The largest database of trusted experimental protocols

Zic hilic guard column

Manufactured by Merck Group
Sourced in United Kingdom

The ZIC-HILIC guard column is a laboratory equipment product designed to protect the main analytical column from contamination and extend its lifetime. It uses zwitterionic hydrophilic interaction liquid chromatography (ZIC-HILIC) technology to separate and retain analytes.

Automatically generated - may contain errors

8 protocols using zic hilic guard column

1

Quantitative LC-MS/MS Analysis of Malonate and Succinate

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS/MS analysis of malonate and succinate was performed using an LCMS-8060 mass spectrometer (Shimadzu, UK) with a Nexera UHPLC system (Shimadzu, UK). Samples were stored in a refrigerated autosampler (4 °C) upon injection of 5 μl into a 15 μl flowthrough needle. Separation was achieved using a SeQuant ZIC-HILIC column (3.5 μm, 100 Å, 150 × 2.1 mm, 30 °C column temperature; Merck Millipore, UK) with a ZIC-HILIC guard column (200 Å, 1 × 5 mm). A flow rate of 200 μl/min was used with mobile phases of: A) 10 mM ammonium bicarbonate (pH unchanged); and B) 100% acetonitrile. A gradient of 00.1 min, 80% MS buffer B; 0.1–4 min, 80–20% B; 4–10 min, 20% B; 10–11 min, 20–80% B; and 11–15 min, 80% B was used. The mass spectrometer was operated in negative ion mode with multiple reaction monitoring (MRM), and spectra were acquired using the LabSolutions software (Shimadzu, UK), with compound quantities calculated from relevant standard curves in MS extraction buffer (50% (v/v) methanol, 30% (v/v) acetonitrile, 20% (v/v) MS-grade water) compared with 1 nmol of relevant internal standard either [13C3]-malonate or [13C4]-succinate for malonate and succinate, respectively (Supplementary Fig. 1).
+ Open protocol
+ Expand
2

Quantitative LC-MS/MS Analysis of Malonate and Succinate

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS/MS analysis of malonate and succinate was performed using an LCMS-8060 mass spectrometer (Shimadzu, UK) with a Nexera UHPLC system (Shimadzu, UK). Samples were stored in a refrigerated autosampler (4 °C) upon injection of 5 μl into a 15 μl flowthrough needle. Separation was achieved using a SeQuant ZIC-HILIC column (3.5 μm, 100 Å, 150 × 2.1 mm, 30 °C column temperature; Merck Millipore, UK) with a ZIC-HILIC guard column (200 Å, 1 × 5 mm). A flow rate of 200 μl/min was used with mobile phases of: A) 10 mM ammonium bicarbonate (pH unchanged); and B) 100% acetonitrile. A gradient of 00.1 min, 80% MS buffer B; 0.1–4 min, 80–20% B; 4–10 min, 20% B; 10–11 min, 20–80% B; and 11–15 min, 80% B was used. The mass spectrometer was operated in negative ion mode with multiple reaction monitoring (MRM), and spectra were acquired using the LabSolutions software (Shimadzu, UK), with compound quantities calculated from relevant standard curves in MS extraction buffer (50% (v/v) methanol, 30% (v/v) acetonitrile, 20% (v/v) MS-grade water) compared with 1 nmol of relevant internal standard either [13C3]-malonate or [13C4]-succinate for malonate and succinate, respectively (Supplementary Fig. 1).
+ Open protocol
+ Expand
3

LCMS-based Succinate Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS/MS analysis of succinate was performed using an LCMS-8060 mass spectrometer (Shimadzu, UK) with a Nexera X2 UHPLC system (Shimadzu, UK). Samples were stored in a refrigerated autosampler (4 °C) before injection of 5 μl using a 15 μl flowthrough needle. Separation was achieved using a SeQuant® ZIC®-HILIC column (3.5 μm, 100 Å, 150 x 2.1 mm, 30 °C column temperature; MerckMillipore, UK) with a ZIC®-HILIC guard column (200 Å, 1 x 5mm). A flow rate of 200 μl/min was used with mobile phases of Buffer A: 10 mM ammonium bicarbonate and Buffer B: 100% acetonitrile. A gradient of 0-0.1 min, 80% MS buffer B; 0.1-4 min, 80%-20% B; 4-10 min, 20% B, 10-11 min, 20%-80% B; 11-15 min, 80% B was used. The mass spectrometer was operated in negative ion mode with multiple reaction monitoring (MRM) and spectra were acquired using Labsolutions software (Shimadzu, UK), with compound quantities calculated from relevant standard curves in MS extraction buffer and comparing against [13C4]-succinate internal standard.
+ Open protocol
+ Expand
4

Quantification of Succinate and Malonate

Check if the same lab product or an alternative is used in the 5 most similar protocols
To quantify succinate and malonate a LCMS-8060 mass spectrometer (Shimadzu, UK) with a Nexera X2 UHPLC system (Shimadzu, UK) was used as described before [29 (link)]. Briefly, metabolites were separated using a SeQuant® ZIC®-HILIC column (3.5 μm, 100 Å, 150 x 2.1 mm, 30 °C column temperature; MerckMillipore, UK) with a ZIC®-HILIC guard column (200 Å, 1 x 5 mm) at a flow rate of 200 μl/min with mobile phases of A) 10 mM ammonium bicarbonate and B) 100% acetonitrile. The MS was operated in negative ion mode with multiple reaction monitoring (MRM) and spectra acquired using Labsolutions software (Shimadzu, UK). Succinate and malonate were quantified using standard curves relative to [13C4]-succinate or [13C3]-malonate internal standard.
+ Open protocol
+ Expand
5

