The largest database of trusted experimental protocols

Kinetex c18 100

Manufactured by Phenomenex
Sourced in United States

The Kinetex C18 100 Å is a high-performance liquid chromatography (HPLC) column. It features a fully porous silica particle with a 100 Ångström pore size and a C18 bonded phase. The column is designed for the separation and analysis of a wide range of organic compounds.

Automatically generated - may contain errors

9 protocols using kinetex c18 100

1

Isolation and Purification of Columbamides from Microcystis

Check if the same lab product or an alternative is used in the 5 most similar protocols
About 5.5 g (wet weight) of M. bouillonii PNG were repetitively extracted with CH2Cl2/MeOH (2:1, v:v) yielding 389 mg of extract. Subsequently the extract was fractionated by silica gel vacuum liquid chromatography (VLC) using a stepwise gradient solvent system of increasing polarity beginning from 100% hexanes to 100% EtOAc to 100% MeOH yielding nine sub-fractions. The fraction eluting with 40% EtOAc and 60% hexanes (fraction D, 3.8 mg) was separated further using a 100 mg reversed-phase C18 solid phase extraction cartridge (Bond Elut-C18 OH, 100 mg, 1 mL, Agilent Technologies) and a stepwise gradient solvent system of 25%, 50%, 75%, 100% CH3CN/H2O, respectively. The 75% eluting fraction was further purified using reversed-phase HPLC: Phenomenex Kinetex C-18 100 Å 100 x 10 mm column with a 4 mL/min flow and a linear gradient from 75% A (A: CH3CN, B: H2O) for 3 min to 95% A in 24 min. Afterwards, the column was equilibrated back to starting conditions. The yield of columbamide A was 2 mg, columbamide B 1 mg and columbamide C 0.5 mg.
+ Open protocol
+ Expand
2

HPLC Profiling of E-cymosum Extracts

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC profiles were developed using an Agilent 1260 HPLC instrument with the specification mentioned in Section 2.1. The separation was carried out using a Phenomenex® Kinetex C18 100 Å (50 × 2.1 mm id., 5 μm) reversed-phase column. The column temperature was kept at 35 °C. Working solutions of samples were prepared by dissolving 10, 1 and 1 mg of extracts of E-cymosum (WE and BuS), isolated compounds, and standards, respectively, with the appropriate solvent for each (1mL of H2O for WE and 1 mL of MeOH for BuS, compounds and standards). All samples were filtered on membrane filters (PTFE, 0.20 μm) and injected (2 μL). Elution was carried out with water containing 0.1% (v/v) formic acid as solvent A and acetonitrile (MeCN) as solvent B, starting with a mixture of 99% of A and 1% of B, increasing the amount of solvent B as follows: 25% at 14 min, 30% at 14–18 min, 35% at 18–22 min, 95% at 22–27, holding this mixture for a minute and returning to the initial conditions at 30 min. The flow rate was 0.35 mL/min. For UV detection, the wavelength program was set at an acquisition of λ 240, 254, 280, 320 and 365 nm; the UV spectra were recorded from 230 at 400.
+ Open protocol
+ Expand
3

HPLC-UV/VIS-DAD Analysis of Lipophilic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC-UV/VIS-DAD chromatograms were collected using the following equipment purchased from Shimadzu Corporation (Kyoto, Japan): control unit CBM-20A, two pumps LC-20AD (high-pressure gradient), degasser DGU-20A5R, autosampler SIL-20AC HT, column oven CTO-20AC, and detector SPD-M20A. Column: Phenomenex Kinetex C18 100, dimensions: 4.6 mm × 150 mm. The system was working under the control of LabSolutions software (Shimadzu Corporation) Version 5.71SP2. The mobile phase was water (A) and acetonitrile (B) with a flow rate—of 1 mL/min. The injection volume was set to 10 μL for the standards and 1 μL for the extracts. Each run started at 10% of B and increased to 12% in 7 min and then 100% in 25 min. The column temperature was set at 30 °C. In order to purge the highly lipophilic compounds present in the extracts, the flow was held from 25 min to 45 min at 100% B, at 60 °C. Then, the B concentration was linearly lowered to 10%, and the temperature was cooled down to 30 °C to restore initial conditions. Every development was triplicated, and the presented results are averages from the obtained results (Figure S1). The RSD values of retention times and peak areas were lower or equal to 5%. Diode-array detection was set to collect data at 218 nm and 254 nm.
+ Open protocol
+ Expand
4

