The largest database of trusted experimental protocols

19 protocols using timstof flex

1

LCMS-based Actinobacteria Profiling and Extraction

Check if the same lab product or an alternative is used in the 5 most similar protocols
High resolution LC-MS/MS data were collected on a Bruker timsTOF fleX in positive mode. The detection window was set from 100 to 2000 Da. An Agilent Poroshell 120 Å 2.1 × 50 mm UPLC column was used with a flow rate of 0.5 mL/min where solvent A was MilliQ H2O w/ 0.1% formic acid and solvent B was acetonitrile w/ 0.1% formic acid. Data were acquired in Compass Hystar 6.0 and otofControl 6.2. The following gradient was used: hold 5% B from 0 to 2 min, 5 – 100% B from 2 to 12 min, and hold 100% B from 12 to 14 min, followed by a 1 min equilibration time. For each sample, 2 μL of a 1 mg/mL solution was injected in partial loop mode with a loop volume set to 50 μL. The ESI conditions were set with the capillary voltage at 4.5 kV. For MS/MS, dynamic exclusion and the top nine precursor ions from each MS1 scan were subject to collision energies scaled according to mass and charge state for a total of nine data dependent MS/MS events per MS1 via Auto MS/MS mode. The datasets generated to prioritize Actinobacteria strains, evaluate Actinobacteria growth partners, test the performance of plug extractions, and test the performance of colony removal with nitrocellulose membranes will be referred to as Dataset 2, 3, 4, and 5, respectively, from here on out.
+ Open protocol
+ Expand
2

Mass Spectrometry Imaging of Hippocampal Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
Mass spectra were acquired on a timsTOF fleX (Bruker Daltonics, Billerica, MA) in negative ion mode. Using FlexImaging 5.1 software, hippocampal sections were outlined as individual ROIs with a pixel size of 20 μm. Most metabolites were acquired in a single imaging run (m/z range 50 -1000); however, a separate method was developed to target small metabolites not identified well by the first method, such as hypoxanthine (m/z range 50 -400). In both methods, each MSI pixel consisted of 1000 laser shots. The laser Smartbeam parameter was set to single with a frequency of 10 kHz. Based on the mass range and selectivity of biomolecules for the two MALDI methods, the instrument parameters were optimized, e.g., quadrupole, ion transfer funnels, collision cell, and focus pre-TOF.
Before each acquisition, the instrument was calibrated by MALDI using the three 15 N standards spiked into the matrix, as described above. For the small molecule method, a small spot of lactate, pyruvate, and 15 N 1 -glutamate was dried onto the ITO slide before matrix application, and these three ions were used for calibration of the lower m/z range.
+ Open protocol
+ Expand
3

Spatial Multiomics Analysis by timsTOF

Check if the same lab product or an alternative is used in the 5 most similar protocols
After spraying the slides with matrix solution, slides were loaded in a slide holder from Bruker, and we ran the slides in timsTOF flex (Bruker). Slides were run in both positive and negative ion modes to facilitate comparison. The tissue was scanned with both MS and TIMS settings at a resolution of 20 µm. The MS settings were: scan range 20–2500 m/z in positive MS scan mode. The TIMS settings were: 1/K0 0–8 − 1.89 V×s/cm2, ramp time of 200 ms, acquisition time of 20 ms, duty cycle = −10%, and ramp rate of 4.85 Hz.
+ Open protocol
+ Expand
4

MALDI-MS Imaging of Hippocampus

Check if the same lab product or an alternative is used in the 5 most similar protocols
Mass spectra were acquired on a timsTOF fleX (Bruker Daltonics, Billerica, MA) in negative ion mode. Using FlexImaging 5.1 software, hippocampal sections were outlined as individual ROIs with a pixel size of 20 μm. Most metabolites were acquired in a single imaging run (m/z range 50 – 1000); however, a separate method was developed to target small metabolites not identified well by the first method, such as hypoxanthine (m/z range 50 – 400). In both methods, each MSI pixel consisted of 1000 laser shots. The laser Smartbeam parameter was set to single with a frequency of 10 kHz. Based on the mass range and selectivity of biomolecules for the two MALDI methods, the instrument parameters were optimized, e.g., quadrupole, ion transfer funnels, collision cell, and focus pre-TOF.
Before each acquisition, the instrument was calibrated by MALDI using the three 15N standards spiked into the matrix, as described above. For the small molecule method, a small spot of lactate, pyruvate, and 15N1-glutamate was dried onto the ITO slide before matrix application, and these three ions were used for calibration of the lower m/z range.
+ Open protocol
+ Expand
5

