The largest database of trusted experimental protocols

Syncronis c18 column

Manufactured by Thermo Fisher Scientific
Sourced in United States, Japan, Germany

The Syncronis C18 column is a reversed-phase liquid chromatography column designed for the separation and analysis of a wide range of compounds. It features a spherical, fully porous silica-based stationary phase with a C18 alkyl bonding, providing high separation efficiency and reproducibility.

Automatically generated - may contain errors

50 protocols using syncronis c18 column

1

Separation and Purification of Phenolic Acids

Check if the same lab product or an alternative is used in the 5 most similar protocols
The separation of phenolic acids of the ME was performed with an HPLC (HITACHI L-6200, Hitachi High Technologies Com., Tokyo, Japan) system with a Syncronis C18 column (5 μm, 4.6 mm × 250 mm; Thermo Fisher Scientific Inc., Waltham, MA, USA). The mobile phase (1.0 mL min−1) consisted of A: 0.1% formic acid (FA; Thermo Fisher Scientific Inc., Rockford, IL, USA); and B: 100% methanol. The gradient conditions (A:B) were programmed as 0–10 min, 5–25% A and 95–75% B; 10–20 min, 25–30% A and 75–70% B; 20–40 min, 30–60% A and 70–40% B. The column was maintained at 25 °C, and the eluents were detected by UV280. Instrument control and data analysis were conducted using ECL2000 software (Analab Corp., Taipei, Taiwan).
The functional phenyl compound 2,4-dimethoxy-6-methylbenzene-1,3-diol (DMMB) was purified by preparative HPLC (PrimelineTM Gradient pump 500G; Analytical Scientific Ins., El Sobrante, CA, USA). The DMMB was separated on a preparative InspireTM C18 column (21.2 mm × 250 mm; Dikma Technologies Inc., Lake Forest, CA, USA) with 40% methanol (10.4 mL min−1). The fraction containing DMMB was detected by UV280 and collected using an Advantec SF2120 fraction collector (Advantec Toyo Kaisha, Ltd., Tokyo, Japan) simultaneously. The collected fractions were concentrated using a rotary evaporator and a pressured nitrogen gas-blowing concentrator.
+ Open protocol
+ Expand
2

HPLC-MS/MS Analysis of Bioactive Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Solutions of LSB, MLB, and ZLB were filtered through a 0.22 μm polyvinylidene difluoride (PVDF) membrane filter (Pall Corporation, Glen Cove, NY, USA) and subjected to HPLC analysis. The HPLC system coupled to a model 600E photodiode array detector (Waters Corporation, Milford, MA, USA) was performed using a Syncronis C18 column (4.6 × 250 mm inner diameter, 5 µm, Thermo Scientific, Waltham, MA, USA) with an eluting gradient as follows: 5% acetonitrile for 0–5 min; linear gradient from 5% to 70% acetonitrile for 5–35 min; 70% acetonitrile for 35–45 min. The acetonitrile solution used in the eluting gradient contained 0.5% acetic acid. The ultraviolet (UV) absorbance was detected at 280 nm. Mass spectrometric analysis was performed on a linear trap quadrupole tandem mass spectrometer (Thermo Electron, San Jose, CA, USA) equipped with an electrospray ionization interface and connected to a Surveyor LC system (Thermo Electron, Waltham, MA, USA) with a 5 μL sample loop. The analytes were separated under the same condition used for HPLC analysis. Negative ESI mode was firstly scanned ranging from m/z 150 to 1500. The other scans were set as the data-dependent MSn scan using the high-purity helium (>99.99%) as the collision gas and the relative collision energy of 30%.
+ Open protocol
+ Expand
3

Quantitative Analysis of Phenolics in N. rtanjensis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Phenolics in the methanol extract of N. rtanjensis were chromatographically separated on a Syncronis C18 column (100 × 2.1 mm) with 1.7 μm particle size (ThermoFisher Scientific, USA), thermostated at 40 °C. The mobile phase consisted of (A) 0.1% acetic acid in water and (B) acetonitrile (MS grade, FisherScientific, UK), which were applied in the following gradient elutions: 5% B in the first minute, 5–95% B from 1.0 to 16.0 min, from 95% to 5% B for 16.0–16.2 min, and 5% B until the 20th min. The flow rate was set to 0.3 mL min−1, and the injection volume was 5 μL. Settings of the TSQ Quantum Access Max QQQ mass spectrometer for the time-selected reaction monitoring (tSRM) experiment were as previously described in Čolić et al. [33 (link)]. Xcalibur software (version 2.2) was used for instrument control, data acquisition, and analysis. The total amount of compounds in samples was calculated based on the calibration curve of pure compounds and expressed as μg 100 mg−1 dry extract.
+ Open protocol
+ Expand
4