Characterization of Red Wine Pigments

Check if the same lab product or an alternative is used in the 5 most similar protocols
The use of an HPLC with a diode array detector and electro spray ionization coupled with a mass spectroscopy (DAD-ESI/MS) was used to identify and characterize malvidin-3-O-glucoside, vitisin B, malvidin-3-O-glucoside acyl derivatives and polymeric pigments with malvidin-3-O-glucoside; the HPLC was a Agilent Technologies™ 1100 (Palo Alto, CA, USA) chromatograph with a column RP KineteX C18 100Å (100 × 4.6 mm; 2.6 μm) (PhenomeneX, Torrance, CA, USA). The eluents used comprised eluent A (water/formic acid 95:5 v/v) and eluent B (methanol/formic acid 95:5 v/v), with the following gradient of eluent B (0.8 mL/min): from 20% to 50% from 0 min to 27 min; 50% from 27 min to 28 min, and finally, from 50% to 20% from 28 min to 29 min until a steady state was reached. According to Loira et al. [7 (link)], malvidin-3-O-glucoside has been used as an eXternal standard at a wavelength of 525 nm for the quantification of all pigments, while the identification was carried out with mass spectrometry positive scanning from 100 to 1500 m/z from 0 min to 23 min. The detection limit was set to 0.1 mg/L [41 (link)].
+ Open protocol
+ Expand
5

Biotinylated Amyloid-β Peptide Synthesis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Freshly purified Aβ(M1–40) monomers (120 μM in 50 mM Tris pH 8.5) were supplemented with 150 mM NaCl, 5 mM CaCl2, and 1 mM TCEP and reacted overnight at room temperature with 50μM SrtAβ and 1 mM GGGK(Btn). After desalting in a 3 kDa molecular weight cutoff spin filter, the reaction mixture was lyophilized and then dissolved in 7 M guanidium chloride, 50 mM Tris pH 7.5, 2 mM EDTA and ran on a Kinetex C18 100 Å (150×30 mm, 5 μm, Phenomenex) column. The acetonitrile concentration was increased from 10 to 35% over the first 5 minutes, 35 to 38% over the next 6 minutes, and then from 38 to 90% over the next 5 minutes. The major peak eluted at 12.8 minutes. This was confirmed to be Aβ(M1–37- GGGK(Btn)) by LC/MS (m/z = 4731.98 observed, 4730.27 expected) and the product was lyophilized and stored at −20 °C for later use.
+ Open protocol
+ Expand
6

Biotinylated Amyloid-β Peptide Synthesis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Freshly purified Aβ(M1–40) monomers (120 μM in 50 mM Tris pH 8.5) were supplemented with 150 mM NaCl, 5 mM CaCl2, and 1 mM TCEP and reacted overnight at room temperature with 50μM SrtAβ and 1 mM GGGK(Btn). After desalting in a 3 kDa molecular weight cutoff spin filter, the reaction mixture was lyophilized and then dissolved in 7 M guanidium chloride, 50 mM Tris pH 7.5, 2 mM EDTA and ran on a Kinetex C18 100 Å (150×30 mm, 5 μm, Phenomenex) column. The acetonitrile concentration was increased from 10 to 35% over the first 5 minutes, 35 to 38% over the next 6 minutes, and then from 38 to 90% over the next 5 minutes. The major peak eluted at 12.8 minutes. This was confirmed to be Aβ(M1–37- GGGK(Btn)) by LC/MS (m/z = 4731.98 observed, 4730.27 expected) and the product was lyophilized and stored at −20 °C for later use.
+ Open protocol
+ Expand
7

