Example 4
Antibody solutions containing romosozumab PARG (SEQ ID NO: 8) C-terminal variant or wild-type romosozumab are measured using a cone and plate. The solutions are concentrated up to 120 mg/mL according to approximate volume depletion, and final concentrations are determined (±10%) using the proteins absorbance at 280 nm (after dilution to end up within 0.1-1 absorbance units (AU)) and a protein specific extinction coefficient. Viscosity analysis is performed on a Brookfield LV-DVIII cone and plate instrument (Brookfield Engineering, Middleboro, Mass., USA) using a CP-40 spindle and sample cup or an ARES-G2 rheometer (TA Instruments, New Castle, Del., USA) using a TA Smart Swap 2 degree cone/plate spindle. All measurements are performed at 25° C. and controlled by a water bath attached to the sample cup. Multiple viscosity measurements were collected, manually within a defined torque range (10-90%) by increasing the RPM of the spindle. Measurements are averaged in order to report one viscosity value per sample to simplify the resulting comparison chart.