The largest database of trusted experimental protocols

Triple quadrupole 7000c

Manufactured by Agilent Technologies
Sourced in Germany, United States

The Triple Quadrupole 7000C is a high-performance mass spectrometer designed for analytical applications. It features a triple quadrupole configuration, providing enhanced selectivity and sensitivity for quantitative analysis. The core function of the 7000C is to accurately measure and analyze the chemical composition of samples.

Automatically generated - may contain errors

8 protocols using triple quadrupole 7000c

1

GC-MS/MS Analysis of Haloacetic Acids

Check if the same lab product or an alternative is used in the 5 most similar protocols
Analytical determination of methyl esters of HAAs was performed using a 7890B GC connected in series to a 7000C triple quadrupole (Agilent Technologies). Ionization was carried out in the electron ionization mode. One µL of the derivatized extract was injected in splitless mode using a7638B automated injector (split flow=50 mL/min, splitless time=1.5 min). GC separation of the analytes was achieved using a capillary The analyzer was operated in selected reaction monitoring (SRM) mode, using nitrogen (1.5 mL/min) as the collision gas. A minimum of two SRM transitions was acquired per analyte (see Table 4). Figure 4 shows the total ion chromatogram obtained after the analysis of a standard calibration solution at a concentration of 10 µg/mL. Mass acquisition was performed using MSD ChemStation and data analysis was done with Mass Hunter B.08 (Agilent Technologies).
+ Open protocol
+ Expand
2

Comprehensive Analytical Approach for Pesticides

Check if the same lab product or an alternative is used in the 5 most similar protocols
Two different analytical approaches had to be employed to cover all analytes. Medium to highly polar compounds were analyzed according to Barbieri et al. (2020) (link) Nonpolar pesticides were analyzed following a previously optimized method based on liquid-liquid extraction (LLE) and gas chromatography-tandem mass spectrometry (GC-MS/MS) detection (Peris and Eljarrat, 2020) . Water samples (50 mL) were manually extracted twice with 25 mL of EtAc/chloroform (1:1) mixture by classical LLE in a 100 mL separatory funnel. The extract obtained was evaporated under a gentle stream of nitrogen: firstly, to an approximate volume of 1 mL using a Turbovap (Biotage, Sweden), and then, to dryness using a needle evaporator (Reacti-Vap III, Pierce, USA). The dried extract was reconstituted in 50 µL of EtAc (1000x concentration factor). GC-MS/MS determination was performed using a 7890B GC system coupled to a 7000C triple quadrupole (Agilent Technologies, Santa Clara, CA, USA) detector. For chromatographic separation, a DB-5MS column (30 m x 250 μm x 0.25 μm) was used.
Table S2 in SM summarizes the analytical limits of detection (LODs) and quantification (LOQs) achieved for the target pesticides with the methodologies employed. Table S3 in SM reports the main LC-MS/MS and GC-MS/MS acquisition parameters.
+ Open protocol
+ Expand
3

GC-MS/MS Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC/MS. GC-MS/MS method was performed on a 7890B gas chromatography (Agilent Technologies, Waldbronn, Germany) coupled to a triple quadrupole 7000C (Agilent Technologies, Waldbronn, Germany) equipped with a High sensitivity electronic impact source operating in positive mode [60 (link)].
+ Open protocol
+ Expand
4

GC-MS/MS Quantitation of Biological Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC-MS/MS method was performed on a 7890B gas chromatography (Agilent Technologies, Waldbronn, Germany) coupled to a triple quadrupole 7000C (Agilent Technologies, Waldbronn, Germany) equipped with a High sensitivity electronic impact source (EI) operating in positive mode [45 (link)]. The scan mode used was the MRM for biological samples. Peak detection and integration of analytes were performed using the Agilent Mass Hunter quantitative software (B.07.01), exported as tables and processed with R software (version 4.0.3) and the GRMeta package (Github/kroemerlab).
+ Open protocol
+ Expand
5

GC-MS/MS Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC-MS/MS method was performed on a 7890A gas chromatography (Agilent Technologies, Santa Clara, CA, USA) coupled to a triple quadrupole 7000C (Agilent Technologies,) equipped with a high sensitivity electronic impact source (EI) operating in positive mode. Detailed analytical methods are described in [26 (link)].
+ Open protocol
+ Expand
6

GC-MS/MS quantification of analytes

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC-MS/MS method was performed on a 7890B gas chromatography coupled to a triple quadrupole 7000C (Agilent Technologies, Waldbronn, Germany) equipped with a High sensitivity electronic impact source (EI) operating in positive mode. Front inlet temperature was 250°C, injection was performed in splitless mode. Transfer line and ion-source temperature were 250°C and 230°C, respectively. Septum purge flow was fixed at 3 mL/min, purge flow to split vent operated at 80 mL/min during 1 min and gas saver mode was set to 15 mL/min after 5 min. Helium gas flowed through column (J&WScientificHP-5MS, 30m x 0.25 mm, i.d. 0.25 mm, d.f., Agilent Technologies Inc.) at 1 mL/min. Column temperature was held at 60°C for 1 min, then raised to 210°C (10°C/min), followed by a step to 230°C (5°C/min) and reached 325°C (15°C/min), and be hold at this temperature for 5 min.
Collision gas was nitrogen. Scan mode used was MRM for biological samples. Peak detection and integration of analytes were performed using Agilent Mass Hunter quantitative software (B.07.01).
+ Open protocol
+ Expand
7

GC-MS/MS for Compound Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC‐MS/MS method was performed on a 7890B gas chromatograph (Agilent Technologies, Waldbronn, Germany) coupled to a triple quadrupole 7000C (Agilent Technologies, Waldbronn, Germany) equipped with a high sensitivity electronic impact source (EI) operating in positive mode (Viltard et al., 2019 (link)).
+ Open protocol
+ Expand
8

GC-MS/MS Analysis of Biological Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC-MS/MS method was performed on a 7890B gas chromatography coupled to a triple quadrupole 7000C (Agilent Technologies, Waldbronn, Germany) equipped with a High sensitivity electronic impact source (EI) operating in positive mode.
Front inlet temperature was 250°C, injection was performed in splitless mode. Transfer line and ion-source temperature were 250°C and 230°C, respectively. Septum purge flow was fixed at 3 mL/min, purge flow to split vent operated at 80 mL/min during 1 min and gas saver mode was set to 15 mL/min after 5 min.
Helium gas flowed through column (J&WScientificHP-5MS, 30m x 0.25 mm, i.d. 0.25 mm, d.f., Agilent Technologies Inc.) at 1 mL/min. Column temperature was held at 60°C for 1 min, then raised to 210°C (10°C/min), followed by a step to 230°C (5°C/min) and reached 325°C (15°C/min), and be hold at this temperature for 5 min.
Collision gas was nitrogen. Scan mode used was MRM for biological samples. Peak detection and integration of analytes were performed using Agilent Mass Hunter quantitative software (B.07.01).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!