The largest database of trusted experimental protocols

26 protocols using lts350

1

Optoelectronic Characterization of Oxide Films

Check if the same lab product or an alternative is used in the 5 most similar protocols
A sample was placed in a heating–cooling stage Linkam LTS350 that enables a precise temperature control in a wide range of temperatures. A 2-channel function generator Tektronix AFG 3022C was used to apply a linearly increasing voltage pulse on ITO and trigger a light pulse from the ITO side. The current transients were readout from Au electrode on the oscilloscope Tektronix TDS 2024C with adjustable input resistance. For an excitation LED light pulse with a radiant power ~ 100 μW and a wavelength of 595 nm, the attenuation depts in Lead Oxide is 90 μm40 , which is much larger than the detector’s thickness of 14 μm. This provides a homogeneous generation of charge carriers through the bulk of the film. The range of electric fields used in these measurements was limited by the bandwidth of the function generator.
+ Open protocol
+ Expand
2

Thermal Sublimation of Smectic-A Liquid Crystals

Check if the same lab product or an alternative is used in the 5 most similar protocols
Silicon wafers were cleaned using acetone and methanol, followed by rinsing with deionized water. The cleaned silicon substrates were spin-coated with PEI (Aldrich; MW: 60,000) for tangential anchoring of the LC. A crystalline powder of Y002 at the silicon substrate was heated to the isotropic phase (200 °C) on a hot stage (LINKAM LTS350) regulated by a temperature controller (LINKAM TMS94). The sample was cooled to 190 °C at a rate of 10 °C min−1. The sublimation experiments were performed in the range between 190 and 130 °C with an increment of 10 °C for different sintering times, ranging from 5 to 106 s. Thermal treatment was performed in the air. The instantaneous rates for mass loss percentage at the temperatures 190, 180, 160 and 130 °C, were determined to be 1.64, 0.82, 0.30 and 0.02% per °C, respectively, Supplementary Fig. 9. The volume available for the SmA molecules to sublimate is orders of magnitudes larger than the volume of the original SmA film (cubic decimetres versus cubic millimetres). Following the thermal treatment, the samples were cooled down to the room temperature, preserving the SmA structure (as verified by X-ray diffraction). The surface morphology was visualized by sputter-coating platinum (Pt) film and imaging the latter under a SEM.
+ Open protocol
+ Expand
3

Planar Liquid Crystal Cell Alignment

Check if the same lab product or an alternative is used in the 5 most similar protocols
The cells were aligned in a planar fashion by using rubbed layers of PI-2555. The directions of rubbing at two opposite plates are antiparallel. The glass plates contain transparent indium tin oxide electrodes of area 5 × 5 mm2. The temperature of the cell is controlled with a Linkam LTS350 hot stage and a Linkam TMS94 controller. The AC electric field was applied using a waveform generator (Stanford Research Systems, Model DS345) and amplifier (Krohn-hite Corporation, Model 7602).
+ Open protocol
+ Expand
4

Polarized Light Microscopy of Oleogels

Check if the same lab product or an alternative is used in the 5 most similar protocols
Polarized light microphotographs (PLM) of the oleogels were obtained using a polarizing light microscope (Olympus BX51; Olympus Optical Co., Ltd., Tokyo, Japan) equipped with a color video camera (KP-D50; Hitachi Digital, Tokyo, Japan) and a heating/cooling stage (TP94; Linkam Scientific Instruments, Ltd., Surrey, England) connected to a temperature control station (LTS 350; Linkam Scientific Instruments, Ltd.) and a liquid nitrogen tank. A drop of the alkyl ester-MSG solution was applied on a glass slide and then set in the heating/cooling stage of the microscope. We applied the same time-temperature program as described in section 2.3, but here after arriving at a temperature 10 • C above the corresponding T Cr the sample was smear on the slide surface using another preheated glass slide at a 45 • angle. Then, the stage was closed and the cooling continued until achieving -5 • C. PLM of the corresponding microstructure were obtained after 2 min at this temperature.
+ Open protocol
+ Expand
5

Birefringence Measurement via Polarizing Microscopy

Check if the same lab product or an alternative is used in the 5 most similar protocols
Optical studies were performed using a polarizing microscope (Axioskop-40) and a Linkam LTS 350 temperature chamber. The investigations were carried out in a quasi-static regime with an accuracy of ±0.1 K in the range of 90-400 K. The birefringence was measured using a Berek compensator (Leica) with an accuracy of ≈0.00001.
+ Open protocol
+ Expand
6

