The largest database of trusted experimental protocols

49 protocols using quantibrite pe beads

1

Quantification of Monocyte HLA-DR Expression in Sepsis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Quantification of mHLA-DR on circulating monocytes from patients with sepsis was performed using a standardized flow cytometry assay (Demaret et al., 2013 (link)). Anticoagulated whole blood was stained with antibodies to HLA-DR and CD14 mixture (QuantiBrite anti-HLA-DR PE/anti-monocyte CD14 PerCP-Cy5.5, Becton Dickinson, CA), prior to lysis, washing, and acquiring of HLA-DR expression on monocytes using BD FACSCanto (BD Biosciences, Franklin Lakes, NJ). To quantify, the median fluorescence intensity of the entire monocyte population was then transformed to number of antibodies per cell (AB/C) using calibrated PE beads (BD QuantiBRITE PE BEADS, Becton-Dickinson San Jose, CA). Results were analyzed using FlowJo (TreeStar Inc, Ashland, OR). and mixed with calibrated PE beads (BD QuantiBRITE PE BEADS, Becton Dickinson, CA).
+ Open protocol
+ Expand
2

Quantifying PE-Antibody Molecules

Check if the same lab product or an alternative is used in the 5 most similar protocols
Quanti-BRITE PE Beads (Becton Dickinson, Cat # 340495, Lot # 30746, Lot # 77602) were used in accordance with the protocol [10 ] to relate the measured signal to the number of PE-antibody molecules. The signal per PE-antibody molecule was determined in each experiment.
+ Open protocol
+ Expand
3

Inducible CAIX Expression in Cells

Check if the same lab product or an alternative is used in the 5 most similar protocols
Skrc-59 Tet-On CAIX cells were counted and seeded at a concentration of 3*105 cells/mL into 6-well plates. The respective doxycycline concentration (0.1–1,000 ng/mL) was added and at the indicated time points and the samples were collected by detaching the cells with trypsin (Corning). The samples were incubated with human Fc blocking solution (1:500) for 10 min at RT, then washed and stained with anti-CAIX-PE antibody (clone REA658, Miltenyi Biotec) for 15 min at RT. After three washing steps, the samples were fixed with 2% paraformaldehyde (PFA) for 15 min at RT and were analyzed together with QuantiBrite PE beads (Becton Dickinson) using flow cytometry (LSRFortessa, BD Biosciences). Data was analyzed using FlowJo software (FlowJo LLC).
+ Open protocol
+ Expand
4

FACS Analysis of Cell Surface Markers

Check if the same lab product or an alternative is used in the 5 most similar protocols
For FACS analysis adherent cells at about 70% confluence were detached using 0.1 mM EDTA in PBS (Sigma Aldrich, Darmstads, Germany), centrifuged and re-suspended in PBS containing 0.2% BSA. Cell aliquots (2.5 × 105 cells) were treated with primary monoclonal antibody PE conjugate, (Millipore, Burlington, MA, USA) or isotype control (Santa Cruz Biotechnology, Dallas, TX, USA), at the same concentration (14 µg/mL), whole a reaction volume of 50 µL for 30 min at 4 °C. After washing, the cells were analyzed by using a flow cytometer equipped with a 488 nm argon laser (FACScan, Becton Dickinson, Franklin Lakes, NJ, USA). A total of 20,000 events per sample were collected. Integrin quantification with Quantibrite PE beads (BectonDickinson Biosciences) was evaluated as previously described [46 (link)]. Values of fluorescence intensity were obtained from the histogram statistic of CellQuest software.
+ Open protocol
+ Expand
5

CD20 Expression Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
For CD20 ABC quantitation, 4 set of precalibrated Quantibrite PE beads and Quantibrite CD20 PE antibody (Becton Dickinson, San Jose) were used. Quantibrite CD20 PE antibody is produced to have 1 fluorochrome per antibody.
+ Open protocol
+ Expand
6

Quantifying Variability in Flow Cytometry

Check if the same lab product or an alternative is used in the 5 most similar protocols
Variance in flow cytometric measurements was determined from a 5 × 3 × 3 experiment, testing the effect of the donor (biological variability), instrument (technical variability), and day (environmental variability) on measured fluorescence intensities. PB was collected from 5 healthy donors on 3 non-consecutive days. The samples were stained with the 4C MRD panel and analyzed on three different instruments. The antibody binding capacity of the PE-tagged targets CD56, CD22, CD38, and CD79b was determined from the consecutive measurements of three of the five donors using Quantibrite PE beads (BD BioSciences, Heidelberg, Germany) according to the manufacturer’s instructions. The MFIs for different populations were used to evaluate the effect of the selected parameter on overall variability using the eta squared value from ANOVA. F-statistics were determined to test models using only the donor as a factor against models using all factors.
+ Open protocol
+ Expand
7

