The largest database of trusted experimental protocols

Agilent 1260 series hplc system

Manufactured by Agilent Technologies
Sourced in United States, Germany, Canada

The Agilent 1260 series HPLC system is a high-performance liquid chromatography system designed for analytical and preparative applications. It features a modular design with interchangeable components, including pumps, autosampler, column compartment, and detectors. The system is capable of performing a range of chromatographic separations and analyses.

Automatically generated - may contain errors

82 protocols using agilent 1260 series hplc system

1

Isolation and Purification of Bioactive Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Ea73 strain was activated and fermented under the optimum fermentation conditions. After the fermentation, the broth was broken by ultrasonic wave (40 Hz, 20 min), and the organic phase was collected by multiple extraction with ethyl acetate, and concentrated to dry at 45°C with rotary evaporator. The extractum was subjected to silica gel column chromatography with stepwise elution of chloroform:methanol (30:1, 20:1, 10:1, 6:1, 1:1; v/v). All eluted fractions were separately collected and concentrated using an evaporator. The fraction (20:1) was separated by gel filtration on Sephadex LH-20 column using running phase of 100% methanol. The eluted components were further purified by prep-high-performance liquid chromatography (HPLC) (Agilent 1260 series HPLC system):preparative reversed-phase column (10 μm, 250 mm × 20 mm), at a flow rate of 10 ml/min, methanol:H2O (25: 75, v/v), and UV detection at 210 nm. The purity of the separated compounds was further detected by HPLC (Agilent 1260 series HPLC system):C18 column (5 μm, 4.6 × 150 mm) at a flow rate of 1.0 ml/min, methanol-water (20–100% methanol in water over 8.0 min followed by 100% methanol to 13.0 min), and UV detection at 210 nm.
+ Open protocol
+ Expand
2

Quantification of Ginsenosides in Panax Notoginseng

Check if the same lab product or an alternative is used in the 5 most similar protocols
PNS samples (S27243) and standard solutions of notoginsenoside R1 (B21099) and ginsenoside (Rg1, B21057; Re, B21055; Rb1, B21050; Rd, B21054) were purchased from Yuanye Bio-Technology Co. Ltd. (Shanghai, China). A 25 mg sample of PNS was placed in a 10 mL volumetric flask. Approximately 8 mL of 70% methanol was added, followed by ultrasonic treatment (640 W, 40 kHz) for 30 min. Then diluted to the mark with 70% methanol and mixed thoroughly.
HPLC were performed on the Agilent HPLC 1260 series system (Agilent, CA, USA). Samples were separated using C18 reversed phase Eclipse XDB column (5 μm, 4.6 mm × 250 mm). A flow rate of 1.3 mL/min and sample injection volume of 10 μL were employed. The detection wavelength was set at 203 nm, and the column temperature was maintained at 25 °C. The mobile phase comprised acetonitrile (A) and water (B), with a linear gradient established as follows: 0–20 min for 20% A, 20–55 min for 20% A to 46% A, and 55–60 min for 46% A.
+ Open protocol
+ Expand
3

Quantification of Polyphyllins in Herbs

Check if the same lab product or an alternative is used in the 5 most similar protocols
After cleaning with ultrapure water, the tissues prepared for phytochemical investigation were separately dried to achieve constant weight in the oven at 40℃ and they then were grinded into powder. The extraction and quantification of polyphyllins were performed as previously reported [24 ]. In detail, 0.5 g of the powder sample was exhaustively mixed with 70% EtOH (25 mL). The mixture was heated under reflux 30 min and added 70% EtOH for the weight loss after cool down. The supernatant was then filtered to get sample solution after centrifugation. Quantitative analysis of polyphyllins (Polyphyllin I, II, VI, and VII) were performed on an Agilent HPLC 1260 series system (Agilent, USA) and the retention time of the four types of polyphyllins were compared to that of the relative standard substances. The chromatographic separation was performed at temperature 30℃ using acetonitrile (A) and water (B) with gradient elution procedure (0–40 min, linear gradient 30–60% A, linear gradient 70–40% B; 40–50 min, linear gradient 60–30% A, linear gradient 40–70% B) as the mobile phase at a flow rate of 1.0 mL/min [52 (link)].
+ Open protocol
+ Expand
4

