The largest database of trusted experimental protocols

Nicolet 6700 fourier transform infrared spectrometer

Manufactured by Thermo Fisher Scientific
Sourced in United States

The Nicolet 6700 Fourier transform infrared (FTIR) spectrometer is a laboratory instrument designed for infrared spectroscopy analysis. It measures the absorption and transmission of infrared radiation by a sample, producing a spectrum that can be used to identify and quantify the chemical compounds present.

Automatically generated - may contain errors

20 protocols using nicolet 6700 fourier transform infrared spectrometer

1

Comprehensive Characterization of Material Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
The Scanning electron microscope (SEM) and Transmission electron microscope (TEM) studies were carried out in a Zeiss Sigma 300 and a FEI TF20, respectively. The nitrogen adsorption-desorption isotherm (BET) was measured on the ASAP 2460 surface area and porosity analyzer. Thermal gravimetric analyses (TGA) was performed on a TGA 5500 thermoanalyzer, the sample was heated in a continuous-flow of N2 from 25 up to 800 °C with 10 °C/min. X-ray diffraction (XRD) analysis was performed by Bruker D8A X-ray diffractometer using Cu-Kα radiation (tube voltage: 40 kV; current: 40 mA; scan angle 2θ range: 10–80°; scan rate: 5°/min. Fourier transform infrared spectroscopy (FT-IR) was performed on a Thermo Scientific Nicolet 6700 Fourier transform infrared spectrometer, and the scanning range was 4000 cm−1–400 cm−1.
+ Open protocol
+ Expand
2

Infrared Spectroscopy of Clay Minerals

Check if the same lab product or an alternative is used in the 5 most similar protocols
Nicolet 6700 Fourier transform infrared spectrometer (Thermo Fisher Scientific, Waltham, MA, USA). was used to measure the infrared spectra of the clays The spectra were then recorded in a wavenumber range from 400 to 4000 cm−1 with 32 scans per spectrum and a resolution of 4 cm−1 [2 (link)].
+ Open protocol
+ Expand
3

Comprehensive Characterization of Nanomaterials

Check if the same lab product or an alternative is used in the 5 most similar protocols
Transmission electron microscope (TEM) images were taken on a transmission electron microscope (Tecnai G2 20 TWIN, FEI, USA) operated at an acceleration of 10 kV by dropping solution onto a carbon-coated copper grid. Dynamic light scattering (DLS) and zeta potential were measured at 25°C on a Zetasizer Nano ZS90 analyzer (Malvern Instrument Ltd, UK). Brunauer-Emmett-Teller (BET) and Barrett-Joyner-Halenda (BJH) methods were used to test the surface area, the pore size distribution, and the pore volume. Fourier transform infrared (FT-IR) spectra were conducted on Thermofisher Nicolet 6700 fourier transform infrared spectrometer with KBr pellets. Raman spectra were performed on XploRA laser Raman spectrometer (HORIBA JobinYvon, France) with excitation of 532 nm. Confocal laser scanning microscopy (CLSM) images were performed on Nikon C2+ laser scanning confocal microscope (Nikon, Japan). Photothermal effect was tested on an 808 nm consecutive NIR laser with spot size of 5 mm×6 mm (Changchun New Industries Optoelectronics Technology Ltd., People’s Republic of China). Thermal images were taken on an infrared camera thermographic system (Infra Tec, VarioCAM research)
+ Open protocol
+ Expand
4

Multi-Analytical Characterization of Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
Raman spectra were recorded on a SENTERRA
Raman microscope (Bruker) system with a 532 nm wavelength laser. Infrared
(IR) spectra were obtained with a Nicolet 6700 Fourier transform infrared
spectrometer (Thermo Scientific). Scanning electron microscopy (SEM)
images were acquired on a field-emission SEM (JSM 7401F; JEOL) equipped
with an Oxford INCA energy dispersive X-ray spectroscopy (EDX) system.
Transmission electron microscopy (TEM) images were obtained on a Philips
EM400T microscope operated at 120 kV. X-ray powder diffraction (XRD)
patterns were obtained on a Scintag PAD X X-ray diffractometer using
Cu Kα radiation. X-ray photoelectron spectroscopy (XPS) measurements
were carried out on a Vacuum Generators Escalab MK II X-ray photoelectron
spectrometer. All XPS spectra were referenced to the binding energy
of the C 1s of adventitious carbon at 285.0 eV. Elemental analysis
was conducted by acid digestion of solid samples, followed by quantification
using a Varian Vista AX inductively coupled plasma atomic emission
spectrometer. The optical absorption properties of the materials were
obtained on a Cary 50 Bio spectrophotometer. A Barrelino diffuse reflectance
probe was used to collect UV–visible spectra, using BaSO4 as a standard.
+ Open protocol
+ Expand
5