In Situ Protein Identification in Breast Cancer Tissue

Check if the same lab product or an alternative is used in the 5 most similar protocols
Tentative identification of the m/z features was carried out by MS/MS measurement on whole-mount breast cancer tissue sections. The measurements were performed in situ using a Rapiflex in LIFT mode (m/z 772.4 and 1198.7) and a timsTOFflex (m/z 1428.7 and 1495.7) mass spectrometer (Bruker Daltonics). Laser power for fragmentation was set at 70–80% in positive ionization mode, 2000 shots at a laser frequency of 5 kHz, and a beam scan of 25 µm2. For the identification, the MS/MS spectra (Figures S1 and S2) were submitted to MASCOT MS/MS Ion Search, where the SwissProt database was searched to match tryptic peptide sequences to the respective intact proteins, defining homo sapiens as the taxonomic class. The MS/MS spectrum search parameters included a mass tolerance of 1 Da, MS/MS tolerance of ±1 Da, up to two missed cleavages, methionine oxidation, protein N-terminal acetylation, and proline oxidation as variable modifications.
+ Open protocol
+ Expand
6

MALDI-IMS-MS Negative Ion Mode Acquisition

Check if the same lab product or an alternative is used in the 5 most similar protocols
MALDI-IMS-MS negative ion mode data were acquired on a timsTOF fleX (Bruker Daltonics GmbH & Co. KG) using timsControl 4.1 and flexImaging 7.2. The imaging run was exported from flexImaging 7.2 via “Save run file as…” to make the geometry files available for MS2 measurements.
R2c_1a-d: Imaging data were acquired with a laser field size of 30 µm. The imaging raster size was 50 µm to keep material for subsequent MS2 acquisition (see Supplementary Fig. S13a). Resulting in a final image resolution of 50 µm. The laser frequency was 10,000 Hz with 100 shots and 1 burst per pixel. The laser parameters were set to ‘Custom’ (Application), a 0% (Power Boost), single (Smart Beam), enabled (Beam Scan), and 26 µm (Scan Range). The mass range was set to 100–1500 m/z, the 1/K0 range was set to 0.65–1.75 Vs/cm² and a 150 ms ramp time. Tune parameters were set to 50 V (MALDI plate offset), −70 V (Deflection delta), 500 Vpp (Funnel 1 RF), 0.0 eV (isCID), 350 Vpp (Funnel 2 RF), 350 Vpp (Multipole Vpp), 10 eV (Collision energy), 1100 Vpp (Collision cell RF), 5 eV (Ion energy), 100 m/z (Low mass), 80 µs (Transfer time), 5 µs (Pre pulse storage). TIMS parameters were set to 20 V (Δt1), 120 V (Δt2), −70 V (Δt3), −100 V (Δt4), 0 V (Δt5), −100 V (Δt6), −220 V (Collision cell in). After MS1 acquisition, data analysis was performed in MZmine 3.
+ Open protocol
+ Expand
7

Trapped Ion Mobility-Quadrupole-TOF Mass Spectrometry

Check if the same lab product or an alternative is used in the 5 most similar protocols
A trapped ion mobility-quadrupole-time-of-flight (TIMS-Q-TOF) instrument (timsTOFflex, Bruker Daltonics, Bremen, Germany) was used. The instrument was equipped with an ESI-to-MALDI switchable dual source. According to the manufacturer, the AP interface of all current Bruker instruments is identical to the above, and thus, the APFD source is fully compatible among these. The mass spectrometer was controlled by the Bruker timsControl software (V 2.0) and data analysis was performed using the Bruker DataAnalysis software (V 6.0).
External mass calibrations were established in ESI mode by using Agilent Tune Mix (G1969-85,000) for the m/z 100–2500 range [43 (link), 44 ]. Mass accuracy was in the order of 3 ppm.
+ Open protocol
+ Expand
8

