The largest database of trusted experimental protocols

173 protocols using aquilion one

1

Contrast-Enhanced CT Imaging Protocols

Check if the same lab product or an alternative is used in the 5 most similar protocols
The CT examinations were performed either on a 320-row multidetector computed tomography (MDCT) scanner (Aquilion ONE, Canon Medical Systems, Tochigi, Japan) or a 192 × 2 slice dual-source CT (DSCT) scanner (Somatom Force, Siemens Medical Solutions, Forchheim, Germany). Tube voltage was set at 80–120 kV and tube current at 40–100 mA according to the patients’ age. Automated tube-current modulation was used with the dual-source CT. The contrast agent used was Omnipaque 300 mg/mL (GE Healthcare, Marlborough, Massachusetts, USA).
For non-gated CT chest, abdomen and pelvis with contrast, 1.5–2 mL/Kg of contrast was administered with an injection rate of 0.5–1.5 mL/sec depending on catheter size and patient size. Scan coverage was from just above the lung apices to just below the symphysis pubis.
For chest CT for suspected pulmonary embolism, 2–2.5 mL/Kg of contrast is administered with an injection rate of 2.5–5 mL/sec depending on catheter size and patient size. Scan coverage was from just above the lung apices to just below the diaphragm. On the Canon Aquilion ONE scanner, a volumetric gated scan mode was used for z-axis coverage of less than 16 cm and a helical protocol was used for a z-axis coverage greater than 16 m. Turbo Flash mode was used with the Siemens Somatom Force dual-source scanner.
+ Open protocol
+ Expand
2

Multidetector CT Scanner Calibration Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
CT was performed using seven multidetector CT scanners (Siemens Somatom Definition Edge, Definition Flash, and Force, Siemens Healthineers, Erlangen, Germany; Philips Brilliance 64 and Brilliance 16, Philips Healthcare, Best, Netherlands; Canon Aquilion One, Canon Medical Systems, Otawara, Japan; GE Revolution EVO, Madison, WI, GE Healthcare). CT was performed using a peak tube voltage of 120 kVp, a slice thickness of 2.0–5.0 mm, and adaptive tube current in helical mode. CT scans were performed using a bone kernel or a soft tissue kernel depending on the imaging purpose. All CT machines were calibrated daily with external air. The HU value of air should remain relatively constant, but it can vary slightly depending on factors such as temperature and humidity. Therefore, a calibration process is performed by conducting a scan using only air, without any specific object or phantom inside the CT bore, to adjust the HU value. Two-dimensional reconstructions were acquired in the coronal and/or sagittal planes with a bone or soft tissue kernel and a thickness of 2.0–5.0 mm.
+ Open protocol
+ Expand
3

Whole-Body Phantom CT Scanning Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
The whole-body phantom was scanned with two CT units (a Canon Aquilion One (Canon, Tokyo, Japan) with a frontal plus a lateral localizer radiograph and a Siemens Definition AS (Siemens, Munich, Germany) with a frontal localizer radiograph) using clinical routine CT protocols of the neck to abdomen/pelvis. The use of two localizer radiographs in Canon CT systems enables angular modulation based on attenuation information [17 (link)], while in Siemens’ CareDose 4D (Siemens, Munich, Germany), the angular modulation (the attenuation in the Lat view) is estimated from a single AP radiograph [18 (link)]. The exam protocols consist of three partially-overlapping helical scans, one mainly of the neck, one mainly of the thorax, and a final one extending from the lower parts of thorax to the pelvis (‘abdomen’). Each of these is performed with dose modulation. Three complete CT examinations were performed with the protocol having the arms/hands positioned differently during localizer radiograph imaging: (i) above the head, (ii) alongside the trunk, and (iii) alongside the trunk with hands on the abdomen. In subsequent CT scans, the arms were not included, but the patient would be asked to raise their arms during the CT scan of the chest to pelvis and to lower their arms during the CT scan of the neck.
+ Open protocol
+ Expand
4

Low-Dose Dual-Energy CT Imaging of Wrists

Check if the same lab product or an alternative is used in the 5 most similar protocols
All patients underwent a low-dose dual-energy CT scan of both wrists (one patient had only one wrist examined) in a 320-row CT scanner (Canon Aquilion One and Canon Aquilion One Vision, Canon Medical Systems, Otawara, Japan), using a standardized scan protocol. Tube settings were as follows: 80 kVp with 90 to 170 mA and 135 kVp with 15 to 30 mA. We used the volume mode with a z-axis coverage of 16 cm without table movement and with the fastest tube rotation time. To establish a conventional CT image visualization, virtual blended 120 kVp equivalent images were computed as well as collagen maps using a vendor software (Canon Medical Systems, Otawara, Japan) on the CT console and applying a dual-energy gradient of 1.1 for collagen on the three-material decomposition software. Readers were blinded to all patient data, which was ensured by pseudonymization with dedicated software (Horos v.2.2.0, The Horos Project).
+ Open protocol
+ Expand
5

3D Knee Joint Morphometry Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Vernier calipers were used to measure the following indicators on the 3D printed model of knee joint: anterior and posterior diameter and the left and right diameter of the internal and external condyles of the femur (APDLF: Anterior and posterior diameter of lateral femoral condyle; APDFC: Anterior and posterior diameter of the femoral medial condyle; LRDMLFC: Left and right diameters of medial and lateral femoral condyles). Left and right diameters and anteroposterior diameters of tibial plateau (LRTP: Left and right diameters of tibial plateau; APTP: anterior and posterior diameters of tibial plateau). The upper and lower meridian and left and right diameters of the patella (ULP: Upper and lower meridian of the patella, LRDP: Left and right diameters of the patella). The anteroposterior diameter, left and right diameter of the femur, the anteroposterior diameter of the tibial plateau, the left and right diameter, the distance from the intercondylar ridge to the tibial tuberosity, the lower femur angle, and the upper tibial angle were measured on the CT images of the knee joint (Canon Workstation, Canon Aquilion One, Japan). The above parameters play an important role in the selection of the internal fixation model. Three measurements were taken, and the average value was used for analysis.
+ Open protocol
+ Expand
6