HPLC Analysis of Arbutin Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC analyses were performed according to Rathi et al., 2019 [36 (link)], with some modifications, as described below. The quantification was conducted on Waters HPLC system, consisting of Waters 1525 Binary pumps with Waters 2487 Dual λ Absorbance Detector (Waters, Milford, MA, USA), equipped with a reverse-phase Kinetex® C18, 100 Å (150 × 4.6 mm, 5 µm) core-shell column (Phenomenex, Torrance, CA, USA), operating at 26 °C. The mobile phase consisted of water (solvent A) and acetonitrile (solvent B), with a flow rate 0.5 mL/min and gradient elution as follows: 0–2 min 99% A; 2–6 min decreased to 40% A; 6–15 min gradually increased back to 99% A arbutin was detected at 285 nm. Six solutions of the arbutin, with different concentrations ranging from 25 to 200 μg/mL dissolved in methanol, were used to construct a linear calibration curve.
+ Open protocol
+ Expand
8

Metabolite Profiling by UPLC-PAD and UHPLC-ESI-QToF-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
A chemical profile was obtained using ultra-high-performance liquid chromatography coupled with photodiode array detection (UPLC-PAD) on an ACQUITY UPLC system (Waters, Milford, MA, USA), which consisted of a photodiode array detector, quaternary pump, degasser, column oven, and autosampler. The separation of metabolites was performed using Phenomenex® Kinetex C18 100 Å (75 mm × 2.1 mm; 2.6 µm). The mobile phase was formic acid 0.1% (component A) and acetonitrile (component B) as follows: 0–8 min, 16% B; 8–10 min, 16–20% B; 10–12 min, 20% B; 12–15 min, 20–22% B; 15–18 min, 22–27% B; 18–20 min, 27–30% B; 20–23 min, 30–35% B; 23–28 min, 35% B; 28–34 min, 35–90% B; and then back to 16% B in 1 min [52 (link)]. The samples were injected automatically at 3 µL. The column temperature was maintained at 30 °C. The flow rate was set at 0.4 mL/min. The photo diode array (PDA) detection was set at 340 nm. Ultra-high-performance liquid chromatography–electrospray ionization–quadrupole time-of-flight–mass spectrometry (UHPLC-ESI-QToF-MS) analysis was performed on Nexera LCMS 9030 Shimadzu Scientific-Instruments (Columbia, MD, USA) equipment. The chromatographic conditions were the same as those described previously: mass spectrum detection for negative ion mode. The capillary potentials were set at +3 kV, drying gas temperature 250 °C, and the flow rate of drying gas 350 L/min.
+ Open protocol
+ Expand
9

Metabolite Profiling by UPLC-PAD and UHPLC-ESI-QToF-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
A chemical profile was obtained using ultra-high-performance liquid chromatography coupled with photodiode array detection (UPLC-PAD) on an ACQUITY UPLC system (Waters, Milford, MA, USA), which consisted of a photodiode array detector, quaternary pump, degasser, column oven, and autosampler. The separation of metabolites was performed using Phenomenex® Kinetex C18 100 Å (75 mm × 2.1 mm; 2.6 µm). The mobile phase was formic acid 0.1% (component A) and acetonitrile (component B) as follows: 0–8 min, 16% B; 8–10 min, 16–20% B; 10–12 min, 20% B; 12–15 min, 20–22% B; 15–18 min, 22–27% B; 18–20 min, 27–30% B; 20–23 min, 30–35% B; 23–28 min, 35% B; 28–34 min, 35–90% B; and then back to 16% B in 1 min [52 (link)]. The samples were injected automatically at 3 µL. The column temperature was maintained at 30 °C. The flow rate was set at 0.4 mL/min. The photo diode array (PDA) detection was set at 340 nm. Ultra-high-performance liquid chromatography–electrospray ionization–quadrupole time-of-flight–mass spectrometry (UHPLC-ESI-QToF-MS) analysis was performed on Nexera LCMS 9030 Shimadzu Scientific-Instruments (Columbia, MD, USA) equipment. The chromatographic conditions were the same as those described previously: mass spectrum detection for negative ion mode. The capillary potentials were set at +3 kV, drying gas temperature 250 °C, and the flow rate of drying gas 350 L/min.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!