The largest database of trusted experimental protocols

2998 photodiode array

Manufactured by Waters Corporation
Sourced in United States

The 2998 Photodiode Array is a high-performance liquid chromatography (HPLC) detector manufactured by Waters Corporation. It utilizes a photodiode array to provide simultaneous multi-wavelength detection for HPLC analysis. The device is designed to accurately measure the absorbance of analytes eluting from an HPLC column over a wide range of wavelengths.

Automatically generated - may contain errors

9 protocols using 2998 photodiode array

1

Characterization of Chalcone 2HMC

Check if the same lab product or an alternative is used in the 5 most similar protocols
All commercially available reagents were purchased from Sigma-Aldrich (St. Louis, MO, USA). 1H NMR (Proton Nuclear Magnetic Resonance) and 13C NMR (Carbon-13 Nuclear Magnetic Resonance) spectra were recorded on a Bruker Avance III (500 MHz) spectrometer (Bruker Optik GmbH, Ettlingen, Germany). Infrared (IR) spectra were performed on a Bomem M102 spectrometer (Bomem Inc., Vanier, Quebec, Canada). The melting point was measured using a melting point apparatus (Karl Kolb GmbH & Co., Dreieich, Germany). The purity of the compound was determined by high performance liquid chromatography (HPLC) using the Waters Alliance–2695 apparatus, XTerra C18 (5 μm, 4.6 mm × 150 mm) column, and Waters 2998 Photodiode Array (PAD) detector at 330 nm for quantification of chalcone 2HMC. The injector was programmed to inject a volume of 10 μL. The mobile phase was CH3CN:aqueous buffer containing 0.1% trichloroacetic acid (55:45).
+ Open protocol
+ Expand
2

UHPLC-MS Analysis of Complex Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
A Waters ACQUITY Arc (U)HPLC system (Milford, MA, USA) equipped with an automatic sample manager (FTN—R), a quaternary solvent manager—R, a Waters 2998 photodiode array (PDA) detector and a Waters empower software (version 7.0.3471.914) workstation was used. MS analysis was performed by a Thermo Fisher Orbitrap Exploris 120 mass spectrometer (Waltham, MA, USA). The extraction of the sample was sonicated using a DT1028H sonicator (Bandelin electronic GmbH & Co. KG, Berlin, Germany). For processing of the raw samples, a TDL-5—A centrifuge (Shanghai Anheng Scientific Instrument Factory, Shanghai, China) and an XQ100 grinder (Shanghai Guangsha Industry and Trade Co., Ltd., Shanghai, China) were employed. Samples were dried using a DHG—9140A oven from Shanghai Jinghong Experimental Equipment Co., Ltd. in China. A KERN ABT100—5M analytical balance (KERN and SOHN GmbH, Balingen, Germany) was used to weigh the samples.
+ Open protocol
+ Expand
3

HPLC Analysis Using Waters Alliance

Check if the same lab product or an alternative is used in the 5 most similar protocols
The analytical HPLC is Waters Alliance (e2695 separations module), equipped with 2998 Photo diode Array (PDA). Data acquisition and control were carried out using Empower 3 chromatography data software (Waters, Germany). The chromatography was performed under reverse phase conditions using a TSQ Quantum Access MAX (Thermo Scientific, San Jose, CA, USA) which includes a Dionex Pump with degasser module, an Accela PDA detector and an Accela Autosampler.
+ Open protocol
+ Expand
4

Multimodal Analytical Approach for Compound Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
Gas Chromatography (Agilent 6890N Network) associated with a Mass Spectrometer (Agilent 5975 Inert Mass Selective Detector) and Flame Ionization Detector with the separation column DB 5MS 29.8 m × 0.32 mm × 0.5 µm was used for concentration determinations. Helium was used as carrier gas at a flow rate of 1 mL∙min−1 and oven program was: Forty degrees celsius during 2 min; heating ramp of 5 °C∙min−1 up to 320 °C and hold for 20 min. The FID detector was operated at 320 °C. The mass spectrometer was operated at 220 °C; scans were carried out between 35 and 620 m/z.
For large solutes, liquid chromatography (Agilent) associated with a UV detector (Waters 2998 Photodiode Array) with the separation column Poroshell 120 EC-C18 2.7 µm, 4.6 mm × 100 mm was used. An isocratic program with a mixture of 5 % of tetrahydrofuran and 95 % of acetonitrile was applied for 20 min with a flow rate of 1 mL∙min−1 at 40 °C.
+ Open protocol
+ Expand
5

