The largest database of trusted experimental protocols

1100 series hplc instrument

Manufactured by Agilent Technologies
Sourced in Germany, United States

The Agilent 1100 Series HPLC instrument is a high-performance liquid chromatography system designed for analytical and preparative applications. It features a modular design, allowing for the configuration of various components such as pumps, autosamplers, detectors, and column compartments to meet specific analytical requirements. The 1100 Series HPLC instrument provides accurate and precise separation, detection, and quantification of a wide range of analytes in complex samples.

Automatically generated - may contain errors

17 protocols using 1100 series hplc instrument

1

HPLC-UV Fingerprinting of Pharmaceutical Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
All experiments were performed on an Agilent 1100 Series HPLC instrument (Waldbronn, Germany) equipped with a binary pump (G1312A model), an automatic sample injector (WPALS G1367A model), a diode-array detector (G1315B model), and a PC with the Agilent Chemstation software. HPLC-UV fingerprints were obtained on a Kinetex® C18 reversed-phase column (100 × 4.6 mm i.d., 2.6 µm partially porous particle size) from Phenomenex (Torrance, CA, USA) under different gradient elution conditions given below.
+ Open protocol
+ Expand
2

Spectroscopic Characterization of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
1H NMR 13C NMR, and 19F NMR spectra were recorded at 400, 101, and 376 MHz on a Bruker Avance NEO spectrometer at the University of Kentucky NMR Center. Chemical shifts were internally referenced to solvent signals; (1H NMR DMSO-d6 at δ = 2.50 ppm, 13C NMR DMSO-d6 at δ = 39.52 ppm) and externally referenced to CFC13δ = 0.00 ppm for 19F NMR. Elemental analysis for C and H was performed at Atlantic Microlabs, Inc. (Norcross, GA) with a combustion temperature of 1400 °C.
Liquid chromatography mass spectra (LC-MS) were obtained using an Agilent 1200 HPLC with a a direct flow injection with a HPLC auto sampler without a column (injection volume: 40 μL, flow rate: 0.2 mL/min). ESI positive mode was taken with a source temperature of 120 °C, desolvation temperature of 300 °C, Capillary Vat 3.5 kV while Cone was set at 35.
In addition to spectroscopic characterization, the purity of all compounds was assessed by RP-HPLC using an Agilent Technologies 1100 series HPLC instrument and an Agilent Phase Eclipse Plus C18 column (4.6 mm × 100 mm; 3.5 μm particle size). All compounds were found to be ≥ 97% pure.
+ Open protocol
+ Expand
3

Reversed-Phase Chromatography of Labeled Proteins

Check if the same lab product or an alternative is used in the 5 most similar protocols
Basic reversed-phase chromatography was conducted on Agilent Zorbax Extend-C18 column (2.1 × 150 mm; 5 μm particle size), using an Agilent 1100 Series HPLC instrument.
The labeled protein samples were loaded at a flow rate of 2 μL/min onto an Acclaim PepMap 100 RP-C18 column (100 μm × 2 cm; Thermo Fisher Scientific, Waltham, MA, USA) and then separated using an Acclaim PepMap RSLC RP-C18 column (75 μm × 15 cm; Thermo Fisher Scientific, Waltham, MA, USA). For the analysis of peptide fractions, mobile phase A was 99.9% water + 0.1% formic acid, and mobile phase B was 80% acetonitrile + 19.9% water + 0.1% formic acid. Full MS scans were acquired in the mass range of 350–1500 m/z with a mass resolution of 60,000 and the automatic gain control (AGC) target value was set at 3 × 106. The 20 most intense peaks in MS were fragmented with higher-energy collisional dissociation (HCD) with a collision energy of 32. MS/MS spectra were obtained with a resolution of 15,000 with an AGC target of 2 × 105 and a max injection time of 40 ms. The dynamic exclusion was set for 30 s and run under positive mode.
+ Open protocol
+ Expand
4

