The largest database of trusted experimental protocols

Agilent 8800 icp ms ms

Manufactured by Agilent Technologies
Sourced in Japan

The Agilent 8800 ICP-MS/MS is a triple quadrupole inductively coupled plasma mass spectrometer. It is designed for highly sensitive and selective trace element analysis.

Automatically generated - may contain errors

10 protocols using agilent 8800 icp ms ms

1

Determination of Ultra-Trace Gold

Check if the same lab product or an alternative is used in the 5 most similar protocols
(Ultra-)trace element determination of Au was carried out using an Agilent 8800 ICP-MS/MS instrument (ICP-QQQ, Agilent Technologies, Japan). The sample introduction system comprises a concentric nebulizer (400 µL min−1) mounted onto a Peltier-cooled (2 °C) Scott-type spray chamber. This instrument is equipped with a tandem mass spectrometry configuration consisting of two quadrupole units (Q1 and Q2) and a collision/reaction cell (CRC) located in-between both quadrupole mass filters (Q1-CRC-Q2). All measurements were performed in MS/MS mode (on-mass approach) with the collision/reaction cell (CRC) operated in “vented” (no gas) mode.
+ Open protocol
+ Expand
2

ICP-MS Quantification of Analytes

Check if the same lab product or an alternative is used in the 5 most similar protocols
The ICP–MS instrument used for the quantification studies with conventional SEC and TFC was an iCAP-QMS Thermo Scientific (Bremen, Germany) with oxygen (purity 5.0, Linde Gas GmbH, Vienna, Austria) as reaction gas and Qtegra Intelligent Scientific Data Solution (version 2.4.1800.33) for the data treatment. For the quantification in the centrifugal ultrafiltration studies and for UHPLC SEC, the Agilent 8800 ICP–MS/MS (Agilent Technologies, Tokyo, Japan) was used with Agilent MassHunter software package (Work- station Software, Version C.01.03, 2016) for the data treatment. The ICP–MS parameters are summarized in Table 1.

ICP–MS operation parameters

Instrumentation ICP–MSiCAP-QMS Thermo ScientificICP–MS/MS Agilent 8800
NebulizerPFA-STMicroMist
Spray chamberCyclonicScott double-pass
Nebulizer gas flow1.01 L/min1.05 L/min
Aux. gas0.99 L/min0.90 L/min
Plasma gas14 L/min15 L/min
Reaction gas0.370 mL/min0.300 mL/min
ICP RF Power1550 W1550 W
m/z measured194.97, 47.97195, 48
+ Open protocol
+ Expand
3

Maternal Blood Hg Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
A 100 µL aliquot of maternal blood cell fraction from Timepoint 2 was diluted with 100 µL of deionized water and diluted with a basic solution containing 25 µg/L of iridium as internal standard, 10 µg/L gold, 0.5 g of EDTA in 1% v/v ammonia hydroxide, 2.5% butanol, and 0.05% v/v Triton X100. The resultant 50-fold dilution was then analyzed for total Hg using inductively coupled plasma mass-spectrometry (Agilent 8800 ICP-MS/MS). Helium was used as collision gas to remove interferences. Two sources of external quality controls were used (Seronorm™ and Clinchek™) at three levels and injected after every 10 samples. The analytical acceptability range was defined as within 20 % of the reference value according to manufacturer guidelines. Duplicates of each sample were run, and for every 10 samples a sample was randomly selected and spiked with the analytes as an additional QC check. Complete details of quality assurance and quality controls, including instrumental limits of detection and quantitation (LOD, LOQ), and Hg recovery are described in the supplementary material (Tables A6, A7, and A8).
+ Open protocol
+ Expand
4