Accelerated Aluminum Corrosion Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The electrochemical corrosion rate of aluminum in deionized water was slow. To more quickly obtain the corrosion surface and corrosion products of aluminum, the samples for scanning electron microscopy (SEM), energy dispersive spectroscopy (EDS), Fourier transform infrared (FTIR) and powder X-ray diffraction (XRD) were subjected to accelerated corrosion by potentiostatic anodic oxidation at 0.5 V (vs. Pt) for 0.5 h [17 (link),18 (link)]. Large area platinum electrodes were used as the cathodes, 1 cm × 1 cm aluminum foils after polishing and washing were used as the anode, and the flowing deionized water with 3 m s-1 was used as the electrolyte at 10, 20, 30, 40, and 50 °C.
The corrosion product compositions were determined using a D8-Focus X-ray powder diffraction instrument with a Cu target (Bruker, Karlsruhe, Germany). The scanning angle range was from 5 to 80 degrees, and the scan rate was 8° min−1. The chemical bonds of the corrosion products were characterized using a Nicolet 6700 Fourier transform infrared spectrometer (Thermo Fisher, Waltham, USA). SEM images were obtained using an SU8010 ultrahigh-resolution field emission scanning electron microscope capable of high-performance X-ray energy dispersive spectroscopy (Hitachi, Tokyo, Japan).
+ Open protocol
+ Expand
6

Comprehensive Asphaltene Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
Asphaltenes were separated according to ASTM D6560-2000 [40 ]. The elemental composition of the asphaltene, including carbon, hydrogen, sulfur, and nitrogen, was determined using a Vario Micro Cube Elemental Analyzer (Elementar Company, Langenselbold, Germany). The oxygen content was obtained using the subtractive difference method. X-ray photoelectron spectroscopy of the asphaltene was performed using an ESCALAB 250Xi spectrometer (Thermo Fischer Scientific, Waltham, MA, USA). The mean molecular weight and distribution of the asphaltenes were determined using tetrahydrofuran (THF) as the mobile phase on a Waters 1515 gel permeation chromatograph (WATERS, Milford, MA, USA). The 1H NMR and 13C NMR spectra of the asphaltenes were obtained using an Ascend 400MHz nuclear magnetic resonance (NMR) spectrometer (BRUKER, Berlin, Germany). Deuterated chloroform (CDCl3) was used as the solvent, and tetramethylsilane (TMS) served as the internal standard. Fourier-transform infrared spectroscopy (FT-IR) analysis of the asphaltene was performed using a Nicolet 6700 Fourier transform infrared spectrometer (Thermo Fisher Scientific, Waltham, MA, USA).
+ Open protocol
+ Expand
7

FT-IR Analysis of Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
FT-IR analysis was performed by using a ThermoScientific Nicolet 6700 Fourier Transform Infrared Spectrometer (Thermo Fisher Scientific Inc., Waltham, MA, USA).
+ Open protocol
+ Expand
8

Characterization of Hydrogel Morphology

Check if the same lab product or an alternative is used in the 5 most similar protocols
The P(AA-co-MAA) and CTS-g-P(AA-co-MAA) hydrogels
prepared were characterized by recording FTIR spectra
on a Nicolet 6700 Fourier transform infrared spectrometer (Thermo
Fisher Scientific Company, USA). The samples were dried prior to measurement.
The surfaces of the samples were sprayed with gold, and the morphology
of the samples were observed under a Quanta 250 scanning electron
microscope (FEI Company, USA).
+ Open protocol
+ Expand
9

Gamma-Irradiated Polysaccharide FTIR Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
In order to investigate the changes in chemical groups of polysaccharides after gamma irradiation, FTIR spectra of APP were analyzed using a Nicolet 6700 Fourier-transform infrared Spectrometer (Thermo, Waltham, MA, USA). Here, 2 mg samples of native and irradiated polysaccharides were mixed with potassium bromide (KBr) and compacted into pellets for spectrometric measurement. The infrared spectra were collected at wavelengths ranging from 4000 cm−1 to 650 cm−1.
+ Open protocol
+ Expand
10

Characterization of Templated Silica Films

Check if the same lab product or an alternative is used in the 5 most similar protocols
Focused ion beam and scanning electron microscopy (FIB/SEM) experiments were carried out on a FEI Q3D dual beam FIB/SEM system, with 30 kV/3 nA initial voltage/current followed by 8 kV/25 pA final polishing voltage/current for ion beam mode, and 5 kV/24 pA for scanning electron microscopy mode. Transmission electron microscopy images were acquired using a JEOL2010F HRTEM, and Ti-mapping was acquired using the same TEM with a Gatan EELS system. GISAXS was performed using a Bruker Nanostar on samples prepared on Anodisc substrates fabricated as indicated above or on Si substrates prepared as described for Fourier-transform infrared analysis (vide infra). Quartz crystal microbalance analyses were performed using a QCM200-5MHz QCM manufactured by Stanford Research Systems. A home-built, air-tight environmental chamber equipped with gas flow controllers to perform the H2O isotherms. Fourier-transform infrared spectroscopy was performed using a Thermo Scientific Nicolet 6700 Fourier-transform infrared spectrometer. A P123-templated silica film was deposited onto intrinsic, IR transparent single crystal Si substrates (400-µm-thick, double-polished) by spin-coating; this film was then processed in an identical manner as the Anodisc supported P123-templated film described above and loaded with CA.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!