 200 °C with a heating rate of 10 °C/min. The first heating continued up to 190 °C at a rate of 20 °C/min and then the sample was cooled down to −20 °C at a rate of 3 °C/min. The results were processed using the TRIOS v. 5.00.44608 software package (TA Instruments, New Castle, DE, USA).
+ Open protocol
+ Expand
9

Thermal Characterization of Drug-Excipient Interactions

Check if the same lab product or an alternative is used in the 5 most similar protocols
A differential scanning calorimeter (DSC 25, TA instruments, New Castle, DE, USA) was employed to observe the melting and the recrystallization behavior of the drug with the excipients. The samples for the DSC analysis include TA and physical mixture of the lipid phase (with similar ratio as for formulation) melted and solidified. Approx. 5 mg of the samples, individually sealed in the aluminum pans, were placed over the sample platform. The reference pan, empty sealed aluminum pan, was placed on the reference platform. The pans were heated from 25 to 320 °C at the rate of 20 °C/min under nitrogen purge (20 mL/min).
+ Open protocol
+ Expand
10

Thermomolecular Characterization of CNC

Check if the same lab product or an alternative is used in the 5 most similar protocols
The thermomolecular
characteristics were investigated using differential
scanning calorimetry from TA equipment (DSC25). The differential scanning
calorimetric (DSC) analysis of a 5 mg CNC sample was performed at
temperatures ranging from 30 to 350 °C. The heating rate was
10 °C/min in a N2 atmosphere (50 mL/min).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!