The largest database of trusted experimental protocols

Dsc 7 calorimeter

Manufactured by PerkinElmer
Sourced in United States

The DSC-7 calorimeter is a thermal analysis instrument that measures the heat flow associated with material transitions and reactions as a function of temperature and time. It provides quantitative and qualitative information about physical and chemical changes that involve endothermic or exothermic processes or changes in heat capacity.

Automatically generated - may contain errors

9 protocols using dsc 7 calorimeter

1

Comprehensive Characterization of Polymeric Films

Check if the same lab product or an alternative is used in the 5 most similar protocols
All films were characterized in terms of surface, morphological and thermal features. The surface microstructure was evaluated by field emission scanning electron microscopy (FESEM, Supra 25 Zeiss, Baden-Württemberg, Germany). Samples were gold coated with an Agar automatic sputter coater and then analyzed.
Water contact angle (WCA) was assessed by the sessile drop method in air using a KSV CAM101 instrument (KSV Instruments Ltd., Helsinki, Finland).
To determine the thermal transitions a DSC7 calorimeter (Perkin Elmer, Waltham, MA, USA) was used under the following conditions: heating at 20 °C min−1 from −120 °C to 40 °C above fusion temperature (I scan), holding for 3 min, quenching to −80 °C (100 °C min-1) and heating to 40 °C above the melting point at 20 °C min−1 (II scan). X-ray diffraction (XRD) patterns of polymeric films were carried out by using an X’PertPro diffractometer (PANalytical, Malvern, UK) equipped with a fast-solid state X’Celerator detector and a copper target (λ = 0.15418 nm). Data were acquired at each 0.10° step for 100 s in the 5–60° 2θ interval.
+ Open protocol
+ Expand
2

Thermal analysis of polyamide 410

Check if the same lab product or an alternative is used in the 5 most similar protocols
DMA was used to study the phase behavior. A TA Q800 viscoelastometer (New Castle, DE, USA) that provided the loss tangent (tanδ) against temperature was employed. The single cantilever bending mode was selected and constant heating rate and frequency were set at, respectively, 4 °C/min and 1 Hz. The studied temperature interval ranged from −100 °C to 150 °C. Thermal properties related to melting and crystallization were studied by DSC using a Perkin-Elmer DSC-7 calorimeter (Waltham, MA, USA) calibrated using an indium standard as a reference. First heating scan was performed from 30 °C to 300 °C at 20 °C/min, and then cooled at the same rate. The melting and crystallization temperatures (Tm, Tc) were determined, respectively, from the maxima of the corresponding peaks during the heating and cooling scans, and the melting and crystallization enthalpies were determined from the areas under each of these peaks. The crystallinity of PA410 was calculated from the melting and cold crystallization enthalpies, considering an enthalpy of a 100% crystalline ( Hf ) PA410 of 269 J/g [45 (link)].
+ Open protocol
+ Expand
3

Crosslinking Density Determination via DSC and DMA

Check if the same lab product or an alternative is used in the 5 most similar protocols
Differential scanning calorimetry (DSC) was conducted using a Perkin-Elmer DSC-7 calorimeter calibrated using an indium standard as a reference. Samples were taken from previously cured specimens and were heated from 30 °C to 250 °C at a rate of 20 °C/min under a nitrogen atmosphere.
Dynamic mechanical analysis (DMA) was performed on rectangular specimens in a TA Q800 viscoelastometer in single cantilever bending mode. The temperature interval was set from −100 °C to 250 °C, and the heating rate at 4 °C/min. The tests were carried out at a frequency of 1 Hz.
In order to calculate the crosslinking density of the samples, the elasticity theory was used (Equation (1)) [49 (link),55 (link)].
νe=Er3RTr
where Er is the storage modulus in the rubbery state, Tr is the temperature corresponding to the Er value (T = 245 °C) and R is the ideal gas constant (R = 8.314 J/mol K).
+ Open protocol
+ Expand
4

Characterizing Crosslinking Density and Glass Transition in Polymers

Check if the same lab product or an alternative is used in the 5 most similar protocols
The glass transition temperature (Tg) and the crosslinking density (νe) were determined by dynamic mechanical analysis (DMA) in a TA Q800 viscoelastometer (New Castle, DE, USA) in single cantilever bending mode. A temperature range of −100 °C to 250 °C was used at a heating rate of 4 °C/min. The frequency and amplitude used were 1 Hz and 15 μm, respectively. Equation (1) was used to calculate the crosslinking density [63 (link)]: νe=Er3RTr,
where R is the ideal gas constant (8.314 J/mol·K), Er is the storage modulus in the rubbery state, and Tr is the temperature at which Er was taken (245 °C).
Differential scanning calorimetry (DSC) was performed in a Perkin Elmer DSC-7 calorimeter (Waltham, MA, USA) calibrated with an indium standard. The samples, which were taken from the cured specimens, were heated from 30 °C to 250 °C at 20 °C/min under a nitrogen atmosphere.
+ Open protocol
+ Expand
5

