The largest database of trusted experimental protocols

Agilent 1200 infinity series

Manufactured by Agilent Technologies
Sourced in United States, Germany

The Agilent 1200 Infinity series is a modular high-performance liquid chromatography (HPLC) system. It is designed to provide reliable and consistent separation and detection of a wide range of chemical compounds. The system consists of various modules, including pumps, autosamplers, detectors, and a control interface, which can be configured to meet specific analytical requirements.

Automatically generated - may contain errors

23 protocols using agilent 1200 infinity series

1

Drug Release Evaluation in PBS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Drug release tests were carried out at 37°C in phosphate-buffered saline (PBS) under stirring. The released DOX was analyzed with an ultraviolet-visible spectrophotometer (PerkinElmer). The released OXA was measured by inductively coupled plasma atomic emission spectroscopy (Thermo Fisher Scientific). The released R837 was detected by high-performance liquid chromatography (Agilent 1200 Infinity Series), and the released anti-PDL1 was determined with the Rat IgG Total ELISA Kit (eBioscience).
+ Open protocol
+ Expand
2

Oxidative Balance and Endothelial Function

Check if the same lab product or an alternative is used in the 5 most similar protocols
The endpoints of the study were markers of oxidative balance/stress and endothelial dysfunction, which have important pathophysiologic and prognostic roles in atherothrombosis. Thus, through an acute in vivo study6 (link),7 (link), precise and informative insights were provided based on: 1) serum levels of soluble NOX2-derived peptide (sNOX2-dp), 2) serum levels of nitric oxide (NO) bioavailability, 3) serum levels of 8-iso-prostaglandin F2α-III (8-iso-PGF2α-III), 4) serum levels of vitamin E, and 5) brachial FMD.
Specifically, sNOX-2-dp was measured using an enzyme-linked immunosorbent assay as described by Loffredo et al8 (link). Nitric oxide bioavailability was measured with a colorimetric assay kit (Abcam, DRG International), 8-iso-PGF2α-III was appraised with a colorimetric assay kit (Abcam, DRG International), and vitamin E was analyzed using an Agilent 1200 Infinity series high-performance liquid chromatography system (Agilent Technologies, equipped with an Eclipse Plus C18 column), with results presented as the ratio between the concentration of a-tocopherol and serum total cholesterol. Brachial FMD was measured with a standardized and validated procedure8.
+ Open protocol
+ Expand
3

VEGFR1D2 Stability Analysis by LC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
VEGFR1D2 of 20 μm was incubated in phosphate buffer 20 mm and TCEP 5 or 10 mm, pH = 7. All the reactions were monitored by LC‐MS, analyzing protein samples on an Agilent 1200 Infinity Series (Agilent Technologies) equipped with an ESI source and a ToF detector. Protein samples were loaded on an Aeris widepore (150 × 2.1 mm, 3.6 μm) XB‐C8 column (Phenomenex), applying a method with a flow rate of 0.2 mL·min−1 and a linear gradient of CH3CN (0.05% TFA) in H2O (0.05% TFA) from 5% to 70% in 20 min. The protein sample precipitated after 24 h of incubation with 10 mm TCEP was centrifuged, the pellet was resuspended in 50 μL of 8 M guanidinium hydrochloride and 1.5 μL were also analyzed.
+ Open protocol
+ Expand
4

HPLC Analysis of Phytochemicals

Check if the same lab product or an alternative is used in the 5 most similar protocols
The analyses of catechin, 2,3,4,5ʹ-tetrahydroxystilbene-2-O-α-glucoside (TSG), rhein, emodin, and chrysophenol were performed on an Agilent 1200 Infinity series (Agilent Technologies, Santa Clara, CA, USA). Samples were separated on an XTerra RP18 analytical column (4.6 × 250 mm, 5 μm; Waters Corp., Milford, MA, USA). The mobile phase contained 0.5% glacial acetic acid in water (solution A) and 0.5% glacial acetic acid in acetonitrile (solution B). The injection volume was 20 μL, and the flow rate was 1.0 mL/min at 30 °C. The detection wavelength was 280 nm.
+ Open protocol
+ Expand
5

Plasma Homocysteine Analysis by HPLC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The plasma homocysteine was determined by HPLC-MS/MS (Agilent 1200 Infinity Series, Agilent Technologies, Santa Clara, CA, USA; Triple Quad™ 4500, SCIEX, Framingham, MA, USA). The coefficient of variation was of 1.5–1.6%.
+ Open protocol
+ Expand
6

