The largest database of trusted experimental protocols

Powerlab 8 35

Manufactured by ADInstruments
Sourced in Australia, United States, New Zealand, United Kingdom

The PowerLab 8/35 is a data acquisition system designed for recording and analyzing a wide range of physiological signals. It features eight analog input channels, 16-bit analog-to-digital conversion, and a sampling rate of up to 200 kHz per channel. The PowerLab 8/35 is capable of recording various signals, such as biopotentials, force, pressure, and more. It provides a flexible and versatile platform for data collection and analysis in research and educational applications.

Automatically generated - may contain errors

131 protocols using powerlab 8 35

1

Muscle Contractility Measurement via EPS

Check if the same lab product or an alternative is used in the 5 most similar protocols
To measure EPS-induced absolute contractile force of models, we used an ex vivo organ-bath contractility assay for isolated mouse muscles similarly as described30 . Models were mounted on two hooks that were connected to a force transducer (WPI, cat. no. SI-KG2) and a micromanipulator in a horizontal tissue bath (WPI, cat. no. SI-HTB2) using the model’s two holes left from post removal. A current-controlled electrical stimulator (Ugo Basile, cat. no. 3165) delivered unipolar EPS excitation consisting of 400 mA, 1 ms pulses at 25 Hz for 300 ms. Force measurements were amplified by a force transducer amplifier (WPI, cat. no. SI-BAM21-LCB), controlled by a data acquisition system (ADInstruments, cat. no. PowerLab 8/35) and analyzed by LabChart V8 software (ADInstruments). A digital oscilloscope (Fluke ScopeMeter 190–202) was used to confirm the frequency, shape and amplitude of the delivered electrical pulse signals. Caffeine was bought from Sigma (Cat. no. C8960) and Tirasemtiv35 (link) was synthesized at Novartis.
+ Open protocol
+ Expand
2

Autonomic Emotional Reactivity in Gambling

Check if the same lab product or an alternative is used in the 5 most similar protocols
Simultaneously the gambling task, the skin conductance response (SCR) was recorded in a group of 5 patients (age: mean = 49.8; s.d. = 5.6; education level: mean = 16.2; s.d. = 2.5) and 5 controls (age: mean = 47.4; s.d. = 9.2; education level: mean = 16.6; s.d. = 2.2). The SCR recording allows us to have a physiological index of the autonomic emotional reactivity (Dawson et al., 2007 ). AD-Instruments PowerLab 8/35 and ML116 GSR Amplifier (providing low constant-voltage AC excitation (22 mVrms @ 75 Hz)) devices were used as signal amplifier with specific GSR sensors consisting of two bipolar finger electrodes. The sensors were applied on the distal phalanx of the index and middle fingers of the left hand. The signal was sampled at 1 KHz and recorded using the software LabChart 7 (AD-Instruments, Inc.).
+ Open protocol
+ Expand
3

Visceromotor Reflex Evaluation in Rat Colitis

Check if the same lab product or an alternative is used in the 5 most similar protocols
VMR was evoked by repeated colorectal distention (CRD) in lightly anesthetized rats, as described previously [20 (link)]. Firstly, anesthesia was induced by 4% isoflurane and maintained with 2% isoflurane for surgical procedures. A flexible latex balloon (3 cm long, 1.5 cm max diameter) was lubricated and inserted into the distal colon lumen via the anus (the tip of the balloon was 1 cm from the anus), and two electrodes made of Teflon-coated platinum wires were inserted into the external abdominal oblique muscle. After surgery, the anesthesia level was decreased to 1% isoflurane for electromyography (EMG) recording. The balloon was inflated by phasic distension (40, 60, and 80 mmHg, each lasting for 20 s) with a 5 min interval. The EMG signals were amplified (×5000), filtered (30~1000 Hz) by NL900D (Neurolog, Digitimer, America), then transmitted into PowerLab 8/35 (AD Instruments, Australia) and analyzed off-line by LabChart 7.1 software. The EMG reflex responding to each CRD stimulus was repeated 3 times and the outcomes were averaged. After all behavior tests were finished, rats were euthanized and the distal colon was removed for histopathological test to confirm the development of colitis.
+ Open protocol
+ Expand
4

Measuring PPG Signals in Athletes

Check if the same lab product or an alternative is used in the 5 most similar protocols
Each athlete was asked to sit quietly for 3 minutes before the pulse recording and keep relaxing during the whole measurement process. The optical PPG pulses were digitally recorded from the right middle finger using a PowerLab data collection system (ADInstruments Pty Ltd., PowerLab 8/35, Bella Vista, NSW, Australia) at a sampling rate of 1000 Hz. Since there is significant effect of the finger temperature on the PPG pulse shape [22 ], finger skin temperature was ensured to be around 25°C for each athlete. When the PPG pulse waveforms were satisfactory and stably shown on the monitoring screen, they were saved for 30 s to a computer for further off-line analysis.
+ Open protocol
+ Expand
5

