The largest database of trusted experimental protocols

J w vf 5ms column

Manufactured by Agilent Technologies

The J&W VF-5MS column is a gas chromatography (GC) column designed for a wide range of analytical applications. It features a 5% diphenyl - 95% dimethylpolysiloxane stationary phase that provides good separation of a variety of organic compounds. The column is suitable for use in both capillary GC and packed GC systems.

Automatically generated - may contain errors

4 protocols using j w vf 5ms column

1

GC-MS Analysis of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
A Trace GC Ultra instrument (Thermo Fisher Scientific) was used with helium as carrier gas, in programmable temperature vaporization splitless injection mode at a flow rate of 1.5 mL/min. Injector temperature was 250 °C and a volume of 1 µL was injected. The instrument was equipped with a J&W VF-5MS column (45 m × 0.32-mm i.d., film thickness 0.4 µm, Agilent Technologies). The GC temperature program was as follows: time (min)/oven temperature (°C) 0/40, 1/40, 10/130, 15/280 and 20/280. The GC was coupled to a TSQ Quantum mass spectrometer (Thermo Fisher Scientific) operating in MRM mode in positive polarity with an electron energy of 70 eV. The solvent delay time was 4 min. Precursor ions were fragmented using argon as collision gas and a collision energy of 10 eV with monitoring transitions from m/z 125 to 99 and from m/z 99 to 81.
+ Open protocol
+ Expand
2

GC-MS Analysis of TMS-Derivatized Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The GC-MS system used comprised of a Gerstel 2.5.2 autosampler, a 7890A Agilent gas chromatograph and a 5975C Agilent quadrupole mass spectrometer (Agilent, Santa Clara, CA, United States). The mass spectrometer was tuned according to the manufacturer’s recommendations using tris-(perfluorobutyl)-amine (CF43).
Gas chromatography was performed on a 30 m Agilent J & W VF-5MS column with 0.25 μm film thickness and 0.25 mm internal diameter with a 10 m Integra guard column. The injection temperature (Inlet) was set at 250°C, the MS transfer line at 280°C, the ion source adjusted to 230°C, and the quadrupole at 150°C. He was used as the carrier gas at a flow rate of 1 mL min–1.
The analysis of TMS-derivatized samples was performed under the following temperature program; start at injection 70°C, a hold for 1 min, followed by a 7°C min–1 oven temperature ramp to 325°C and a final 6 min heating at 325°C. Mass spectra were recorded at 2.66 scans.s–1 with an 50–600 m/z scanning range.
+ Open protocol
+ Expand
3

GC-MS Analysis of Erucamide

Check if the same lab product or an alternative is used in the 5 most similar protocols
An Agilent 6890 gas chromatograph (GC) (Agilent Technologies, Wilmington, CA, USA) with a mass detector (MS) was used to measure the samples. The MS detector was heated to 230 °C. An Agilent J&W VF-5 ms column (5% phenyl and 95% dimethylpolysiloxane) with 30 m × 0.25 mm i.d., with a diameter of 0.25 m, was used. The oven heating cycle started at 200 °C for 4 min, rose to 280 °C at a rate of 10 °C min−1, and remained there for 7 min. Helium, at 1.0 mL min−1, was the carrier gas (99.996%). The injection system was in splitless mode. One µL of the sample was injected. The GC-MS apparatus was operated and the data were processed with Chemstation software. With these chromatographic parameters, it was possible to obtain a retention time (Tr) of 7.5 min for erucamide.
+ Open protocol
+ Expand
4

GC-MS Analysis of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
A 6890 Series II GC system (Agilent Technologies) equipped with an injector from Gerstel (Mühlheim, Germany) was used with helium as carrier gas, in splitless injection mode (splitless time for 0.75 min) with a constant flow at 1.5 mL/min. The injector temperature was 250 °C, and 1 µL was injected. The instrument was equipped with a J&W VF-5MS column (50 m × 0.32-mm i.d., film thickness 0.4 µm, Agilent Technologies). The GC temperature program was as follows: time (min)/oven temperature (°C) 0/40, 1/40, 10/130, 15/280 and 20/280. The GC was connected to a 5973 N mass spectrometer (Agilent Technologies) operating in selected ion monitoring mode (m/z 158) with positive polarity and an ionization energy at 235 eV using ammonia as reaction gas (flow 21%). The solvent delay time was 4 min. Source temperature was set to 160 °C.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!