The largest database of trusted experimental protocols

1260 als autosampler

Manufactured by Agilent Technologies
Sourced in United States

The 1260 ALS autosampler is a component of Agilent's analytical instrumentation solutions. It is designed to automatically introduce liquid samples into an analytical system, such as a chromatograph or spectrometer, for analysis. The autosampler provides precise and reproducible sample handling to ensure reliable and consistent data acquisition.

Automatically generated - may contain errors

10 protocols using 1260 als autosampler

1

HPLC Analysis of Chemical Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC analyses were performed on an Agilent 1260 Infinity system that includes a 1260 quaternary pump VL, a 1260 ALS autosampler, a 1260 Thermostatted Column Compartment, and a DAD Multiple Wavelength Detector (Agilent Technologies, Santa Clara, CA, USA). Detection wavelengths were set at 220, 230, 254, and 280 nm but only 220 nm was used for analysis. A Zorbax Eclipse XDB-C18 analytical column (5 μm, 4.6 × 150 mm) from Agilent Technologies was used. Mobile phase A consisted of 10 mM aqueous ammonium formate buffer titrated to pH 4.5. Mobile phase B consisted of acetonitrile. The injection volume of samples was 10 μL, flow rate was 1.0 mL/min, and the column temperature was set at 25°C. Samples were prepared by preparing a 1 mg/mL solution in 1:1 A:B. All samples were injected in duplicate with a wash in between each run. Run time was 10 minutes with a mobile phase ratio (isocratic) of 1:1 for A:B. Chromatograms were analyzed using the Agilent ChemStation Software (Agilent Technologies). Purity values were calculated from area under the curves of the absorbance at 220 nm of any resulting peaks.
+ Open protocol
+ Expand
2

Insulin-Loaded Nanoparticle Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
To determine the EE% and LE%, an exact amount (30 mg) of drug-loaded NPs was dispersed in distilled water, the freshly prepared colloidal suspension was centrifuged at 14,000 rpm for 10 min at 4°C and the supernatant was analyzed for the determination of nonencapsulated insulin using HPLC method. The samples were injected to Agilent® 1260 infinity equipped with 1260 Quat pump VL, 1260 ALS auto sampler and 1260 DAD VL detector. The detector was set at 214 nm. C18 column was used for HPLC analysis of insulin using linearly regressed calibration curve. The mobile phase was a mixture of buffered aqueous phase and acetonitrile in a ratio of buffer:acetonitrile (70:30). Buffer was prepared from KH2PO4 (0.1 M) and triethylamine (1%), and the pH was adjusted to 2.8 using phosphoric acid. Flow rate was adjusted to 0.5 mL/min, and the data were captured using Agilent Chemstation® software.
To calculate the EE and LE, the amount of nonencapsulated insulin in the supernatant of the centrifuged drug-loaded NPs suspension was determined. All the experiments were performed in triplicate, and the mean values were used to calculate EE% and LE% according to Equations 1 and 2, respectively:
EE%=Total amount of insulinInsulinin the supernatantTotal amount of insulin(100)
LE%=Total amount of insulinInsulin in the supernatantTotal weight of drug-loaded NPs(100)
+ Open protocol
+ Expand
3

Polymer Molecular Weight Analysis via GPC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Polymer molecular
weights and polydispersities were determined using a DMF GPC instrument
operating at 60 °C that comprised two Polymer Laboratories PL
gel 5 μm Mixed C columns and one PL gel 5 μm guard column
connected in series to an Agilent Technologies 1260 Infinity multidetector
suite (refractive index detector only) and an Agilent Technologies
1260 ISO pump fitted with a 1260 ALS autosampler. The GPC eluent was
HPLC-grade DMF containing 10 mM LiBr and was filtered prior to use.
The flow rate used was 1.0 mL min–1 and DMSO was
used as a flow-rate marker. Calibration was conducted using a series
of 10 near-monodisperse poly(methyl methacrylate) standards (Mn = 625–618000 g mol–1, K = 2.094 × 10–3, α
= 0.642). Chromatograms were analyzed using Agilent Technologies GPC/SEC
software version 1.2.
+ Open protocol
+ Expand
4