Arsenic Accumulation in E. coli Strains

Check if the same lab product or an alternative is used in the 5 most similar protocols
The E. coli strains W3110, WC3110, AW3110, AW3110/pTe2854, AW3110/pTe3384, AW3110/pTe3675, WC3110/pTe2854 and WC3110/pTe3384 were pre-cultured overnight. Then, 0.6 ml of each overnight culture was used to inoculate Erlenmeyer flasks (250 ml) containing 60 ml of LB medium supplemented with 1 mM arsenate or 3 mM arsenite. The inoculated flasks were incubated with continuous stirring (200 rpm) for 5 h at 37°C. After incubation, a volume of 15 ml of culture was filtered through nitrocellulose filters (0.45-μm pore diameter; Whatman) and the cells retained on the filters were washed twice with 0.05 M potassium phosphate buffer (pH 7.6). Filtering and washing were performed using a vacuum pump (Millipore 230V). The filters, which contained the retained cells, were treated with 10 ml of 65% Nitric acid at 70°C for 120 min. Nitric acid (Sigma-Aldrich) was doubly distilled from 65% HNO3 reagent grade with a DST-1000 sub-boiling acid purification system from Savillex (Savillex, Eden Prairie, MN, USA) and diluted with Milli-Q water. Samples were analyzed in an Agilent 8800 ICP-MS/MS (Agilent Technologies) at the General Facilities for Research of the University of Sevilla (http://citius.us.es).
+ Open protocol
+ Expand
5

Trace Element Analysis in Sediments using ICP-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
All measurements were conducted using a tandem mass spectrometer with inductively coupled plasma (ICP-MS/MS) using an optimized method for the measurement of TCEs in sediment digests (Klein et al. 2021 (link)). The ICP-MS/MS system consists of an Agilent 8800 ICP-MS/MS (Agilent 8800, Agilent Technologies, Tokyo, Japan) coupled to an ESI SC-4 DX FAST autosampler (Elemental Scientific, Omaha, Nebraska, USA) equipped with an ultra-low penetration air (ULPA) filtration unit (Elemental Scientific, Omaha, Nebraska, USA). For each measurement, the instrument was optimized on a daily basis using a tune solution containing Li, Co, Y, Ce, and Tl (10 μg L−1). Optimal tuning and instrument parameters can be found in the supplemental information Table A.4. An external calibration, covering a concentration range from 0.1 μg L−1 to 100 μg L−1 for all analytes, was used to allow quantification of each element. Each solution for instrument tuning and quantification was prepared from custom-made multielement standards (Inorganic Ventures, Christiansburg, USA) on a daily basis. A solution containing 10 μg L−1 of Ir and Rh was dosed online to the sample solutions to allow for drift correction. To monitor potential carryover effects, wash blanks of 2% (w/w) HNO3 were measured after each sample triplicate.
+ Open protocol
+ Expand
6

Centrifugal Ultrafiltration of Platinum Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Centrifugal ultrafiltration was carried out using Amicon Ultra 0.5-mL centrifugal regenerated cellulose membrane filters, with a 10-kDa molecular weight cut-off (MWCO) (Millipore Co, Carrigtwohill, Co. Cork, Ireland). The samples were ultracentrifuged with a Hermle Z 466 K device with a precooled rotor (4 °C) at a G-force of 12,500g. For preconditioning of the membrane filters and ultracentrifugation of the drug solutions, a centrifugation time of 60 min and 15 min was used, respectively. The filters used for centrifugal ultrafiltration were preconditioned using the solvents with the highest recovery for the drugs, respectively. For KP2156, 50 mM CH3COONH4, pH = 6.0 was employed, whereas for KP2157 FCS was used. Flow injection measurements for platinum quantification were performed using an Agilent 1260 Infinity Bio-inert HPLC system (Agilent Technologies, Waldbronn, Germany) coupled to an Agilent 8800 ICP–MS/MS (Agilent Technologies, Tokyo, Japan). The HPLC system was operated with CH3COONH4 (50 mM, pH = 6) as eluent, a flow rate of 250 μL/min, and an injection volume of 10 μL. The ICP–MS operation parameters are summarized in Table 1.
+ Open protocol
+ Expand
7