LCMS-Based Metabolite Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples were analysed using an LCMS-8060 mass spectrometer (Shimadzu, UK) with a Nexera X2 UHPLC system (Shimadzu, UK). Sample separation was achieved using a SeQuant®ZIC®-HILIC column (3.5 μg, 100 Å, 150 × 2.1 mm, 30 °C column temperature; MerckMillipore, UK) with a ZIC®-HILIC guard column (200 Å, 1 × 5 mm). A flow rate of 200 μl/min was used with mobile phases of 10 mM ammonium bicarbonate (pH unchanged) and B) 100% acetonitrile. The mass spectrometer was operated in negative ion mode with multiple reaction monitoring (MRM). Sample spectra were acquired using Labsolutions software (Shimadzu, UK) and the peak area for each compound of interest measured relative to the internal standard. Sample concentrations were then calculated from a standard curve of known compound concentrations produced by LC-MS/MS in a similar manner.
+ Open protocol
+ Expand
6

Succinate Quantification by LC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC–MS/MS analysis of succinate was performed using an LCMS-8060 mass spectrometer (Shimadzu, UK) with a Nexera X2 UHPLC system (Shimadzu). Samples were stored in a refrigerated autosampler (4°C) before injection of 5 μL using a 15 μL flowthrough needle. Separation was achieved using a SeQuant® ZIC®-HILIC column (3.5 μm, 100 Å, 150 mm × 2.1 mm, 30°C column temperature; MerckMillipore, UK) with a ZIC®-HILIC guard column (200 Å, 1 mm × 5 mm). A flow rate of 200 μL/min was used with mobile phases of buffer A: 10 mM ammonium bicarbonate and buffer B: 100% acetonitrile. A gradient of 0–0.1 min, 80% MS buffer B; 0.1–4 min, 80–20% B; 4–10 min, 20% B, 10–11 min, 20–80% B; 11–15 min, 80% B was used. The mass spectrometer was operated in negative ion mode with multiple reaction monitoring and spectra were acquired using LabSolutions software (Shimadzu), with compound quantities calculated from relevant standard curves in MS extraction buffer and comparing against [13 (link)C4]-succinate internal standard.
+ Open protocol
+ Expand
7

LC-MS/MS Quantification of Malonate

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS/MS measurement
of malonate was performed using an LCMS-8060 mass spectrometer (Shimadzu,
UK) with a Nexera UHPLC system (Shimadzu, UK), as described previously.31 (link) Samples were stored in a refrigerated autosampler
(4 °C) until injection of 5 μL into a 15 μL flowthrough
needle. A SeQuant ZIC-HILIC column (3.5 μm, 100 Å, 150
× 2.1 mm, 30 °C column temperature; Merck Millipore, UK)
with a ZIC-HILIC guard column (200 Å, 1 × 5 mm) was used
for liquid chromatography. A flow rate of 0.2 mL/min was used with
mobile phases of (A) 10 mM ammonium bicarbonate (pH unchanged) and
(B) 100% acetonitrile. A gradient of 0–0.1 min, 80% MS buffer
B; 0.1–4 min, 80–20% MS buffer B; 4–10 min, 20%
MS buffer B; 10–11 min, 20–80% MS buffer B; and 11–15
min, 80% MS buffer B was used. The mass spectrometer was operated
in negative ion mode with multiple reaction monitoring (MRM), and
spectra were acquired using Labsolutions software (Shimadzu, UK),
with malonate levels calculated from a standard curve in MS extraction
buffer (50% (v/v) methanol, 30% (v/v) acetonitrile, and 20% (v/v)
MS-grade water) compared to 1 nmol of MS internal standard (13C3-malonate).
+ Open protocol
+ Expand
8

Glycopeptide and O-Glycan Analysis via LC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS was performed on an Agilent Technologies
(Santa Clara, CA) Infinity 1290 LC system coupled via a dual-source
AJS electrospray interface to an Agilent Technologies 6560B ESI Ion
Mobility Q-TOF. Glycopeptide standards were analyzed with a SeQuant
ZIC-HILIC column (20 × 2.1 mm2, 3.5 μm particles;
Merck, Darmstadt, Germany), with 0.1% (v/v) formic acid as eluent
A and ACN as eluent B, using a linear gradient from 90–50%
B over 5 min and maintaining 50% B for 8 min. MS was performed in
positive ion mode for glycans 1 and 3 and
negative ion mode for glycans 2 and 4. O-glycans, released from BSM oxidatively or by β-elimination,
were separated with a ZIC-HILIC column (150 × 2.1 mm2, 3 μm particles) with a ZIC-HILIC guard column (20 ×
2.1 mm2, 3 μm particles; Merck, Darmstadt, Germany)
using 5 mM ammonium formate (pH 6.5) as eluent A and ACN as eluent
B. Chromatographic separation was achieved using 85% B for 5 min,
followed by a linear gradient to 50% B over 25 min at 0.2 mL/min. O-glycans released by β-elimination were additionally
analyzed with the same gradient using 10 mM ammonium bicarbonate (pH
7.8) as eluent A to preserve sulfated moieties better. MS analysis
was performed in negative ion mode with a capillary voltage of 3500
V, nozzle voltage of 2000 V, nebulizer pressure of 40 psi, drying
gas flow rate of 300 °C at 8 L/min, and sheath gas temperature
of 300 °C at 11 L/min.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!