MALDI-QTOF Imaging of Extracellular Matrix

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples were prepared as previously described for targeted collagen and extracellular matrix peptide imaging55 (link),101 (link)–104 (link). Briefly, deglycosylated samples105 ,106 (link) were antigen retrieved with 10 mM Tris, pH 9, autosprayed (M5, HTX-Technologies) with 0.1 µg/µL collagenase type III (Worthington) dissolved in ammonium bicarbonate pH 7.4, 1 mM CaCl2, and incubated at 38.5 °C with ≥85% humidity for 5 h. The matrix α-Cyano-4-hydroxycinnamic acid (CHCA, Sigma-Aldrich) was dissolved in 1.0% trifluoracetic acid (Sigma), 50% acetonitrile (LC-MS grade, Fisher Chemical) and autosprayed (M5, HTX Technologies) onto tissue. Tissues were imaged on a MALDI-QTOF (timsTOF-flex, Bruker) in positive ion mode over m/z range 700–2500. Laser was adjusted to 20 µm2 and each pixel consisted of 300 laser shots. Images were collected with a laser step size of 60 µm. Data was visualized in SCiLS Lab Software (v2022b, Bruker) and processed for image segmentation and principal component analysis. Peak data were exported by mean spectrum processed as peak maximum and further statistical comparisons were done using Metaboanalyst 5.0 and GraphPad Prism 9.0. Exported peak intensities were visualized as heatmaps using the TM4 MultiExperiment Viewer suite107 (link) with clustering by Manhattan metric and single linkage.
+ Open protocol
+ Expand
9

MALDI-MSI Metabolite Imaging in Hippocampus

Check if the same lab product or an alternative is used in the 5 most similar protocols
Mass spectra were acquired on a timsTOF fleX (Bruker Daltonics, Billerica, MA) in negative ion mode. Using FlexImaging 5.1 software, hippocampal sections were outlined as individual ROIs with a pixel size of 20 μm. Most metabolites were acquired in a single imaging run (m/z range 50–1000); however, a separate method was developed to target small metabolites not identified well by the first method, such as hypoxanthine (m/z range 50–400). In both methods, each MSI pixel consisted of 1000 laser shots. The laser Smartbeam parameter was set to single with a frequency of 10 kHz. Based on the mass range and selectivity of biomolecules for the two MALDI methods, the instrument parameters were optimized, e.g., quadrupole, ion transfer funnels, collision cell, and focus pre-TOF.
Before each acquisition, the instrument was calibrated by MALDI using the three 15N standards spiked into the matrix, as described above. For the small molecule method, a small spot of lactate, pyruvate, and 15N1-glutamate was dried onto the ITO slide before matrix application, and these three ions were used for calibration of the lower m/z range.
+ Open protocol
+ Expand
10

MALDI-ToF MS Acquisition for Biomolecule Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Data acquisition was performed on a timsToF fleX (Bruker Daltonics, Bremen, Germany) in positive ion mode in the range of m/z 900–3200.
Spectra were recorded using 600 shots per pixel with a laser repetition rate of 10 kHz using a 50 × 50 µm step size. Transfer parameters were as follows: Funnel 1 RF 500 Vpp; Funnel 2 RF 500 Vpp; Multipole RF 500 Vpp; Deflection Delta 70 V; MALDI Plate Offset: 50 V. Quadrupole parameters: Ion energy 5.0 eV; Low Mass m/z 900. Focus PreToF Parameters: Transfer Time 145 µs; PrePulse Storage 28 µs.
External calibration was done in the electrospray mode using ESI-Low Concentration Tuning Mix (Agilent Technologies, Santa Clara, CA, USA)
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!