The largest database of trusted experimental protocols

23 protocols using 5 m guard column

1

Metabolite Profiling via GC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The samples from Intracellular, extracellular, and biomass were derivatized using the methyl chloroformate (MCF) method as previously described (Smart et al. 2010 (link)). The chemical derivatives were analyzed by a system of Agilent GC7890B coupled to an MSD5977A mass selective detector (EI) set at 70 eV. The ZB-1701 GC capillary column (30 m × 250 µm id × 0.15 µm with 5 m guard column, Phenomenex) was used for metabolite analysis. The parameter analysis was previously described (Smart et al. 2010 (link)).
+ Open protocol
+ Expand
2

GC-MS Analysis of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
We analysed the samples on a GC-MS system—a GC7890B coupled to an MSD5977A operated at 70 eV electron ionization (Agilent Technologies)—equipped with a ZB-1701 column, 30 m × 0.25 mm × 0.15 μm with 5 m guard column (Phenomenex), at a constant flow rate of 1.0 mL/min of helium. Based on a published protocol39 (link), samples (1 µL) were injected under pulsed splitless mode (for trace analyses) with the inlet at 290 °C (180 kPa for 1 min, 50 mL/min purge flow after 1 min). The GC oven temperature was initially held at 45 °C for 2 min, after which it was raised to 180 °C at 9 °C/min and held for 5 min. Subsequently, the temperature was raised to 220 °C at 40 °C/min and held for 5 min. The temperature was raised again to 240 °C at 40 °C/min and held for 11.5 min. Finally, the temperature was raised to 280 °C at 40 °C/min and held for 10 min. The interface was kept at 250 °C and the quadrupole at 130 °C. The detector operated in scan mode, started after 5.5 min with a mass range between 38–550 atomic mass units at 2.9 scans/s. Negative controls were injected at the beginning of each run and QCs were run every seventh sample.
+ Open protocol
+ Expand
3

Plasma Metabolite Profiling by GC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The extracted samples were chemically derivatized via the methyl chloroformate (MCF) method based on the recommendations published by Smart et al. [20 (link)]. All cord plasma samples were analyzed in a single batch and derivatized compounds were examined by an Agilent GC7890B system using a ZB-1701 GC capillary column. An MSD5977A mass selective detector with the electron impact voltage set to 70 eV was applied to analyze the compounds. The GC column used for metabolite separation was the ZB-1701 GC capillary column (30 m × 250 μm id × 0.15 μm with a 5 m guard column, Phenomenex). The GC temperature was set up according to the protocol of Han et al. [21 (link)].
+ Open protocol
+ Expand
4

GC-MS analysis of derivatized hair metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
Dried hair extracts were resuspended in 200 μL of sodium hydroxide (1 M) and methyl chloroformate derivatization was performed to make the compounds more volatile for GC-MS analysis, based on the protocol published by Smart et al (Smart et al. 2010 (link)). All neonatal hair samples were analysed in a single batch and derivatized compounds were separated by a GC7890 chromatography system using a ZB-1701 GC capillary column. (30 m × 250 μm id × 0.15 μm with a 5 m guard column, Phenomenex) and analyzed by a MSD5975 mass spectrometer (Agilent, California, USA) with electron impact ionization via electron emission at 70 eV. The GC temperature program and MS parameters were set up according to the protocol described in Smart et al (Smart et al. 2010 (link)).
+ Open protocol
+ Expand
5

GC-MS Analysis of MCF Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
The dried pellets were resuspended in 200 µl of sodium hydroxide (1 mol/l) and then transferred into silanized glass tubes. The MCF derivatization was performed in accordance with the protocol published in Smart et al (37 (link)). The derivatives of the MCF metabolites were analyzed using an Agilent GC7890B system (Agilent Technologies, Inc.) coupled to a MSD5977A mass spectrometer (Agilent Technologies, Inc.) with the electron impact ionization set at 70 eV. The gas chromatograph used to separate the metabolites was a Zebron ZB-1701 Capillary GC column (30 mx250 µm id ×0.15 µm with a 5 m guard column; Phenomenex); 1 µl of derivatized sample was injected into the GC inlet that operated at 290°C in a split-less mode under 180 kPa for 1 min. Details of GC and MS parameters were also set up according to the methodology reported in Smart et al (37 (link)). Helium gas flow rate was constantly controlled at 1 ml/min. The GC oven temperature was initiated at 45°C for 2 min and raised to 180°C at 9°C/min for 5 min. Subsequently, the oven temperature was raised at 40°C/min to 220°C and held for 5 min. The temperature was then increased again at 40°C/min to 240°C and held for 11.5 min. The last temperature increase was at 40°C/min to 280°C and held for 2 min. The transfer interface between GC and MS was setup at 250°C, the MS ion source at 250°C and the quadrupole at 130°C.
+ Open protocol
+ Expand
6