Differential Scanning Calorimetry of Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
Differential scanning calorimetry measurements were carried out
using a Perkin-Elmer DSC7 calorimeter with an intracooler cooling
unit at −20 °C (ethylene glycol–water, 1:1, V/V,
cooling mixture). Measurements were carried out at different scanning
rates |β| = 2, 10, and 20 °C min–1 under
a nitrogen gas flow of 20 mL min–1. Samples with
approximate mass between 2 and 5 mg were placed in 10 or 50 μL
aluminum pans, with similar empty ones used as reference. Temperature
calibration was performed with high grade standards:26 (link),27 (link) biphenyl (CRM LGC 2610, Tfus = (68.93
± 0.03) °C) and indium (Perkin-Elmer, x = 0.9999, Tfus = 156.6 °C). Indium
was also used for enthalpy calibration (ΔfusHm = 3286 ± 13 J mol–1). Benzoic acid (CRM LGC 2606, Tfus =
(122.35 ± 0.02) °C) was used to check the calibration. Pyris
software version 3.50 was used for instrument control and result analysis.
The reported first-order phase transition temperatures correspond
to the onset of the peaks. The glass transition temperatures were
determined from the midpoints of the characteristic baseline step
changes.
+ Open protocol
+ Expand
6

Thermal Behavior Analysis of 3D Printed PLA and NFRCs

Check if the same lab product or an alternative is used in the 5 most similar protocols
A differential scanning calorimetry (DSC) analysis of printed PLA and NFRCs was performed using a DSC 7 calorimeter (Perkin-Elmer, Waltham, MA, USA) to study NFRCs’ melting behavior and crystallinity. Only the first heating was performed from room temperature to 220 °C at a heating rate of 5 °C/min under an inert nitrogen atmosphere. The cold crystallization temperature Tcc and melting temperature Tm were recorded during the first heating. The degree of crystallinity Xc was calculated during the first heating to determine the effect of the printing process on the crystallinity of the printed PLA and NFRCs. A similar approach has been used previously, in which only the first heating curves were recorded to correlate the degree of crystallinity resulting from the 3D printing process with the mechanical properties without eliminating the thermal history [48 (link),49 (link)]. Xc was calculated using the following equation: Xc=ΔHmΔHccΔHmo×WPLA
where ΔHm is the melting enthalpy, ΔHcc is the cold crystallization enthalpy, ΔHmo=93 J/g is the melting enthalpy of 100% crystalline PLA [50 (link)], and WPLA is the weight fraction of PLA in the sample.
+ Open protocol
+ Expand
7

Thermal Characterization of PA410 Polymer

Check if the same lab product or an alternative is used in the 5 most similar protocols
The phase behavior was studied by dynamic mechanical analysis (DMA) using a TA Q800 viscoelastometer (New Castle, DE, USA) that provided the loss tangent (tanδ) against temperature. The scans were carried out in single cantilever bending mode at a constant heating rate of 4 °C/min and a frequency of 1 Hz, from −100 to 150 °C. The melting and crystallization behavior of the materials was studied by DSC using a Perkin–Elmer DSC-7 calorimeter (Waltham, MA, USA), which was calibrated using an indium standard as a reference. The samples were first heated from 30 to 300 °C at 20 °C/min and then cooled at the same rate. The melting and crystallization temperatures (Tm, Tc) were determined, respectively, from the maxima of the corresponding peaks during the heating and cooling scans, and the melting and crystallization enthalpies were determined from the areas of each of these peaks. The degree of crystallization of PA410 was calculated from the melting and cold crystallization enthalpies, taking the enthalpy of a 100% crystalline ( ΔHf ) PA410 as 269 J/g [5 (link)].
+ Open protocol
+ Expand
8

Comprehensive Materials Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
DMA was performed
using a PerkinElmer Pyris Diamond DMA analyzer with a sample size
of 5.0 cm × 1.0 cm × 0.2 cm. The storage modulus E′ and tan δ were determined as the sample
was subjected to the temperature scan mode at a programmed heating
rate of 5 °C/min at a frequency of 1 Hz. The test was performed
using a bending mode with an amplitude of 5 μm. UV spectra were
obtained using UV WinLab Lambda 25. Thermogravimetric analysis was
performed using a PerkinElmer Pyris 1 at a heating rate of 20 °C/min
in an atmosphere of nitrogen or air. DSC scans were obtained using
a PerkinElmer DSC 7 calorimeter in a nitrogen atmosphere at a heating
rate of 10 °C/min. NMR measurements were recorded using a Varian
Inova 600 NMR spectrometer in DMSO-d6,
and the chemical shift was calibrated by setting the chemical shift
of DMSO-d6 at 2.49 ppm. IR spectra were
obtained from at least 32 scans in the standard wavenumber range of
667–4000 cm–1 using a PerkinElmer RX1 infrared
spectrophotometer.
+ Open protocol
+ Expand
9

Thermal Analysis of Phase Change Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
Thermal analyses of the samples were carried out on a DSC-7 calorimeter (Perkin Elmer Inc., Norwalk, USA) from -20 to 20ºC under a nitrogen atmosphere using a refrigerating cooling accessory (Intracooler 2) (Perkin Elmer Inc., Norwalk, USA). The scanning rate was 2ºC/min in order to minimize the influence of this parameter in the thermal properties. The amount of material used for the DSC experiments was adjusted so as to have a theoretical PCM content of 1-2 mg approximately. The enthalpy results obtained were, thus, corrected according to this PCM content. All tests were carried out in triplicate.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!