HPLC Analysis of Oolong Tea Constituents

Check if the same lab product or an alternative is used in the 5 most similar protocols
Tea infusions were prepared by adding 20 mL of boiling water to 1 g of fresh, old or baked oolong teas for 15 min. After cooling to room temperature, the brew was filtered through a 0.22 μm polyvinylidene difluoride (PVDF) membrane filter (Pall Corporation, Glen Cove, NY) for the following analysis. Chemical constituents in oolong tea infusions were analyzed by HPLC system coupled to a 600E photodiode array detector (Waters Corporation, Milford, MA), and separation was performed on the Syncronis C18 column (4.6 × 250 mm inner diameter, 5 μm, Thermo Scientific, Waltham, MA, USA). The separated condition of HPLC analysis was modified according to Shih et al. [19 (link)]. The mobile phase consisted of (A) water containing 0.5% acetic acid and (B) acetonitrile. The program for gradient elution started at 95% solvent A and 5% solvent B, increased linearly to 77% solvent A and 23% solvent B in 70 min. The column was maintained at room temperature and the injection volume was 5 μL at a flow rate of 1 mL/min. The ultraviolet (UV) absorbance detection wavelength was set at 280 nm. Epigallocatechin-3-gallate (EGCG), caffeine, gallic acid (GA), and 5-galloylquinic acid (5GA) shown in the HPLC profiles of this study were identified according to the same procedure as described previously [10 (link)].
+ Open protocol
+ Expand
5

Quantification of PTX and IR780 in Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
The amount of PTX in the samples were determined through ultra-performance liquid chromatography-mass spectrometry (UPLC-MS/MS, ThermoFisher Scientific). Thermo Syncronis C18 column (50 × 2 mm, 1.7 μm) was used at a column temperature of 40 °C. Acetonitrile/0.1% formic acid (68:32, v/v) was used as the eluent at a flow rate of 0.3 mL/min. An electrospray ionization source was used for positive-ion detection. Settings for the remaining parameters are presented in Additional file 1: Table S4. The endogenous substances in biological samples did not interfere with the detection of PTX (Additional file 1: Figure S8), and specificity was deemed good. To assay the biological samples, a linear relationship was established between the peak area and the concentration of PTX with a range of 2–5000 ng/mL (r = 0.9999), with an extraction recovery of more than 90%.
IR780 was detected using a UV-Vis Cary 8454 instrument (Agilent Technologies Inc., Santa Clara, CA, USA) at a wavelength of 780 nm. The IR780 concentration in the range of 0.25–5 µg/mL showed a linear relationship (r = 0.999) with the absorbance, and IR780 was recovered at a rate of > 90%.
+ Open protocol
+ Expand
6

Quantifying Ginsenosides in Herbal Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
The content of notoginsenoside R1 and ginsenosides Rg1 and Rb1 in the samples was determined using HPLC and the method reported in the Chinese Pharmacopoeia (2015). HPLC analyses were performed on an Ultimate 3000 series system (Thermo Fisher Scientific, China) consisting of a quaternary pump, DAD detector, and autosampler, and the data were analyzed using Chromeleon 7 software. A Thermo Syncronis C18 column (250 × 4.6 mm, 5.0 µm) was adopted for the separation. The mobile phase consisted of A (pure water) and B (acetonitrile). The gradient mode was as follows: 0–12 min, 19% A; 12–60 min, 19% ∼ 36% A. The flow rate was 1.0 mL/min, and the detection wavelength was set to 203 nm. The column temperature was 25°C, and the sample injection volume was set to 20 µL.
+ Open protocol
+ Expand
7