HPLC Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC analyses were performed on an Agilent 1260 Series (Agilent Technologies) equipped with a 1260 Quat pump VL quaternary pump, 1260 ALS autosampler, 1260 TCC column thermostat, 1260 DAD VL diode array detector and a Hypersil BDS C18 column (4.6 mm × 100 mm; 3.5 µm particle size); injection volume was 10 µL and the wavelength of detection was set at 254 nm. The mobile phases were (A) 0.5% acetic acid in water and (B) methanol. Gradient elution was used from 0% to 100% B for 40 min, 100% B for 10 min. The column was equilibrated with 100% A for 10 min prior to each analysis. The flow rate was set at 1.0 mL/min at 25 °C.
+ Open protocol
+ Expand
5

Quantification of Azithromycin by HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Concentration of azithromycin was measured by a validated high performance liquid chromatography (HPLC) method (53 (link)). Briefly, the HPLC system consisted of 1260 Quat Pump, 1290 Thermostate, 1260 ALS autosampler, 1260 TCC thermostatic column compartment, 1260 VWD variable wavelength detector and an Eclipse Plus column (C18, 150 × 4.60 mm, 5 μm) (all were supplied by Agilent, Waldbronn, Germany). The mobile phase was composed of: (A) 20 mM potassium dihydrogen phosphate (pH was adjusted to 7 with 10% w/v sodium hydroxide) and methanol (B). The isocratic elution program used for azithromycin detection was 80% A and 20% B v/v for 15 min at the flow rate of 1.0 mL/min at a wavelength of 210 nm. The retention time for azithromycin was 12 mins. A calibration curve of azithromycin was linear (r2 > 0.999) in the required concentration range (0.0125–0.5 mg/mL).
+ Open protocol
+ Expand
6

HPLC Optimization and Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
For initial experiments (optimisation of the HPLC conditions), an LC-10 HPLC system (Shimadzu, Nakagyo-ku, Japan) configured as described previously [39 (link)] was used. The column oven was operated at 25 °C and the UV detector was set to 284 nm. Columns tested included a Cyclobond I 2000 5 µm 250 × 4.6 mm column (Sigma-Aldrich, St. Louis, MO, USA), a Lichrospher 100 CN 5 µm 250 × 4 mm column (Merck Millipore, Burlington, MA, USA), a Nucleodur 100-5 NH2 5 µm 125 × 4.6 mm column and a Nucleodex beta-OH 5 µm 200 × 4 mm column (both from Macherey-Nagel, Düren, Germany). A flow rate of 0.8 ml L−1 was used. The eluents (isocratic elution) were formed using a gradient mixer by mixing eluent A (ACN/water 80/20 (v/v)) and eluent B (pure ACN) at the required ratios. For subsequent experiments, a series 1260 Infinity HPLC system (Agilent, Santa Clara, CA, USA) equipped with a 1260 VL quaternary pump, a 1260 ALS autosampler, a 1260 TCC column oven (operated at 25 °C) and either a 1260 MWD VL UV/VIS detector (set to 284 nm) or a 1260 DAD VL diode array detector, was used. For the 1260 system, pre-mixed ACN/water 92/8 (v/v) was used as eluent in the isocratic mode. Chloroform was found to not be retained on the Cyclobond I 2000 stationary phase. Thus, this compound was used to determine t0, the dead volume of the system, which was required for calculation of the capacity factors (k′).
+ Open protocol
+ Expand
7

Size Exclusion Chromatography in DMF

Check if the same lab product or an alternative is used in the 5 most similar protocols
Size exclusion chromatography (SEC) measurements were performed in DMF at 60 °C with a PSS SecCurity system (Agilent Technologies 1260 Infinity, Santa Clara, CA, USA). Sample injection was performed by a 1260-ALS autosampler (Agilent) at 60 °C. GRAM columns (PSS) with dimensions of 300 × 80 mm, 10 μm particle size, and pore sizes of 100, 1000, and 10,000 Å were employed. The DRI Shodex RI-101 detector (ERC, Kawaguchi, Japan) and UV−vis 1260-VWD detector (Agilent) were used for detection. Calibration was achieved using PS standards provided by Polymer Standards Service.
+ Open protocol
+ Expand
8