HPLC-MS Analysis of Bile Acids

Check if the same lab product or an alternative is used in the 5 most similar protocols
MS analysis of bile acids was performed on an Agilent 1,260 series HPLC system (Agilent Technologies, Santa Clara, CA, United States) equipped with a quaternary pump, degasser, auto sampler and thermostatically controlled column compartment. Chromatographic separation was performed on an Agilent Poroshell 120 EC-C18 (3.0 mm × 150 mm, 2.7 micron) at 37°C. The structures of bile acids are shown in Figure 1 and Table 1. The mobile phase consisted of 5 mmol/L ammonium acetate containing 0.1% formic acid (solvent A) and methanol (solvent B). A gradient elution program was used as follows: linear gradient from 0 to 2 min, 25% B–46% B; 2–27 min, 46% B–70% B; 27–35 min, 70% B−95% B; 35–38 min, 95% B–98% B. The flow rate was 0.3 mL/min and the injected sample volume was 3 µL. Detection was achieved using an Agilent triple quadrupole 6,470 mass spectrometer (MS) with an electrospray ionization source. The main working parameters for MS were set as follows: drying gas (N2) flow rate, 10 L/min; drying gas temperature, 300°C; nebulizing gas (N2) pressure, 45 psi; capillary voltage, 3,000 V; quadrupole temperature, 300°C. Fragment ion spectra were recorded by negative electrospray ionization in the multiple reaction monitoring mode (MRM) and MS parameters, as shown in Table 2.
+ Open protocol
+ Expand
5

Quantitative Analysis of Analytes via HPLC-ELSD

Check if the same lab product or an alternative is used in the 5 most similar protocols
The quantitative assay was performed on an Agilent 1260 series HPLC system equipped with an ELSD (Agilent Technologies, Palo Alto CA, USA), which was controlled by Agilent ChemStation software (B.04.03-SP1). The separation of analytes was conducted on a Waters XBridge HILIC column (4.6 × 150 mm, 3.5 μm) with a flow rate of 1.0 mL/min at 30°C. The mobile phases were acetonitrile (A) and water (B) with a gradient elution of 88% A at 0–1 min, 88–78% A at 1–10 min, 78–65% A at 10–20 min, 65–88% A at 20–20.1 min, and 88% A at 20.1–35 min. The injection volume was 5 μL. The drift tube temperature of ELSD was 50°C and the nitrogen cumulative flow rate was 1.0 mL/min.
+ Open protocol
+ Expand
6

HPLC Analysis of MDA Compound

Check if the same lab product or an alternative is used in the 5 most similar protocols
The separation and analysis of MDA were conducted on an Agilent 1260 series HPLC system (Agilent Technologies, USA) equipped with a G1311B quaternary pump, a G1329B autosampler, a G1316A thermostated column compartment, a G1321B fluorescence detector (FLD) and an Agilent ZORBAX SB-C18 column (250 mm × 4.6 mm, 5 μm).
Eluent A and B were water and acetonitrile, respectively. The separation of MDA was accomplished with gradient elution: 0–10 min, 20% B to 45% B; 10–15 min, 45% B to 100% B. The flow rate was constant at 1.0 mL/min. The fluorescent detection wavelengths were set at 370 nm for excitation and 420 nm for emission. The column temperature was kept at 30 °C.
+ Open protocol
+ Expand
7

Quantitative Analysis of Glyphosate in Urine

Check if the same lab product or an alternative is used in the 5 most similar protocols
200 μL of urine sample was transferred into a 15-mL polypropylene (PP) tube and spiked with the labeled internal standard mixture (2-13C, 99%; 15N, 98+% Glyphosate), allowed to stand at room temperature for 30 min and then diluted to 1.0 mL (5-fold dilution) with 1% formic acid in water. The diluted sample was vortexed for 1 min, centrifuged and filtered through a nonsterile regenerated cellulose (RC) membrane filters (0.2 μm; Phenomenex, Inc., Torrance, CA, USA). The filtrate then transferred into an auto sampler vial for LC-MS/MS analysis. An Agilent 1260 Series HPLC system (Agilent Technologies Inc., Santa Clara, CA, USA) coupled with an ABSCIEX 4500 QTRAP mass spectrometer (Applied Biosystems, Foster City, CA, USA) running under negative mode electrospray ionization was used for the analysis. An anion-exchange column, Dionex IonPac AS 21 (2 mm × 250 mm, 7 μm) was employed for the separation of target chemicals under isocratic elution condition with a mobile phase consist of 1% formic acid in water and acetonitrile (95:5) mixture. Isotopic dilution mass spectrometry with MRM mode of analysis was used for selective quantification of the target chemicals. The QA/QC protocols included matrix spike (mean spike recoveries (n=5) for glyphosate was 109.5%) and procedural blanks (non-detected or
+ Open protocol
+ Expand
8