TCAS Quantitative Determination by HPLC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The quantitative determination of TCAS was conducted using a new method. A Thermo TSQ Vantage triple quadrupole mass spectrometer (Thermo Fisher Scientific, San Jose, CA, USA) equipped with a Waters ACQUITY Ultra Performance HPLC system and a Thermo Syncronis C18 column (100 mm × 2.1 mm, 1.9 µm I.D.) was used in this study. The mobile phase consisted of 0.1% formic acid water solution (A) and acetonitrile containing 0.1% formic acid (B). The gradient program used was as follows: 0 min 10% B, 0–1 min 20% B, 1–1.5 min 85% B, 1.5–2 min 85% B, and 2–3 min 10% B, for a total duration of 4.5 min. The column temperature was set at 25 °C and the flow rate was 0.25 mL·min−1. For quantification, the instrument was operated in the ESI positive ion multiple reaction monitoring (SIM) mode with the following optimized MS/MS conditions: transfer capillary temperature of 350 °C; spray voltage of 3500 V. The selected transitions were 271.6 for TCAS and 260 for PRO (IS). The linear quantitation ranges for TCAS ranged from 50 to 500 μg/mL. The lower limit of detection of TCAS was 50 μg/mL.
+ Open protocol
+ Expand
8

HPLC Determination of Organic Pollutants

Check if the same lab product or an alternative is used in the 5 most similar protocols
The stock solution (LTL, HDQ, QRT, P-NP, 1000 mg L−1) was prepared in MeOH, and stored at 4 °C. The working solutions were diluted with MeOH to obtain the desired concentration. The standard solution was adjusted to the corresponding pH with NaOH and H3PO4.
Using an LC-20A HPLC system (Shimadzu, Japan) consisted of an ultraviolet detector (SPD-20A), an auto sample injector (SIL-20A), a liquid delivery pump (LC-20AT), a column oven (CTO-20A), and the LabSolutions workstation (Shimadzu, Japan) for HPLC analysis. A syncronis C18 column (4.6 × 250 mm, 5 μm, Thermoscientific, USA) as the chromatographic separation. The gradient mobile phases using 0.5% H3PO4 in DDW (A) and MeOH (B), filtered by a 0.45 μm millipore filter and a Model DOA-P504-BN pump (IDEX, USA) was applied to degas for 20 min. The ratio of the mobile phase was 1 : 9 (v/v) of pump A to pump B, the column temperature was 35 °C, the flow rate of 1.0 mL min−1, the injection volume of 10 μL, and the detection wavelengths of 360 nm.
+ Open protocol
+ Expand
9

Oolong Tea Chemical Profiling

Check if the same lab product or an alternative is used in the 5 most similar protocols
Tea infusions were prepared by adding 20 mL of boiling water to 1 g of various oolong teas for 10 minutes. After cooling to room temperature, the brew was filtered through a 0.22 μm polyvinylidene difluoride (PVDF) membrane filter (Pall Corporation, Glen Cove, NY, USA) for the following analysis. Chemical constituents in oolong tea infusions were analyzed by HPLC system coupled to a 600E photodiode array detector (Waters Corporation, Milford, MA, USA), and separation was performed on the Syncronis C18 column (4.6 × 250 mm inner diameter, 5 μm, Thermo Scientific, Waltham, MA, USA). The separated condition of HPLC analysis was prepared according to Lo et al [21 (link)]. The mobile phase consisted of (A) water containing 0.5% acetic acid and (B) acetonitrile. The gradient was as follows: 0–60 minutes, linearly gradient from 10% to 30% B; 61–70 minutes 30% B; and 71–100 min, linearly gradient from 30% to 10% B. The column was maintained at room temperature and the injection volume was 10 μL at a flow rate of 1 mL/ min. The ultraviolet (UV) absorbance detection wavelength was set at 280 nm.
+ Open protocol
+ Expand
10

RP-HPLC Peptide Analysis and MALDI-TOF MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Analytical RP‐HPLC was performed using a Thermo Scientific™ Dionex™ UltiMate™ 3000 UHPLC system (Thermo Fisher Scientific, Germering, Germany) and a Syncronis C‐18 column (100 Å, 5 μm, 250 × 4.6 mm, Thermo Fisher Scientific) at a flow rate of 1.5 mL/min. The UV detection was set at 220 nm. The elution system was (A) 0.06% (v/v) TFA in water, and (B) 0.05% (v/v) TFA in ACN. The peptides were dissolved in ACN/H2O (10:90, v/v) containing 0.1% TFA. Analytical chromatograms were obtained with the following gradient: 1% B for 8 min, then to 50% B in 35 min. Mass spectra were recorded on an Autoflex Speed MALDI‐TOF mass spectrometer (Bruker Daltonics, Bremen, Germany) by using α‐cyano‐4‐hydroxycinnamic acid as matrix.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!