The largest database of trusted experimental protocols

Uplc system

Manufactured by Waters Corporation
Sourced in United States, United Kingdom, Japan

The UPLC system is a high-performance liquid chromatography (HPLC) instrument developed by Waters Corporation. It utilizes ultra-high pressure liquid chromatography (UPLC) technology to achieve rapid and efficient separation of complex samples. The UPLC system is designed to provide improved resolution, sensitivity, and analysis speed compared to traditional HPLC methods.

Automatically generated - may contain errors

190 protocols using uplc system

1

Quantifying N-Linked Glycans in Eculizumab

Check if the same lab product or an alternative is used in the 5 most similar protocols
The identification and relative percentage of N-linked glycan in eculizumab was determined by hydrophilic interaction UPLC (HILIC-UPLC). Eculizumab (SB12 and RP) samples were treated with PNGase-F enzyme to release N-glycan after denaturation. The released N-glycan was separated from proteins using cold ethanol, and dried completely. 2-aminobenzamide (2-AB)-labeled N-glycan was gradually separated using an Acquity UPLC BEH glycan column (Waters), and quantified on a fluorescence detector connected to a Waters UPLC system. For identification, N-glycan was labeled with procainamide and identified on LC-ESI-MS connected to a Waters UPLC system.
+ Open protocol
+ Expand
2

Profiling Phenolic Compounds by UHPLC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The phenolic profile of the extract was characterized by using a UHPLC-DAD-ESI-QToF/MS, an ACQUITY UPLCTM system coupled to a DAD and a SYNAPTTM G2 HDMS (Waters, Milford, MA, USA). The separation was carried out using a reversed-phase ACQUITY UPLC BEH C18 column (100 × 2.1 mm, 1.7 µm) with a pre-column of the same material (VanGuardTM) (Waters, Milford, MA, USA) and a method described by Garrido et al. [24 (link)] with minor modifications. The separation was carried out using 0.1% (v/v) of acetic acid in water and 0.1% (v/v) of acetic acid in methanol as mobile phases. The injection volume was 5.0 µL. Flavan-3-ols were recorded at 280 nm, as were hydroxycinnamic acids at 320 nm and flavonols at 370 nm. Mass spectral data were recorded in positive and negative ion modes.
+ Open protocol
+ Expand
3

UPLC-QTOF-MS Profiling of Biological Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
The chromatographic separation was achieved on an Acquity UPLCTM System coupled to a Xevo G2 Q-TOF/MS with a Waters UPLC BEH C18 column (2.1 × 100 mm I.D., 1.7 µm; Waters Corp., Milford, MA, USA) at a column temperature of 45 °C. The mobile phase was composed of 0.2% formic acid aqueous solution (A) and MeOH (B) with the gradient set as follows: 0–1.0 min, 95–95% B; 1.0–2.0 min, 95–2% B; 2.0–13.0 min, 2–2% B; 13–13.5 min, 2–95% B; 13.5–15 min, 95–95% B. The flow rate was 0.40 mL/min, and the injection volume was 2 μL. The autosampler temperature was conditioned at 4°C.
Electrospray ionization (ESI) in positive ion (ESI+) mode and negative ion (ESI−) mode was applied for high-resolution MS detection. The mass range was set at m/z 50–1200 Da. The optimized operating parameters were set as follows: ion spray voltage of 3.0 kV, cone voltage of 25 V, cone gas flow of 50 L/h, source temperature of 120 °C, dry gas (N2) flow of 10 mL/min, atomization temperature of 450 °C, and 400 °C for ESI+ and ESI−. MS data were recorded in MSE mode. The accurate mass and composition of the relative target ions were calculated with MassLynx V 4.0 software (Waters Corp., Milford, MA/USA).
+ Open protocol
+ Expand
4