Liver Metastasis Perfusion Imaging

Check if the same lab product or an alternative is used in the 5 most similar protocols
All patients received CTP of the liver one day before initial planned liver surgery with the use of a 320-slice CT scanner (Aquilion ONE, Canon Medical Systems). Dynamic imaging was only performed within the 16 cm detector range of the scanner. A non-contrast helical CT was performed of all patients in order to ensure the coverage of the metastases within the 16 cm scan range for further dynamic imaging. A scanning protocol was chosen combined with a diagnostic venous phase liver CT and perfusion imaging in one session, e.g., diagnostic images were acquired between the dynamic time points. The CT had an X-Ray tube voltage of 100 kV. The milliampere-seconds was set at 50 mAs or Automatic Exposure Control for respectively the dynamic and diagnostic imaging. Rotation time was 0.5 seconds with a pitch of 0. Patients received 150 ml of iodine contrast at a concentration of 300 mg/ml at an injection rate of 5 ml/s (Iomeron® Bracco UK Ltd.) followed by a saline flush (Mallinckrodt /Covidien OptiVantage Power Injector). Imaging was performed during 110 seconds with 22 time points starting 7 seconds after contrast injection. Median DLP was 775 mGy.cm (range 347–1367).
+ Open protocol
+ Expand
7

CT Imaging Protocol for Abdomen Scanning

Check if the same lab product or an alternative is used in the 5 most similar protocols
All patients underwent CT examination with Aquilion ONE (Canon Medical Systems, Otawara, Japan). CT scanning parameters were as follows: tube voltage of 120 kVp; tube current of automatic tube current modulation was used with standard deviation (SD) set at 13.0, the helical pitch of 0.8125:1, and gantry rotation time of 0.5 s. Iodinated contrast media at 600 mgI/kg was intravenously injected using a power injector, and the portal venous phase was obtained 70-90 s after starting the injection. Raw CT data were processed using two different reconstruction algorithms: FBP with the reconstruction kernel of FC03 and DLR (advanced intelligent clear-IQ Engine with body sharp standard). The following image reconstruction parameters were same for FBP and DLR: field of view of 350 mm (adjusted to body size) and slice thickness/interval of 3/3 mm. CT images were anonymized and exported from the PACS in Digital Imaging and Communications in Medicine format.
+ Open protocol
+ Expand
8

Evaluating Radiographic Findings for Tuberculosis

Check if the same lab product or an alternative is used in the 5 most similar protocols
A 320-detector row CT scanner (Aquilion ONE, Canon Medical Systems, Tochigi, Japan) was used. Scans were performed using 1.0-mm thick sections of contiguous images from the apex to the lung base. Images were obtained at a window setting of –600 (level) and 1500 (width). If the patient underwent a CT scan before referral to our hospital, the CT scan features were evaluated using the images obtained from the referring institutions. The finding of calcification on chest X-ray (lung field or pleura) or chest CT (lung field, mediastinal lymph node, or pleura) were determined at the mediastinal window. Fibrotic lesions were not assessed in this study because fibrotic change is more frequently observed on chest CT than on chest X-ray in various types of lung diseases including interstitial pneumonia, smoking-related lung diseases, and age-related interstitial changes, and this would be less specific for previous TB infection in elderly people. Chest images were independently evaluated by respiratory medicine specialists (IY and KK) who were blinded to the clinical information. Any disagreement between the presence of these findings in each case was resolved by a review conducted by the same two physicians in order to reach a consensus.
+ Open protocol
+ Expand
9

Comparative CT Imaging of Iodinated Agents

Check if the same lab product or an alternative is used in the 5 most similar protocols
CT images of iomeprol (400 mgI/mL), iopamidol (300 mgI/mL; Iopamiron 300, Bayer, Osaka, Japan) and IPL (220.7 mgI/mL) were obtained using a cylindrical water-filled phantom. Three test tubes (inner diameter, 8 mm) filled with each of iomeprol (400 mgI/mL), iopamidol (300 mgI/mL), and IPL were placed in the phantom and scanned by a 320 area detector CT scanner (Aquilion ONE, Canon Medical Systems, Japan) at applied voltage of each of 100 kVp and 120 kVp. The imaging conditions were as follows: tube current, 50 mA; rotation time, 1.0 s; pitch factor, 0.813; and 80 × 0.5 mm detectors. The obtained filtered back-projection images were reconstructed with a display field of view of 100 mm, nominal slice thickness of 5 mm, and abdominal standard reconstruction kernel FC13. CT attenuation values were measured and compared among the iodinated solutions.
+ Open protocol
+ Expand
10

Multidetector CT Scanner Protocols

Check if the same lab product or an alternative is used in the 5 most similar protocols
Computed tomography images were acquired by a 320-row multidetector row volume CT scanner (Aquilion ONE; Canon Medical Systems, Otawara, Japan), a third-generation dual-source CT scanner (SOMATOM Force; Siemens Healthineers, Erlangen, Germany), and a 256-slice multidetector row CT scanner (Brilliance iCT; Philips Healthcare, Best, the Netherlands). The scan parameters and reconstruction technique from each CT scanner are shown in Table 3.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!