Purification of Crude Peptides by HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
The mobile phases that were used for the HPLC analysis are A: H2O + 0.1% TFA and B: CH3CN + 0.1% TFA. The crude peptide is purified by a preparative method on a Waters device (pump module: 2535 Quaternary Gradient module; UV detector: Waters 2998 Photodiode Array), with an XBridge BEH 300A Prep C18 column OBD 5 μm 19 × 250 mm, at a flow rate of 13 mL/min. Analytical HPLC is performed on a Waters device (pump module: e2695 separation module; UV detector: Waters 2998 Photodiode Array) with an XBridge BEH 300A C18 column (5 μm, 4.6 mm × 250 mm, Waters). The indicated retention times are given in minutes. Gradient used was 0–50% B in 50 min with a wash phase at 80% B for 10 min and a column equilibration phase at 100% A for 10 min.
+ Open protocol
+ Expand
6

HPLC-MS Analysis and Purification of PCBLs

Check if the same lab product or an alternative is used in the 5 most similar protocols
Analysis was performed on a Waters platform, composed of a Waters 2695 HPLC unit equipped with a Waters 2998 photodiode array (PDA) detector and a Luna silica column (4.6 mm i.d. ×250 mm, 5 µm particle size; Phenomenex, Torrance, CA, USA). The unit was coupled with a Bruker Esquire 3000 Plus ion trap mass spectrometer (Bruker–Franzen Analytik GmbH, Bremen, Germany) equipped with ESI. The PCBLs was dissolved in methanol to prepare a 20 mg/mL solution. The mobile phases consisted of hexane/methanol/ethyl acetate (10∶3∶1, v/v/v) (A) and hexane/methanol/ethyl acetate (1∶3∶1, v/v/v) (B). Separations were done by linear gradient at 37°C at 1 mL/min flow rate as follows: 0–90 min, 83.6% A; 90–130 min, 83.6–19.4% A; 130–150 min, 19.4% A. The PDA detector was set to 280 nm and scanning was done from 200 to 400 nm.
Preparation was performed on a Phenomenex (Torrance, CA, USA) Luna silica preparative column (21.2 mm i.d. ×250 mm) with a 5 µm particle size at 37°C. A Shimadzu system equipped with a CBM-20A module, an SIL-10AP autosampler, an SPD-20A UV–vis detector, and two LC-8A pumps was used. Elution was the same as the method described above. The flow rate was 21.6 mL/min and the absorption wavelength was at 280 nm. On a given run, 1 mL (200 mg/mL) extract was applied. The fractions were collected and evaporated under vacuum.
+ Open protocol
+ Expand
7

HPLC Analysis of Compound Separation

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC measurements were carried out on a Waters Alliance 2695 HPLC system (America) equipped with a 2998 Photo-Diode Array. Chromatographic separations were performed using a Diamonsil C18 (250 mm×4.6 mm, 5 μm), connected with a C18 guard column (10 mm×4.6 mm, 5 μm). A linear gradient elution of eluent A (acetonitrile) and B (0.5% acetic acid; v/v) was used for the separation. The elution program was performed as follows: 19% A at 0–10 min, 19%–50% A at 10–20 min, 50%–80% A at 20–30 min, and 80%–100% A at 30–40 min.
+ Open protocol
+ Expand
8

HPLC Analysis of Phytochemicals

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC analyses were carried out on a Waters e2695 HPLC system equipped with a 2998 photo diode array (PDA) detector (Waters Corp., Milford, MA, USA) using a Phenomenex Luna C18 column (5 μm, 250 × 4.6 mm). Mobile phase A was acetonitrile and mobile phase B was 0.01% (v/v) phosphoric acid aqueous solution. Linear gradient elution was used as follows: 8% A in 0–20 min, 8–24% A in 20–50 min, 50% A in 50–60 min, 50–90% A in 60–68 min. The flow rate was 1.0 mL/min and the column temperature was 27°C. The injection volume was 10 μL. The UV wavelengths were 274 nm for gallic acid, catechin, albiflorin, paeoniflorin, ethyl gallate, galloylpaeoniflorin, pentagalloylglucose, benzoic acid, benzoylpaeoniflorin, and paeonol and 257 nm for oxypaeoniflorin.
+ Open protocol
+ Expand
9

HPLC Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC analyses were conducted on a Waters e2695 HPLC system equipped with a 2998 photodiode array (PDA) detector (Waters Corp., Milford, MA) using a C18 column (AkzoNobel Kromasil 100-5C18, 250 × 4.6 mm). Mobile phase A was acetonitrile, and mobile phase B was a formic acid aqueous solution with the formic acid concentrations shown in Table 1. The mobile phase gradient and the flow rate were set as shown in Tables 1 and2. After 80 min of gradient elution, 5 min of column wash gradient (90% acetonitrile) was used to elute the compounds remaining in the column. The column temperature was set at 35°C, and the injection volume was 10 μL. UV spectra in the range of 250 -400 nm with a resolution of 1.2 nm were collected by the PDA detector.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!