Pl50 instrument

Manufactured by Agilent Technologies

The PL50 instrument is a laboratory equipment product offered by Agilent Technologies. It is designed to perform core functions related to the analysis and characterization of samples. The detailed technical specifications and intended use of this product are not available in this response.

Automatically generated - may contain errors

3 protocols using pl50 instrument

1

Polymer Characterization by SEC

Check if the same lab product or an alternative is used in the 5 most similar protocols
2 mg mL−1 solution of each polymer in DMF containing 0.1% (w:v) LiBr were incubated for 16 h at room temperature. Analyte samples were filtered through a nylon membrane with 0.22 μm pore size before injection. An Agilent PL50 instrument equipped with differential refractive index (DRI) and UV detectors was used for all the measurements. The system was equipped with two x PolarGel M columns (300 × 7.5 mm) and a PolarGel 5 µm guard column, connected in series. The eluent was DMF containing 0.1% LiBr. All experiments were performed at a flow rate of 1 ml min−1 at 50 °C. Experimental molar mass (Mn,SEC) and dispersity (Đ) values of synthesized polymers were determined by comparison with poly(methyl methacrylate) standards (Agileny EasyVials) using Agilent GPC/SEC software. All data were elaborated using Origin.
+ Open protocol
+ Expand
2

Polymer Molecular Weight Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
GPC was measured using an Agilent PL50 instrument with a differential refractive index detector. The instrument contained two PolarGel H columns (300 mm × 7.5 mm) and a PolarGel 5 μm guard column. DMF with 0.1% LiBr additive was used as the eluent. The system ran at 1 mL min−1 (50 °C), with an injection volume of 100 μL. The samples were prepared by filtering them through 0.22 μm pore size PTFE membranes, before injection. Agilent EasyVial poly(methyl methacrylate) standards were used to calibrate the instrument and output data were analysed using Agilent GPC/SEC software.
+ Open protocol
+ Expand
3

Gel Permeation Chromatography Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
GPC was measured using an Agilent PL50 instrument with a differential refractive index detector. The instrument contained two PolarGel H columns (300 mm × 7.5 mm) and a PolarGel 5 μm guard column. DMF with 0.1% LiBr additive was used as the eluent. The system ran at 1 mL min−1 (50 °C), with an injection volume of 100 μL. The samples were prepared by filtering them through 0.22 μm pore size PTFE membranes, before injection. Agilent EasyVial poly(methyl methacrylate) standards were used to calibrate the instrument and output data were analyzed using Agilent GPC/SEC software.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.