The largest database of trusted experimental protocols

93 protocols using spectralis sd oct

1

Macular OCT Scans for POAG and Chiasmal Compression

Check if the same lab product or an alternative is used in the 5 most similar protocols
OCT scans carried out as part of routine clinical care were retrospectively identified for patients with POAG and chiasmal compression. Macular volume scans of both eyes obtained with Spectralis SD-OCT (Heidelberg Engineering, Heidelberg, Germany) that passed OSCAR-IB quality control criteria were included.25 (link)
For healthy controls, the OCT measurements were performed prospectively with a Spectralis SD-OCT with the eye-tracking function enabled, using acquisition software V.6.7.13.0. Macular volume scans (1024 A-scans, 37 B-scans volume=15×15°, automatic real-time function = 25) centred on the fovea with the high-resolution setting enabled were performed, with subsequent scans performed on follow-up mode.
+ Open protocol
+ Expand
2

Retinal Nerve Fiber Layer and Ganglion Cell Measurements

Check if the same lab product or an alternative is used in the 5 most similar protocols
Retinal examinations were performed using a Heidelberg Engineering Spectralis SD-OCT (Heidelberg Engineering, Heidelberg, Germany) with automatic real-time (ART) function. Scan quality was checked using the OSCAR-IB Criteria (Schippling et al., 2015 (link)) with the Advised Protocol for OCT Study Terminology and Elements APOSTEL recommendations (Cruz-Herranz et al., 2016 (link)). 3.4 mm ring scans around the optic nerve head were used to measure the peripapillary retinal nerve fibre layer (12°, 1536 A-scans 16 ≤automated real time tracking (ART) ≤ 100). The ganglion cell inner plexiform (GCIP) layer volume was measured using a 6 mm diameter cylinder around the fovea from macular volume scans (25° × 30°, 61 vertical B-scans, 768 A-scans per B-scan, ART = 15). Layer segmentation was performed using Eye Explorer 1.9.10.0, viewing module 6.3.4.0 (Heidelberg Engineering, Heidelberg, Germany).
+ Open protocol
+ Expand
3

Retinal Nerve Fiber Layer Thickness Measurements

Check if the same lab product or an alternative is used in the 5 most similar protocols
The RNFL thickness measurements were derived using the Heidelberg Engineering Spectralis SD-OCT (Heidelberg Engineering Inc., Heidelberg, Germany). Spectralis OCT provides an automatic real-time function with an eye-tracking system that can increase image quality. A circular scan was manually placed in the center of the optic disc while the eye-tracking system was activated. The peripapillary RNFL thickness of the scans was determined automatically based on the software algorithms. The RNFL was also automatically segmented into 6 (superonasal, nasal, inferonasal, inferotemporal, temporal, and superotemporal segments). Then the computed average RNFL thickness values for the overall RNFL thickness and for each sector were presented in the Spectralis OCT RNFL printouts. As suggested by the manufacturer, scans with signal strength of <15 dB (range, 0–40 dB) were excluded from the analysis.
+ Open protocol
+ Expand
4

Retinal Layer Thickness Assessment using SD-OCT

Check if the same lab product or an alternative is used in the 5 most similar protocols
The subjects in this study underwent a SD-OCT examination using the Heidelberg Engineering Spectralis SD-OCT (Heidelberg Engineering, Heidelberg, Germany, Spectralis EYE Explorer version 1.10.20). This SD-OCT utilizes a scanning superluminescence diode to emit a scan beam with a wavelength of 870 nm to acquire 40,000 A scans/s with a depth resolution of 7 μm, from which various retinal layers can be identified, and objectively and precisely assessed. The SD-OCT device also combines SD-OCT and confocal infrared laser ophthalmoscope, which provides a reference infrared fundus image. The SD-OCT parameters calculated for this study were mean RNFL thickness around the optic nerve in nasal (NAS), inferior (INF), temporal (INF), and superior (SUP) quadrant of both eyes. The mean retinal thickness was measured in μm per quadrant of the peripapillary area. All scans were performed by experienced ophthalmologists (AP, FP, and GC). The images were subsequently reviewed for acceptable signal strength, correct placement of the scan ring, and appropriate beam placement.
+ Open protocol
+ Expand
5

Retinal Ganglion Cell Complex Changes After Blast TBI

Check if the same lab product or an alternative is used in the 5 most similar protocols
Spectral-domain optical coherence tomography (SD-OCT) analysis was performed 2 months following blast TBI or sham procedure using a Spectralis SD-OCT (Heidelberg Engineering, Vista, CA, USA) imaging system coupled with a 25 D lens for mouse ocular imaging (Heidelberg Engineering). Anesthetized mice were placed on a temperature-controlled platform and pupils were dilated using a 1% tropicamide solution. The cornea was moisturized with a saline solution applied every 20 to 30 seconds during the procedure. Volume scans (49-line dense array) positioned directly over the optic nerve head were performed to quantify the RGC complex (inner plexiform layer + ganglion cell layer + retinal nerve fiber layer) thickness. Scans were analyzed by an individual masked to the treatment of the mouse in the superior and inferior central retina, approximately 150 μm from the peripapillary region. All scans were analyzed by excluding blood vessels from the thickness calculation. Baseline measurements were obtained 1 week prior to blast exposure, with follow-up analysis occurring 2 months post blast exposure.
+ Open protocol
+ Expand
6

