The largest database of trusted experimental protocols

8 protocols using 5977b mass selective detector

1

GC-MS Analysis of Chemical Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The GC–MS analysis was performed using a 7890B gas chromatograph equipped with a 7693 autosampler and a 5977B mass-selective detector (Agilent, Santa Clara, CA, USA). The capillary column used was HP-5MS, 30 m × 0.25 mm × 0.25 mm, from Agilent (Santa Clara, CA, USA). Helium was used as carrier gas at 1.5 mL/min. The split–splitless injector was operated in pulsed pressure splitless mode as follows: initial pressure 0.2 MPa (30 p.s.i.) for 1.3 min, decreased to constant flow. The purge valve was opened after 1.5 min. The injection volume was 5 µL. The temperatures of the GC system were the following: injector temperature 290 °C; transfer line temperature 280 °C; oven temperature program: 50 °C (1.5 min)–30 °C/min–180 °C–20 °C/min–280 °C (20 min). MS detector (quadrupole) was operated in the EI mode at 70 eV with a mass scan range of 50–450 m/z and the sampling rate of 3.6 scans/s.
+ Open protocol
+ Expand
2

Volatile Compounds in Cooked Sausages

Check if the same lab product or an alternative is used in the 5 most similar protocols
The volatile compounds were determined in the cooked sausages along the storage period (0 and 56 days), according to Dominguez et al. [40 (link)]. The conditioning, extraction, and injection of the samples were carried out with a PAL-RTC (PalSystem, Schlieren, Switzerland) 120 autosampler. The extractions occurred at a temperature of 37 °C for a period of 30 min, after an initial temperature equilibration at the same temperature, for a period of 15 min. After sampling, to determine the volatiles, the fiber was transferred to the injection port of a gas chromatograph (7890B Agilent Technologies, Santa Clara, CA, USA) equipped with a 5977B mass selective detector (Agilent Technologies, Santa Clara, CA, USA) and a DB-624 capillary column (30 m × 0.25 mm i.d., film thickness 1.4 μm; J&W Scientific, Folsom, CA, USA). The compounds were identified by comparing their mass spectra with those contained in the NIST14 library. The results were defined in units of area (AU) × 104/g of sample.
+ Open protocol
+ Expand
3

GC-MS Analysis of Hemp Seed Oil

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC–MS analyses of hemp seed oil and extract were performed using an Agilent Technologies 7890 B (GC) equipped with a 5977B Mass Selective Detector (MS). Separation was achieved on a VF WAXms capillary column (30 m × 250 × 0.25 μm) coupled with helium gas as the carrier at a flow rate of 1 mL min−1. The column temperature was programmed initially at 60 °C, which was increased to 160 °C at 10 °C min−1 and further increased to 250 °C at 2.5 °C min−1, holding time for 15 min. The mass spectrometer was operated in the electron ionization mode at 70 eV with a source temperature of 230 °C, with continuous scanning from 35 to 500 m/z. The chemical constituents were identified by comparing their mass spectral data with those from the Wiley library and NIST14 libraries.
+ Open protocol
+ Expand
4

GC-MS Analysis of Chemical Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC-MS analysis was performed using an Agilent 7890 B gas chromatograph equipped with a 5977B mass selective detector (GC/MSD) (Agilent Technologies, Santa Clara, CA, USA). Chromatographic separation was achieved using a Rxi-5HT capillary column (30 m × 0.25 mm I.D., 0.25 μm film thickness; RESTEK, Bellefonte, PA, USA). An autosampler (Agilent Technologies; 7683B) was used for all experiments. The temperature of the injector was 280 °C. Using split mode, 1 µL of each extract was injected at a ratio of 1/50. The carrier gas was helium C-60 at a constant flow rate of 1 mL/min. The GC oven temperature was initially 60 °C for 1 min, increased to 160 °C in 3 °C/min increments, and then increased to 320 °C in 10 °C/min increments, followed by a hold for 10 min. The mass spectrometer used the electron ionization mode at 70 eV with the ion source temperature set at 250 °C. The running time was 30 min. Scan mode was used in the range of 30–600 m/z with a scan rate of 2.6 scan/s. Agilent Mass Hunter Qualitative Analysis B.04.00 software was used for data analysis. Single compounds were identified by comparing mass spectra with NIST mass spectral libraries (National Institute of Standards, 2020 version) and the Wiley Registry 12th Edition.
+ Open protocol
+ Expand
5

