The largest database of trusted experimental protocols

1260 infinity

Manufactured by Agilent Technologies
Sourced in United States, Germany, France, United Kingdom, Japan, Spain

The 1260 Infinity is a high-performance liquid chromatography (HPLC) system designed by Agilent Technologies. It is a modular system that can be configured with various components to perform a wide range of analytical tasks. The 1260 Infinity system provides precise and reliable liquid chromatography separations, enabling users to obtain accurate and reproducible results in their analytical workflows.

Automatically generated - may contain errors

552 protocols using 1260 infinity

1

HPLC Separation and Quantitation Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC instrument was (Agilent 1260 Infinity, Germany) connected with a preparative pump (G1361A). The HPLC was also connected to quadrupole LC/MS detector (Agilent 6120) and diode array detector VL (Agilent 1260 Infinity, G131SD). In addition, the HPLC was also equipped with thermostated column compartment (Agilent 1260 Infinity, G1316A) and preparative Autosampler (Agilent 1260 Infinity, G2260A). The quantitation and separation were processed on Eclipse Plus C18 column (3.5 µm particle size, 100 mm × 4.6 mm i.d) (USA).

Xz6 bench top laboratory centrifuge with maximum speed to 5000 rpm and time from zero to sixty minutes

Sonix TV series ultrasonicator (USA)

+ Open protocol
+ Expand
2

HPLC Quantification of L-PAPA and Glucose

Check if the same lab product or an alternative is used in the 5 most similar protocols
For the determination of L-PAPA and glucose concentration, supernatants at given time points were isolated by centrifugation (21,000g, 10 min). The samples were stored at −20 °C until HPLC analysis. This was performed on the 1260 Infinity HPLC system (Agilent Technologies). L-PAPA was detected at 210 nm with a diode array detector (DAD, 1260 Infinity, Agilent Technologies). For separation, a Prontosil C18 column (250 × 4 mm, CS-Chromatographie Services, Germany) was used at 40 °C. As a mobile phase, 40 mM Na2SO4 (pH 2.7 adjusted with methane sulfonic acid) was used with a flow rate of 1 mL/min. L-PAPA was quantified using 1.0 mM L-phenylglycine as the internal standard to correct variabilities in analytes and a five-point calibration curve with L-PAPA as an external standard. Glucose quantification was performed by HPLC with a refractive index detector (RID; 1260 Infinity, Agilent Technologies). An Organic Acid column (300 × 8 mm, CS-Chromatographie Services, Germany) was used at 40 °C with 5 mM H2SO4 and a flow rate of 0.8 mL/min. Glucose was quantified using a five-point calibration curve with glucose as an external standard.
+ Open protocol
+ Expand
3

HPLC Analysis of Hydroxycinnamic Acids

Check if the same lab product or an alternative is used in the 5 most similar protocols
Hydroxycinnamic acids and the 4-vinyl derivatives in both the organic solvent and N2HPO4-citric acid buffer phases were analyzed by HPLC (1260 Infinity, Agilent, USA) equipped with a ZORBAX-SB C18 (4.6 × 150 mm, 5 μm in diameter, Agilent, USA) column, a quat pump, and a DAD detector (1260 Infinity, Agilent). The mobile phase acetic acid (0.1%)-methanol (60:40) was used at a flow rate of 1 mL min−1 at 30 °C. The injection volume for all samples was 20 μL. The detection wavelength was set at 280 nm (Hu et al., 2015 (link)). The standard curve was plotted by performing linear regression analysis on the peak area of each concentration, using the concentration of the 4-vinylguaiacol (4-VG) standard sample as the abscissa and the peak area as the ordinate (Fig. S1). It showed the good linear relationship between the peak area and the concentration of the 4-VG solution in the range of 0 mg L−1 to 5 mg L−1.
+ Open protocol
+ Expand
4

Molecular Weight Analysis of Gamma-Irradiated PLA:PCL Nanofibrous Scaffolds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Gel permeation chromatography (GPC) was performed to evaluated molecular weight (Mw) and molecular number (Mn) of PLA:PCL SMPNs after gamma irradiation.
Analyses were performed using Agilent Technologies 1260 Infinity GPC apparatus, a chromatographic system composed by a precolumn (Agilent GPC/SEC Guard Columns) and three Ultrastyragel columns connected in series (7.7 × 250 mm each with different pore diameters of 104 Å, 103 Å, and 500 Å), a pump (Agilent Technologies 1260 Infinity), an IR detector (Agilent Technologies 1260 Infinity), and a data handling software (OpenLab and Cirrus).
PLA:PCL pristine powder and non-irradiated and gamma-irradiated PLA:PCL electrospun nanofibrous scaffolds were solubilized in THF and the solutions obtained (1 mg/mL) were filtered with a Fluoropore 0.45-μm (Millipore) filter before being injected in GPC; THF was the mobile phase and constant flow rate was set at 1 mL/min.
+ Open protocol
+ Expand
5

Quantification of Soluble Sugars

Check if the same lab product or an alternative is used in the 5 most similar protocols
Soluble sugars verbascose, stachyose, raffinose, and sucrose were determined using high-performance liquid chromatography (1260 Infinity, Agilent Technologies, Santa Clara, USA), equipped with a Refractive Index Detector (1260 Infinity RID), according to the method previously described in De Angelis et al. (2021b) . The results were expressed as mg g -1 on a dry matter basis. The analysis was carried out in triplicate.
+ Open protocol
+ Expand
6