Electrochemical Characterization of ZHN-SDS-BP

Check if the same lab product or an alternative is used in the 5 most similar protocols
The deionized water from EASY pure LF, Barnstead, was used to prepare all the solutions. The characterization of ZHN-SDS-BP was carried out with a scanning electron microscope (SEM) and a transmission electron microscope (TEM) using a field emission scanning electron microscopy (FESEM) model Hitachi SU 8020 UHR (Tokyo, Japan). The pH measurements were conducted with a pH meter, an Orion 720A (MI, USA), equipped with a glass electrode. Cyclic voltammetry (CV) and square-wave voltammetry (SWV) were performed with the Potentiostat Gamry Series-G750 (Warminster, PA, USA), and the electrochemical data were obtained using three-electrode systems consisting of an Ag/AgCl electrode model MF-2052 from Bioanalytical System (West Lafayette, IN, USA) with a fiber junction as the reference electrode, a platinum wire as a counter/auxiliary electrode, and a ZHN-SDS-BP-modified multi-walled carbon nanotube (MWCNT) paste electrode as a working electrode. The electrochemical impedance spectroscopy (EIS) measurements were carried out on a Potentiostat/Galvanostat Gamry model Ref 3000 (PA, USA). Convincingly, the real sample validation was conducted by using high-performance liquid chromatography (HPLC, column, C-18; detector, UV-Vis (DAD) 280 nm) with the Agilent 1200 Infinity Series (Waldbronn, Germany).
+ Open protocol
+ Expand
7

Disulfide Bridge Oxidation State

Check if the same lab product or an alternative is used in the 5 most similar protocols
In order to estimate the oxidation state of the disulfide bridge at 353 K, VEGFR1D2 was diluted up to 10 µm in Tris‐HCl 50 mm, pH = 7.0, containing NaCl 150 mm and incubated for 40 min in a thermostated bath set to the above‐mentioned temperature. Then, iodoacetamide (Sigma‐Aldrich, St. Louis, MO, USA), dissolved at a concentration of 500 mm in the same buffer, was added to the heat‐treated protein domain at a final concentration of 14 mm. The alkylation reaction was allowed to proceed at room temperature, for 30 min, in the dark. As positive control, VEGFR1D2 was first reduced with 5 mm DTT at 353 K for 40 min and then alkylated by iodoacetamide as described for the unreduced sample. Analysis of protein samples was assessed by LC‐MS on an Agilent 1200 Infinity Series (Agilent Technologies) equipped with an ESI source and a ToF detector, using an Aeris widepore (150 × 2.1 mm, 3.6 μm) XB‐C8 column (Phenomenex, Bologna, Italy), applying a method with a flow rate of 0.2 mL·min−1 and a linear gradient of CH3CN (0.05% TFA) in H2O (0.05% TFA) from 5% to 70% in 20 min.
+ Open protocol
+ Expand
8

Optimized LC-MS Metabolic Fingerprinting of Plasma Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
Plasma metabolic fingerprinting in the RH study was performed with the Agilent 1200 Series LC system (Agilent Technologies, Waldbronn, Germany) coupled with the Agilent 6224 Series TOF LC/MS system (Agilent Technologies, Waldbronn, Germany). In the PH study, plasma metabolic fingerprinting was conducted with the Agilent 1200 Infinity series (Agilent Technologies, Waldbronn, Germany) coupled with the Agilent Technologies QTOF (6520) mass spectrometry detector. The chromatographic and mass spectrometer parameters of the optimized LC/MS methods were described in detail in the Supplementary Material section. Quality control samples (QCs) were prepared as a pool of equal volume of each plasma samples included in each study. The QCs were analyzed in order to monitor system's and method's stability during the whole sequence run. Detailed clinical information about studied groups in both non-targeted metabolomics studies were described in Tables S1, S2 in the Supplementary Material section.
+ Open protocol
+ Expand
9

Plasma Methylmalonic Acid Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The plasma methylmalonic acid was determined by HPLC-MS/MS (Agilent 1200 Infinity Series, Agilent Technologies, Santa Clara, CA, USA; Triple Quad™ 5500, SCIEX, Framingham, MA, USA). The coefficient of variation was of 2.6–5.0%.
+ Open protocol
+ Expand
10

SEC-HPLC Analysis of ScpA Variant Stability

Check if the same lab product or an alternative is used in the 5 most similar protocols
An SEC-HPLC method was developed to analyze
the stability and homogeneity of ScpA variants using an Agilent 1200
Infinity Series (Agilent Technologies, USA) and comprised a quaternary
pump (G1311B 1260), an ALS autosampler (G1329B 1260), a thermostated
column compartment (G1316A 1260), and an MWD VL diode-array detector
(G1365D 1260). The chromatographic separations were carried out using
an AdvanceBio SEC 300A 2.7 μm 7.8 × 300 mm column. The
column was washed with water and equilibrated with the mobile phase
(150 mM sodium phosphate, pH 7.0) for 2 h with a flow rate of 1 mL/min.
The ScpA samples (5 μL of a 1 mg/mL solution) were injected,
and the eluate was analyzed for 15 min, using both 280 and 220 nm
detection wavelengths. The peak shape and retention times were analyzed
and used for the interpretation of purity. All the measurements were
performed in triplicates and the average spectra used for analysis.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!