The largest database of trusted experimental protocols
Sourced in Germany, United States

The HAAKE is a line of laboratory equipment designed for rheology measurements and analysis. It provides precise control and measurement of the viscosity, flow, and deformation properties of various materials, including fluids, suspensions, and semi-solids. The HAAKE instruments are engineered to deliver reliable and accurate results for research, development, and quality control applications.

Automatically generated - may contain errors

8 protocols using haake

1

Rheological and Compressive Analysis of Composite Hydrogels

Check if the same lab product or an alternative is used in the 5 most similar protocols
The rheological tests of hydrogels were measured at 37 °C using a rotational rheometer (ThermoFisher HAAKE, USA) with a parallel plate of 35 mm diameter and an operating gap distance of 0.5 mm. Time sweep tests were performed using a constant strain of 1 %. Viscosity measurements were performed at a constant strain of 1 %. Recovery properties were characterized by three cycles of repetitive stress at either 5 % strain or at 1 % strain, each lasting for 60s. The compressive strength was measured by a universal testing machine (HY-0230, Shanghai Hengyi, China). Different composite hydrogels were prepared into cylinders (10 mm in diameter, 8 mm in height) and tested with a strain rate of 2 mm/min. The stress at a strain of 50 % was measured for quantification.
+ Open protocol
+ Expand
2

Rapid RBC Measurement via Microfluidics

Check if the same lab product or an alternative is used in the 5 most similar protocols
Measurements of the blood samples were done with the AcCellerator instrument (Zellmechanik Dresden, Dresden, Germany) as described previously18 (link),19 (link). For maximum quality, measurements were done no longer than 4 h after blood collection to minimize cell lysis. In preparation for the measurement, citrate blood was diluted 1:20 with phosphate buffered saline (PBS) containing methylcellulose and viscosity was adjusted to 60 mPa s at 24 °C using a falling ball viscometer (Haake, Thermo Scientific). 1 ml of cell suspension was loaded into a syringe, placed in a syringe pump and connected to the microfluid chip made of polydimethylsiloxane (PDMS) attached to cover glass. The inlet of the chip led to a square channel of 20 × 20 µm and 300 µm length. Pure measurement buffer administered via a second inlet to the microfluidic chip induced a focused constant flow of the cell suspension through the channel. The total flow rate was 0.02 µL/s, of which the sheath flow rate was 0.015 µL/s and the sample flow rate was 0.005 µL/s. Images of the cells were taken at the last third of the channel with a high-speed microscope at 3000 fps in a region of 250 × 80 pixels. Cell analysis was focused on RBCs applying filters for object length and heights from 1.0 µm to 80.0 µm and limited to 10 000 events (Fig. 1b).
+ Open protocol
+ Expand
3

Comprehensive Chemical Characterization of POP

Check if the same lab product or an alternative is used in the 5 most similar protocols
The phenol-sulfuric acid method was used to determine the total polysaccharide content of POP extract, using D-glucose as the standard[22 ]. Protein content of POP extract was determined by Lowery method, using BSA as the standard[23 (link)]. The m-hydroxybiphenyl assay was used to measure uronic acid content of POP extract, using glucuronic acid as reference material [24 ]. The moisture content of POP was measured, referring to the method outlined in Kong et al (22). The pH values of the POP at the 2 mg/mL of POP were determined using a pH meter and the relative viscosity of POP (10 mg/mL) was determined using viscometer (Thermo Scientific HAAKE, 388–0100) at 25°C[25 (link)].
+ Open protocol
+ Expand
4