UPLC-MS/MS Quantification of Strigolactones

Check if the same lab product or an alternative is used in the 5 most similar protocols
Strigolactones were analyzed using an Acquity UPLCTM System (Waters, Milford, MA, USA) coupled to a Xevo® TQ-XS tandem quadrupole mass spectrometer (Waters MS Technologies, Manchester, UK) with an electrospray interface. The separation was performed on an Acquity BEH C18 column (2.1 × 100 mm, 1.7 μm) at flow rate 0.45 mL/min and temperature 45 °C, with 12 min binary gradient elution as follows: 0–0.4 min (85% A), 0.4–5 min (60% A), 5–8 min (35% A), 8–8.7 min (35% A), 8.7–9.5 (5% A), column wash for 0.8 min (5% A) and final column equilibration for 3 min for initial conditions (85% A), where A = 15 mM formic acid/water and B = 15 mM formic acid/acetonitrile (v/v). The eluate was introduced in the ion source of the mass spectrometer and analysed using the following settings: ion source/desolvation temperature (120/550 °C), desolvation/cone gas (nitrogen) flow (1000/150 L/hr), collision gas (argon) flow 0.15 ml/min. Compounds were analyzed in multiple reaction monitoring mode (MRM) and quantified by diagnostic transitions of: 4-deoxyorobanchol (331 > 97), putative methoxy-5-deoxystrigol isomers (361 > 97) using optimized cone voltage (20–22 V) and collision energy (20 eV). The MassLynxTM software, version 4.2 (Waters) was used to operate the instrument, acquire and process the data.
+ Open protocol
+ Expand
5

Phenolic Profiling of Graciano Extract

Check if the same lab product or an alternative is used in the 5 most similar protocols
The phenolic composition of the Graciano extract (GE) was studied by ultra-high performance liquid chromatography and mass spectrometry using an ACQUITY UPLCTM system with a diode array detector and coupled to a quadrupole time of flight mass spectrometer (Waters, Milford, MA, USA). The separation was carried out as described by Asensio-Regalado et al. [15 (link)] Briefly, the mobile phases were 0.1% (v/v) acetic acid in MeOH and 0.1% (v/v) acetic acid in water delivered at a flow rate of 0.35 mL/min; the volume injected was 5.0 µL; flavan-3-ols, hydroxycinnamic acids, and flavanols were detected at 280, 320, and 370 nm, respectively. Positive and negative ion modes mass spectra were recorded.
+ Open protocol
+ Expand
6

UPLC-MS/MS Quantitative Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
A UPLC TM system (Acquity, Waters, Milford, MA, USA) was interfaced to a triple quadrupole mass spectrometer (Xevo TQ-S, Waters Corporation, Manchester, UK). LC separation was performed with a 50 x 2.1mm, 1.7 μm particle size Acquity UPLC BEH C18 analytical column (Waters). The mobile phases employed consisted on water (A) and acetonitrile (B) both with 0.0025% HCOOH, at a flow rate of 0.3 mL•min -1 . The gradient program started with 50% B, increased linearly to 90% of B for 1.5 min and maintained during 1.5 min. Finally the gradient was held to initial conditions in order to re-equilibrate the column. Temperature column was set to 25ºC. 2 μL were selected as injection volume.
In the Selected Reaction Monitoring (SRM) method applied, dwell time values ranging from 5 to 90 ms were used in order to obtain 12 points per peak. Source temperature was set to 150 ºC.
Drying and nebulising gas was nitrogen (Praxair, Valencia, Spain). Desolvation gas flow was set to 1200 L•h -1 and the cone gas to 250 L•h -1 . For operating in MS/MS mode, argon (99.995%; Praxair, Valencia, Spain) was used as collision gas at 0.25 mL•min-1). Capillary voltage and desolvation gas temperature were set at 3.2 kV (1.9 kV in ESI -mode) and 650ºC respectively. TargetLynx (MassLynx v. 4.1, Waters, Manchester, UK) software was used to process the quantitative data.
+ Open protocol
+ Expand
7

