The largest database of trusted experimental protocols

Q200 dsc instrument

Manufactured by TA Instruments
Sourced in United States

The Q200 DSC is a differential scanning calorimetry (DSC) instrument manufactured by TA Instruments. The core function of the Q200 DSC is to measure the heat flow associated with physical and chemical changes in a sample as a function of temperature or time.

Automatically generated - may contain errors

9 protocols using q200 dsc instrument

1

Conformational Analysis of Tissue Scaffolds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Conformational analysis was performed on untreated and decellularized patches to test whether sterilization, decellularization, and different storing conditions could affect the SF structure.
Fourier transform-infrared (FTIR) spectroscopy measurements were performed on a NEXUS (Thermo Nicolet) FTIR spectrometer employing an attenuated total reflectance (ATR) accessory model, Smart Performer. All spectra were obtained with a ZnSe crystal cell as the internal reflection element. Spectra were recorded in the 400 to 4,000 cm-1 wave number range by accumulating 64 scans at a resolution of 4 cm-1, subjected to data smoothing (15 points with the Savitzky-Golay method) and normalized to the 1,452 cm-1 peak before any data processing. Each spectrum was the average of at least three spectra measured in different areas of the sample. The crystallinity index was calculated as the intensity ratio between the two amide III components at 1260 cm-1 and 1230 cm-1 (C.I. = I1260/I1230) [28 (link)].
Differential scanning calorimetry (DSC) measurements were performed with a DSC Q200 instrument (TA Instruments, Waters, Milan, Italy). The scanned temperature ranged from room temperature to 400°C, at a heating rate of 10°C/minute. The analysis was carried out on samples of about 2 to 3 mg weight, in open aluminum pans of 40 μL volume, under N2 gas flux.
+ Open protocol
+ Expand
2

Comprehensive Thermal Analysis of Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
The thermal analysis includes DSC, DSC-TG, in situ FTIR, and DSC-TG-FTIR-MS quadruple technology experiments. Thermal analysis experiments were carried out with model TG-DSC STA 449C instrument (NETZSCH, Germany) and DSC Q200 instrument (TA, America). Operation conditions: sample mass, 0.5 mg; atmosphere, dynamic nitrogen; aluminum cell. IR spectra were recorded on a Nicolet 60SX FTIR spectrometer with HgCdTe detector. In situ FTIR spectroscopy studies were carried out with Nicolet 60 SXR FTIR spectrometer. Operation conditions: sample mass, 0.5 mg; heating rate, 10 °C/min; resolution, 4 cm−1; spectral acquisition rate, 17.8 file/min, 16 scans/file; temperature range, 20~450 °C. Cautions and experimental condition can be found in Supplementary Materials File S1.
+ Open protocol
+ Expand
3

Spectroscopic and Thermal Characterization of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols

1H, 13C and 19F NMR spectra of intermediates were recorded on a JEOL resonance ECZ 400S spectrometer (400 MHz) in CDCl3. Tetramethylsilane (TMS) and CDCl3 were used as internal standards for 1H and 13C NMR, respectively; CF3CO2H was used as an external standard for 19F NMR. FT-IR spectra were recorded on a Nicolet AVATAR-360 FT-IR spectrophotometer with a resolution of 4 cm−1. Number-average molecular weights (Mn) and molecular weight distributions (Mw/Mn) were obtained on a conventional gel permeation chromatography (GPC) system equipped with a Waters 515 Isocratic HPLC pump, a Waters 2414 refractive index detector, and a set of Waters Styragel columns (HR3 (500–30 000), HR4 (5000–600 000) and HR5 (50 000–4 000 000), 7.8 × 300 mm, particle size: 5 μm). GPC measurement was carried out at 35 °C using tetrahydrofuran (THF) as eluent with a flow rate of 1.0 mL min−1. The system was calibrated with linear polystyrene standards. Differential scanning calorimetry (DSC) was performed on a TA Q200 DSC instrument in N2 with a heating rate of 10 °C min−1. Thermogravimetry analysis (TGA) was conducted on a TA Discovery TGA 55 thermal analysis system in N2 with a heating rate of 10 °C min−1. UV/vis spectra were acquired on a Hitachi U-2910 spectrophotometer.
+ Open protocol
+ Expand
4

Thermal Analysis of Amorphous Solid Dispersions

Check if the same lab product or an alternative is used in the 5 most similar protocols
The Tgs of the ASDs were detected by MTDSC using a Q200 DSC instrument and TA Instruments DSC Refrigerated Cooling System (TA Instruments, New Castle, Delaware). Samples of 3–4 mg were heated in aluminum pans with sealed aluminum lids. Nitrogen was used as the purge gas at a flow rate of 20 mL/min. Samples were heated from 0 °C to 110–185 °C at 2 °C/min, with an amplitude of ± 0.318 °C and a modulation period of 60 s. Results were analyzed with the Universal Analysis 2000 software (TA Instruments). The midpoint of the transition was taken as the Tg. Sapphire was used to calibrate the heat capacity, while indium was used for the calibration of enthalpy and temperature. All measurements were carried out in triplicate.
+ Open protocol
+ Expand
5