Purification and Characterization of Indolocarbazole Alkaloid

Check if the same lab product or an alternative is used in the 5 most similar protocols
Crude extracts of co- and mono-cultures (5.0 mg/mL in methanol) were initially screened by HPLC-UV analysis (Agilent Technologies 1260 Infinity) with spectral scanning from 230−320 nm and a linear gradient from 0–100% methanol.
The co-culture crude was then fractionated by vacuum liquid chromatography on silica gel 60 (Acros Organics™ ultra-pure 60A 40-63u), eluting with a gradient of n-hexane−ethyl acetate−MeOH to give 10 subfractions (CC1-CC10). HPLC-UV-HRESIMS analysis revealed that most metabolites that were not observed in the monocultures were in CC2; hence, further purification of this fraction was carried out by reversed-phase HPLC (Agilent Technologies 1260 Infinity) (ACE C18-HL 10 µM 10 × 250 mm) equipped with a diode array detector (DAD). HPLC separation was performed using solvent A (95% H2O, 5% methanol, and 0.1% trifluoroacetic acid) and solvent B (100% methanol) as eluents, with a linear gradient from 20%−100% MeOH over 40 min and a solvent flow rate of 1.5 mL min−1. All of the employed chemicals were HPLC-grade. The separation afforded 4.0 mg of an indolocarbazole alkaloid, BE-13793C 1.
BE-13793C: yellow solid. UV (PDA) λmax: 245, 305, 330, 395nm; IR (neat) νmax (cm−1): 3427, 1740, 1590, 1422, 1343, 1205, 857 nm; 1H, 13C-NMR data, see Table 1; HR ESIMS (positive mode) m/z calculated [M + H]+ = 358.0822; observed [M + H]+ = 358.0826; ∆ = 1.0576 ppm.
+ Open protocol
+ Expand
7

GPC Analysis of Polymer Molecular Weights

Check if the same lab product or an alternative is used in the 5 most similar protocols
Molecular weight distributions were assessed using a DMF GPC instrument operating at 60 °C. The set-up comprised two Polymer Laboratories PL gel 5 μm Mixed-C columns and one PL polar gel 5 μm guard column connected in series to an Agilent Technologies 1260 Infinity multidetector suite and an Agilent Technologies 1260 Infinity pump injection module. The GPC eluent was HPLC-grade DMF containing 10 mM LiBr and was filtered prior to use. The flow rate was 1.0 ml min–1 and DMSO was used as a flow-rate marker. Calibration was conducted using a series of ten near-monodisperse poly(methyl methacrylate) standards ranging from 625 to 618 000 g mol–1. Chromatograms were analyzed using Varian Cirrus GPC software (version 3.3).
+ Open protocol
+ Expand
8

PEG Molecular Weight Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
Before PEGDA synthesis, PEG number-average molecular weights (nominal molecular weights of 3.4, 6, 10, 20, and 35 kDa) were measured via size exclusion chromatography (SEC) carried out on an Agilent system with a 1260 Infinity isocratic pump, degasser, and thermostat column chamber containing a guard column (table S2). Specimens were dissolved in chloroform (2 to 6 mg/ml) at room temperature and syringe-filtered. Injections (100 μl) were passed through an Agilent 5-μm MIXED-D column with an operating range of 200 to 400,000 g mol−1 at a flow rate of 0.5 ml/min at 30°C and a mobile phase of chloroform with 50 ppm (parts per million) amylene. Separated components were passed through the Agilent 1260 Infinity refractometer and Agilent 1260 Infinity bio-inert multidetector suite with dual angle static and dynamic light scattering detectors. Number-average molecular weight (Mn) was determined using the Agilent Bio-SEC software relative to PEG standards. For each molecular weight of PEG, three samples were measured with duplicate injections, and the number-average molecular weight was calculated from each measurement (n = 6).
+ Open protocol
+ Expand
9

Vitamin C Analysis by HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC (Agilent Technologies 1260 Infinity, Santa Clara) with RP C18 column at 254 nm followed the standard method [16] (link).
Vitamin C determination HPLC (Agilent Technologies 1260 Infinity, Santa Clara) with ACE 5 C18 column using UV detector at 254 nm followed the method described [17] .
+ Open protocol
+ Expand
10

Quantification of GABA and Glutamate by HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
After treatment with 8-OaS for 3 days, GABA and glutamate were detected using high-performance liquid chromatography (HPLC; Agilent Technologies 1260 Infinity, Agilent Technologies, Wilmington, DE), according to previously described methods (Wang et al. 2017 (link)). Before derivatization, the samples were dissolved with boracic acid buffer (pH 9.0) and centrifuged for 15 min at 3000 r/min at 4 °C. Samples were mixed with 2,4-dinitrofluorobenzene (DNFB) and 0.5 mol/L NaHPO3 buffer for 1 h at 60 °C; then, a phosphate buffer solution (pH 7) was added to stop the reaction. Samples were analyzed using a UV detector (360 nm, Agilent Technologies 1260 Infinity). The mobile phase was KH2PO4 buffer (pH 6.0), acetonitrile, and ddH2O (84:8:8, v/v/v) at a flow rate of 1.0 mL/min. A Themo TC-C18 column (250 × 4.6 mm2; particle size 5 mm) was used. The concentrations were obtained using the LC solution software (Shimadzu) based on standard substances.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!