Hunter software (version 4.4) for instrument control and data processing (Agilent Technologies 2009). Detailed parameters for method quantification, adjustments and validation are outlined in (Webster et al. 2021, under review) and in (Webster et al. 2021a ) for analysis of environmental samples. In brief, Argon as the dilution gas was added before sample introduction into the plasmas.
Helium collision cell mode using the 4 th generation Octopole Reaction System was used for analysis of element isotopes 11 B, 55 Mn, 56 Fe, 59 Co, 60 Ni, 63 Cu, 66 Zn, 95 Mo. Inter-element corrections were made for possible isobaric interferences of high 115 Sn on 115 In used as internal standard. Hydrogen reaction gas was used to measure 78 Se to improve sensitivity and remove the dual-charged interference from 156 Gd observed in some samples during initial investigation. Optimization for sensitivity and low oxide ratios (CeO/Ce < 0.3%) was performed daily.