Automated Synthesis of Peptides

Check if the same lab product or an alternative is used in the 5 most similar protocols
The reagents and solvents were purchased from commercial sources (Fisher) and used directly. The intermediates and products were purified by using an ISCO CombiFlash system and prepacked SiO2 cartridges. Final compounds were purified on preparative high-pressure liquid chromatography (RP-HPLC) was performed on Agilent 1260 Series system. Systems were run with 0–20% methanol/water gradient with 0.1% TFA. NMR spectra were acquired on a Bruker AV500 instrument (500 MHz for 1H-NMR, 126 MHz for 13C-NMR). TLC-MS were acquired using Advion CMS-L MS. Matrix-assisted laser desorption ionization mass spectra (MALDI-MS) data were acquired in positive-ion mode using a Sciex 4800 MALDI TOF/TOF MS. The peptides were synthesized using a CEM Liberty Blue Automated Microwave Peptide Synthesizer with the manufacturers standard coupling cycles at 0.1 mmol scale. The purity of final compounds was confirmed by Waters LC-MS system and/or Agilent 1260 Series HPLC system. Systems were run with 0–40% methanol/water gradient with 0.1% TFA. All the purity of target compounds showed >95% in RP-HPLC.
+ Open protocol
+ Expand
9

HPLC Analysis of HMF and Oxidized Derivatives

Check if the same lab product or an alternative is used in the 5 most similar protocols
HMF and its oxidized derivatives were analyzed using an Agilent 1260 series HPLC system (Agilent, Wald Bronn, Germany) equipped with a Bio-Rad Aminex HPX-87H Column and a UV detector (wavelength = 264 nm). H2SO4 (7.5 mM) was used as the mobile phase. The separation and identification of these chemicals was conducted at 60 °C with a flow rate of 0.6 mL/min. The injection volume was 20 μL. The retention times of HMF, HMFCA, DFF, FFCA and FDCA were 29.74, 20.47, 36.2, 22.42, and 16.15 min, respectively. The titer (mM) of the oxidized products was calculated based on the plotted standard curve. The yield (%) of the oxidized products was calculated using the following formula (1): Yield%=Oxidized productmMSubstrate HMFmM×100%
+ Open protocol
+ Expand
10

HPLC-MS Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC system consisted of an Agilent 1260 Series HPLC system (Agilent Technologies, Palo Alto, CA, USA), a Shimadzu, System Controller, SCL-10A Vp, Pump, LC 10AD Vp Solvent Degasser (DGU14A, Shimadzu, Columbia, MD, USA), and an CTC PAL autosampler (Zwingen, Switzerland). The analytical column was XBridge C18, 50 × 2.1 mm, 5µm (Waters Corp, Milford, MA, USA). The column temperature was maintained at 25 °C. Mobile phase A was water and mobile phase B was 0.1 % formic acid in acetonitrile. The elution pump (Agilent 1260) gradient was: 0–1.0 min 0.0% B at 0.4 mL min−1 flow, 1.1–3.0 min from 50% to 90% B at 0.4 mL min−1 flow rate, 3.1–7.0 min 90% B at 0.8 mL min−1 flow rate and 7.1–10.0 min 0.0% B at 0.8 mL min−1 flow rate. The makeup pump (Shimadzu) ran an isocratic gradient at 50:50 mobile phase A: mobile phase B and at flow rate of 0.2 mL min-1. The switching valve program was: 0–1.8 min at position A (loading sample into column and desalting), 1.8–3.0 min at position B (elution to MS source) and 3.0 to 10.0 min at position A (column flushing and re-equalibration). Injection volume was 50 µL and total run time was 10.0 min.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!