Quantifying Mercury in Biological Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
AAS with reducing vaporization (cold vaporization AAS, CVAAS; mercury analyzer RA-5A, Nippon Instruments Corporation, Tokyo, Japan) was used to measure Hg concentration. The accuracy, precision and repeatability of Hg quantitative analysis by AAS were validated by analyzing NIES CRM No.13: Human Hair (National Institute for Environmental Studies, Ibaraki, Japan). A laser ablation system (NWR 213; Nd: YAG 213 nm; ESI, CA) was coupled to an Agilent 8800 ICP-MS/ MS (Agilent Technologies, Tokyo, Japan) with a tube (Tygon® ND-100-65; Saint-Gobain, Courbevoie, France). LA measurement parameters were calibrated by using NIST 612 (NIST, Gaithersburg, MD, USA). Each parameter was optimized to maximize the signal intensities of m/z 11, 138, and 208. Analysis software iQuant2 (Suzuki et al., 2018) was used to analyze spot and imaging data by LA-ICP-MS. Samples for LA-ICP-MS analysis were embedded with O.C.T. Compound (Sakura Finetek Japan, Tokyo, Japan) and sectioned with a cryostat (Leica CM3050 S Research Cryostat; Leica, Wetzlar, Germany).
+ Open protocol
+ Expand
8

Trace Element Analysis in Biological Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
Digested samples were analyzed using an Agilent 8800 ICP-MS/MS equipped with an octopole reaction system (ORS), situated between two scanning quadrupole mass analyzers (Agilent Technologies, Santa Clara, CA). A 400-μL min−1 MicroMist nebulizer (Glass Expansion, Pocasset, MA) and a Peltier cooled (2 °C) Scott double-pass spray chamber (Agilent Technologies, Santa Clara, CA) were used. Argon (Airgas USA, LLC, Albany, NY 12205) was used as the plasma, dilution, and carrier gas. Horn and nail samples and hair CRMs were analyzed over six runs by ICP-MS/MS.
+ Open protocol
+ Expand
9

Determination of Inorganic Arsenic in Rice

Check if the same lab product or an alternative is used in the 5 most similar protocols
For the determination of i-As in rice, a tandem ICP-MS (Agilent 8800 ICP-MS/MS, Agilent Technologies, Tokyo, Japan), operated by the Agilent MassHunter soware package (Version C.01.05, 2019), was used for the data collection and manual peak integration. The ICP-MS parameters are summarized in Table 1.
The acquired data (as peak areas) were processed using laboratory developed MS Excel™ spreadsheets. Two different types of nebuliser have been used depending on the required ow; Micromist (Crawford Scientic, Strathaven, Scotland) performed better with lower signal RSDs for ows below 900 mL min À1 and Conikal (Crawford Scientic) was used for ows of 1 mL min À1 and above.
+ Open protocol
+ Expand
10

Plutonium Isotope Measurement by ICP-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
A triple quadrupole ICP-MS (Agilent 8800 ICP-MS/MS, Agilent Technologies, Tokyo, Japan) with an octapole collision/reaction cell situated between two quadrupole mass filters (Q1 and Q2) was used to measure plutonium isotopes. A schematic diagram of the ICP-MS instrument is presented in Fig. 1. NH3-He mixture gasses were introduced into the reaction cell at a flow rate of 1.2 mL min -1 for NH3 and 8 mL min -1 for He. The sample uptake rate was 0.4 mL min -1 . The instrument was tuned each time using ~2 µg L -1 uranium and ~1 ng L -1 plutonium ( 242 Pu) standard solution, the optimized instrumental parameters are shown in Table 1. 3.5 mL of sample solution in 0.5 mol L -1 HNO3 containing 1 µg L -1 In(III) (in the form of InCl3) as internal standard was measured. A 0.5 mol L -1 HNO3 solution was used as a washing solution between consecutive samples. No memory effects were observed for consecutive measurement of different level samples. The regression line of measured standard series of 242 Pu was proven to be linear over a large range of 0.01 ng L -1 to 30 ng L -1 (R 2 =0.999).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!