HPLC Fingerprinting of Complex Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
Non-targeted HPLC-UV and HPLC-FLD chromatographic fingerprints were obtained with an Agilent 1100 Series HPLC instrument (Waldbronn, Germany) equipped with an automatic injector, a binary pump, and both diode-array and fluorescence detectors (connected in series). Agilent Chemstation software was employed for instrument control and data processing. Reversed-phase separation using a Kinetex® C18 (100 × 4.6 mm i.d., 2.6 µm partially porous particle size) column (Phenomenex, Torrance, CA, USA) was employed. Separation was carried out by gradient elution employing as mobile phase components a 0.1% formic acid aqueous solution and methanol. The gradient elution conditions are shown in Table 1. An injection volume of 5 µL was used. Acquisition was carried out at 280 nm for UV detection and at 280 nm (excitation) and 350 nm (emission) for FLD detection.
+ Open protocol
+ Expand
5

Optimized HPLC-UV Fingerprinting of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
An Agilent 1100 Series HPLC instrument equipped with a quaternary pump (G1311A), a degasser (G1322A), an autosampler (G1329A), a diode array detector (G1315B), and a PC with the Agilent Chemstation software, all of them from Agilent Technologies (Waldbronn, Germany), was employed. HPLC-UV fingerprints were obtained by reversed-phase mode using a Kinetex C18 porous-shell column (100 mm × 4.6 mm I.D., 2.6 µm particle size) from Phenomenex (Torrance, CA, USA) at room temperature. Chromatographic separation was performed under gradient elution mode, using 0.1% (v/v) formic acid aqueous solution (solvent A) and methanol (solvent B) as mobile phase components, following the next elution program: 0–20 min, linear gradient from 15% to 95% solvent B; 20–30 min, isocratic elution at 95% solvent B; 30–30.1 min, back to initial conditions; and from 30.1–35 min, at 15% solvent B for column re-equilibration. The mobile phase flow rate was 0.4 mL/min, and the injection volume was 5 µL. The HPLC-UV fingerprints were registered at 250 nm.
+ Open protocol
+ Expand
6

Polyphenol Analysis in Leaf Extracts

Check if the same lab product or an alternative is used in the 5 most similar protocols
The polyphenols in different leaves extracts were evaluated by chromatographic technique, using the earlier reported method [32 ]. The chromatographic system consists of an Agilent 1100 series HPLC instrument (Santa Clara, USA) equipped with an MS detector. Analytical separation was carried out in a C18 column (4.6 mm × 100 mm × 5 μm, Agilent Technology) with a flow rate of 0.8/min with two mobile solvent phases (eluent A = 10 mM ammonium acetate and 1% acetic acid in water; eluent B = 1% acetic acid in methanol). The gradient elution was performed as follows: 0.3 min, 15–50% A; 3–5.5 min, 50–90% A; 5.5–9 min, 90% A; 9–9.5 min, 90–15% A, 9.5–10 min, 15% A. The sample injection volume was 20 μL and the temperature of the column was fixed at 40 °C. Compounds were identified with the mass spectra and Rt with the NIST library of standard compounds.
+ Open protocol
+ Expand
7

Gram-scale Peptide Synthesis and Purification

Check if the same lab product or an alternative is used in the 5 most similar protocols
Peptides were synthesized with a CEM LibertyBlue solid phase peptide synthesizer using standard Fmoc chemistry. The sequences used are shown in Table 1. All peptides were C-terminal amidated and N-terminal acetylated. The crude peptides were cleaved with 2.5% each of H2O, Triisopropylsilane (TIS), and 3,6-dioxa-1,8-octanedithiol (DoDT) and 92.5% Trifluoroacetic acid (TFA) (10 mL total for 0.1mM scale synthesis) for 30 min at 37°C. The cleaved peptides were crashed out in cold (−20°C) ether, centrifuged, ether decanted, and left to dry overnight. The resulting crude peptide pellets were dissolved in Milli-Q water at a concentration of ~1 mg/mL, pH adjusted to 7, and dialyzed with Spectra/Por S/P 7 RC dialysis tubing at a 2000 Da cutoff tubing against DI water for 3 days. The purified peptides were frozen and lyophilized for at least 3 days, which resulted in a white cotton-like powder. Peptide mass and purity were verified >85% by an Agilent 1100 series HPLC instrument with an Agilent (Santa Clara, CA) C3 reverse phase column. The molecular weights of the peptides were verified with an Orbitrap Q Exactive LC/MS (Thermo Scientific, Waltham, MA) instrument. To verify commercial manufacturing potential, the peptides were manufactured at gram-scale with similar purity (>90%) by AmbioPharm Inc (Beech Island, SC), a manufacturer of peptide APIs for the US market.
+ Open protocol
+ Expand
8