Measuring Skin Blood Flow and Temperature

Check if the same lab product or an alternative is used in the 5 most similar protocols
The test setting was the same for both protocols. To ensure each participant maintained the same posture throughout the experiment, he/she was asked to sit on a height-adjustable mat table with his/her feet on the LIV platform. The height of the table was adjusted to ensure the angles at ankle and the knee were maintained at 90°. To standardize loading between feet and platform, each participant was instructed to lean on his/her elbows which were placed on the top of the knees. Fig 1 demonstrates the test setting. All participants sat quietly for 10 minutes before the experimental protocol was initiated.
LIV was applied using a platform that vibrated vertically at 30Hz with peak acceleration of 0.4 g (0.22mm peak to peak displacement) for 10 minutes (LiveMD, Marodyne Medical LLC, Tampa, FL). To measure SBF data continuously, an optic probe (CP1T-1000, Moor Instruments, Wilmington, DE) was placed on the right foot. The probe was a combination of laser Doppler flowmetry (LDF) and thermometer. PowerLab 8/35 with LabChart V8 (AD Instruments, Colorado Springs, CO) was used for data acquisition. The probe was placed at the midpoint of the horizontal line drawn between the medial malleolus and the Achilles tendon. Fig 1 demonstrates the placement of the probe for measuring SBF and temperature.
+ Open protocol
+ Expand
6

Synchronized Respiratory and Cardiovascular Monitoring

Check if the same lab product or an alternative is used in the 5 most similar protocols
During each session, breathing was continuously recorded using a spirometer. Cardiovascular signals (ECG) were also continuously recorded at 1000 Hz. RR-interval time series were extracted from raw ECG signals52 (link). Respiration and cardiovascular signals were recorded, digitalized and synchronized using a Power-Lab 8/35 device (Human respiratory kit with spirometer and ECG Bio Amps, ADInstrument®). Calibrations were performed according to the manufacturer’s instructions before each test on each subject.
+ Open protocol
+ Expand
7

In Vitro Cortical Electrophysiology

Check if the same lab product or an alternative is used in the 5 most similar protocols
Cortical brain slices were placed on an interface recording chamber maintained at 34°C superfused with oxygenated aCSF at 2 mL/min. Ascending cortical inputs were stimulated with a tungsten concentric biopolar stimulating electrode (FHC, ME) at the layer VI-white matter boundary. A recording electrode was placed in layer II-III of the cortex. Signals were recorded using glass micropipettes (resistance ≅ 1 MΩ) and acquired with DP-311 amplifier (Warner Instruments, CT) and digitized with a PowerLab 8/35 (AD Instruments, CO) using LabChart software.
+ Open protocol
+ Expand
8

Hemodynamic Study of CCl4-induced Liver Injury

Check if the same lab product or an alternative is used in the 5 most similar protocols
A hemodynamic study was performed two weeks after the last dose of CCl4, a washout period intended to avoid the interference of inflammation associated with acute hepatic injury. The rats were anesthetized with sevoflurane (Abbott Laboratories), and the right common carotid artery and external jugular vein were dissected and canalized using a 24 G Abbocath catheter (B. Braun) and a polyethylene tube (PE50), respectively, to measure the mean arterial pressure (MAP) and the central venous pressure (CVP). Thereafter, a mid-laparotomy was performed and a 24 G Abbocath catheter (B. Braun) was inserted into the ileocolic vein to measure the portal pressure (PP). After 5 minutes of stabilization, blood pressures were registered for 5 minutes using pressure transducers and a multichannel PowerLab 8/35 and Lab Chart Reader software (AD Instruments) for analysis. Body temperature was monitored with a thermometer and maintained at stable levels with a warming pad throughout the experiment.
+ Open protocol
+ Expand
9

Calibrating Response Time Measurements

Check if the same lab product or an alternative is used in the 5 most similar protocols
Vocal responses were recorded and used to verify, offline, the accuracy of the response times reported by Presentation software. A Matlab script that detected a vocal response onset based on the sound envelope of the vocalization showed a high level of concordance with the software voice key. This analysis is located in the Supplementary Material.
Further to this analysis, we calibrated the response times reported by Presentation with PowerLab 8/35 (https://www.adinstruments.com/products/powerlab) that measured the veridical times of the stimulus and response relative to each other. We used this to adjust the two response modalities, subtracting 75.83 ms from the button press and 67.43 ms from the vocal responses. Please refer to “Appendix A” for a description and graphical representation of this calibration procedure.
+ Open protocol
+ Expand
10

Measuring Portal Pressure in Mice

Check if the same lab product or an alternative is used in the 5 most similar protocols
Mice were anaesthetized with a mixture of ketamine (Ketaset, 100 mg/mL; Zoetis, Berlin, Germany) and xylazine (Xylazin, 20 mg/mL; WDT, Garbsen, Germany). To evaluate portal hypertension, the portal pressure was measured in each mouse. The portal vein was cannulated with a 25-gauge safety-multifly needle (Sarstedt, Nümbrecht, Germany), fixated with a vascular clamp and connected to a highly sensitive pressure transducer. The external zero reference point was placed at the midportion of the animal. The measurements were recorded for at least 1 min on a multichannel computer-based recorder using PowerLab 8/35 and the LabChart Software (Powerlab; ADInstruments, Dunedin, New Zealand). The final value for portal pressure of one biological replicate was determined as the mean in the recorded phase.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!