Quantitative and Qualitative OCT Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
We used a Spectralis SD-OCT (Heidelberg Engineering GmbH, Heidelberg, Germany). At each visit, the following scans were acquired in all eyes: a high-definition horizontal fovea-centered cross line B-scan at 30° and a horizontal macula raster consisting of 49 B-scans 120 µm spaced over an area of 20°.The ‘‘Thickness Map” function was used to measure automatically the mean CMT, based on the mean retinal thickness within a circular area of 0.5 mm radius from the foveal center.
Two masked observers (R.R. and G.F.) independently evaluated OCT images quantitatively and qualitatively. A third observer (A.R.) resolved any case of disagreement. Segmentation errors were manually corrected with built-in software when needed.
+ Open protocol
+ Expand
7

Quantitative Retinal Imaging in Rats

Check if the same lab product or an alternative is used in the 5 most similar protocols
Subjects received SD-OCT of both eyes using a Spectralis SD-OCT (Heidelberg Engineering, Vista, CA, USA) imaging system (Heidelberg Engineering). Rats were anesthetized (as described above), pupils were dilated using a 1% tropicamide solution, and the cornea was moisturized with a saline solution. Volume scans (49-line dense array, 15 A-scans per B-scan, 20° scan angle, 20° × 25° scan area) positioned directly over the optic nerve head were taken for each eye. The retinal ganglion cell complex (RGCC) thickness + the retinal nerve fiber layer (RNFL) was quantified as the sum of the thickness of the retinal nerve fiber, ganglion cell, and inner plexiform layers. Using unlabeled data, two B-scans per eye were analyzed by two experimenters by measuring the overall thickness at five equidistant points along the scan and excluding blood vessels from the calculation.
+ Open protocol
+ Expand
8

Comprehensive Ophthalmic Imaging Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Each subject was required to have at least one good-quality Spectralis SD-OCT (Heidelberg Engineering Inc., Heidelberg, Germany) circular scan, SD-OCT ONH cube scan, SD-OCT ONH radial scan, and SD-OCT macular cube scan acquired on the same day. Spectralis OCT uses a dual-beam SD OCT, a confocal laser-scanning ophthalmoscope with a wavelength of 870 nm, and an infrared reference image to obtain images of ocular microstructures. The instrument has an acquisition rate of 40,000 A-scans per second. Spectralis OCT incorporates a real-time eye-tracking system that couples confocal laser-scanning ophthalmoscope and SD OCT scanners to adjust for eye movements and to ensure that the same location of the retina is scanned over time. The image acquisition protocols included (1) high-resolution RNFL circle scan, which consists of 1536 A-scan points from a 3.45-mm circle centered on the optic disc, (2) high-resolution cube scan centered on the optic disc (73 B-scans with 768 A-scans each), (3) enhanced depth imaging (EDI) scan centered on the optic disc (48 B-scans with 1024 A-scans each), and (4) high-resolution cubic scan centered on the fovea (73 B-scans with 768 A-scans each). Quality assessment of OCT scans was evaluated by Imaging Data Evaluation and Assessment (IDEA) Center experienced examiners masked to the subject's results of the other tests.
+ Open protocol
+ Expand
9

Wide-field Retinal Imaging Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
All patients underwent both UWFFAF imaging and pseudo-color (red and green only) imaging using the Optos ultra–wide-field imaging device (Optos, Dunfermline, United Kingdom). The UWFFAF images were acquired using a scanning laser ophthalmoscope technology with a green laser of 532 nm as the excitation light and with signal emitted from the retina detected from a raster scan after passing through an emission filter with a bandpass of 570 nm to 780 nm. OCT scans of the retina were obtained using the Spectralis SD-OCT (Heidelberg Engineering, Heidelberg, Germany).
+ Open protocol
+ Expand
10

Longitudinal SD-OCT and BCVA Evaluation

Check if the same lab product or an alternative is used in the 5 most similar protocols
SD-OCT was performed using Cirrus HD-OCT (Carl Zeiss Meditec AG, Jena, Germany) or Spectralis SD-OCT (Heidelberg Engineering GmbH, Heidelberg, Germany) pre-operatively and post-operative on day 1, week 1, months 1, 3, 6, 12 and 24. ETDRS BCVA was measured on the respective follow-up dates and converted to the logarithm of the minimum angle of resolution (logMAR) for statistical analysis. Outcomes included anatomical changes based on CMT and functional changes based on BCVA. Proportional early CMT and BCVA changes form the pre-operative baseline values were also evaluated as a percentage of the final respective values at 24 months follow-up and as a percentage of the healthy fellow eye CMT and BCVA.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!