Enhancing PLLA/PDLA Blend Thermal Stability

Check if the same lab product or an alternative is used in the 5 most similar protocols
PLLA,
PDLA, and ADP were placed in the vacuum oven for 24 h at 80
°C to remove moisture. The mixtures containing 1:1 (mass ratio)
of PLLA and PDLA with different mass fractions of ADP were added to
an internal mixer (Haake, Thermo Fisher Scientific) for 5 min of blending
at 190 °C. The rotation speed is at 60 rpm. The obtained white
powder PLLA/PDLA/ADP samples is dried in a vacuum oven at 100 °C
for 12 h and then injected into the mold using a plastic injection
molding machine (JPH30). The temperature of the first zone of the
screw is 250 °C, and the temperatures of following zones are
255, 265, 255, and 250 °C, respectively. The mold temperature
is room temperature. The preparation for PLLA samples and PLLA/PDLA
(PLLA/PDLA = 1:1) samples are carried out the same way without adding
ADP or PDLA. For studying the heat resistance of pristine PLLA, PLLA/PDLA,
and PLLA/PDLA/ADP15 samples, the powder obtained from internal mixer
was injected [Thermo Fisher Scientific (China) Co., Ltd.] into the
mold with optimal mold temperatures at 120 °C (30 min), 130 °C
(2 min), and 170 °C (2 min), respectively (Table 1).
+ Open protocol
+ Expand
5

Structural Rheology of Si-HA Hydrogels

Check if the same lab product or an alternative is used in the 5 most similar protocols
Si-HA molecular weight (kDa) 51 51
Si-HA degree of substitution (%) 30 20
Table 1. Si-HA characteristics (molecular weight and degree of substitution) function of the emulsion used.
The structural integrity of emulsion after the formulation process was evaluated using digital microscope analysis (VHX-7000, Keyence, Bois-Colombes, France).
Gel Time Evaluation.
The sol-gel transition was analyzed by performing multiwave frequency sweeps on a stresscontrolled MARS rheometer (HAAKE, ThermoFisher Scientific, USA) with a cone/plate 60 mm 1° titanium geometry. A range of 5 frequencies between 0.
+ Open protocol
+ Expand
6

Rheological Characterization of Hydrogels

Check if the same lab product or an alternative is used in the 5 most similar protocols
Rheological measurements were performed by means of a controlled-rate viscometer (Thermo Scientific HAAKE, RotoVisco 1, Karlsruhe, Germany) equipped with a plate/plate combination (Ø 20 mm serrated PP20S sensor system). The temperature was set at 32 (±0.5) °C by means of a Peltier temperature controller. The rheological properties of hydrogel samples were measured within the shear rate range of 10–500 s−1. The shear stress vs. shear rate dependencies measured at increasing and then immediately decreasing shear rate were used to determine the thixotropic properties.
+ Open protocol
+ Expand
7

Measurement of Myocyte Volume Changes

Check if the same lab product or an alternative is used in the 5 most similar protocols
Myocyte volume was measured as described previously.1 (link) Isolated myocytes were placed on an inverted microscope stage (model IX51; Olympus, Tokyo, Japan). After a 5-minute stabilization, the chamber was perfused (3 mL/min). Chamber temperature was controlled by a water bath system (37°C; HAAKE; Thermo Electron Karlsruhe, Karlsruhe, Germany or 9°C; Polystat; Cole-Parmer, Vernon Hills, Ill). Cell images were displayed on a video monitor using a charge-coupled device camera (IonOptix, Westwood, Mass). Digital images of viable cells were captured using a video frame grabber (Scion, Frederick, Md) every 5 minutes (Figure 1). Relative cell volume change was determined as described previously.1 (link)
+ Open protocol
+ Expand
8

Spring Tensile Testing Methodology

Check if the same lab product or an alternative is used in the 5 most similar protocols
After clinical use, displacement control was used to test each spring in tension (PCT) (Model 4411; Instron, Norwood, MA, USA). While submersed in a double-chambered device at 37°C, the springs were pre-equilibrated for a minimum of 7 min prior to testing between two 0.032” stainless steel hooks (Figure 2b). The temperature in the inner bath was maintained by pumping water from a temperature-controlled water bath (HAAKE™; Thermo Electron, Karlsruhe, Germany) through the outer bath. The controller thermocouple was located in the inner bath. The springs were stretched at 5 mm/min to the maximum activation length determined by the DMA pretesting, and the load was recorded as a function of the activation length. When the maximum activation length was reached, the crosshead was reversed, and the spring was allowed to relax at the same rate.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!