Thermal Analysis of PLA Composites

Check if the same lab product or an alternative is used in the 5 most similar protocols
DSC was performed using a Q200 DSC instrument (TA Instruments, New Castle, DE, USA) equipped with a liquid nitrogen cooling accessory. The samples were first heated from 0 °C to 200 °C at a rate of 10 °C/min, followed by an isothermal step for 1 min to eliminate previous thermal history, and then cooled to 0 °C at a rate of 10 °C/min. After the isothermal step for 1 min, the second heating was conducted between 0 °C and 200 °C at a heating rate of 10 °C/min. The measurements were performed in a nitrogen atmosphere at a flow rate of 50 mL/min using aluminium standard pans with lids. The sample mass was between 5 and 6 mg. The cold crystallization temperature (Tcc), melting temperature (Tm), enthalpy of cold crystallization (ΔHcc), and enthalpy of melting (ΔHm) were determined. The degree of crystallinity (Xcc) was estimated according to the following equation: Xcc(%)=ΔHccΔH0×XPLA×100(%)
where ΔHcc refers to the cold crystallization enthalpy of the PLA and composites; ΔH0 refers to the enthalpy value of 100% crystalline PLA, which is 93.6 J/g; and XPLA refers to the weight ratio of PLA in the composites.
+ Open protocol
+ Expand
6

Thermal Analysis of Nanoassemblies

Check if the same lab product or an alternative is used in the 5 most similar protocols
To further confirm the incorporation of the drug
into the nanoassemblies, we carried out a DSC analysis. Each sample
(3.2 mg) was weighed into a hermetic aluminum sample pan and then
sealed tightly with a lid. Samples were then loaded into a Q200 DSC
instrument (TA Instruments). Prior to insertion of the sample, the
instrument was cooled to 0 °C under nitrogen. The thermal phase
transition of each sample was recorded up to 280 °C. Each sample
was run three times to ensure reproducibility.
+ Open protocol
+ Expand
7

Thermal Analysis of Terpene Homopolymers

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 8

The polymer samples were characterized using differential scanning calorimetry using a TA instruments Q200 DSC instrument. The DSC plots for the citronellol homopolymer, geraniol homopolymer, and linalool homopolymer are presented in FIGS. 28-30 respectively. The glass transition temperatures of these homopolymers are −50.2° C. for the citronellol homopolymer, −40.82° C. for the geraniol homopolymer, and −33.35° C. for the linalool homopolymer

+ Open protocol
+ Expand
8

Characterization of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Chemical reagents and solvents were purchased from Aladdin (Shanghai, China) and used as supplied without further purification. 1H NMR and 13C NMR spectra were obtained using a JEOL GSX 600 MHz nuclear magnetic resonance (NMR) spectrometer in a d6-DMSO solution by using tetramethylsilane as internal standard. Fourier transform infrared (FT-IR) spectra were recorded on a Nicolet-5700 FTIR spectrometer using pressed KBr pellets in the wavelength range of 4000 cm−1 to 400 cm−1. Elemental analyses (C, H, and N) were conducted using a Vario EL CUBE device. DSC was performed by a Q200 DSC instrument (TA Instruments, United States) at a heating rate of 10 K min−1 under flowing high-purity nitrogen. TGA was performed with an SDT Q600 TGA instrument (TA Instruments, United States) at a heating rate of 10 K min−1 under flowing high-purity nitrogen.
+ Open protocol
+ Expand
9

Characterization of Pb-based Metal-Organic Framework

Check if the same lab product or an alternative is used in the 5 most similar protocols
The FT-IR spectrum was recorded on Nicolet 380 FT-IR spectrophotometer (Thermo Fisher Nicolet, Waltham, MI, USA) employing a KBr matrix with a resolution of 4 cm−1, in the wavelength range of 400 cm−1 to 4000 cm−1. Elemental analysis was performed on a Vario ELCUBE Elemental Analyzer (Elementar, Hanau, Germany). DSC was performed by a Q200 DSC instrument (TA Instruments, New Castle, DE, USA) at a heating rate of 5, 10, 15 and 20 K·min−1, respectively, in flowing high-purity nitrogen. Approx. 1.0 mg sample was sealed in aluminum pans in the temperature range of 313 to 773 K for DSC experiments. The sensitivity to impact stimuli was determined by fall hammer apparatus according tothe standard staircase method (GJB of China). With a step of 0.04 cm, 50 ± 1 mg of test specimens were used and a 10-kg drop weight was allowed to fall freely from different heights. The results were reported in terms of height for 50% probability of explosion (h50%). The picture of micron sized crystal MOF was obtained by a microscope (×1000). Molecular structure packing diagram and coordination polyhedron geometry of [Pb(bta)·2H2O]n were drawn by Mercury and Diamond software (Diamond 3.1, Crystal Impact GbR, Bonn, Germany).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!