Thermal Characterization of Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
The
thermal behavior of samples was characterized using a Leica DM4500
polarizing optical microscope with ×5, ×10, ×20, and
×40 objectives, a Linkam LTS350 heating stage, a CI94 temperature
controller, and a LN2 cooling system. A small amount of the sample
was placed atop a glass microscope slide and covered with a thin cover
glass. POM analyses were performed using three heating–cooling
cycles: first, a heating–cooling cycle to allow the mixing
of components and the potential kinetically trapped states to transition
to their thermodynamic equivalents. The thermotropic behavior was
studied during the second heating–cooling cycle, after which
the samples were allowed to stand at room temperature to study the
metastable phases.
+ Open protocol
+ Expand
7

Perovskite Device Electrical Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
5 nm Cr and 20 nm Au layers were sequentially deposited on perovskite by thermal evaporation. The channel was defined by an Al wire with diameter of 20 μm serving as shadow mask. The samples were loaded on Linkam LTS350 stage purged with dry nitrogen. DC bias was applied, and current was measured at dark using Keithley 4200-SCS parameter analyzer with 4200-PA remote preamplifier. The voltage was turned on after device initialization and turned off as the ion accumulation was saturated, while the current over time was monitored. The samples were relaxed over 30 min after each test till they recovered from the polarized state.
+ Open protocol
+ Expand
8

Impedance Spectroscopy of Ionic Transport

Check if the same lab product or an alternative is used in the 5 most similar protocols
Ionic transport properties were studied using
a HF2IS impedance
spectroscope (Zurich Instruments). The samples were measured in MicruX
ED-IDE1-Au-interdigitated electrodes (averaged cell constant 0.0276
cm–1 from the manufacturer) that were glued onto
microscope slides with copper wires directly soldered onto the electrode
cells. A small amount of sample was placed on the electrode, and the
sample was heated up to its flowing state to fill the cell. The filled
sample cell was dried in a vacuum oven overnight at 25–40 °C
prior to measurement. At the start of the measurement, the cell was
heated to 150 °C and stabilized for 20 min. The ionic transport
measurements were then performed during cooling scans from 150 to
0 °C at a rate of 1 °C min–1, and the
equilibration time was 3 min at the measurement temperature to ensure
a thermal equilibrium. A Linkam LTS350 heating plate, a CI94 temperature
controller, and a LN2 cooling system were used. In the EIS measurement,
a 4-terminal measurement setup with an AC excitation of 100 mV was
applied, and the impedance spectroscopy was performed over the 10
MHz to 10 Hz frequency range. The analysis of EIS data was performed
by using ZView software (Scribner Associates). Additional information
on the EIS can be found in Supporting Information Section S7.
+ Open protocol
+ Expand
9

Planar Alignment ITO Cell Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
The cell is composed of two glass substrates coated with indium tin oxide (ITO), which serve as the transparent electrode of active area 5 × 5 mm2. The alignment layers PI-2555 coated on the surface of ITO were rubbed to provide a planar alignment. The temperature of the cell is controlled with a Linkam LTS350 hot stage and a Linkam TMS94 controller. The AC voltage is applied using a waveform generator (Stanford Research Systems, Model DS345) and an amplifier (Krohn-hite Corporation, Model 7602).
+ Open protocol
+ Expand
10

Impedance Spectroscopy of Perovskite Devices

Check if the same lab product or an alternative is used in the 5 most similar protocols
To investigate environment dependence of transport mechanisms, impedance spectroscopy (IS) measurements were performed on the device using a Zahner IM6 electrochemical workstation. To avoid the physical or chemical damage‐induced degradation in the device, Au plated and spring loaded round pogo pins on a PCB substrate were used for reliable electrical contact. The device was sealed inside a chamber (Linkam LTS350). For initial measurement in dark condition, N2 gas was purged into the chamber overnight. The impedance spectra was checked constantly and the time‐ and environmental‐dependent impedance responses were measured within 1 h. The data point was captured every 2 min during the measurements. For environmentally dependent transport mechanisms under illumination, the measurements were performed under 1 Sun illumination using a Newport solar simulator. Impedance spectroscopy was recorded inside the sealed chamber under different gaseous conditions at a constant flow rate of around 100 mL min−1. The light illumination time was limited to 20 min as continuous illumination causes significantly excessive degradation of the perovskite thin films.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!