Exionlc infinity series uhplc system

Manufactured by AB Sciex

The ExionLC Infinity series UHPLC System is a high-performance liquid chromatography (HPLC) instrument designed for ultra-high-performance liquid chromatography (UHPLC) applications. The system is capable of operating at high pressures, enabling the use of smaller particle size columns for improved resolution and faster separation times.

Automatically generated - may contain errors

3 protocols using exionlc infinity series uhplc system

1

LC-MS/MS Analysis of Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
The ExionLC Infinity series UHPLC System (AB Sciex) with the Kinetex C18 column (2.1 × 100 mm, 1.7 μm; Phenomen) was used for LC/MS analyses. The mobile phase A consisted of 40% water and 60% acetonitrile with 10 mmol/L ammonium formate, and the mobile phase B consisted of 10% acetonitrile and 90% isopropanol, with 10 mmol/L ammonium formate. The elution gradient was set to 0–12.0 min, 40–100% B; 12.0–13.5 min, 100% B; 13.5–13.7 min, 100–40% B; 13.7–18.0 min, 40% B. The column temperature was set to 45 ℃. The injection volume was 2 μL (pos) and 6 μL (neg), respectively.
The TripleTOF 5600 Mass Spectrometer was used to acquire the MS/MS spectra on an information-dependent basis (IDA) with the acquisition software (Analyst TF 1.7, AB Sciex) to evaluate the full scan survey MS data. The most intensive 12 precursor ions with intensity > 100 were selected for MS/MS at collision energy (CE) of 45 eV. The ESI source conditions were set as follows: Gas 1 as 60 psi, Gas 2 as 60 psi, Curtain Gas as 30 psi, source temperature as 600℃, declustering potential as 100 V, and ion spray voltage floating (ISVF) as 5000 V or − 3800 V in positive or negative modes, respectively16 (link).
+ Open protocol
+ Expand
2

UHPLC-QTOF-MS for Metabolite Profiling

Check if the same lab product or an alternative is used in the 5 most similar protocols
Ultra (high) performance liquid chromatography (UHPLC) separation was carried out using an ExionLC Infinity series UHPLC System (AB Sciex), equipped with a UPLC BEH Amide column (2.1 × 100 mm, 1.7 μm, Waters). The mobile phase consisted of 25 mmol/L ammonium acetate and 25 mmol/L ammonia hydroxide ( CNW Technologies) in water (pH = 9.75) (A) and acetonitrile (B). The analysis was carried out with an elution gradient. The column temperature was 25℃. The autosampler temperature was 4 ℃, and the injection volume was 2 μL (pos) or 2 μL (neg), respectively.
TripleTOF 5600 mass spectrometry (AB Sciex) was used for its ability to acquire MS/MS spectra on an information-dependent basis (IDA) during LC/MS experiments. In this mode, the acquisition software (Analyst TF 1.7, AB Sciex) continuously evaluates the full scan survey MS data as it collects and triggers the acquisition of MS/MS spectra depending on preselected criteria. In each cycle, the most intensive 12 precursor ions with intensity above 100 were chosen for MS/MS at a collision energy (CE) of 30 eV. In the detection process, standard products and blank controls were strictly used for quality control, and data quality control was also carried out. The UHPLC-QTOF-MS analysis was performed at Biomarker Technologies Corporation, Beijing, China (https://international.biocloud.net/).
+ Open protocol
+ Expand
3

UHPLC-MS/MS Metabolomic Profiling Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
The UHPLC separation was carried out using a ExionLC Infinity series UHPLC System (AB Sciex), equipped with a Kinetex C18 column (2.1 * 100 mm, 1.7 μm, Phenomen). The mobile phase A consisted of 40% water, 60% acetonitrile, and 10 mmol/L ammonium formate. The mobile phase B consisted of 10% acetonitrile and 90% isopropanol, which was added with 50 mL 10 mmol/L ammonium formate. The analysis was carried with elution gradient as follows: 0~12.0 min, 40%~100% B; 12.0~13.5 min, 100% B; 13.5~13.7 min, 100%~40% B; 13.7~18.0 min, 40% B. The column temperature was 45 ℃.The injection volume was 2 μL (pos) or 2 μL (neg), respectively.
The TripleTOF 5600 mass spectrometer was used for MS/MS spectra detection on an informationdependent basis (IDA). In this mode, the most intensive 12 precursor ions with intensity above 100 were chosen for MS/MS at collision energy (CE) of 45 eV (12 MS/MS events with accumulation time of 50 msec each). ESI source conditions were set as following: Gas 1 as 60 psi, Gas 2 as 60 psi, Curtain Gas as 30 psi, Source Temperature as 600 ℃, Declustering potential as 100 V, Ion Spray Voltage Floating (ISVF) as 5000 V or -3800 V in positive or negative modes, respectively (22) .
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!