UPLC-MS/MS Quantification of Thiols in Baijiu

Check if the same lab product or an alternative is used in the 5 most similar protocols
The thiols were separated by using an UPLC system (Waters, Milford, USA) equipped with a vacuum degasser, a binary solvent manager, and an autosampler. As stationary phase an analytical column (waters BEH C18, 100 × 2.1 mm, 1.7 μm) was used. Flow-rate was 0.3 mL/min and the composition of eluents was: solvent A (0.1% formic acid in water) and solvent B (0.1% formic acid in acetonitrile). The linear gradient for solvent B was as follows: 0 min, 15%; 13 min, 22%; 14 min, 30%; 18 min, 35%; 18.5 min, 100%; 21.5 min, 100%; and 22 min, 15%. The injection volume was set at 10 μL.
The derivatization Baijiu sample was analyzed by means of UPLC–MS/MS using precursor ion scan screening for compounds releasing the diagnostic ion (m/z X→143.5) (24 (link)). In the source, cone voltage of 23 V and collision energy of 20 eV was applied. The eleven derivatives were quantified in MRM mode by monitoring their corresponding precursor ion, product ion, cone voltage, and collision energy, respectively (Table 1). Each Baijiu sample was tested in three different sessions to obtain an average value.
+ Open protocol
+ Expand
8

Red Ginseng UPLC-QToF-MS Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
For the UPLC-QToF-MS analysis, the powdered red ginseng samples were extracted in a manner similar to our previous studies using ultrasonic extraction [13 (link)]. In total, 300 mg of each powered red ginseng was weighed in a centrifuge tube (15 mL, PP-single use; BioLogix Group, Jinan, Shandong, China) and shaken vigorously after the addition of 6 mL of methanol. The extract was then placed in an ultrasonic cleaner (60 Hz; Wiseclean, Seoul, Korea) for 30 min. The solution was centrifuged (Legand Mach 1.6R; Thermo, Frankfurt, Germany) at a speed of 3000 rpm for 10 min, and an aliquot of the supernatant solution was filtered (0.2 µm; Acrodisk, Gelman Sciences, Ann Arbor, MI, USA) and injected into the UPLC system (Waters Co., Milford, MA, USA).
+ Open protocol
+ Expand
9

UPLC-MS Analysis of Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
UPLC/MS analysis was performed on a UPLC system coupled with XEVO G2 Q-TOF mass spectrometer via an ESI source (Waters Corp., Milford, MA). For UPLC separation, 2 μL of sample solution was injected into an ACQUITY HSS T3 C18 column (100 × 2.1 mm, 1.7 μm, Waters). The mobile phase consisted of ACN (A) and water containing 0.1% (v/v) formic acid (B). Linear gradient elution was applied (0–5 min, 5–30% A; 5–10 min, 30–40% A; 10–20 min, 40–65% A; 20–25 min, 65–90% A) at a flow rate of 0.4 mL/min. The column temperature was 45°C. For MS detection, accurate mass was maintained by the LockSpray interface of sulfadimethoxine (309.0658 [M-H]). The operating parameters in negative ion mode were as follows: capillary voltage, 3.0 kV; cone voltage, 30 V; desolvation gas flow rate, 750 L/h; source temperature, 120°C; desolvation temperature, 350°C. MS data were acquired in centroid mode and processed by MassLynx software (Waters, version 4.1).
+ Open protocol
+ Expand
10

UPLC-MS Analysis of Compound Peaks

Check if the same lab product or an alternative is used in the 5 most similar protocols
The identification of component peaks was carried out on an ultra‐performance liquid chromatography (UPLC) system (Waters, Milford, MA, USA) equipped with a photo‐diode array detector. Aliquots (2.0 μl) of test sample were then injected into an analytical column (BEH C18, 2.1 × 100 mm, 1.7 μm, Waters) at a flow rate of 0.4 ml/min. The mobile phase consisted of water containing 0.1% formic acid (A) and acetonitrile containing 0.1% formic acid (B). The linear gradient was as follows: 0–1 min, 10% B; 1–5 min, 10–40% B; 5–9 min, 40–60% B; 9–11.2 min, 60–100% B; 11.2–13.2 min, 100% B; 13.2–15 min; return to 10% B. The quadrupole time‐of‐flight mass spectrometer (Q‐ToF Premier™, Waters) was operated in both positive‐ and negative‐ion mode under the following conditions: capillary voltage, 2.5 kV; cone voltage, 40 V; source temperature, 110°C; desolvation temperature, 350°C. A sprayer with a reference solution of leucine‐enkephalin ([M + H]+m/z 556.2771 and [M–H]m/z 554.2615) was used as the lock mass.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!