GC-MS Analysis of Derivatized Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
The derivatized samples were analyzed using an Agilent GC7890 system linked to a MSD5975 with electron impact ionization (70 eV). The gas capillary column was a ZB-1701 (30 m × 250 μm id × 0.15 μm with 5 m guard column, Phenomenex). The parameters of the GC oven and MS were operated in accordance with Smart et al.20 (link)’s published protocol. The samples were injected into a pulsed splitless mode inlet at 290 °C with the flow of helium gas at 1 mL.min−1. The temperatures of the auxiliary, MS quadrupole, and MS source were 250 °C, 230 °C, and 150 °C respectively. The mass range was detected from 30 um to 550 µm. Scan speed was set to 1.562 µ.s−1 and the solvent delay was applied until 5.5 min. The estimated time required for sample preparation, derivatisation, and GC-MS acquisition is shown in Supplementary Table S6.
+ Open protocol
+ Expand
7

GC-MS Analysis of Hair Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
The MCF derivatized extracts were analyzed in an Agilent C7890B gas chromatograph coupled to a MSD5975 mass spectrometer. The mass spectrometer setting was at 70 eV. All hair samples were analyzed in a single batch and the gas chromatographic column used for analysis was a zb-1701 gas chromatographic capillary column constituted of 30 m × 250 μm id × 0.15 μm with a 5 m guard column (Phenomenex). GC–MS analysis parameters were based on the parameters reported in Smart et al. (2010) (22 (link)) and Han et al. (2019) (23 (link)). 1 μL of sample was injected into the inlet using an autosampler with the inlet temperature at 260°C and in the pulsed splitless mode at 180 kPa for 1 min. The helium gas flow through the GC-column was set at a constant flow of 1.0 mL/min. The GC oven temperature was maintained at 45°C for 2 min, and increased at 9°C/min to 180°C held for 5 min. The temperature gradually increased at 40°C/min to 220°C. After 5 minutes, the temperature increased at 40°C/min to 240°C for 11.5 min. Finally, the temperature was increased at 40°C/min to 280°C, and maintained at 280°C for 2 min. The temperature of the interface was kept at 250°C, and the quadrupole temperature was kept at 230°C. The mass spectrum started after 5.5 min with a mass range of 38-550amu, and the scanning time was 0.1 s.
+ Open protocol
+ Expand
8

GC-MS Analysis of Metabolite Extracts

Check if the same lab product or an alternative is used in the 5 most similar protocols
Dried intracellular and extracellular metabolite extracts were resuspended in 200 μl of sodium hydroxide (1 M). The MCF derivatization step was performed according to the method published by Han et al. [20 (link)]. All MCF-derivatized samples were analyzed using an Agilent GC7890B chromatography system coupled to a MSD5975 mass spectrometer (Agilent, California, USA) operating at 70 eV. The GC column was a ZB-1701 GC capillary column (30 m × 250 μm id × 0.15 μm with a 5 m guard column, Phenomenex, USA). GC-MS parameters and temperature programs were set according to the protocol reported in Han et al. [20 (link)].
+ Open protocol
+ Expand
9

GC-MS Analysis of Polar Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
Polar metabolites were derivatised for 90 min at 55°C with 20 μL of methoxyamine (c = 20 mg/mL) in pyridine under continuous shaking and subsequently for 60 min at 55°C with 20 μL of MTBSTFA w/1% TBDMCS. GC-MS analysis was performed using an Agilent 7890B GC coupled to an Agilent 5977A Mass Selective Detector (Agilent Technologies). A sample volume of 1 μL was injected into a Split/Splitless inlet, operating in splitless mode at 270°C. Gas chromatograph was equipped with a 30 m (I.D. 250 μm, film 0.25 μm) ZB-35MS capillary column with 5 m guard column (Phenomenex). Helium was used as carrier gas with a constant flow rate of 1.2 mL/min. GC oven temperature was held at 100°C for 2 min and increased to 300 °C at 10 °C/min and held for 4 min. Total run time was 26 min. Transfer line temperature was set to 280°C. Mass selective detector (MSD) was operating under electron ionisation at 70 eV. MS source was held at 230°C and the quadrupole at 150°C. For precise quantification of the MID, measurements were performed in selected ion monitoring mode. Target ions (m/z) and Dwell times are shown in Table S1.
The MetaboliteDetector software package (v. 3.220180913) was used for mass spectrometric data post processing, quantification, MID calculations, correction of natural isotope abundance, and determinations of fractional carbon contributions.96 (link)
+ Open protocol
+ Expand
10

MCF Derivatization and GC-MS Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
All prepared extracts were chemically modified to lower their boiling point by MCF derivatization, based on the method published in Nature protocols (36 (link)). The volatile compounds were then separated by ZB-1701 GC capillary column (30 m × 250 μm id × 0.15 μm with 5 m guard column, Phenomenex, CA, USA) and detected by GC-MS (Agilent 7890B-5977A) with electron impact ionization via electron emission at 70 eV. The GC-MS parameters were operated following the procedure in previous research (37 (link)). The GC-MS inlet was set at 290°C with the pulsed splitless mode, 1 ml/min in the flow rate of the helium carrier. The temperature was controlled at 280°C, 230°C, and 150°C of auxiliary, MS quadrupole, and MS source respectively. The mass range was detected between 30 μm to 550 μm, with a scan speed of 1.562 μ/s and the mass spectrometry detector turned on after 5.5 min.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!