Quantitative Analysis by LC-TOF/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The samples were analyzed by LC-TOF/MS using an Agilent Technologies 1200 Infinity Series LC system (Santa Clara, CA, USA), consisting of 1260 quaternary pump, a 1260 ALS autosampler coupled to an electrospray ionization (ESI) source, and a 6224 TOF/MS, controlled by the Mass Hunter Acquisition software version B.06.01. The chromatographic separation was accomplished using an Agilent Poroshell 120 SB-C18 2.7 μm particle size, 4.6 mm × 50 mm (inner diameter × length) column, and as mobile phase, a mixture of 80% acetonitrile and 20% water containing 0.1% trifluoroacetic acid at a flow rate of 0.3 mL/min was used. For MS, the ESI was used in the positive mode with the following settings: dual spray needles for continuous infusion of reference mass solution, drying gas flowing at 3.0 L/min, nebulizer pressure of 15 psig, capillary voltage of 3500 V, and fragment or voltage of 175 V. The TOF was tuned and calibrated using Agilent ESI-TOF calibration and tuning mix. The data acquisition mass range was 100 to 1000 m/z at 10,002 transients/scan and 1 scan/s.
+ Open protocol
+ Expand
9

Analytical Techniques for Metabolite Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
Propane levels were determined by headspace injection using an Agilent 490 Micro GC, containing an Al 2 O 3 /KCl column, a thermal conductivity detector (TCD) and a heated injector (110 1C; 100 ms injection) using helium as the carrier gas (10.2 psi). During continuous monitoring mode, fermenter exhaust gases were dried by passage through an ice-cooled condenser prior to entering the Micro GC. Compounds were separated isothermally (100 1C) over 120 s under static pressure conditions, with a sampling frequency of 100 Hz.
Aqueous culture metabolites (glycerol and butyric acid) were analysed by HPLC using an Agilent 1260 Infinity HPLC with a 1260 ALS autosampler, TCC SL column heater and a 1260 refractive index detector (RID). Cell-free culture supernatant samples (10 mL injection) were analysed isocratically on an Agilent Hi-Plex H column (300 Â 7.7 mm; 5 mM H 2 SO 4 ) at 60 1C with a flow rate of 0.7 mL min À1 for 40 minutes. Analyte concentrations determined by Micro GC or HPLC were calculated by comparing the peak areas to a standard curve generated under the same running conditions.
+ Open protocol
+ Expand
10

Quantification of Arbutin in Extracts

Check if the same lab product or an alternative is used in the 5 most similar protocols
Five mg of lyophilisate was dissolved in 1.5 mL of ultrapure water, and centrifuged at 13,000×g at room temperature. Clear supernatant was decanted, mixed with a standard solution of arbutin (80 µg mL -1 ) in 1:1 (v/v) ratio, f i l t e r e d t h r o u g h a s y r i n g e f i l t e r ( 1 3 m m , polytetrafluoroethylene (PTFE) membrane 0.45 μm; Supelco, Bellefonte, PA, USA) and 20 μL of extract was injected into the HPLC system.
An Agilent 1260 Infinity HPLC system was consisted of a 1260 Quat Pump (G1311B), 1260 ALS Autosampler (G1329B), 1260 TCC Column heater (G1316A), and DAD 1260 VL+ (G1315C) detector (Agilent Technologies, Santa Clara, CA, USA). Separations were performed using a Zorbax Eclipse Plus C18 column, 4.6×150 mm in size, with a 1.8 µm particle diameter (Agilent, Technologies, Santa Clara, CA, USA). The mobile phase was (A) 0.1 % formic acid and (B) methanol. The gradient program was as follows: 0-2 min 95 % A, 2-6 min 95 % A to 85 % A, 6-8 min 85 % A to 75 % A, 8-12 min 75 % A to 30 % A, 12-14 min 30 % A to 5 % A, 14-20 min 5 % A, 20-25 min 5 % A to 95 % A. The flow rate was 0.5 mL min -1 . Arbutin was detected and quantified at 290 nm. Quantification was performed using the calibration curve of the arbutin standard, concentration range 20-150 µg mL -1 .
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!