HPLC Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC analyses were performed on an Agilent 1100 series HPLC instrument, equipped with a quaternary pump, an autosampler, a column oven, and a photodiode array detector (Agilent, Waldbronn, Germany). Optimum separation was achieved on a YMS-Pack Pro C18 RS (150 × 4.6 mm, 3 µm) column from YMC, guarded with an in-line filter and using a mobile phase consisting of 0.9% (v/v) formic acid and 0.1% (v/v) acetic acid in water (A) and methanol (B). The applied gradient was as follows: 0% B at 0 min, 0% B at 15 min, 10% B at 23 min, 15% B at 30 min, and 98% B at 35 min, and held at this composition for 5 min (total runtime of 40 min); for the next 15 min the column was equilibrated under initial conditions. Flow rate, temperature, and injection volume were adjusted to 0.55 mL· min−1, 25 °C, and 4.0 μL. The detection wavelength was set to 330 nm.
+ Open protocol
+ Expand
9

Quantitative Analysis of Sagunja-tang

Check if the same lab product or an alternative is used in the 5 most similar protocols
For quantitative analysis, five of the reference compounds solutions, glycyrrhizin and liquiritin (1,000 μg/mL), were prepared in 100% methanol and stored at 4°C. The standard solutions were prepared by six concentrations of diluted solutions (methanol). All calibration curves were attained by assessing the peak areas at six concentrations in the range of 10–500 μg/mL for all reference compounds. The linearity of the peak area (y) versus concentration (x, μg/mL) curve for each component was used to calculate the contents of the main Sagunja-tang. 102.8 mg of Sagunja-tang extract powder was subsequently resuspended in 10 mL distilled water for high-performance liquid chromatography (HPLC) analysis.
The contents of glycyrrhizin and liquiritin in the SGJT water extract were analyzed using a 1100 series HPLC instrument (Agilent Technologies, USA) with a Gemini C18 column (4.6 × 250 mm, 5 μm; Phenomenex, USA). The mobile phase consisted of the solvents, distilled water (A) and acetonitrile with 0.1% formic acid (B). The following gradient was used: 0 min, A : B 80 : 20 (v/v); 10 min, A : B 60 : 40; 15 min, A : B 40 : 60; 30 min, A : B 15 : 85; and 40 min, A : B 0 : 100. The mobile phase flow rate was 1.0 mL/min, the column temperature was 30°C, the injection volume was 10 uL, and UV detection was at 254 nm (glycyrrhizin) and 280 nm (liquiritin).
+ Open protocol
+ Expand
10

Quantitative HPLC Analysis of Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
Sample reactions were analyzed to quantitate metabolites by absorbance. Analytes were separated on a Waters XBridge BEH C18 3.5 μM column (4.6 mm × 100 mm) using an Agilent 1100 series HPLC instrument equipped with a VWD detector (280 nm). The mobile phase consisted of solvent A (10% solvent B, 90% water) and solvent B (0.1% formic acid/acetonitrile). A gradient method started at 80% solvent A for the first minute, then decreased to 70% for the next 6 min. The method held 70% A for 3 min before decreasing to 15% A over 11 min, then returning to 80% over 3 min and holding for the final 2 min. The flow rate was 1.2 mL/min for a total run time of 25 min. Parent drug and metabolite responses were normalized to the internal standard acenocoumarol and quantitated using a standard curve generated with 5OH-APINACA. The absorbance response corresponds to the indazole ring, which remained unmodified in these reactions [28 (link)], so that the relative absorbance response for all analytes were approximately the same, making inference for quantitation purposes possible.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!