Quantification of Monocyte HLA-DR Expression

Check if the same lab product or an alternative is used in the 5 most similar protocols
Expression of human leucocyte antigen-antigen D-related (HLA-DR) on CD14+ monocytes was measured by flow cytometry according to the manufacturer’s recommendations (Quantibrite HLA-DR/Monocyte antibody cocktail, BD Bioscience, Heidelberg, Germany). Briefly, 50 μL of EDTA-anticoagulated whole blood was incubated with 20 μl of the antibody cocktail and incubated for 30 min. Afterwards, erythrocyte lysis was performed by adding 450 μl FACS Lysing solution (BD Bioscience, Heidelberg, Germany) and further incubation for 15 minutes. Measurement was immediately done on a FACSCalibur flow cytometer (BD Bioscience). In addition, a 4-point calibration curve (Quantibrite PE Beads, BD Bioscience) was also measured daily to enable the transformation of the measured sample values to molecules of HLA-DR.
+ Open protocol
+ Expand
8

CD19 and CD22 Expression in Leukemia Cell Lines

Check if the same lab product or an alternative is used in the 5 most similar protocols

EXAMPLE 2

This example demonstrates the expression of CD22 and CD19 antigens on leukemia cell lines.

Human leukemia cell lines (REH, SEM, NALM6, KOPN8, Daudi, Raji, and K562) were evaluated for the expression level of CD19 and CD22 on the cell surface using QUANTI-BRITE PE beads (BD Biosciences) and PE-labeled anti-CD19 and anti-CD22 antibody (Table 2). “Receptor Number Per Cell” indicates the approximate absolute number of molecules per cell on each of the indicated cell lines. Data were calculated by determining antibodies bound per cell (ABC) using the CELLQUEST software (BD) data analysis tools in accordance with the manufacturer's instructions.

TABLE 2
Leukemia Cell LineReceptor Number Per Cell
REH CD1915,100
SEM CD1950,800
NALM-6 CD1950,500
KOPN-8 CD1960,800
Daudi CD1915,000
Raji CD1950,000
K562 CD19<100
REH CD227,000
SEM CD227,000
NALM-6 CD228,000
KOPN-8 CD2215,300
Daudi CD228,000
Raji CD2260,800
K562 CD22<200

+ Open protocol
+ Expand
9

Quantifying Monocyte HLA-DR Expression by Flow Cytometry

Check if the same lab product or an alternative is used in the 5 most similar protocols
To quantify the expression of HLA-DR on monocyte surface, flow cytometry (FACSVerse flow cytometer, BD Bioscience, Heidelberg, Germany) was conducted after incubating 50 μl ETDA-anti-coagulated whole blood with 20 μl anti-HLA-DR antibody (Quantibrite anti-HLA-DR/Monocyte antibody, BD Bioscience, Heidelberg, Germany) and after lysis of erythrocytes (FACS Lysing solution BD Bioscience, Heidelberg, Germany) as previously described and according to manufacturer`s instructions [18 (link)]. HLA-DR surface expression was measured as median fluorescence intensity after gating out monocytes based on CD14 expression (S10 Fig).
For quantification of HLA-DR molecules on cell surface (`counts per monocyte´), determined sample values were converted using daily measured 4-point calibration curves (Quantibrite PE Beads, BD Bioscience, Heidelberg, Germany).
+ Open protocol
+ Expand
10

Quantifying CD33 Surface Expression

Check if the same lab product or an alternative is used in the 5 most similar protocols
Tumor line CD33 surface expression was determined by flow cytometry. The anti-CD33 antibody clones AC104.3E3 (Miltenyi Biotec), WM53 and HIM3-4 (both from BioLegend) were used. Isotype-controls were used for negative gating. Estimation of CD33 site density on cell lines of interest was performed using QuantiBRITE PE beads (BD Biosciences, San Jose, CA) using the antibodies bound per cell (ABC) method as per manufacturer's protocol. Beads conjugated with PE molecules at four different intensities were used to generate a standard curve, and tumor cells stained with anti CD33 antibodies conjugated to PE beads were acquired under identical settings. The intensity of PE staining for each cell line was then used to extrapolate the number of PE molecules per tumor cell.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!