GC-MS Analysis of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The GC/MS analysis was performed on an Agilent 7820A GC System Plus gas chromatograph coupled with 5977B Mass Selective detector and flame-ionization detector (Agilent Technologies, Palo Alto, CA, USA). A fused silica capillary column, a mid-polar DB-17HT (J&W Scientific, Folsom, CA, USA) with 60 m column length, 0.25 mm i.d., 0.25 μm film thickness, was used. The oven temperature was programmed from 60 °C (2.5 min held) to 100 °C at a rate of 5 °C/min, from 100 to 225 °C at a rate of 2.5 °C/min and from 225 to 275 °C at a rate of 5 °C, 10 min held at the final temperature was applied. Helium (99.999%) was used as a carrier gas at a constant flow rate of 0.8 mL/min. The split ratio was 1:125, the inlet temperature was set to 260 °C and the transfer line temperature was 280 °C. Mass selective detector operated in electron impact ionization (EI) mode at 70 eV electron energy, the ion source temperature was set to 230 °C and the quadrupole temperature was 150 °C. The mass scan range was 45–1050 m/z.
+ Open protocol
+ Expand
6

GC-MS Analysis of Chemical Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC–MS analysis was performed using an Agilent 7890 B gas chromatograph equipped with a 5977B mass selective detector (GC/MSD) (Agilent Technologies, Santa Clara, CA, USA). Chromatographic separation was achieved using a Rxi-5HT capillary column (30 m × 0.25 mm I.D., 0.25 μm film thickness; RESTEK, Bellefonte, PA, USA). An automatic autosampler (Agilent 7683B) was used for all experiments. The temperature of the injector was 280 °C. One microliter of each extract was injected in split mode with a ratio of 1/50. The carrier gas was helium C-60 at a constant flow of 1 mL/min. The GC oven temperature was initially 60 °C for 1 min, increased to 160 °C in 3 °C/min increments, and then increased to 320 °C in 10 °C/min increments, followed by a hold for 10 min. The mass spectrometer was tuned on electron ionization mode at 70 eV with the ion source temperature set at 250 °C. The running time was 30 min. Scan mode was used in the range of 30–600 m/z with a scan rate of 2.6 scan/s. The Agilent Mass Hunter Qualitative Analysis B.04.00 software was used for data analysis. Single compounds were identified by comparing mass spectra with NIST mass spectral libraries (National Institute of Standards, 2005 version) and the Wiley Registry 8th Edition.
+ Open protocol
+ Expand
7

GC-MS Analysis of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC-MS analysis was carried out using a 7890B GC equipped with a 5977B mass selective detector (7890B–5977B, Agilent Technologies, Inc., Santa Clara, CA, U.S.A.). A non-polar HP-5MS capillary column (30 m × 250 μm × 0.25 μm, Agilent) was employed for separation. Helium (99.9%) served as the carrier gas at a constant flow rate of 2.0 mL min−1 through the column. The GC oven temperature was initially set at 40 °C (maintained for 3 min), then ramped up to 200 °C at a rate of 4 °C min−1 (held for 2 min), followed by a further increase to 250 °C at a rate of 10 °C min−1 with a final hold of 5 min. Mass spectra were acquired in electron ionization (EI) mode (70 eV) in full scan mode, with a mass range of m/z 35–450 at a scanning rate of 1.8 scans per s. The GC injection port temperature was set at 250 °C, the ion source temperature at 230 °C, and the quadrupole temperature at 150 °C. Sample injection was performed in splitless mode with a solvent delay of 1.5 min.
+ Open protocol
+ Expand
8

GC-MS Analysis of Volatile Compounds in Fruit Peel

Check if the same lab product or an alternative is used in the 5 most similar protocols
Volatile compounds in the peel were extracted using the same method as non-volatiles, except dichloromethane (VWR, Pennsylvania, USA) was used as the extraction solvent instead of methanol. The chromatographic and spectrometric parameters were adapted from Goh et al. 2019 and operated on a 7890B GC system coupled with a FID and the 5977B mass selective detector (all from Agilent Technologies, California, USA) [22 ]. For each analysis, a 1 μL splitless injection was used, and 2-octanol (VWR, Pennsylvania, USA) was used as an internal standard. Spectra of compounds detected were matched against an in-house library and the NIST 14 library, and linear retention indices were determined using C7-40 alkane standards (Supelco, Pennsylvania, USA). Data analysis was carried out on the MSD Chemstation software (ver. F.01.03.2357) (Agilent Technologies, California, USA). PCA biplots were constructed using concentrations of key volatiles as well as volatiles categorised by functional group using RStudio (Version 1.3.1093) and accompanying packages (ggbiplot).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!